§ 606.60 Equipment.

FederalRegulations

Ask Donna

How this section applies to your facts.

Title 21—Food and Drugs > CHAPTER I—FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES > SUBCHAPTER F—BIOLOGICS > PART 606—CURRENT GOOD MANUFACTURING PRACTICE FOR BLOOD AND BLOOD COMPONENTS > Subpart D—Equipment

This text was captured on Sep 22, 2026. It is a snapshot, not a live feed, so check the official code before relying on it.

Text

(a) Equipment used in the collection, processing, compatibility testing, storage and distribution of blood and blood components shall be maintained in a clean and orderly manner and located so as to facilitate cleaning and maintenance. The equipment shall be observed, standardized and calibrated on a regularly scheduled basis as prescribed in the Standard Operating Procedures Manual and shall perform in the manner for which it was designed so as to assure compliance with the official requirements prescribed in this chapter for blood and blood products.

(b) Equipment that shall be observed, standardized and calibrated with at least the following frequency, include but are not limited to:

Equipment

Performance check

Frequency

Frequency of calibration

Temperature recorder

Compare against thermometer

Daily

As necessary.

Refrigerated centrifuge

Observe speed and temperature

Each day of use

Do.

Hematocrit centrifuge

Standardize before initial use, after repairs or adjustments, and annually. Timer every 3 mo.

General lab centrifuge

Tachometer every 6 mo.

Automated blood-typing machine

Observe controls for correct results

Each day of use

Hemoglobinometer

Standardize against cyanmethemoglobin standard

......do

Refractometer

Standardize against distilled water

......do

Blood container scale

Standardize against container of known weight

......do

As necessary.

Water bath

Observe temperature

......do

Do.

Rh view box

......do

......do

Do.

Autoclave

......do

Each time of use

Do.

Serologic rotators

Observe controls for correct results

Each day of use

Speed as necessary.

Laboratory thermometers

Before initial use.

Electronic thermometers

Monthly.

Vacuum blood agitator

Observe weight of the first container of blood filled for correct results

Each day of use

Standardize with container of known mass or volume before initial use, and after repairs or adjustments.

(c) Equipment employed in the sterilization of materials used in blood collection or for disposition of contaminated products shall be designed, maintained and utilized to ensure the destruction of contaminating microorganisms. The effectiveness of the sterilization procedure shall be no less than that achieved by an attained temperature of 121.5 °C (251 °F) maintained for 20 minutes by saturated steam or by an attained temperature of 170 °C (338 °F) maintained for 2 hours with dry heat.

[40 FR 53532, Nov. 18, 1975; 40 FR 55849, Dec. 2, 1975, as amended at 45 FR 9261, Feb. 12, 1980; 57 FR 11263, Apr. 2, 1992; 57 FR 12862, Apr. 13, 1992]

This is a copy of a public record, reproduced as it was published. It is not legal advice, and it may not be the version a court would rely on. Check the official source before you cite it.