RADIATION CONTROL - REGISTRATION OF RADIATION MACHINES, FACILITIES AND SERVICES

6 CCR 1007-1 Part 02

[Editor’s Notes follow the text of the rules at the end of this CCR Document.]

_________________________________________________________________________

Adopted by the Board of Health February 21, 2024, effective date April 14, 2024

PART 2:

REGISTRATION OF RADIATION MACHINES, FACILITIES AND SERVICES

2.1

Purpose and Scope.

2.1.1

Authority

2.1.1.1 Rules and regulations set forth herein are adopted pursuant to the provisions of sections

25 1 108, 25 1.5 101(1)(l), and 25-11-104, CRS.

2.1.2

Basis and Purpose.

2.1.2.1 A statement of basis and purpose of these regulations accompanies this part and

changes to this part. A copy may be obtained from the Department.

2.1.3

Scope.

2.1.3.1 This part provides for:

(1)

Registration of facilities;

(2)

Certification of radiation machines;

(3)

Registration of persons providing radiation machine services including assembly,

installation, maintenance and repair;

(4)

Registration of qualified inspectors and qualified experts; and

(5)

Approval of radiation safety officers, mammographers and other operators.

2.1.4

Applicability.

2.1.4.1 The requirements and provisions of this part apply to each person who uses, operates,

services or certifies radiation machines and to each registrant or applicant for registration

subject to this part unless specifically exempted.

2.1.4.2 The provisions of this part are in addition to (and not in substitution for) other applicable

provisions in Parts 1, 4, 5, 6, 7, 8, 9, 10, 24 and other parts of these regulations.

2.1.5

Published Material Incorporated by Reference.

2.1.5.1 Throughout this Part 2, federal regulations, state regulations, and standards or guidelines

of outside organizations have been adopted and incorporated by reference. Unless a

prior version of the incorporated material is otherwise specifically indicated, the materials

incorporated by reference cited herein include only those versions that were in effect as

of the most recent effective date of this Part 2 (April, 2024), and not later amendments or

editions of the incorporated material.

2.1.5.2 Materials incorporated by reference are available for public inspection, and copies

(including certified copies) can be obtained at reasonable cost, during normal business

hours from the Colorado Department of Public Health and Environment, Hazardous

Materials and Waste Management Division, 4300 Cherry Creek Drive South, Denver,

Colorado 80246. Additionally, https://cdphe.colorado.gov/hm/radregs identifies where the

incorporated federal and state regulations are available to the public on the internet at no

cost. A copy of the materials incorporated in this Part is available for public inspection at

the state publications depository and distribution center.

2.1.5.3 Availability from Source Agencies or Organizations.

(1)

All federal agency regulations incorporated by reference herein are available at

no cost in the online edition of the Code of Federal Regulations (CFR) hosted by

the U.S. Government Printing Office, online at

https://www.govinfo.gov/app/collection/cfr/.

(2)

All state regulations incorporated by reference herein are available at no cost in

the online edition of the Code of Colorado Regulations (CCR) hosted by the

Colorado Secretary of State’s Office, online at

https://www.sos.state.co.us/CCR/NumericalDeptList.do#1000.

(3)

Copies of the standards or guidelines of outside organizations are available at no

cost or for purchase from the source organizations listed below.

(a)

American Registry of Radiologic Technologists

1255 Northland Drive

St. Paul, MN 55120-1155

Phone (651) 687-0048

https://www.arrt.org/

2.2

Definitions.

2.2.1

Definitions of general applicability to these regulations are in Part 1, section 1.2.

2.2.2

As used in Part 2, each term below has the definition set forth.

“ARRT” means the American Registry of Radiologic Technologists, 1255 Northland Drive, St.

Paul, MN 55120, Phone (651) 687-0048, web site: https://www.arrt.org/.

“ASRT” means the American Society of Radiologic Technologists.

“Assembler” means any person engaged in the business of assembling, replacing, or installing

one or more components into a radiation machine system or subsystem.

“Calibration” means to adjust and/or determine the:

(1)

Response or reading of an instrument relative to a series of conventionally true

values; or

(2)

Strength of a radiation source relative to a standard or conventionally true value.

“Certification Evaluation” (CE) means the evaluation of a radiation machine at a facility by a

qualified inspector or the Department for the purpose of ascertaining the performance of the

radiation machine system and/or facility in order to determine conformance with these

regulations.

“Certified Nuclear Medicine Technologist” means an individual who is currently registered in

nuclear medicine with the NMTCB or ARRT, designated CNMT or R.T.(N), respectively.

“Computed tomography” (CT) means the production of a tomogram by the acquisition and

computer processing of x-ray transmission data. For the purposes of Part 2, the requirements

stated for computed tomography machines do not apply to:

(1)

“Volumetric Dental Imaging Systems”; or

(2)

Digital breast tomosynthesis.

“Direct supervision” means the supervisor is present in the facility and immediately available to

furnish assistance and direction to the supervisee throughout the performance of a procedure.

(1)

The direct supervisor is not required to be present in the room when the

procedure is performed.

“Dual-energy X-Ray Absorptiometry” (DXA, previously DEXA) means an imaging technique using

radiation machines for quantifying bone density, used in the diagnosis and management of

osteoporosis.

“Facility” means, for purposes of Part 2, the location within one building (or vehicle, or under one

roof, or at one address) and under the same administrative control, at which a radiation machine

is or was installed, operated and/or located.

“FDA” means the United States Food and Drug Administration.

“Fluoroscopy” means a technique for generating x-ray images and presenting them

simultaneously and continuously as visible images.

“Industrial Radiography” means an examination of the structure of materials by the nondestructive

method of utilizing ionizing radiation to make radiographic images.

“Inter-comparison” means the direct comparison, in accord with 2.4.4.5, of two instruments

designed to measure the same physical quantity.

“Limited-scope operator” (LSO) means an individual who has taken and passed a required test

and has approval by the Department pursuant to 2.4.5.1 to operate x-ray systems and to conduct

specified radiographic examinations of the chest, extremities, skull, hip/pelvis and spine/sacrum

“MQSA” means Mammography Quality Standards Act.

“NIST” means the National Institute of Standards and Technology.

“NMAA” means a Nuclear Medicine Advanced Associate working as a mid-level provider under

the supervision of a licensed physician. The NMAA must be a Certified Nuclear Medicine

Technologist registered as an R.T.(N) or CNMT.

“NMTCB” means the Nuclear Medicine Technology Certification Board, Inc, 3558 Habersham at

Northlake, Building I, Tucker, GA 30084-4009, web site: https://www.nmtcb.org/.

“Operator” means an individual adequately trained in accordance with these regulations in the

purpose and experienced in the practice of performing a radiographic examination.

“Performance adjustment” means the adjustment or repair of a function (not including the setting

of operator-selectable functions, such as time, mA and/or kVp for an individual exposure) of an x

ray machine or imaging system that is required to bring the machine into compliance with these

regulations and the specifications.

“Personal supervision” is as defined in Part 1 of the regulations.

“Provisional mammographer” means an individual who is in-training to become a Qualified

mammographer and meets the requirements of Appendix 2M.

“Provisional qualified inspector” (PQI) means an individual who meets the applicable

requirements of Section 2I.2 of Appendix 2I and has current Department approval in a designated

specialty to perform evaluations of radiation machines, facilities, and operators for compliance

with these regulations while under the supervision of a qualified inspector.

“QE(R)” means a qualified expert medical physicist approved to design or evaluate shielding for

radiation machines used in the healing arts.

“QE(S)” means a qualified expert physicist approved to design or evaluate shielding for radiation

machines used for non-healing arts purposes.

“QE(T)” means a qualified expert medical physicist approved to design or evaluate shielding for

radiation machines used in radiation therapy.

“Qualified expert” (QE) means an individual who meets the applicable requirements of Appendix

2B or 2C and has current Department approval as QE(S), QE(R), or QE(T) to evaluate radiation

shielding design and recommend radiation safety practices, as provided in 2.4.3.

“Qualified inspector” (QI) means an individual who meets the applicable requirements of

Appendix 2I and has current Department approval in a designated specialty to perform

evaluations of radiation machines, facilities, and operators for compliance with these regulations,

as provided in 2.4.4.

“Qualified mammographer” means a mammographer who meets the applicable requirements of

2.4.5.4(1) and 2.4.5.4(2).

“Qualified trainer” (QT) means an individual whose training and experience adequately prepares

the individual to carry out specified training assignments as illustrated in Appendix 2J.

“Radiology Practitioner Assistant” means an individual who is currently registered as RPA by the

Certification Board for Radiology Practitioner Assistants and are designated RPA (CBRPA).

“Radiographic Examination” means performing a procedure, including selection of exposure

settings, positioning the x-ray system and the patient, and initiating and terminating the exposure.

“Radiologic technologist” means an individual who is currently registered in radiologic technology

with the ARRT. See “R.T.(CT)(ARRT)”, “R.T.(R)(M)(ARRT)”, “R.T.(N)(ARRT)”, “R.T.(R)(ARRT)”,

and “R.T.(T)(ARRT)”.

“Registered Radiologist Assistant” means an individual who is certified by the ARRT as a

Registered Radiologist Assistant designated as R.R.A.(ARRT).

“Registered medical physicist” (RMP) means an individual who meets the applicable

requirements of Appendix 2I and has current Department approval to perform medical physics

activities, including shielding design, performing radiation surveys, and providing consultation for

radiation protection and quality assurance and clinical medical physics for radiation therapy,

computed tomography, mammography and/or other healing arts facilities.

“R.T.(CT)(ARRT)” means an individual who is certified and registered by the ARRT with a

specialty post-secondary certification in computed tomography. (Note: Since CT certification is a

post-secondary registration and has several primary paths, the “(R)” is not included as it may vary

between individuals depending on their primary certification.)

“R.T.(R)(M)(ARRT)” means an individual who is certified and registered by the ARRT in

radiography with a specialty certification in mammography.

“R.T.(N)(ARRT)” means an individual who is certified and registered by the ARRT in nuclear

medicine technology.

“R.T.(R)(ARRT)” means an individual who is certified and registered by the ARRT in radiography.

“R.T.(T)(ARRT)” means an individual who is certified and registered by the ARRT in radiation

therapy.

“Service company” means a person who is engaged (or offers to engage) in the business of

selling, leasing, transferring, lending, assembling, installing, maintaining, repairing, storing,

trading out, disabling, or disposing of radiation machines and their related components, or is

engaged in the business of furnishing or offering to furnish radiation machine servicing or

services.

Service technician” means an individual who is employed by a service company to perform

radiation machine servicing or services.

“Shielding design” means physical specifications, such as room layout, floor plan, construction

materials, and equipment configuration, to demonstrate compliance with the radiation limits set

forth in Part 4 of these regulations.

“Volumetric dental imaging system” means an x-ray machine that produces, for oral and

maxillofacial structures, a three-dimensional tomographic data set or a time sequence of three-

dimensional tomographic data sets. A dental x-ray machine only capable of producing a two-

dimensional image is not considered to be a volumetric dental imaging system. For the purposes

of Part 2, the requirements stated for “computed tomography” machines do not apply to

“Volumetric Dental Imaging Systems”.

EXEMPTIONS FROM THE REGULATORY REQUIREMENTS

2.3

Exemptions.

2.3.1

Electronic equipment that is not designed primarily to produce radiation is exempt from the

registration and notification requirements of Part 2, provided that the dose equivalent rate

averaged over an area of 10 cm2 does not exceed 5 µSv (0.5 mrem) per hour at 5 cm from any

accessible surface of such equipment.

2.3.2

Radiation machines in transit or in storage incident to transit are exempt from the requirements of

Part 2.

2.3.3

Domestic television receivers, computer monitors, and similar devices are exempt from the

requirements of Part 2.

2.3.4

A radiation machine that is out of service yet kept at a facility is exempt from the registration and

certification evaluation requirements of Part 2 provided:

2.3.4.1 The radiation machine has been made physically inoperable by inactivating or

dismantling the electrical circuitry such that the radiation machine is not capable of

producing radiation, and

2.3.4.2 The Department has received documentation of 2.3.4.1 on Form R 61, “Disposition of a

Radiation Machine”, or equivalent form, that is signed by a registered service technician.

2.3.5

An electron microscope or electron microprobe is exempt from Part 2 provided that:

2.3.5.1 A survey shows compliance with 2.3.1; or

2.3.5.2 The device is not capable of exceeding an operating voltage of 50,000 electron volts.

2.3.6

The legal owner of electronic equipment which meets the requirements of 2.3.1 but which is not

specifically exempted under 2.3.2, 2.3.3, and 2.3.4 shall maintain for the lifetime of the equipment

radiation measurement results or certification from the manufacturer or a qualified expert

indicating that the equipment complies with the exposure rates specified in 2.3.1.

REQUIREMENTS FOR DEPARTMENT APPROVAL AND/OR REGISTRATION

2.4

State of Colorado Authorization or Approval Recognized by the Department is Required

for Each Category Designated in This Section.

2.4.1

Registration of a Facility.

2.4.1.1 Each person possessing or in the process of coming into the possession of a radiation

machine facility shall:

(1)

Be registered with the Department prior to using a radiation producing machine

at the facility;

(2)

Before the facility registration expiration date, at least every twelve (12) months,

submit a complete application for registration on the applicable Department R-4

series Form, and include all of the information required by the form and any

accompanying instructions. The facility shall:

(a)

Designate a radiation safety officer who meets the applicable

requirements of Appendix 2A to be responsible for overall radiation

protection for the facility; and

(b)

Document that a written shielding design has been:

(i)

Completed in accordance with Parts 6, 8, or 9 of these

regulations, as applicable, prior to any radiation machine

installation; and

(ii)

Retained on file at the facility for the life of the facility.

(c)

Pay the radiation machine facility registration fee for radiation control

services indicated by Part 12, Category 26. The radiation machine facility

registration fee is not required for registration updates required by 2.4.6.5

unless the update is submitted less than thirty (30) days prior to the

registrant’s expiration date.

2.4.1.2 As prescribed by 6.3.3.4 for a healing arts screening program, registrants shall complete

and submit a Healing Arts Screening application including all of the information required

by Part 6, Appendix 6F.

2.4.1.3 In addition to the other requirements of 2.4, any research using radiation machines on

living humans shall be approved by an Institutional Review Board (IRB).

2.4.2

Registration as a Service Company.

2.4.2.1 Each person who is engaged (or offers to engage) in the business of selling, leasing,

transferring, lending, assembling, installing, maintaining, repairing, storing, trading out,

disabling or disposing of radiation machines and their related components, or is engaged

in the business of furnishing or offering to furnish radiation machine servicing or services

in this State, shall be registered with the Department prior to performing such activities.

2.4.2.2 Each Service Company shall complete the Form R-60 series application for registration

with all of the information required by the Department indicated on the form and all

accompanying instructions, together with the fee required by Part 12, Category 22.

2.4.2.3 Each applicant for registration under 2.4.2 shall specify:

(1)

The service category for which registration is being requested, including but not

limited to:

(a)

Selling, leasing, transferring, lending, assembling, installing, maintaining,

trading out, disabling or disposing of radiation machines and associated

radiation machine components; and

(b)

Servicing of radiation machines and associated radiation machine

components, to include preventative maintenance, performance

adjustment, calibration, or repair.

(2)

The name and qualifications of each service technician who will provide service,

including:

(a)

A management attestation that the technician’s training and experience

was evaluated and meets the requirements of Appendix 2H; and

(b)

A management attestation that each service technician has been

instructed in, and demonstrates an understanding of the requirements of:

(i)

These regulations; and

(ii)

The Federal Performance Standard (21 CFR Chapter I,

Subchapter J; and

(3)

An attestation that the type of personnel dosimetric monitoring in use meets the

requirements of 4.17 and 4.18; and

(4)

An attestation that all calibration and testing instruments are adequate to ensure

that machine performance and manufacturer’s specifications will be met.

(5)

Each service company registrant under 2.4.2 shall notify the Department when

any service technician is no longer authorized to provide radiation machine

services for the registrant.

(a)

The registrant will be assessed the acceptance review fee required by

Part 12, Category 24 when adding a technician, unless technicians are

added during a registration renewal.

2.4.2.4 Service Company registration will be for a one (1) year period.

2.4.3

Registration as a Qualified Expert.

2.4.3.1 Each individual who designs or evaluates protective shielding around a radiation area so

the area meets the public exposure requirements of Part 4, shall be registered with the

Department as a qualified expert designated QE(R), QE(S) or QE(T).

(1)

Each individual who designs or evaluates shielding for a radiation machine

regulated by Parts 8 or 9 and not used in the healing arts shall be registered with

the department as a QE(S) and meet the requirements of Appendix 2C.

(2)

Each individual who designs or evaluates shielding for a radiation machine used

in the healing arts as regulated by Part 6, but not used in radiation therapy, shall

be registered with the department as a QE(R) and meet the requirements of

Appendix 2B

(3)

Each individual who designs or evaluates shielding for a radiation machine used

in radiation therapy as regulated by Part 24, shall be registered as a QE(T) and

meet the requirements of Appendix 2B.

2.4.3.2 Each Qualified Expert shall complete the applicable Form R-68 series application for

registration and include all of the information required by the form and any accompanying

instructions, together with the fee required by Part 12, Category 22.

2.4.3.3 Qualified Expert registration shall be for a one (1) year period.

2.4.4

Registration as a Qualified Inspector.

2.4.4.1 Each individual who performs a certification evaluation of a radiation machine or an

evaluation of a facility shall be registered with the Department as a qualified inspector

who meets the criteria established in Appendix 2I.

2.4.4.2 Each individual who performs a certification evaluation on mammography, fluoroscopy or

computed tomography machines used in the healing arts or, evaluates the quality

assurance programs of digital imaging systems used in the healing arts shall be

registered with the department as a qualified inspector with approval in the Registered

Medical Physicist category.

(1)

Individuals who perform a certification evaluation on Volumetric Dental Imaging

Systems shall be registered with the department as a qualified inspector with

approval in “Volumetric Dental Imaging Systems”.

2.4.4.3 Each individual who performs registered medical physicist duties required by Part 24

shall be registered with the department as a qualified inspector with approval in the

radiation therapy Registered Medical Physicist category.

2.4.4.4 Each Qualified Inspector shall complete the applicable Form R-53 series application for

registration and include all of the information required by the form and any accompanying

instructions, together with the fee required by Part 12.

2.4.4.5 Qualified Inspector registration shall be for a period of one (1) year.

2.4.4.6 Certification evaluation measurements shall be made with instruments that are

sufficiently sensitive to determine compliance with these regulations.

(1)

The instruments shall be maintained and used in good working order.

(2)

The instruments shall be calibrated at least every two (2) years, or in accordance

with the manufacturer's recommendation, whichever is more frequent, or after

any repair that could affect the calibration of the instrument.

(3)

Calibrations shall be NIST-traceable where such traceability is feasible.

(4)

Procedures for instrument calibration done by inter-comparison with a suitable

and appropriately calibrated instrument must be approved by the department.

(a)

The comparison shall be between an instrument that has a current

calibration traceable to NIST and an instrument for which a calibration

factor is to be determined.

(b)

The comparison shall be made using the actual physical quantity to be

routinely measured (for example, radiation energy/quality or visible light

spectrum) and shall be compared in the same physical geometry.

(c)

The procedure(s) for inter-comparison shall be documented and

available for review by the department.

(5)

In addition to the requirements in 2.4.4.6, instruments used for the certification

evaluation report to measure the air kerma or air kerma rate of mammography

machines shall be calibrated with an accuracy of ± six (6) percent (95 percent

confidence level) in the mammography energy range.

2.4.5

Registration of specific radiation machine operators.

Except as otherwise specified in these regulations, registration with the Department is not

required for an individual who holds a current, valid national registry in radiography, nuclear

medicine technology, radiation therapy, computed tomography or mammography as issued by

the ARRT or NMTCB (with specialty certification in Computed Tomography) or other nationally

recognized registry specifically accepted by the Department. Additional requirements may be

applicable in accordance with Appendix 2E, Appendix 2G, Appendix 2M, or Appendix 2O. All

other non-physician individuals operating x-ray imaging systems on living humans who are not

nationally registered or certified by ARRT or NMTCB must meet the requirements specified in the

regulations and shall register with the Department, when applicable.

2.4.5.1 Limited Scope Operator.

(1)

Each individual operating an x-ray system on living humans in the State of

Colorado, shall be registered as a Limited Scope Operator consistent with

2.4.5.1(2), except for:

(a)

Those individuals subject to 2.6.1.5, 2.6.1.6, 2.6.1.7, 2.6.1.8, 2.6.1.10,

2.6.1.11, and 2.6.1.12, or

(b)

Those individuals having current registration with the American Registry

of Radiologic Technologists in radiography.

(2)

Registration

(a)

The applicant for LSO registration must complete the requirements of

2D.2.1, 2D.2.2 and 2D.2.3 in a structured and documented training

program in order to apply for registration as a Limited Scope Operator.

(b)

Each Limited Scope Operator shall complete an application with all of the

information required by the form and instructions, together with the fee

required by Part 12, Category 24.

(i)

The Form R-70 series application shall be used to initiate the

registration process.

(ii)

The Form R-71 series application shall be used to confirm the

completion of the requirements of 2D.2.1, 2D.2.2 and 2D.2.3.

(c)

Application for registration as a Limited Scope Operator shall be made

within one year upon completion of the requirements of 2D.2.1 and within

ninety (90) calendar days upon completion of the requirements of 2D.2.2

and 2D.2.3.

(d)

If an applicant cannot achieve a passing score per 2D.2.4 within three

attempts, the applicant must restart the training required by 2D.2.1,

2D.2.2, and 2D.2.3.

(e)

Registrants must meet the requirements of 2D.2.5 in order to renew the

Limited Scope Operator registration.

(i)

The Form R-90 series application shall be used to renew the

registration for a Limited Scope Operator.

2.4.5.2 Computed Tomography Operator

(1)

Each individual operating a computed tomography system on living humans shall

meet the requirements of Appendix 2E.

2.4.5.3 Bone Densitometry Equipment Operator (BDEO).

(1)

Each individual operating a dual-energy x-ray absorptiometry system used on a

living human shall be registered as a Bone Densitometry Equipment Operator,

except for:

(a)

Those individuals registered with the American Registry of Radiologic

Technologists as a radiologic technologist, nuclear medicine technologist

or radiation therapist; or

(b)

Those individuals registered with the Nuclear Medicine Technology

Certification Board (NMTCB) as a certified nuclear medicine

technologist.

(2)

Registration

(a)

The applicant must complete the requirements of 2F.2.1, 2F.2.2, and

2F.2.3 in a structured and documented training program in order to apply

for registration as a Bone Densitometry Equipment Operator.

(b)

Applicants with International Society of Clinical Densitometry (ISCD)

certification must, at a minimum, document the completion of the

requirements of 2F.2.1.1 through 2F.2.1.3.

(i)

ISCD-certified applicants have met the requirements of 2F.2.1.4

through 2F.2.1.9, 2F.2.2 and 2F.2.3 and are exempt from the

requirements of 2F.2.4

(c)

Application for the Bone Densitometry Equipment Operator registration

shall contain all of the information required by the form and instructions,

together with the fee required by Part 12, Category 24.

(i)

The Form R-80 series application shall be used to initiate the

registration process.

(ii)

The Form R-81 series application shall be used to confirm the

completion of the requirements of 2F.2.1, 2F.2.2 and 2F.2.3.

(d)

Application for registration as a Bone Densitometry Equipment Operator

shall be made within one year upon completion of the requirements of

2F.2.1 and within ninety (90) calendar days upon completion of the

requirements of 2F.2.2 and 2F.2.3

(e)

If an applicant cannot achieve a passing score per 2F.2.4 within three

attempts, the applicant must restart the training required by 2F.2.1,

2F.2.2 and 2F.2.3.

(f)

Bone Densitometry Equipment Operator registration is issued for a

period of three years.

(g)

Registrants must meet the requirements of 2F.2.5 in order to renew the

Bone Densitometry Equipment Operator approval.

2.4.5.4 Mammographer

Any individual performing mammography shall:

(1)

Be certified by the ARRT in Mammography (R.T.(R)(M)(ARRT)); and

(2)

Meet the qualifications of and maintain the education and experience

requirements for MQSA under 21 CFR Part 900.12(a)(2);

Or

(3)

Register as a provisional mammographer, meet the requirements of Appendix

2M, and be considered to be in-training until the requirements of 2.4.5.4(1) and

2.4.5.4(2) are met.

2.4.5.5 Fluoroscopy operator

(1)

On or after January 1, 2021, each individual operating a fluoroscopy imaging

system on living humans shall be registered with the department as a

fluoroscopy operator consistent with 2.4.5.5(2) or 2.4.5.5(3), except for:

(a)

A physician who has an active license from the applicable State of

Colorado licensure board consistent with the requirements of Section

2.6.1.2; or

(b)

A Registered Radiologist Assistant or Radiology Practitioner Assistant

(RPA) who meets the requirements of Appendix 2G; or

(c)

An individual with a current R.T.(R), R.T.(CV), R.T.(CI), R.T.(VI), or

R.T.(T) registration.

(2)

Individuals whose training and experience has been evaluated by the department

in writing prior to January 1, 2021, as having met the training and experience

requirements of Appendix 2O:

(a)

Need not complete the training or testing requirements of Appendix

2O.1; and

(b)

Shall be required to obtain and maintain registration in accordance with

2.4.5.5(3)(b) through 2.4.5.5(3)(f) on or after January 1, 2021.

(3)

Registration

(a)

In order to apply for registration as a fluoroscopy operator, the applicant

for fluoroscopy operator registration must complete the requirements of

Appendix 2O in a structured and documented training program that

meets the requirements of ARRT or another program as authorized by

the regulations or as approved in writing by the department.

(b)

Each fluoroscopy operator shall complete an R-50 series application

form with all of the information required, together with the fee required by

Part 12, Category 24.

(i)

The Form R-50 series application form shall be used to confirm

the completion of the requirements of Appendix 2O.

(c)

Except for those individuals meeting the requirements of 2.4.5.5(2),

application for registration as a fluoroscopy operator shall be made within

one year following completion of the training requirements of Appendix

2O.

(d)

If an applicant cannot achieve a passing score on the applicable national

registration exam per Appendix 2O, section 2O.1.3.1 or 2O.1.3.2 within

three attempts, the applicant must restart the training required by

Appendix 2O.

(e)

A fluoroscopy operator registration is valid for a two year period.

(f)

Registrants must meet the requirements of 2O.2 in order to renew the

fluoroscopy operator registration.

(i)

The Form R-50 series application form shall be used to renew

the fluoroscopy operator registration every two years.

(g)

Reciprocal recognition of a registration or license specifically authorizing

fluoroscopy use and granted by another state or organization shall be

submitted to the Department for review and evaluation on an individual

case-by-case basis.

(h)

Department registered fluoroscopy operators shall operate machines

within their respective scope of practice, training, and experience.

2.4.6

General Requirements Applicable to Issuance and Maintenance of Department Registrations.

2.4.6.1 The application to be registered in the State of Colorado shall be submitted on the

appropriate Department form(s) and shall contain all information required by the

Department as indicated on the form(s) and all accompanying instructions.

2.4.6.2 Upon a determination that an applicant meets the requirements of the regulations, the

Department shall issue a Notice of Registration.

2.4.6.3 The Department may incorporate in the Notice of Registration at the time of issuance, or

thereafter by appropriate rule, regulation, or order, such additional requirements and

conditions with respect to the registrant's activities as the Department deems appropriate

or necessary.

2.4.6.4 Approval to conduct or perform activities in accordance with the registration requirements

of these regulations shall be:

(1)

For a period of two (2) years, except as otherwise specified by these regulations

or the Department; and

(2)

Limited to the category or categories of activities specifically designated in the

Notice of Registration.

2.4.6.5 The registrant shall notify the Department in writing within thirty (30) calendar days of

making any change of information contained in the application for registration and/or the

Notice of Registration.

2.4.6.6 Except as provided by 2.4.6.7, each Notice of Registration shall expire at the end of the

month in the year stated therein.

2.4.6.7 In any case in which a registrant, not less than thirty (30) calendar days prior to the

expiration of the registrant’s authorization, has filed an application in proper form for

renewal or for a new registration authorizing the same activities, such existing

authorization shall not expire until final action by the Department.

2.4.6.8 The Department will not review or otherwise process a new application or application for

renewal for which no fee is received.

(1)

All application fees are non-refundable.

2.4.6.9 The Department may deny, withdraw, limit or qualify its approval of any person to perform

activities upon determining that such action is necessary in order to prevent undue

hazard to health and safety, or for other reasonable cause.

2.4.7

Providing Notice of Registrant’s Rights

2.4.7.1 Whenever a business relationship exists between the qualified inspector and a registered

service company, a “Notice of Registrant's Rights” Form R-65 shall be provided to the

registered facility prior to beginning the service or evaluation, including:

(1)

When a qualified inspector is also registered to perform services and servicing;

(2)

When a qualified inspector is also a qualified expert; and

(3)

When a qualified inspector, a qualified expert and/or a services and servicing

provider is a member of the same corporation, partnership or other formal

business relationship.

2.4.8

No person, in any advertisement, shall refer to the fact that the person is registered with the

Department pursuant to the provisions of 2.4.1, 2.4.2, 2.4.3, 2.4.4, and 2.4.5 and no person shall

state or imply that the quality of conduct or performance of any activity under such registration

has been approved or endorsed by the Department.

CERTIFICATION EVALUATION

2.5

Certification Evaluations.

2.5.1

Frequency of Certification Evaluations.

2.5.1.1 Each radiation machine registrant shall have its radiation machine(s) and facility

evaluated by a Department-approved qualified inspector annually, except as provided in

2.5.1.2 through 2.5.1.5.

(1)

Each certification evaluation shall determine if the machine is safe for each

intended use and is in compliance with the specifications of the equipment

manufacturer and these regulations.

(2)

Each certification evaluation subsequent to the initial certification evaluation shall

be completed in or prior to the same calendar month as the previous certification

evaluation.

(3)

The calendar month of a certification evaluation of a machine in any month prior

to the month in which it is due shall become the calendar month in which the

subsequent certification is due.

(4)

A certification evaluation conducted after the month in which it was due shall not

change the month in which subsequent certification evaluations are due.

2.5.1.2 Each non-healing-arts x-ray imaging machine or system regulated by Parts 5, 8 or 9 shall

be inspected at least every two (2) years. These include, but are not limited to, x-ray

machines used for industrial radiography, nondestructive analysis, forensics or non-

human security screening, foodstuff, packaging or equipment inspections or

measurements.

2.5.1.3 Each bone densitometry, dental, podiatry or veterinary radiation machine shall be

inspected at least every three (3) years, except that:

(1)

Each radiographic x-ray machine used in non-intraoral dentistry or podiatry that

is capable of continuously variable kilovoltage peak (kVp) or continuously

variable milliamperage (mA) or continuously variable collimation shall be

inspected annually.

(2)

Each machine used in podiatry that is capable of operating at more than 30 mA

shall be inspected annually.

(3)

Each volumetric dental imaging system or computed tomographic system for

human use shall be inspected annually.

(4)

Each portable hand-held instrument used for any purpose on living humans shall

be inspected annually.

TABLE 2-1: SUMMARY OF FREQUENCY OF RADIATION MACHINE CERTIFICATION EVALUATIONS

Category

Frequency of

certification

evaluation

Excluding systems used in veterinary medicine, and unless otherwise

specified in this Table 2-1, each:

•

General use x-ray system;

•

CT (Computed Tomography) system;

•

Fluoroscopy system;

•

Dental Cone Beam Computed Tomography (CBCT) system;

•

Volumetric dental imaging system;

•

Hand-held x-ray imaging systems for human use;

•

Podiatry system used at more than 30 mA;

•

Non-intraoral dentistry or podiatry x-ray system capable of continuously

variable kilovoltage peak (kVp) or continuously variable milliamperage

(mA) or continuously variable collimation;

•

Therapy systems for human or veterinary use;

Every one (1) year

Category

Frequency of

certification

evaluation

•

Security scanner x-ray systems used on living humans;

•

All systems identified above entering the state under reciprocity for more

than 180 days.

Each industrial (non-healing-arts) x-ray imaging machine or system regulated

under Parts 5, 8 or 9 including:

•

Security scanners for non-living human use;

•

X-ray fluorescence (XRF) systems;

•

Industrial radiography/Non-destructive testing;

•

Forensics;

•

Tissue specimen imaging systems;

•

Scanning systems for food production or packaging inspection.

•

Therapy systems for non-healing arts use.

Every two (2)

years

Except as otherwise specified in this Table 2-1, each:

•

Bone densitometry (DXA) system;

•

Dental system;

•

Podiatry system used at less than or equal to 30 mA;

•

Veterinary system, including hand-held units.

Every three (3)

years

2.5.1.4 Except as otherwise specified in regulation, each radiation machine shall have a

certification evaluation performed within ninety (90) calendar days of:

(1)

The initial installation of a new radiation machine, a radiation machine that is new

to the facility, or a radiation machine that is relocated to a new area or room of an

existing facility; or

(2)

Any service after initial installation that could potentially affect radiation output

(dose indices) or technique settings, including but not limited to the repair or

replacement of high voltage generators, tube heads, consoles or image receptor

systems.

(3)

Receipt of a new radiation machine that does not require a physical installation,

including hand-held x-ray systems, or portable or fixed x-ray systems that are

battery operated or that plug into an electrical outlet.

2.5.1.5 The following radiation machines shall have a certification evaluation performed within

ninety (90) calendar days of installation and prior to being used to perform any

examination on living humans:

(1)

Each initial (new) installation of a mammography imaging system. The evaluation

must be performed by a registered medical physicist authorized in

mammography;

(2)

Each initial (new) installation of a Computed Tomography (CT) system,excluding

volumetric dental imaging systems, dental CBCT systems, and digital breast

tomosynthesis systems. The evaluation must be performed by or under the

personal supervision of a registered medical physicist authorized in CT.

2.5.1.6 Any radiation machine and/or facility not inspected in accordance with 2.5.1.1 through

2.5.1.5, or otherwise determined to be out of compliance with these regulations, shall be

subject to a Department enforcement inspection and subject to the fees specified in Part

12.

2.5.2

Procedures for Certification Evaluations by Qualified Inspectors.

2.5.2.1 Each qualified inspector who performs a certification evaluation of a radiation machine

and facility evaluation shall use procedures that are sufficient to determine compliance

with these regulations.

2.5.2.2 If a radiation machine fails to meet any requirement specified by these regulations,

including manufacturer's required specifications, the qualified inspector shall notify the

owner (registrant) or operator immediately and shall notify the department within three

days after the determination.

2.5.2.3 If the radiation machine is determined to be unsafe (as provided in Part 6 and described

in Appendix 6D), the qualified inspector shall affix to such radiation machine system, in a

location clearly visible to the operator and patient, if applicable, an “Unsafe for Use” label

authorized and issued by the Department, indicating, as applicable, that such machine is

not authorized for human, animal or other use.

2.5.2.4 Reporting and Labeling Procedures.

(1)

Each qualified inspector shall provide an accurate and complete Certification

Evaluation Report to the registrant and to the Department on Form R 59-1, “X ray

Machine Certification Evaluation Report,” in accordance with the instructions

contained in that form.

(a)

A clear and legible report may be substituted for Form R 59-1, provided

that it is in the same format and provides all of the information required

by Form R 59-1.

(b)

Violations of the regulations not related to the performance of the specific

radiation machine(s) shall be reported to the registrant and Department

using Form R 59-2, “X-ray Facility Compliance Evaluation Report,” in

accordance with the instructions contained in that form.

(c)

Report(s) required by 2.5.2.4(1) shall indicate full or partial compliance

and any specific violation of these regulations.

(d)

Report(s) required by 2.5.2.4(1) shall include recommendations for

corrective actions by the registrant (if applicable) to assist in achieving

full compliance or improving radiation safety and the quality of the

imaging process.

(e)

The Department shall be notified within three (3) business days of

radiation machine violations. Report(s) required by 2.5.2.4(1) that does

not indicate violations shall be received by the Department no later than

fifteen (15) calendar days after the inspection date, unless otherwise

authorized by the Department.

(2)

A certification label issued by the Department shall be affixed in a location clearly

visible to the machine operator and patient, if applicable, when it is determined

that the machine requirements of these regulations are fully met.

(a)

For a machine that was found to be in full compliance, the certification

label shall be affixed no later than fifteen (15) calendar days (unless

otherwise authorized by the Department) after the inspection date.

(b)

For a noncompliant machine, the certification label shall be affixed no

later than fifteen (15) calendar days (unless otherwise authorized by the

Department) after the date that full compliance was achieved.

(3)

Each qualified inspector shall ensure that the following documentation is

provided to the Department to confirm that each violation was corrected as

required by 2.6.3.1 and/or 2.6.4.1 within thirty (30) calendar days of the date of

inspection.

(a)

For a noncompliant machine for which full compliance has been

achieved, the completed documentation (on Form R 59-1 or equivalent)

shall be received by the Department no later than fifteen (15) calendar

days after the date that compliance was achieved.

(b)

For a noncompliant facility, the completed documentation (on Form R 59-

2 or equivalent) shall be received by the Department no later than fifteen

(15) calendar days after the date that full compliance was achieved.

(4)

Concealing, defacing or altering of Department-issued certification labels is

prohibited.

(5)

Repeated failure by a qualified inspector, to affix certification labels or to

complete certification evaluation reports in a timely manner as provided in 2.5.2.4

shall be subject to review and audit as provided in 2.9 and also subject to the non

routine inspection fee as provided in Part 12.

2.6

Facility Registrant Responsibilities.

2.6.1

The registrant shall allow only individuals who are adequately trained in radiation safety to

operate the machine and perform a radiographic examination. Training shall include instruction

on the specific x-ray system to be used and review of the applicable and critical requirements of

the operator manual.

2.6.1.1 The facility registrant shall evaluate and document the qualifications of each individual

permitted to operate any radiation machine at the facility.

(1)

Each operator shall meet all radiation safety training and experience

requirements of the respective State of Colorado professional licensure board, as

applicable, and any applicable requirements of this Part 2.

(2)

The registrant shall maintain a list of all operators of any radiation machine used

by the facility registrant.

(a)

For fluoroscopy equipment used in examination of a living human, a list

of operators and individuals providing supervision of operators shall be

maintained.

(b)

The list of all operators and supervisors shall be updated at least

annually as part of the radiation safety program required by Part 4,

Section 4.5.

(3)

Records of evaluations shall:

(a)

Include current certifications and qualifications;

(b)

Be updated annually by the facility; and

(c)

Be produced for examination upon request during any inspection

conducted under the requirements of these regulations.

2.6.1.2 A physician, chiropractor, dentist, podiatrist, or veterinarian who meets the applicable

requirements of Part 6, Section 6.3.1.6(1) and these regulations, is considered to have

demonstrated adequate training in radiation safety and the safe and effective use of the

radiation machine (consistent with 2.6.1.5) and may operate radiation machines as part

of a medical, chiropractic, dental, podiatric or veterinary practice, respectively.

2.6.1.3 For a radiologist assistant “adequately trained” shall mean that the individual is qualified

as provided in Appendix 2G.

2.6.1.4 For any radiographic x-ray system used on a living human (consistent with 2.6.1.2,

2.6.1.3 and 2.6.1.5 through 2.6.1.14), “adequately trained” shall mean that the individual

meets the requirements of Appendix 2D.

(1)

Limited-scope x-ray machine operator approval is limited to imaging procedures

for x-ray examination of the skull, chest, hip/pelvis and spine/sacrum, upper

extremities and lower extremities, and abdomen.

(2)

A limited-scope x-ray machine operator shall not perform radiologic procedures

involving the administration or utilization of contrast media, bone densitometry,

fluoroscopic, mammography, computed tomography, or radiation therapy

procedures.

2.6.1.5 For fluoroscopy equipment used in examination of a living human, “adequately trained”

shall mean that, in addition to meeting all applicable requirements in 2.4.5.5, 2.6.1.1

through 2.6.1.4, and Appendix 2O:

(1)

Each individual who either supervises a fluoroscopy procedure or operates a

fluoroscopy imaging system shall have adequate training in its safe operation.

This training shall be documented and include the following:

(a)

Basic properties of radiation;

(b)

Biological effects of x-ray;

(c)

Principles and safe operation of the specific fluoroscopic x-ray system(s)

to be used;

(d)

Dose management including dose reduction techniques, monitoring, and

recording;

(e)

Applicable requirements of these regulations.

After January 1, 2022, the training required by 2.6.1.5 shall also include:

(f)

Radiation protection methods for patients and staff;

(g)

Units of measurement and dose, including DAP (dose-area product)

values and air kerma;

(h)

Factors affecting fluoroscopic outputs;

(i)

High level control options; and

(j)

Fluoroscopic and fluorographic (radiation) outputs of each mode of

operation on the system(s) to be used clinically.

2.6.1.6 For mammography equipment used in radiography of the human breast, “adequately

trained” shall mean that the individual operator meets the requirements of 2.4.5.4(1) and

2.4.5.4(2).

(1)

Registered provisional mammographers may operate machines and perform

radiographic examinations under supervision while in-training as specified in

Appendix 2M.

2.6.1.7 For any computed tomography (CT) system used on a living human (excluding

Volumetric Dental Imaging Systems, CBCT systems, and systems used for digital breast

tomosynthesis) “adequately trained” shall mean that the individual operator meets the

requirements of Appendix 2E.

2.6.1.8 For any bone densitometry equipment used in examination of a living human,

“adequately trained” shall mean that the individual operator meets the requirements of

Appendix 2F.

2.6.1.9 For radiographic equipment used in the practice of medicine, “adequately trained” shall

mean that the individual operator meets all applicable requirements of the Colorado

medical board.

2.6.1.10

For radiographic equipment used in chiropractic, “adequately trained” shall mean

that the individual operator meets all applicable requirements of the Colorado Board of

Chiropractic Examiners and Rule 19 of 3 CCR 707-1.

2.6.1.11

For radiographic equipment used in dentistry, including Volumetric Dental

Imaging Systems, “adequately trained” shall mean that the individual operator meets all

applicable requirements of the Colorado Dental Board and Rule X of 3 CCR 709-1.

2.6.1.12

For radiographic equipment used in podiatry, “adequately trained” shall mean

that the individual operator meets all applicable requirements of the Colorado Podiatry

Board and 3 CCR 712-1.

2.6.1.13

For radiographic equipment used in veterinary medicine, “adequately trained”

shall mean that the individual operator meets all applicable requirements of the Colorado

Board of Veterinary Medicine and 4 CCR 727-1.

2.6.1.14

An individual, enrolled in an ARRT-recognized program or graduated from such a

program, may operate radiation machines so long as the individual works under the direct

supervision of a radiologic technologist or other qualified trainer and has documentation

of having completed education and experience equal to that specified in the program.

(1)

A graduate from an ARRT-recognized program is granted ninety (90) calendar

days from the date of graduation to schedule, take and pass the ARRT radiologic

technology registry examination.

(2)

During the 90-day period allowed by 2.6.1.14(1), the graduate is considered to

satisfy Appendix 2D requirements.

(3)

A student or graduate who fails to pass the registry examination has not met the

requirements of Appendix 2D and shall not operate any radiation machine

system on a living human unless otherwise authorized by the Department.

2.6.1.15

For radiation machines used in non-healing-arts applications, “adequately

trained” shall mean that the individual operator meets the requirements of Appendix 2N.

(1)

For industrial radiography, the requirements in Part 5 apply, as stated in 2N.1.

(2)

The requirements of 2N.2 apply to all non-healing-arts applications subject to

Part 8 (including but not limited to analytical, forensic, morgue, and homeland

security uses) but not subject to Part 5.

2.6.1.16

For assembly, installation and repair of radiation machines, “adequately trained”

shall mean that the individual service technician meets the requirements of Appendix 2H.

2.6.1.17

Department recognition of training as adequate pursuant to 2.6.1.3 through

2.6.1.16 shall pertain only to the areas of training and experience specifically identified in

these regulations.

2.6.1.18

The Department may, upon application or upon its own initiative, accept as being

adequate:

(1)

Documented combinations of radiation safety training and experience; or

(2)

Equivalent approval by another state or agency.

2.6.2

The facility registrant shall ensure that all required certification and compliance evaluations are

performed as required by 2.5.2 in accordance with the instructions that accompany Form R 59-1,

“X-ray Machine Certification Evaluation Report” and Form R 59-2, “X-ray Facility Compliance

Evaluation Report.”

2.6.2.1 Upon receipt of a Form R 59-1 signed by a registered qualified inspector, the facility shall

complete the certification evaluation process with that qualified inspector unless

department approval is granted or required to have the certification evaluation done by a

different qualified inspector.

2.6.3

For each radiation machine finding of noncompliance (Form R 59-1), the facility registrant shall:

2.6.3.1 Correct any failure of a radiation machine or imaging system to meet the requirements of

these regulations or manufacturer’s required specifications, within thirty (30) calendar

days or as otherwise specified by the Department, in particular as identified on Form R

59-1, “X ray Machine Certification Evaluation Report.”

2.6.3.2 Not use a radiation machine that has been determined to be unsafe for use, as

determined by the criteria in Part 6, Appendix 6D, until subsequent certification by a

Department-approved qualified inspector or the Department.

2.6.3.3 Permit only a person who has provided evidence of current registration with the

Department in accordance with 2.4.2 to provide radiation machine servicing or services.

2.6.3.4 Notify the qualified inspector who issued the Certification Evaluation Report when the

radiation machine violations have been corrected.

(1)

A copy of the Certification Evaluation Report, Form R 59-1, with the service

repair certification signed and dated by the person providing service, shall be

provided to the qualified inspector who initiated the certification evaluation..

(2)

A copy of any service report shall be provided to the qualified inspector upon

request as evidence of completed corrective action.

2.6.3.5 Retain documentation that each indicated violation has been corrected to bring the

machine into compliance in accordance with Section 2.6.6.

2.6.4

For each finding of facility noncompliance (Form R 59-2), the registrant shall:

2.6.4.1 Correct any violation within thirty (30) calendar days of each finding of facility

noncompliance (Form R 59-2) or as otherwise specified by the Department.

2.6.4.2 Provide documentation to the Department to confirm that each indicated violation has

been corrected to bring the facility into compliance.

(1)

For any item identified for correction on Form R 59-2, “X-ray Facility Compliance

Evaluation Report”, provide a copy of the Form R 59-2 with the “Registrant’s

Certification of Correction” section signed and dated by the registrant or

registrant’s agent.

2.6.5

Except as otherwise specified in Part 6 and Part 24 of these regulations, each registrant shall

follow all applicable manufacturer’s recommended equipment maintenance and quality assurance

procedures.

2.6.6

Record Retention and Reports.

2.6.6.1 The registrant shall maintain each diagnostic image in a medical record for each patient

as specified by the applicable State of Colorado professional licensure board; absent an

applicable board specification, record retention shall be for a period not less than ten (10)

years or any period of minority or incompetency.

2.6.6.2 The registrant shall maintain for the duration of the registration, records of each shielding

design, and each radiation survey required by 6.9.4.1, performed for the facility.

(1)

Upon any transfer of ownership, such shielding design(s) and survey records

shall also be transferred to the new owner.

2.6.6.3 The registrant shall maintain for the duration of the registration, until a machine is retired

from service, the operator and service manual(s) provided by the manufacturer, if

available.

(1)

If the operator manual is not obtainable from the manufacturer, such a manual of

written operating procedures shall be developed and maintained by the

registrant, including:

(a)

A description, including purpose and function, of each control panel

knob, button, and meter;

(b)

Techniques for collimation and centering of the beam to the image

receptor;

(c)

The function of all locks and detents; and

(d)

Emergency shutdown instructions.

2.6.6.4 The registrant shall maintain for inspection for a period of three (3) years for each x-ray

imaging or image processing system (six years for a facility or machine inspected only

every three years) records of:

(1)

Operator certifications;

(2)

Operator training;

(3)

Service and repair reports;

(4)

Radiation machine disposition

(5)

Radiation machine inspection certification evaluation reports;

(6)

Facility compliance evaluation reports; and

(7)

Notices of violation.

2.6.7

For each certification label issued by a qualified inspector, facility registrants shall pay the label

fee required by Part 12, Category 25. Facility registrants who fail to pay the label fee may be

subject to review, audit, and non-routine inspection fees in accordance with Section 2.9.

2.7

Service Company Registrant Responsibilities.

2.7.1

No person shall certify or declare that a radiation machine or component is ready for its intended

use, until:

2.7.1.1 The shielding design has been completed as required by 6.3.2, as documented by a

comment on Form FDA 2579 (for machines used in the healing arts) or a signed and

dated notification to the Department; and

2.7.1.2 The machine or component meets the manufacturer specifications and the requirements

of these regulations; and

2.7.1.3 The registrant has been provided, by the vendor, assembler or services and servicing

personnel, the following:

(1)

All guidance documents, including instruction manuals, manufacturer

specifications and information notices, that are applicable to each newly installed

radiation machine system or component; and

(2)

The Colorado x-ray facility registrant’s responsibilities list, as posted on the

department website.

2.7.2

Any person who sells, leases, transfers, lends, assembles, installs, trades out or repairs any

radiation machine, or component, which affects radiation output or technique setting in this State

shall notify the Department in writing within fifteen (15) calendar days of each transaction subject

to this section with the following information:

2.7.2.1 The full name and address of each person who has received the radiation machine or

component and the specific location within the facility; and

2.7.2.2 Specific details about the system or sub-system, including the manufacturer, model, and

serial number of each radiation machine or component transferred; and

2.7.2.3 The date of transfer, assembly, or installation of each radiation machine or component;

and

2.7.2.4 A completed Form FDA 2579 or a signed and dated affirmation that all instruction

manuals, written instructions and regulations applicable to the newly installed radiation

machine system or components have been delivered to the registrant.

2.7.3

A report of assembly (Form FDA 2579 or equivalent) shall be submitted to the Department within

fifteen (15) calendar days following completion of the assembly or installation.

2.7.3.1 The assembly or installation is considered completed when the unit has properly been

made operational and is ready for its intended use.

2.7.3.2 Form FDA 2579 or an equivalent report suffices in lieu of any reports required in 2.7.2.

2.7.4

As required by the Department on a Certification Evaluation Report, Form R 59-1, a service

company technician who performs a radiation machine repair shall:

2.7.4.1 Sign the service repair certification section of the Certification Evaluation Report, Form R

59-1 issued by the qualified inspector who performed the evaluation; and

2.7.4.2 Provide a written detailed description of the service to the registered facility within one (1)

business day.

2.7.5

A service technician who performs any activity that could potentially affect the radiation machine

output, cause a change to the clinical technique settings of the radiation machine, or affect image

quality shall provide a written detailed description of all service to the registered facility within one

business day of the service.

2.7.6

Any person who disables a radiation machine in order to meet the requirements of 2.3.4 shall be

registered with the Department as a Service Company.

RECIPROCITY

2.8

Out-of-State Radiation Machines.

2.8.1

Subject to these regulations, any person who desires to bring radiation machines into this state

for temporary use is hereby granted authorization to conduct activities using these machines for a

period not to exceed a total of 180 days in any calendar year, provided that:

2.8.1.1 The out-of-state registration, and/or other documents authorizing the use of radiation

machines issued by the agency having jurisdiction where the out-of-state registrant

maintains an office for directing the registered activity and at which radiation safety

records are normally maintained, does not limit the activity authorized by such document

to specified installations or locations; and

2.8.1.2 The person proposing to bring such machines into Colorado shall give written notice to

the Department at least fifteen (15) calendar days before such machine is to be used in

the state, unless otherwise authorized by the Department as provided in 2.8.2. The notice

shall be made using the Department's “X-ray Reciprocity Request” Form R-200 and shall

include all information required by that form.

(1)

As part of this notice, the person requesting reciprocity shall certify that:

(a)

A copy of all applicable parts of these regulations shall be available at

each use location in State of Colorado;

(b)

Each machine has been evaluated and determined to be in compliance

with these, or equivalent, regulations; and

(c)

The operation of each radiation machine shall be in accordance with the

applicable requirements of these regulations.

(2)

In the case of a request to perform a healing arts screening program within the

State, submit a completed Form R-300, “Application for Registration – Healing

Arts Screening,” with the reciprocity request, including all of the information

required, pursuant to Part 6, Appendix 6F, by the form and any accompanying

instructions.

(3)

In the case of a request to perform mammography screening within the State, a

copy of the facility's mammography certificate issued by the FDA (21 CFR Part

900.11(a)) and applicable American College of Radiology credentials shall be

included with the reciprocity request.

(4)

The person requesting reciprocity shall also supply such other information as the

Department may request.

2.8.1.3 The out-of-state registrant complies with all applicable regulations of the Department; and

2.8.1.4 The out-of-state registrant shall at all times during work at any work location within the

State have available the pertinent documentation as required by these regulations,

including:

(1)

Pertinent registration documentation;

(2)

Written authorization from the Department for in-state activities;

(3)

Applicable sections of these regulations as certified pursuant to 2.8.1.2(1)(a);

(4)

Documentation that each radiation machine has been evaluated in accordance

with these regulations, or other state regulations which are equivalent; and that

(a)

The machines comply with the manufacturer's required specifications;

(b)

The evaluations are current, having been performed within one year prior

to entry into the State as required in 2.5; and

(5)

In the case of mammography-related functions, a copy of the mammography

certificate issued by the FDA, applicable American College of Radiology

credentials, quality control records, personnel records, and the most recent

medical physicist survey.

2.8.2

Based upon an application that includes documentation of why it is not possible or is an undue

hardship to provide fifteen (15) calendar days notice, the Department may:

2.8.2.1 Grant permission to proceed sooner; or

2.8.2.2 Waive the requirement for filing additional written notifications during the remainder of the

calendar year following the receipt of the initial notification from a person engaging in

activities pursuant to 2.8.1.

2.8.3

While in the State of Colorado, all radiation machines are subject to inspection and may require a

certification evaluation by a qualified inspector who is registered with the Department.

2.8.4

The out-of-state registrant shall notify the Department within one hour after arrival at the actual

work location within the State and shall notify the Department within one hour after any change of

work location within the State.

2.8.5

If multiple individuals work concurrently at more than one work location under an approval

granted pursuant to 2.8.1, each day worked per location shall be counted separately toward the

limit of 180 cumulative total days per calendar year.

2.8.6

The Department may revoke, limit, or qualify its approval for the use of radiation machines in the

State upon determining that the approval was based on false or misleading information submitted

to the Department or that such action is necessary in order to prevent undue hazard to public

health and safety or property.

2.8.7

Each person operating a radiation machine within the State under reciprocity in areas of exclusive

federal jurisdiction shall comply with the applicable federal requirements.

ENFORCEMENT

2.9

Department Review of Performance.

2.9.1

The Department as appropriate shall:

2.9.1.1 Notify the registrant or person operating a radiation machine, as appropriate, regarding

inadequate action on any item of violation;

2.9.1.2 Determine a schedule for correction of each violation and specifying a date by which

compliance must be achieved;

2.9.1.3 Confirm and verify by inspection a corrective action by a registrant or person operating a

radiation machine, as appropriate, to assure compliance with these regulations; and

2.9.1.4 Assess a non-routine inspection fee provided in Part 12, at the programmatic hourly rate,

for the inspection of a radiation machine system or facility, if:

(1)

The registrant or person operating a radiation machine, as appropriate, fails to

fulfill the requirements of these Regulations; or

(2)

Any item of violation has not been corrected in accordance with the compliance

schedule established in 2.9.1.2.

2.9.2

The Department shall periodically review and audit:

2.9.2.1 The compliance of any person registered under 2.4 with these Regulations;

2.9.2.2 The competency of each service technician in meeting standards and requirements for

adequate service company performance;

2.9.2.3 The performance of each qualified inspector, in particular:

(1)

Adequacy of inspections;

(2)

Competency in determining radiation machine system or facility compliance with

these regulations; and

(3)

Completeness and accuracy of findings on Form R 59-1 or R 59-2;

2.9.2.4 The performance of each qualified expert and/or registered medical physicist, in

particular:

(1)

Adequacy of shielding design reports; and

(2)

Competency in performing activities in accordance with these regulations.

2.9.3

The Department shall notify the registrant of any failure to meet a performance standard or

requirement of the regulations that is identified as a result of the review or audit.

2.9.4

The Department shall determine a schedule for actions required, specifying the date by which

adequacy or competency shall be demonstrated.

2.9.5

For any failure to demonstrate adequacy or competency in accordance with the compliance

schedule established in 2.9.4, the Department will assess a non-routine inspection fee at the

programmatic hourly rate for Department effort to enforce compliance with a performance

standard or requirement of the regulations.

2.9.6

The Department may deny, withdraw, limit or qualify its approval of any person to perform

activities upon determining that such action is necessary in order to prevent undue hazard to

health and safety, or for other reasonable cause.

2.9.7

A registrant that fails to comply with these regulations including 2.4.5 and 2.4.6 shall be subject to

revocation as provided in 2.10.

MODIFICATION AND REVOCATION OF REGISTRATION

2.10

The terms and conditions of all registrations/certificates shall be subject to amendment, revision,

or modification or the registration/certificate may be suspended or revoked by reason of

amendments to the Act, or by reason of rules, regulations, and orders issued by the Department.

PART 2, APPENDIX 2A: RADIATION MACHINE RADIATION SAFETY OFFICER (RSO) ADEQUATE

RADIATION SAFETY TRAINING AND EXPERIENCE

Each individual who performs the duties of a Radiation Safety Officer for a facility using radiation

machines shall meet the following education and experience requirements:

2A.1

For non-healing arts facilities (such as those governed by Part 8, “Radiation Safety Requirements

for Radiation Generating Machines Not Used in the Healing Arts”, and Part 9, “Radiation Safety

Requirements for Particle Accelerators Not Used in the Healing Arts”):

2A.1.1 Has current Department approval as a Qualified Expert, or

2A.1.2 Has current Department approval as a registered medical physicist, or

2A.1.2 Has satisfactorily completed a baccalaureate or higher degree in natural or physical

science, health physics, radiological sciences, nuclear medicine, nuclear engineering, or

2A.1.3 Has completed a structured educational program that included classroom training in the

responsibilities of an RSO, including but not limited to:

2A.1.3.1

Establishing and overseeing operating and safety procedures that

maintain radiation exposures as low as reasonably achievable (ALARA), and to

review them regularly to ensure that the procedures are current and conform with

these regulations;

2A.1.3.2

Ensuring that individual monitoring devices are properly used by

occupationally exposed personnel, that records are kept of the monitoring

results, and that timely notifications are made as required by Part 4;

2A.1.3.3

Investigating and reporting to the agency each known or suspected case

of radiation exposure to an individual or radiation level detected in excess of

limits established by these regulations and each theft or loss of source(s) of

radiation, determining the cause, and taking steps to prevent its recurrence;

2A.1.3.4

Having a thorough knowledge of management policies and

administrative procedures of the registrant and keeping management informed

on a periodic basis of the performance of the registrant's radiation protection

program, if applicable;

2A.1.3.5

Assuming control and having the authority to institute corrective actions

including shutdown of operations when necessary in emergency situations or

unsafe conditions;

2A.1.3.6

Maintaining records as required by these regulations; and

2A.1.3.7

Ensuring that personnel are adequately trained and complying with these

regulations, the conditions of the certificate of registration, and the operating and

safety procedures of the registrant; or

2A.2

For a healing arts facility not using fluoroscopy, computed tomography, or radiation therapy

machines, unless otherwise provided or prohibited by these regulations:

2A.2.1 Has department approval as a registered medical physicist; or

2A.2.2 Is a physician, chiropractor, dentist, podiatrist or veterinarian with a current active license

from the appropriate State of Colorado professional licensure board and is performing

RSO duties within their scope of practice; For dental facilities using a Volumetric Dental

Imaging System, a dentist with a current active license from the Colorado Board of Dental

Examiners may perform the duties of a Radiation Safety Officer;

or

2A.2.3 Meets the applicable operator requirements of 2.6.1.3 through 2.6.1.13; and has

completed a structured educational program that includes ionizing radiation safety; or

2A.3

For a healing arts facility using fluoroscopic or computed tomography machines, unless otherwise

provided or prohibited by these regulations:

2A.3.1 Has department approval as a registered medical physicist; or

2A.3.2 Is a physician or veterinarian who has a current active license from the appropriate State

of Colorado professional licensure board; or

2A.4

For a healing arts facility using radiation therapy machines, unless otherwise provided or

prohibited by these regulations:

2A.4.1 Has department approval as a radiation therapy registered medical physicist, or

2A.4.2 Is a physician or veterinarian who has a current active license from the appropriate State

of Colorado professional licensure board and is performing RSO duties within their scope

of practice, or

2A.5

Has prior Department approval pursuant to another part of these regulations as an authorized

RSO

PART 2, APPENDIX 2B: QE(R) AND QE(T) ADEQUATE TRAINING AND EXPERIENCE

2B1

Each Qualified Expert who designs or evaluates shielding for a radiation machine used in the

healing arts as regulated by Part 6, but not used in radiation therapy, and is designated as a

QE(R), or each Qualified Expert who designs or evaluates shielding for a radiation machine used

in radiation therapy, and is designated as a QE(T) shall:

2B.1.1 Have current certification in health physics or a subfield of medical physics by:

2B.1.1.1

The American Board of Medical Physics; or

2B.1.1.2

The American Board of Health Physics; or

2B.1.1.3

The Canadian College of Medical Physics; or

2B.1.1.4

The American Board of Radiology in a radiological physics category; or

2B.1.1.5

American Board of Nuclear Medicine Science; or

2B.1.2 Has current certification in an equivalent specialty board recognized by the Department,

and;

2B.1.2.1

Has provided written documentation that the individual:

(1)

Holds a master or doctorate degree from an accredited college or

university in physics, biophysics, radiological physics, health physics, or

medical physics; and

(2)

Has satisfactorily completed 2 years of training and work experience

acceptable to the Department that includes one year of documented, full-

time training in the appropriate field under the supervision of a qualified

expert.

PART 2, APPENDIX 2C: QE(S) – ADEQUATE TRAINING AND EXPERIENCE

2C.1

Each Qualified Expert who designs or evaluates shielding for a radiation machine not used in the

healing arts, designated as QE(S), shall:

2C.1

Have current certification in health physics or a subfield of medical physics by:

2C.1.1 The American Board of Medical Physics; or

2C.1.2 The American Board of Health Physics; or

2C.1.3 The Canadian College of Medical Physics; or

2C.1.4 The American Board of Radiology in a radiological physics category; or

2C.1.5 American Board of Nuclear Medicine Science; or

2C.2

Has current certification in an equivalent specialty board recognized by the Department,

and;

2C.2.1 Has provided written documentation that the individual:

2C.2.1.1

Holds a master or doctorate degree from an accredited college

or university in physics, biophysics, radiological physics, health physics,

or medical physics; and

2C.2.1.2

Has satisfactorily completed 2 years of training and work

experience acceptable to the Department that includes one year of

documented, full-time training in the appropriate field under the

supervision of a qualified expert;

PART 2, APPENDIX 2D: X-RAY SYSTEM OPERATOR ADEQUATE RADIATION SAFETY TRAINING

AND EXPERIENCE, INCLUDING LIMITED SCOPE X RAY MACHINE OPERATOR (LSO)

Each operator of a radiation machine used for healing arts purposes on living humans other than in

dentistry, chiropractic or podiatry, shall meet the following education and experience requirements:

2D.1

Is certified or registered by:

2D.1.1 The American Registry of Radiologic Technologists as a Radiologic Technologist; or

2D.1.2 A specialty board determined by the department to have substantially equivalent

requirements for certification as the American Registry of Radiologic Technologists,

Or

2D.2

Is certified by the Department as a State of Colorado-registered limited scope operator, to

conduct only those radiographic examinations specified in Section 2.6.1.4 and having

satisfactorily completed:

2D.2.1 At least 80 hours of didactic training providing the minimum hours of instruction in the

specific subjects listed in 2D.2.1.1 through 2D.2.1.6:

2D.2.1.1

Basic X-Ray Physics—20 hours

(1)

Structure of matter and the atom

(2)

General description of production of x-rays

(3)

X-ray emission, quantity and quality

(4)

Function of filtration and effects it has on x-ray beam collimation

(5)

Types of function of beam limiting devices

(6)

Design, features and functions of x-ray tubes

(7)

Circuitry of the x-ray machine

2D.2.1.2

Radiobiology—3 hours

(1)

Effects of ionizing radiation on the human body

(2)

Molecular and cellular radiobiology

(3)

Factors that cause somatic and genetic damage

2D.2.1.3

Radiation Protection—6 hours

(1)

ALARA

(2)

Shielding materials

(3)

Radiation quantity and units of measurement

(4)

Basic interactions of x-rays with matter

(5)

Primary and secondary scatter

(6)

Importance of time, distance, shielding

(7)

Maximum permissible doses: occupational and public

(8)

Patient protection

2D.2.1.4.

Principles of Exposure—15 hours

(1)

Factors that control and influence radiographic quality

(2)

Properties of x-rays

(3)

Size distortion

(4)

Shape distortion

(5)

kVp, mAs, time

(6)

AEC and manual

(7)

Grids

(8)

Collimation

(9)

Intensifying screens

(10)

X-ray films and holders

(11)

Artifacts

(12)

Inverse square law

2D.2.1.5

Procedures and Processing—4 hours

(1)

Film storage and handling

(2)

Manual, automatic processing film processing and troubleshooting

(3)

Computed Radiography (CR)

(4)

Digital Radiography (DR)

(5)

PACs

(6)

Quality assurance / quality control

2D.2.1.6

Anatomy and Positioning—32 hours

(1)

Chest—4 hours

(2)

Extremity—12 hours

(3)

Spine—8 hours

(4)

Skull—8 hours;

and

2D.2.2 At least 480 hours of clinical training during which time the individual may perform x-ray

examinations under direct supervision of a qualified trainer, including:

2D.2.2.1

At least 320 hours experiential training at a clinic; and

2D.2.2.2

No more than 160 hours of laboratory training (exclusive of the didactic

hours required by 2D.2.1.1 through 2D.2.1.6);

and

2D.2.3 Performance of the following imaging procedures (at least 84 examinations in total, with

record of each examination kept on file):

2D.2.3.1

Ribs—4 examinations;

2D.2.3.2

Hand—4 examinations;

2D.2.3.3

Wrist—4 examinations;

2D.2.3.4

Forearm—4 examinations;

2D.2.3.5

Elbow—4 examinations;

2D.2.3.6

Humerus—4 examinations;

2D.2.3.7

Shoulder—4 examinations;

2D.2.3.8

Clavicle—4 examinations;

2D.2.3.9

Femur—4 examinations;

2D.2.3.10

Tibia – Fibula—4 examinations;

2D.2.3.11

Ankle—4 examinations;

2D.2.3.12

Foot—4 examinations;

2D.2.3.13

Sinuses—4 examinations;

2D.2.3.14

Skull—4 examinations;

2D.2.3.15

Facial Bones—4 examinations;

2D.2.3.16

C-Spine—4 examinations;

2D.2.3.17

Thoracic Spine—4 examinations;

2D.2.3.18

Lumbar Spine—4 examinations;

2D.2.3.19

Chest—4 examinations;

2D.2.3.20

Hip / Pelvis—4 examinations;

2D.2.3.21

Abdomen—4 examinations.

and

2D.2.4 A passing score on the American Registry of Radiologic Technologists (ARRT)

examination for the Limited Scope of Practice in Radiography. A passing score is:

2D.2.4.1

A score of at least 75% correct on the Core Module, and

2D.2.4.2

An average score of at least 75% correct on the Radiographic

Procedures Modules for Chest, Extremities, Skull/Sinuses, and Spine.

2D.2.5 And, has maintained a minimum of twenty-four (24) hours of continuing education every

two years in the areas of radiology, radiation safety, radiography and similar fields. This

education shall:

2D.2.5.1

Conform to guidelines equivalent to the most current revision of the

ARRT Continuing Education Requirements for Renewal of Registration;

PART 2, APPENDIX 2E: COMPUTED TOMOGRAPHY (CT) ADEQUATE RADIATION SAFETY

TRAINING AND EXPERIENCE

Each operator of a computed tomography system on living humans shall hold a current, valid registry in

Radiography, Nuclear Medicine, or Radiation Therapy issued by ARRT, NMTCB, or, where the operator

has obtained written approval from the Department, another nationally recognized registry organization

not listed herein, shall meet the following requirements:

2E.1

Certification:

2E.1.1 For general imaging computed tomography procedures, each operator is certified;

2E.1.1.1

By the ARRT in computed tomography, R.T.(CT); or

2E.1.1.2

By the Nuclear Medicine Technology Certification Board (NMTCB) in

computed tomography, NMTCB(CT);

Or

2E.1.2 For nuclear medicine (hybrid or fusion imaging) computed tomography procedures such

as PET-CT or SPECT-CT, is certified;

2E.1.2.1

by the ARRT in nuclear medicine as R.T.(N); or

2E.1.2.2

by the NMTCB as CNMT; or

2E.1.2.3

by the NMTCB as NMAA; or

2E.1.2.4

in accordance with 2E.1.1.

Or

2E.1.3 For simulation or localization computed tomography procedures associated with radiation

therapy, is certified;

2E.1.3.1

by the ARRT in Radiation Therapy, R.T.(T); or

2E.1.3.2

in accordance with 2E.1.1.

Or

2E.1.4 Is certified by a specialty board determined by the department to have substantially

equivalent requirements for certification in computed tomography as the American

Registry of Radiologic Technologists.

Or

2E.2

Prior to August 1, 2017, and holds a current, valid registry as an R.T.(R) and was also registered

with the Department as a Computed Tomography Operator.

PART 2, APPENDIX 2F: BONE DENSITOMETRY (BD) ADEQUATE RADIATION SAFETY TRAINING

AND EXPERIENCE

Each operator of a dual-energy x-ray absorptiometry system used on a living human shall meet the

following education and experience requirements:

2F.1

Is certified or registered:

2F.1.1 As R.T.(R), R.T.(M), R.T.(N), R.T.(T), or CNMT; or

2F.1.2 By The International Society for Clinical Densitometry (ISCD), combined with or including

the didactic radiation safety training in 2F.2.1.1, 2F.2.1.2 and 2F.2.1.3; or

2F.1.3 By A specialty board determined by the department to have substantially equivalent

requirements for certification;

Or

2F.2

Is accepted by the Department as having satisfactorily completed:

2F.2.1 At least 30 hours of didactic training recognized by the Department that provided the

minimum hours of instruction (as part of, or in addition to, specialty certificate and

equipment operation training) in the specific subjects listed in 2F.2.1.1 through 2F.2.1.9:

RADIATION SAFETY:

2F.2.1.1

Basic X-Ray Physics—2 hours

(1)

Structure of matter and the atom

(2)

General description of production of x-rays

(3)

X-ray emission, quantity and quality

(4)

Function of filtration and effects it has on x-ray beam collimation

(5)

Types of function of beam limiting devices

(6)

Design, features and functions of x-ray tubes

(7)

Circuitry of the x-ray machine

2F.2.1.2

Radiobiology—2 hours

(1)

Effects of ionizing radiation to the human body

(2)

Molecular and cellular radiobiology

(3)

Factors that cause somatic and genetic damage

2F.2.1.3

Radiation Protection—5 hours

(1)

ALARA

(2)

Shielding materials

(3)

Radiation quantity and units of measurement

(4)

Basic interactions of x-ray with matter

(5)

Primary and secondary scatter

(6)

Importance of time, distance, shielding

(7)

Maximum permissible dose: occupational and public

(8)

Patient protection

(a)

Patient instruction

(b)

Comparison levels of radiation

(i)

Natural background radiation

(ii)

Central DXA

(iii)

Peripheral DXA

2F.2.1.4

Basic Concepts—8 hours

(1)

Osteoporosis

(a)

World Health Organization definition and diagnostic criteria

(b)

Primary vs. secondary

(c)

Type I (postmenopausal) vs. Type II (senile)

(d)

Risk factors

(i)

Controllable (smoking, calcium intake, estrogen,

medications)

(ii)

Uncontrollable (heredity, race, gender, age, medical

conditions)

(2)

Bone physiology

(a)

Functions of bone

(i)

Structural support and protection

(ii)

Storage of essential minerals

(b)

Types of bone

(i)

Cortical

(ii)

Trabecular

(c)

Bone remodeling cycle

(i)

Resorption / formation

(ii)

Osteoblasts/osteoclasts

(d)

Bone health

(i)

Nutrition

(ii)

Exercise

(3)

BMD testing methods (anatomical sites scanned, key advantages and

disadvantages)

(a)

Dual-energy X-ray Absorptiometry (DXA)

(b)

Single X-ray Absorptiometry (SXA)

(c)

Quantitative Ultrasound (QUS)

(d)

Radiographic Absorptiometry (RA)

(4)

Measuring BMD

(a)

Basic statistical concepts

(i)

Mean

(ii)

Standard deviation

(iii)

Coefficient of variation

(b)

Reporting patient results

(i)

BMD formula

(ii)

Z–score

(iii)

T–score

2F.2.1.5

Equipment Operation & Quality Control—6 hours

(1)

Computer console

(a)

Major components

(b)

File management

(2)

Fundamentals of x-ray energy production

(a)

Properties of x-ray beam: quality (kVp), quantity (mA),

duration/time (s)

(b)

Filters and collimators

(c)

X-ray energy production: single; dual

(3)

Types of DXA systems

(a)

Pencil beam systems

(b)

Fan beam systems

(c)

Cone beam systems

(4)

Quality control

(a)

Equipment safety (electrical, pinch points, emergency stop)

(b)

Use of phantoms and/or calibration

(c)

Troubleshooting

(i)

Shift or drift

(ii)

Pass / fail

(d)

Record maintenance

(5)

Determining quality in BMD

(a)

Precision (definition)

(b)

Accuracy (definition)

(c)

Factors affecting accuracy and precision

(i)

Scanner

(ii)

Operator

(iii)

Patient

2F.2.1.6

DXA Scanning of Finger and Heel (OS CALCIS)—1 hour

(1)

Anatomy

(a)

Regions of interest

(b)

Bony landmarks

(c)

Radiographic appearance

(2)

Scan acquisition

(a)

Patient instructions

(b)

Patient positioning

(c)

Evaluating pre-set scan parameters

(3)

Scan analysis: BMD, T score, Z score

(4)

Common problems

(a)

Nonremovable artifacts

(b)

Fractures or pathology

2F.2.1.7

DXA Scanning of Forearm—2 hours

(1)

Anatomy

(a)

Regions of interest

(b)

Bony landmarks

(c)

Radiographic appearance

(d)

Adjacent structures

(2)

Scan acquisition

(a)

Patient instructions

(b)

Patient positioning

(c)

Evaluating pre-set scan parameters

(3)

Scan analysis

(a)

Accurate ROI placement

(b)

BMC, area, and BMD

(c)

T-score, Z-score

(4)

Common problems

(a)

Poor bone edge detection

(b)

Nonremovable artifacts

(c)

Variant anatomy

(d)

Fractures or pathology

(5)

Follow-up scans

(a)

Unit of comparison: BMD, T-score

(b)

Reproduce baseline study

2F.2.1.8

DXA Scanning of Lumbar Spine—2 hours

(1)

Anatomy

(a)

Regions of interest

(b)

Bony landmarks

(c)

Radiographic appearance

(d)

Adjacent structures

(2)

Scan acquisition

(a)

Patient instructions

(b)

Patient positioning

(c)

Evaluating pre-set scan parameters

(3)

Scan analysis

(a)

Accurate ROI placement

(b)

BMC, area, and BMD

(c)

T-score, Z-score

(4)

Common problems

(a)

Poor bone edge detection

(b)

Nonremovable artifacts

(c)

Variant anatomy

(d)

Fractures or pathology

(5)

Follow-up scans

(a)

Unit of comparison: BMD, T score

(b)

Reproduce baseline study

2F.2.1.9

DXA Scanning of Proximal Femur—2 hours

(1)

Anatomy

(a)

Regions of interest

(b)

Bony landmarks

(c)

Radiographic appearance

(d)

Adjacent structures

(2)

Scan acquisition

(a)

Patient instructions

(b)

Patient positioning

(c)

Evaluating pre-set scan parameters

(3)

Scan analysis

(a)

Accurate ROI placement

(b)

BMC, area, and BMD

(c)

T-score, Z-score

(4)

Common problems

(a)

Poor bone edge detection

(b)

Nonremovable artifacts

(c)

Variant anatomy

(d)

Fractures or pathology

(5)

Follow-up scans

(a)

Unit of comparison: BMD, T-score

(b)

Reproduce baseline study;

and

2F.2.2 At least 480 hours of clinical training during which time DXA examinations are performed

only under direct supervision of a Colorado qualified bone densitometry equipment

operator or other qualified trainer:

2F.2.3 Performance of the following imaging procedures (at least 30 examinations in total, with

record of each examination kept on file):

2F.2.3.1

DXA scanning of the forearm—10 examinations;

2F.2.3.2

DXA scanning of the lumbar spine—10 examinations;

2F.2.3.3

DXA scanning of the proximal femur—10 examinations;

and

2F.2.4 A passing score on the American Registry of Radiologic Technologists (ARRT) Bone

Densitometry Equipment Operator Examination.

and

2F.2.5 Has maintained a minimum of eighteen (18) hours continuing education every three

years, documented by certificate(s) or other attestation(s) of satisfactory completion.

PART 2, APPENDIX 2G: RADIOLOGIST ASSISTANT (RA) ADEQUATE RADIATION SAFETY

TRAINING AND EXPERIENCE

Any person who acts as a Radiologist Assistant or Radiologist Practitioner Assistant shall be an individual

who is 18 years of age and has provided written documentation as evidence of:

2G.1

Current certification as both R.T.(R) and a

2G.1.1 Registered Radiologist Assistant (R.R.A.(ARRT)); or

2G.1.2 Radiology Practitioner Assistant (RPA) prior to January 1, 2008;

And

2G.2

Having:

2G.2.1 Met the specific qualifications in education recognized by the ARRT, ASRT, ACR, or

equivalent nationally recognized entity; and

2G.2.2 Been trained and worked under the direction of a radiologist.

PART 2, APPENDIX 2H: ADEQUATE EDUCATION AND TRAINING TO PERFORM RADIATION

MACHINE ASSEMBLY, INSTALLATION AND/OR REPAIR

Any individual who performs radiation machine assembly, installation or service shall meet the following

educational and experience requirements:

2H.1

Completion of a structured educational program that includes training in radiation machine safety,

assembly, installation and service, including, but not limited to:

2H.1.1 A baccalaureate degree in electrical engineering with specialized training in radiation

producing devices; or

2H.1.2 A one-year associate degree in biomedical equipment repair; or

2H.1.3 Equivalent manufacturer, military or other technical school training;

and

2H.2

For each service category requested:

2H.2.1 At least six (6) months of supervised, documented training on assembly, installation and

service of the applicable radiation machine.

PART 2, APPENDIX 2I: QUALIFIED INSPECTOR (QI) ADEQUATE RADIATION SAFETY TRAINING

AND EXPERIENCE

As provided by 2.4.4, approval of registration as a qualified inspector shall be given to an individual who:

2I.1

Has provided written documentation that the individual:

2I.1.1

Holds an associates or higher degree in physics, applied physics, biophysics, biophysical

engineering, medical physics, radiologic physics, health physics, or equivalent, from an

accredited college or university; and

2I.1.2

Has experience with each category of radiation machine for which approval is requested,

including, but not limited to:

2I.1.2.1 Measuring ionizing radiation;

2I.1.2.2 Evaluating radiation machines and components;

2I.1.2.3 Evaluating facility radiation safety programs;

2I.1.2.4 Image processing;

2I.1.2.5 The applicable requirements of these regulations; and

2I.1.2.6 Digital imaging and image processing system software and hardware, when

applicable and available; and

2I.1.3

The experience duration required by 2I.1.2 will be in combination with the education

requirements from 2I.1.1 as follows:

2I.1.3.1 One year with a masters or doctorate degree; or

2I.1.3.2 Two years with an arts or sciences baccalaureate degree; or

2I.1.3.3 Three years with an Associate Degree; and

2I.1.4

The experience required by 2I.1.2 shall be acquired:

2I.1.4.1 Within the 7 years preceding the date of application; or

2I.1.4.2 Through documented subsequent continuing education and experience within 7

years preceding the date of the application.

2I.2

Approval for registration as a Provisional Qualified Inspector shall be given to an individual who

has met the requirements of 2I.1.1 and has:

2I.2.1

Provided training program documentation describing how the Provisional Qualified

Inspector will meet the requirements of 2I.1.2, 2I.1.3 and 2I.1.4. The training program

documentation shall:

2I.2.1.1 Require direct supervision of the Provisional Qualified Inspector during the

evaluation of at least the initial five (5) radiation machines for each category

inspected by the Provisional Qualified Inspector; and

2I.2.1.2 Identification of the Qualified Inspector(s) who will provide the Provisional

Qualified Inspector with general supervision until the requirements of 2I.1.2,

2I.1.3 and 2I.1.4 are met.

2I.2.2

At the time when the requirements of 2I.1.2, 2I.1.3 and 2I.1.4 are met, the Provisional

Qualified Inspectors must apply for registration as a Qualified Inspector.

2I.2.3

Registered Qualified Inspectors may apply for approval as a Provisional Qualified

Inspector for new categories that are being requested.

2I.3

In addition to the requirements of 2I.1, approval for registration in the Registered Medical

Physicist category shall be give to an individual who:

2I.3.1

Is certified by:

2I.3.1.1 The American Board of Radiology in Radiological Physics, Diagnostic Medical

Physics, Diagnostic Radiological Physics, or Medical Nuclear Physics; or

2I.3.1.2 The American Board of Medical Physics in Diagnostic Radiological Physics; or

2I.3.1.3 The Canadian College of Physicists in Medicine in Radiological Physics; or

2I.3.1.4 The American Board of Radiology in Nuclear Medical Physics, American Board

of Medical Physics in Nuclear Medicines Physics, or American Board of Science

in Nuclear Medicine in Nuclear Medicine Physics and Instrumentation and who

shall be limited to RMP certification evaluation activities associated with CT or

hybrid (positron emission tomography/CT and single-photon emission

computerized tomography/CT systems only; or

2I.3.1.5 An equivalent specialty board or certification approved in writing by the

department.

2I.3.2

Approval for registration as a Provisional Registered Medical Physicist shall be given to

an individual who is in the process of certification to meet 2I.3.1 and has:

2I.3.2.1 Passed the initial testing requirements of the respective certifying organization;

and

2I.3.2.2 Provided training program documentation describing how the Provisional

Registered Medical Physicist will be supervised. The training program

documentation will include:

(a)

The names of the Registered Medical Physicist(s) who will provide

general, direct or personal supervision as the individual works to meet

the requirements of their certifying organization; and

(b)

A list of specific duties, and the level of supervision for each duty, that

the Provisional Registered Medical Physicist will perform.

2I.4

In addition to the requirements of 2I.1 and 2I.3, approval for registration in the Mammography

category shall be approved for a Registered Medical Physicist who:

2I4.1

Has the following combination of initial training and experience:

2I.4.1.1 A master's degree or higher in a physical science from an accredited institution

with no less than 20 semester hours in physics; and

2I.4.1.2 Have 20 contact hours of specialized training in conducting mammography

facility evaluations; and

2I.4.1.3 Experience of conducting evaluations of at least one mammography facility and a

total of at least ten (10) mammography units under the following conditions;

(a)

No more than one evaluation of a specific unit within a period of sixty

(60) calendar days can be counted towards the total mammography unit

survey requirement; and

(b)

This experience must be accomplished under the direct supervision of a

Registered Medical Physicist with approval in the Mammography

category;

2I.4.2

And the following continuing education and experience:

2I.4.2.1 At least fifteen (15) documented hours of continuing education in mammography

which are no more than thirty-six months old;

(a)

Medical physicists failing to maintain the continuing education

requirements of 2I.4.2.1 must meet 2I4.2.1 requirements prior to

independently conducting evaluations of mammography facilities.

2I.4.2.2 Surveys of at least six (6) mammography units operated in at least two (2)

mammography facilities within the immediately previous twenty-four (24) months;

(a)

Medical physicists failing to maintain the continuing experience

requirements of 2I.4.2.2 must meet 2I.4.2.2 requirements while under the

direct supervision of a Registered Medical Physicist with approval in the

Mammography category.

2I.4.2.3 Before a medical physicist may begin independently performing mammographic

evaluations of a new mammographic modality, that is, a mammographic modality

other than one for which the physicist received training to qualify under 2I.4.1, the

physicist must receive at least 8 hours of training in evaluating units of the new

mammographic modality.

2I.5

In addition to the requirements of 2I.1, approval for registration as a Registered Medical Physicist

for the Therapeutic Radiation Machines category shall be given to an individual who:

2I.5.1

Is certified by:

2I.5.1.1 The American Board of Radiology in Therapeutic Medical Physics, Therapeutic

Radiological Physics or Radiological Physics; or

2I.5.1.2 The American Board of Medical Physics in Radiation Oncology Physics; or

2I.5.1.3 The Canadian College of Physicists in Medicine in Radiation Oncology Physics;

or

2I.5.1.4 A equivalent specialty board or certification approved by the department.

PART 2, APPENDIX 2J: QUALIFIED TRAINER (QT) ADEQUATE RADIATION SAFETY TRAINING

AND EXPERIENCE

Any person who acts as a qualified trainer shall be an individual who:

2J.1

Has training and experience commensurate with criteria and standards for the radiation machine

application(s) that adequately prepare the individual to carry out the specified training

assignment(s).

2J.1.1 An interpreting physician, radiologic technologist or medical physicist who is approved

under MQSA program requirements is considered a qualified trainer for the respective

competency.

2J.1.2 A physician, radiologic technologist, or operator who is approved pursuant to 2.6.1 is

considered a qualified trainer for the respective competency.

2J.1.3 Other examples of an individual who might be considered by the Department to be a

qualified trainer for the purpose of providing training to meet the requirements of this part

include, but are not limited to:

(1)

A trainer in a post-secondary-school training institution; or

(2)

A manufacturer's representative; or

(3)

An individual approved as a RMP in the relevant specialty area; or

(4)

A program director or faculty member of a CAMPEP (Commission on

Accreditation of Medical Physics Education Programs) or AGCME (American

College of Graduate Medical Education) medical physics residency program.

PART 2, APPENDIX 2K: AUTHORIZED USER (24.3.3) FOR RADIATION THERAPY (24.7 OR 24.8)

ADEQUATE RADIATION SAFETY TRAINING AND EXPERIENCE

Any person who acts as an Authorized User for any therapeutic radiation machine subject to Part 24 shall

be a physician who has a current active State of Colorado license and:

2K.1

Has provided evidence of current certification in:

2K.1.1 Radiology or therapeutic radiology by the American Board of Radiology; or

2K.1.2 Radiation oncology by the American Osteopathic Board of Radiology; or

2K.1.3 Therapeutic radiology by the Royal College of Physicians and Surgeons of Canada; or

2K.1.4 Radiology, with specialization in radiotherapy, by the British Royal College of Radiology,

as a British “Fellow of the Faculty of Radiology” or “Fellow of the Royal College of

Radiology”; or

2K.1.5 Radiation therapy by a recognized specialty board that requires each candidate for

certification to:

2K.1.5.1

Satisfactorily complete a certification process that includes training

equivalent to that required in 2K.2.1 and supervised practical experience

equivalent to that required by 2K.2.2; and

2K.1.5.2

Pass an examination, administered by diplomates of the specialty board,

that tests knowledge and competence in radiation safety, treatment planning,

quality assurance, and human use of therapeutic radiation machines; or

2K.2

Has satisfied the following criteria:

2K.2.1 Satisfactory completion of 700 hours in basic techniques applicable to the use of a

therapeutic radiation machine unit, including:

2K.2.1.1

At least 200 hours of classroom and laboratory training in the following

areas:

(1)

Radiation physics and instrumentation;

(2)

Radiation protection;

(3)

Mathematics pertaining to the use and measurement of radioactivity; and

(4)

Radiation biology; and

2K.2.1.2

At least 500 hours of work experience, involving:

(1)

Reviewing full calibration measurements and periodic quality assurance

checks;

(2)

Evaluating prepared treatment plans, calculation of treatment times, and

patient treatment settings;

(3)

Using administrative controls to prevent reportable medical events;

(4)

Implementing emergency procedures to be followed in the event of the

abnormal operation of a therapeutic radiation machine unit or console;

and

(5)

Checking and using of radiation survey meters; and

2K.2.2 Completion of 3 years of supervised clinical experience in radiation therapy, including:

2K.2.2.1

An approved formal training program, approved by the Residency

Review Committee of the Accreditation Council for Graduate Medical Education

or the Council on Postdoctoral Training of the American Osteopathic Association;

and

2K.2.2.2

Supervised clinical experience, under the supervision of an authorized

user who meets the requirements of this Appendix 2K, or equivalent

requirements, to include:

(1)

Examining individuals and reviewing their case histories to determine

their suitability for therapeutic radiation machine treatment, and any

limitations and/or contraindications;

(2)

Selecting proper dose and how it is to be administered;

(3)

Calculating the therapeutic radiation machine doses and collaborating

with the authorized user in the review of patients' progress and

consideration of the need to modify originally prescribed doses and/or

treatment plans as warranted by patients' reactions to radiation; and

(4)

Post-administration follow-up and review of case histories.

2K.3

Training and experience required by Appendix 2K shall have been obtained:

2K.3.1 Within the 7 years preceding the date of license application; or

2K.3.2 Through documented subsequent continuing education and experience.

PART 2, APPENDIX 2L: RADIATION THERAPIST (24.3.5) ADEQUATE RADIATION SAFETY

TRAINING AND EXPERIENCE

Any person who operates a radiation therapy machine on living humans shall be an individual who:

2L.1

Has provided evidence of:

2L.1.1 Successful completion of a training program in radiation therapy which has resulted in a

certificate, associate degree, or baccalaureate degree in a radiologic technology program

that complies with the requirements of:

2L.1.1.1

The Joint Review Committee on Education in Radiologic Technology; or

2L.1.1.2

An accreditation organization recognized by the Council for Higher

Education Accreditation as an accrediting agency, other organizations

recognized by the United States Department of Education (USDE) or the Council

For Higher Education Accreditation (CHEA) to accredit educational programs in

radiation therapy; and

2L.1.2 Accreditation as a radiation therapist by, and having continued to maintain registration by

meeting the requirements of, The American Registry of Radiologic Technologists

(ARRT), or

2L.1.3 Accreditation by a specialty board recognized by the Department as equivalent to ARRT.

2L.2

Has maintained a minimum of twenty-four (24) hours of continuing education every two years in

the areas of radiology, radiation safety, radiography and similar fields. This education shall be

documented by certificate(s) or other attestation(s) of satisfactory completion.

PART 2, APPENDIX 2M: REQUIREMENTS FOR REGISTRATION AS A PROVISIONAL

MAMMOGRAPHER

Any individual who performs mammography and does not meet the requirements of 2.4.5.4(1):

2M.1

Shall have completed or be currently enrolled in a structured and documented training program

that requires:

2M.1.1 Forty (40) hours or more of documented training that includes breast anatomy and

physiology, positioning and compression, quality assurance/quality control techniques,

and imaging of patients with breast implants; and

2M.1.2 Eight (8) hours or more documented training in each mammography modality to be used

by the technologist in performing mammography examinations;

And

2M.2

Shall, prior to performing mammograms on living humans, register with the department as a

Provisional Mammographer.

2M.2.1 Each applicant for a provisional mammographer registration shall submit the Form R-64

series application and shall include all information required by the department as

indicated on the form(s) and all accompanying instructions.

2M.2.2 The provisional mammographer registration is issued for a period of one year and may

be renewed one time.

And

2M.3

While in training, shall perform at least 100 mammography examinations on patients under the

supervision of a qualified mammographer as follows:

2M.3.1 The initial 25 mammography examinations shall be performed under the personal

supervision of a qualified mammographer.

2M.3.2 All remaining mammography examinations after the initial 25 shall be performed under

the direct supervision of a qualified mammographer.

2M.3.2 All mammography examinations required by 2M.3.1 and 2M.3.2 shall be documented.

Documentation shall include the name of the supervised individual (individual in-training),

the type of exam/modality, the facility name, the examination date, and the name of the

supervising qualified mammographer or individual.

PART 2, APPENDIX 2N: INDUSTRIAL RADIATION MACHINE OPERATOR ADEQUATE RADIATION

SAFETY TRAINING AND EXPERIENCE

Any person who operates an analytical, industrial or other non-healing-arts radiation generating machine

shall be an individual who:

2N.1

For industrial radiography, has complied with all applicable training and experience requirements

of Part 5 and these regulations.

2N.2

For all non-healing-arts applications subject to Part 8 (including but not limited to analytical,

forensic, morgue, and homeland security uses) but not subject to Part 5, has provided written

documentation as evidence of:

2N.2.1 At least eight (8) hours of general training and experience in radiation safety acceptable

to the Department, except as follows:

2N.2.1.1

One (1) hour for any hand-held non-healing-arts radiation generating

machine; or

2N.2.1.2

One (1) hour for any cabinet or self-contained airport or port-of-entry x

ray machine or system; or

2N.2.1.3

Sufficient training and experience acceptable to the Department.

2N.2.2 The training required by 2N.2.1 shall include radiation safety training specific for each

radiation machine used, and demonstration of an understanding thereof, including

instruction in the:

2N.2.2.1

Proper operating procedures for the equipment, having read the

operating manual;

2N.2.2.2

Identification of radiation hazards associated with the use of the

equipment;

2N.2.2.3

Significance of the various radiation warning, safety devices, and

interlocks incorporated into the equipment, or the reasons they have not been

installed on certain pieces of equipment, and the extra precautions required in

such cases;

2N.2.2.4

Recognition of symptoms of an acute localized exposure; and

2N.2.2.5

Proper procedures for reporting an actual or suspected exposure; and

2N.2.3 Has subsequent documented annual training.

PART 2, APPENDIX 2O: FLUOROSCOPY IMAGING SYSTEM OPERATOR ADEQUATE RADIATION

SAFETY TRAINING AND EXPERIENCE

Except for those individuals exempted in 2.4.5.5(1), any person who operates a fluoroscopic machine or a

machine capable of fluoroscopic imaging while in fluoroscopic mode for clinical purposes, shall be limited

to a licensed Physician Assistant, licensed Advanced Practice Registered Nurse, or a nationally certified

and registered Cardiovascular Lab Specialist who is at least 18 years of age working within their scope of

practice, and:

2O.1

Meets the following requirements:

2O.1.1 Has completed a course that includes at least forty (40) hours of education on topics that

include, but are not limited to, radiation physics, radiation biology, radiation safety and

radiation management applicable to fluoroscopy;

And

2O.1.2 Has completed forty (40) hours of clinical experience in the use of fluoroscopy for

guidance in diagnostic and therapeutic procedures under the personal supervision of a

Colorado licensed physician;

And

2O.1.3 Meets the requirements of 2O.1.3.1 or 2O.1.3.2 or 2O.1.3.3, as follows:

2O.1.3.1

Is a Physician Assistant or Advanced Practice Registered Nurse who

has received a passing score on the American Registry of Radiologic

Technologists (ARRT) fluoroscopy operators examination.

Or

2O.1.3.2

Is registered through Cardiovascular Credentialing International (CCI) as

a Registered Cardiovascular Invasive Specialist (RCIS) or a Registered

Electrophysiology Specialist (RCES);

Or

2O.1.3.3

Is registered with another organization that has been specifically

approved in writing by the department.

And

2O.1.4 Is registered with the department in accordance with Section 2.4.5.5.

And

2O.2

Maintains their department fluoroscopy operator registration by submitting the registration

renewal application and required fee along with the following:

2O.2.1 Physician Assistants and Advanced Practice Registered Nurses shall submit a current

active state of Colorado license issued by the Colorado Department of Regulatory

Agencies.

2O.2.2 Nationally certified/registered Cardiovascular Lab Specialists shall submit a copy of their

active national certification/registration.