6 CCR 1007-1 Part 09
[Editor’s Notes follow the text of the rules at the end of this CCR Document.]
_________________________________________________________________________
Adopted by the Board of Health October 16, 2024, effective date December 15, 2024
PART 9
RADIATION SAFETY REQUIREMENTS FOR PARTICLE ACCELERATORS NOT
USED IN THE HEALING ARTS
9.1
Purpose and Scope.
9.1.1
Authority.
9.1.1.1 Rules and regulations set forth herein are adopted pursuant to the provisions of sections
25 1 108, 25-1.5-101(1)(l), and 25-11-104, CRS.
9.1.2
Basis and Purpose.
9.1.2.1 A statement of basis and purpose accompanies this part and changes to this part. A copy
may be obtained from the Department.
9.1.3
Scope.
9.1.3.1 This part establishes procedures for the registration and the use of particle accelerators
not used in the healing arts.
9.1.4
Applicability.
9.1.4.1 The requirements and provisions of Part 9 apply to each registrant or applicant for
registration subject to Part 9 unless specifically exempted from Part 9.
9.1.4.2 The provisions of this part are in addition to (and not in substitution for) other applicable
provisions of Part 1, 2, 4, 10 or other parts of these regulations.
(1)
Registrants engaged in industrial radiographic operations are subject to the
applicable requirements of Part 5.
(2)
Registrants engaged in use of particle accelerators in the healing arts are subject
to the requirements of Part 24.
(3)
Registrants whose operations result in the production of radioactive material are
subject to the applicable requirements of Part 3.
9.1.5
Published Material Incorporated by Reference.
9.1.5.1 Throughout this Part 9, federal regulations, state regulations, and standards or guidelines
of outside organizations have been adopted and incorporated by reference. Unless a
prior version of the incorporated material is otherwise specifically indicated, the materials
incorporated by reference cited herein include only those versions that were in effect as
of the most recent effective date of this Part 9 (December 2024), and not later
amendments or editions of the incorporated material.
9.1.5.2 Materials incorporated by reference are available for public inspection, and copies
(including certified copies) can be obtained at reasonable cost, during normal business
hours from the Colorado Department of Public Health and Environment, Hazardous
Materials and Waste Management Division, 4300 Cherry Creek Drive South, Denver,
Colorado 80246. Additionally, https://cdphe.colorado.gov/hm/radregs identifies where the
incorporated federal and state regulations are available to the public on the internet at no
cost. A copy of the materials incorporated in this Part is available for public inspection at
the state publications depository and distribution center.
9.1.5.3 Availability from Source Agencies or Organizations.
(1)
All state regulations incorporated by reference herein are available at no cost in
the online edition of the Code of Colorado Regulations (CCR) hosted by the
Colorado Secretary of State’s Office, online at
https://www.sos.state.co.us/CCR/NumericalDeptList.do#1000.
9.2
Definitions.
9.2.1
Definitions of general applicability to these regulations are in Part 1, section 1.2.
REGISTRATION PROCEDURE
9.3
Registration.
9.3.1
No person shall receive, possess, use, transfer, own, or acquire a particle accelerator except as
authorized in a registration issued pursuant to Part 2.
9.3.2
In addition to the requirements of Part 2, a registration application for use of a particle accelerator
will be approved only if the Department determines that:
9.3.2.1 The applicant is qualified by reason of training and experience to use the accelerator in
question for the purpose requested in accordance with this part and Parts 4 and 10 in
such a manner as to minimize danger to public health and safety or property;
9.3.2.2 The applicant's proposed or existing equipment, facilities, and operating and emergency
procedures are adequate to protect health and minimize danger to public health and
safety or property;
9.3.2.3 The issuance of the registration will not be inimical to the health and safety of the public,
and the applicant satisfies any applicable special requirement in 9.4;
9.3.2.4 The applicant has appointed a Radiation Safety Officer in accordance with Part 2,
Appendix 2A;
9.3.2.5 The applicant and the applicant's staff have substantial experience in the use of particle
accelerators and training sufficient for application to their intended uses;
9.3.2.6 The applicant has established a radiation safety committee to approve, in advance,
proposals for uses of particle accelerators, as required by the regulations. Where a
radiation safety committee has not been established, the facility shall have written
procedures to approve, in advance, proposals for use of particle accelerators; and
9.3.2.7 The applicant has an adequate training program for operators of particle accelerators.
9.3.3
If radioactive materials are also present at the facility, the requirements of Part 9 apply, as
appropriate, and shall be coordinated with the applicable license or application for a license.
RADIATION SAFETY REQUIREMENTS FOR THE USE OF PARTICLE ACCELERATORS
9.4
Use.
9.4.1
A registrant shall use the accelerator in accordance with the manufacturer's radiation safety and
operating instructions.
9.5
Limitations.
9.5.1
No registrant shall permit any individual to act as an operator of a particle accelerator until such
individual has:
9.5.1.1 Been instructed in radiation safety and shall have demonstrated an understanding
thereof;
9.5.1.2 Received copies of and instruction in this part and the applicable requirements of Parts 4
and 10, pertinent registration conditions and the registrant's operating and emergency
procedures, and shall have demonstrated understanding thereof; and
9.5.1.3 Demonstrated competence to use the particle accelerator, related equipment, and survey
instruments which will be employed.
9.5.2
The radiation safety officer shall have the authority to terminate the operations at a particle
accelerator facility if such action is deemed necessary to minimize danger to public health and
safety or property.
9.6
Shielding and Safety Design Requirements.
9.6.1
A qualified expert, registered with the Department in accordance with Appendix 2B or Appendix
2C, shall be consulted in the design of a particle accelerator installation and called upon to
perform a radiation survey when the accelerator is first capable of producing radiation.
9.6.2
Each particle accelerator installation shall be provided with such primary and secondary barriers
as are necessary to assure compliance with Part 4, sections 4.6 and 4.14.
9.7
Particle Accelerator Controls and Interlock Systems.
9.7.1
Instrumentation, readouts, and controls on the particle accelerator control console shall be clearly
identified and easily discernible.
9.7.2
Each entrance into a target room or other high radiation area shall be provided with a safety
interlock that shuts down the machine under conditions of barrier penetration.
9.7.3
Each safety interlock shall be on a circuit which shall allow it to operate independently of all other
safety interlocks.
9.7.4
All safety interlocks shall be designed so that any defect or component failure in the safety
interlock system prevents operation of the accelerator.
9.7.5
When a safety interlock system has been tripped, it shall only be possible to resume operation of
the accelerator by:
9.7.5.1 Manually resetting the controls at the position where the safety interlock has been tripped
and, lastly, at the main control console; or
9.7.5.2 After the condition causing the interrupt has been corrected.
9.7.6
A scram button or other emergency power cutoff switch shall be located and easily identifiable in
all high radiation areas.
9.7.6.1 Such a cutoff switch shall include a manual reset so that the accelerator cannot be
restarted from the accelerator control console without resetting the cutoff switch.
9.8
Warning Devices.
9.8.1
Each location designated as high radiation area, and each entrance to such location, shall be
equipped with easily observable warning lights that operate when, and only when, radiation is
being produced.
9.8.2
Except in facilities designed for human exposure, each high radiation area shall have an audible
warning device which shall be activated for fifteen (15) seconds prior to the possible creation of
such high radiation area.
9.8.2.1 Such warning device shall be clearly discernible in all high radiation areas.
9.8.3
Barriers, temporary or otherwise, and pathways leading to high radiation areas shall be posted in
accordance with Part 4, section 4.28.
9.9
Operating Procedures.
9.9.1
Each particle accelerator, when not in operation, shall be secured to prevent unauthorized use.
9.9.2
The safety interlock system shall not be used to turn off the accelerator beam except in an
emergency.
9.9.3
Safety Checks.
9.9.3.1 All safety and warning devices, including interlocks, shall be checked for proper operation
at intervals not to exceed three (3) months. Safety checks shall assure, as appropriate,
proper operation of:
(1)
Electrical interlocks at each room entrance;
(2)
Timer, dose terminator, emergency off and door interlocks;
(3)
Beam condition indicator lights on the accelerator unit, on the control panel, and
in the facility;
(4)
Viewing systems;
(5)
Doors from inside and outside the accelerator room; and
(6)
Electrically assisted room doors with the accelerator unit electrical power turned
off.
9.9.3.2 A registrant shall promptly repair any system identified in 9.9.3.1 that is not operating
properly. The accelerator shall not be used until all repairs are completed.
9.9.3.3 Records.
(1)
A registrant shall maintain a record of the results of each safety check required
by 9.9.3.1 for three (3) years at the accelerator facility for inspection by the
Department.
(2)
The record shall include:
(a)
The date of the safety check;
(b)
The manufacturer's name, model number, and serial number for the
accelerator;
(c)
The manufacturer's name, model number, serial number and calibration
date of the instrument used to conduct any measurements;
(d)
Notations indicating the operability of each entrance door interlock, each
electrical or mechanical stop, each beam condition indicator light, the
viewing system, and doors: and
(e)
The signature of the individual who performed the periodic spot checks.
9.9.3.4 If the result of the safety checks required in 9.9.3.1 indicate the malfunction of any
system specified in 9.9.3.1, the registrant shall lock the control console in the “off”
position and not use the unit except as may be necessary to repair, replace, or check the
malfunctioning system.
9.9.4
Electrical circuit diagrams of the accelerator and the associated safety interlock systems shall be
kept current and maintained for inspection by the Department and shall be available to the
operator at each accelerator facility.
9.9.5
If, for any reason, it is necessary to intentionally bypass a safety interlock or interlocks, such
action shall be:
9.9.5.1 Authorized by the Radiation Safety Officer;
9.9.5.2 Recorded in a permanent log and a notice posted at the accelerator control console; and
9.9.5.3 Terminated as soon as possible.
9.9.6
Safety Instructions.
9.9.6.1 A copy of the current operating and the emergency procedures shall be maintained at the
accelerator control panel. These instructions shall inform the operator of:
(1)
The procedure to be followed if the operator is unable to turn the accelerator off
with controls at the control panel or any other abnormal operation occurs; and
(2)
The names and telephone numbers of the authorized users and Radiation Safety
Officer to be immediately contacted if the accelerator or console operates
abnormally; and
9.9.6.2 A registrant shall provide instruction in the topics identified in 9.9.6.1 to each individual
who operates an accelerator and shall provide appropriate refresher training to each
individual operator at intervals not to exceed one (1) year.
9.9.6.3 A registrant shall maintain a record of each individual receiving instructions required by
9.9.6.2, a description of the instruction, the date of instruction, and the name of the
individual who gave the instruction for three (3) years.
9.10
Radiation Monitoring Requirements.
9.10.1 There shall be available at each particle accelerator facility appropriate portable monitoring
equipment which is operable and has been appropriately calibrated for the radiations being
produced at the facility.
9.10.1.1
Such equipment shall be tested for proper functioning prior to each day of
accelerator operation and calibrated at intervals not to exceed one (1) year and after
each servicing and repair.
9.10.2 A radiation protection survey shall be performed and documented by a qualified expert, registered
with the Department, when changes have been made in shielding, operation, equipment, or
occupancy of adjacent areas.
9.10.2.1
A registrant shall maintain a record of the radiation measurements made
following installation of the accelerator for the duration of the registration and when
subsequent changes occur that could potentially affect the radiation levels in adjacent
areas.
9.10.2.2
The record required by 9.10.2.1 shall include:
(1)
The date of the measurements;
(2)
The manufacturer's name, model number and serial number of the accelerator;
(3)
A description of the accelerator configuration including whether there were any
test objects in the accelerator beam;
(4)
The instrument used to measure radiation levels;
(5)
A plan of the areas surrounding the accelerator that were surveyed;
(6)
The measured dose rate at several points in each area expressed in microsievert
(millirem) per hour;
(7)
The calculated maximum level of radiation over a period of one (1) week for each
restricted and unrestricted area; and
(8)
The signature of the Radiation Safety Officer.
9.10.3 The radiation levels in all high radiation areas not in the exposure room shall be continuously
monitored.
9.10.3.1
The monitoring devices shall be electrically independent of the accelerator
control and safety interlock systems and capable of providing a readout at the control
panel.
9.10.4 All area monitors shall be calibrated at intervals not to exceed one (1) year and after each
servicing and repair.
9.10.5 Whenever applicable, periodic surveys shall be made to determine the amount of airborne
particulate radioactivity present.
9.10.6 All surveys shall be made in accordance with the written procedures established by a qualified
expert, registered with the Department, or the Radiation Safety Officer.
9.10.7 Records of all radiation protection surveys, calibrations, and instrumentation tests shall be
maintained for three (3) years at the accelerator facility for inspection by the Department.
9.11
Ventilation Systems.
9.11.1 Ventilation systems shall be provided to ensure that personnel entering any area where airborne
radioactivity may be produced will not be exposed to airborne radioactive material in excess of
those limits specified in Part 4, Appendix 4B, Table 4B1.
9.11.2 A registrant shall not vent, release, or otherwise discharge airborne radioactive material to an
unrestricted area which exceeds the limits specified in Part 4, Appendix 4B, Table 4B2, except as
authorized pursuant to Part 4, section 4.33.
9.11.3 For purposes of 9.11.1 and 9.11.2, concentrations may be averaged over a period not greater
than one (1) year.
9.11.4 Every effort should be made to maintain releases of radioactive material to unrestricted areas as
far below the limits prescribed in 9.11.1 and 9.11.2 as is reasonably achievable.