DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

21 CFR Part 888

[Docket No. 95N-0176]

RIN 0910-ZA12

Orthopedic Devices: Classification and Reclassification of

Pedicle Screw Spinal Systems

Agency: Food and Drug Administration, HHS.

Action: Final rule.

-----------------------------------------------------------------------

SUMMARY: The Food and Drug Administration (FDA) is classifying certain

previously unclassified preamendments pedicle screw spinal systems into

class II (special controls) and reclassifying certain postamendments

pedicle screw spinal systems from class III (premarket approval) to

class II. FDA is taking this action because it believes that special

controls would provide reasonable assurance of safety and

effectiveness. This action is being taken under the Federal Food, Drug,

and Cosmetic Act (the act), as amended by the Medical Device Amendments

of 1976 (the 1976 amendments), the Safe Medical Devices Act of 1990

(the SMDA), and the Food and Drug Administration Modernization Act of

1997 (FDAMA).

EFFECTIVE DATE: August 26, 1998.

FOR FURTHER INFORMATION CONTACT: Aric D. Kaiser, Center for Devices and

Radiological Health (HFZ-410), Food and Drug Administration, 9200

Corporate Blvd., Rockville, MD 20850, 301-594-2036.

SUPPLEMENTARY INFORMATION:

Table of Contents

I. Background

II. Regulatory History of the Device

III. Summary of the Final Rule

IV. Proposed Rule Clarifications

V. Analysis of Comments and FDA's Response

A. Issues Relating to the Recommendations of the Panel, FDA's

Tentative Findings, and Summary of the Data Upon Which FDA's

Findings Were Based

B. Issues Relating to Information Published in the 1994

Supplementary Issue of the Journal Spine (vol. 20S, 1994)

C. Issues Relating to the January 1995, 510(k) Substantial

Equivalence Determination for a Pedicle Screw Spinal System Intended

for Severe Spondylolisthesis

D. Issues Relating to Misstatements or False Statements

Appearing in the Proposed Rule

E. Issues Relating to the FDA's Issuance of Regulations

F. Responses to Comments Which Contained Clinical Data

G. Requests for Additional Pedicle Screw Clinical Trials and

Data Analyses

H. Issues Relating to Indications for Use

I. Issues Relating to Special Controls

J. Other Issues

K. Labeling of Bone Screws

L. Review of New Pedicle Screw Spinal System 510(k)'s

M. Review of New Information Published and Submitted After

Publication of the Proposed Rule: Pedicle Screw and Related

Literature and MedWatch and MDR System Reports

VI. References

VII. Environmental Impact

VIII. Analysis of Impacts

I. Background

The act (21 U.S.C. 331 et seq.), as amended by the 1976 amendments

(Pub. L. 94-295), the SMDA (Pub. L. 101-629), and FDAMA (Pub. L. 105-

115), established a comprehensive system for the regulation of medical

devices intended for human use. Section 513 of the act (21 U.S.C. 360c)

established three categories (classes) of devices, depending on the

regulatory controls needed to provide reasonable assurance of their

safety and effectiveness. The three categories of devices are: Class I

(general controls), class II (special controls), and class III

(premarket approval).

Under section 513 of the act, devices that were in commercial

distribution before May 28, 1976 (the date of enactment of the 1976

amendments), generally referred to as preamendments devices, are

classified after FDA has: (1) Received a recommendation from a device

classification panel (an FDA advisory committee); (2) published the

panel's recommendation for comment, along with a proposed regulation

classifying the device; and (3) published a final regulation

classifying the device. FDA has classified most preamendments devices

under these procedures.

Devices that were not in commercial distribution prior to May 28,

1976, generally referred to as postamendments devices, are classified

automatically by statute (section 513(f) of the act) into class III

without any FDA rulemaking process. Those devices remain in class III

and require premarket approval, unless and until: (1) The device is

reclassified into class I or II; (2) FDA issues an order classifying

the device into class I or II in accordance with new section 513(f)(2)

of the act, as amended by FDAMA; or (3) FDA issues an order finding the

device to be substantially equivalent, in accordance with section

513(i) of the act, to a predicate device that does not require

premarket approval. The agency determines whether new devices are

substantially equivalent to previously offered devices by means of

premarket notification procedures in section 510(k) of the act (21

U.S.C. 360(k)) and part 807 of the regulations (21 CFR part 807).

A preamendments device that has been classified into class III may

be marketed, by means of premarket notification procedures, without

submission of a premarket approval application (PMA) until FDA

promulgates a final regulation under section 515(b) of the act (21

U.S.C. 360e(b)) requiring premarket approval.

Reclassification of classified preamendments devices is governed by

section 513(e) of the act. This section provides that FDA may, by

rulemaking, reclassify a device (in a proceeding that parallels the

initial classification proceeding) based upon ``new information.'' The

reclassification can be initiated by FDA or by the petition of an

interested person. The term ``new information,'' as used in section

513(e) of the act, includes information developed as a result of a

reevaluation of the data before the agency when the device was

originally classified, as well as information not presented, not

available, or not developed at that time. (See, e.g., Holland Rantos v.

United States Department of Health, Education, and Welfare, 587 F.2d

1173, 1174 n.1 (D.C. Cir. 1978); Upjohn v. Finch, 422 F.2d 944 (6th

Cir. 1970); Bell v. Goddard, 366 F.2d 177 (7th Cir. 1966).)

Reevaluation of the data previously before the agency is an

appropriate basis for subsequent regulatory action where the

reevaluation is made in light of newly available regulatory authority

(see Bell v. Goddard, supra, 366 F.2d at 181; Ethicon, Inc. v. FDA, 762

F. Supp. 382, 389-91 (D.D.C. 1991)), in light of changes in ``medical

science.'' (See Upjohn v. Finch, supra, 422 F.2d at 951.) Regardless of

whether data before the agency are past or new data, the ``new

information'' on which any reclassification is based is required to

consist ``valid scientific evidence,'' as defined in section 513(a)(3)

of the act and Sec. 860.7(c)(2) (21 CFR 860.7(c)(2)). (See, e.g.,

General Medical Co. v. FDA, 770 F.2d 214 (D.C. Cir. 1985); Contact Lens

Assoc. v. FDA, 766 F.2d 592 (D.C. Cir.), cert. denied, 474 U.S. 1062

(1985).) FDA relies upon ``valid scientific evidence'' in the

classification process to determine the level of

[[Page 40026]]

regulation for devices. For the purpose of reclassification, the valid

scientific evidence upon which the agency relies must be publicly

available. Publicly available information excludes trade secret and/or

confidential commercial information, e.g., the contents of a pending

PMA. (See section 520(c) of the act (21 U.S.C. 360j(c).)

II. Regulatory History of the Device

Consistent with the act and the regulation, FDA referred the

proposed classification and reclassification of pedicle screw spinal

systems to the Orthopedic and Rehabilitation Devices Panel (the Panel),

an FDA advisory committee, for its recommendation on the requested

classification and change in classification.

The Panel reviewed complication type and rate data present in the

literature, a meta-analysis of the literature; a nationwide,

retrospective Cohort study of patients treated with the devices;\1\ and

a review of publicly released investigational device exemptions (IDE)

data from patients treated with pedicle screw spinal systems. The Panel

recommended that the postamendments pedicle screw spinal systems

intended to treat spinal fracture and degenerative spondylolisthesis of

the thoracic, lumbar, and sacral spine, be reclassified from class III

into class II.

---------------------------------------------------------------------------

\1\ The Cohort study was an open, nonblinded, historical Cohort

study designed to recruit the maximum number of surgeons to provide

clinical data on patients who had undergone spinal fusion surgery.

Three hundred fourteen surgeons were recruited through announcements

at professional society meetings and direct mailings to professional

society memberships. Only clinical data from spinal fusion surgeries

intended to treat degenerative spondylolisthesis or spinal trauma

(fracture) that were performed between January 1, 1990, and December

31, 1991, were used in the analysis. This was done in an effort to

maximize the number of patients with a minimum of 24 months

followup. Data from 3,498 patients were collected.

---------------------------------------------------------------------------

In January, 1995, a manufacturer was able to demonstrate

preamendments status for pedicle screw spinal systems intended to

provide immobilization and stabilization of spinal segments as an

adjunct to spinal fusion in the treatment of grades 3 or 4 severe

spondylolisthesis at the fifth lumbar-first sacral (L5-

S1) spinal level. In an April 1995, homework assignment, FDA

requested that the Panel recommend a classification for this

unclassified preamendments device. The Panel recommended that the

unclassified preamendments pedicle screw spinal systems intended to

provide immobilization and stabilization of spinal segments as an

adjunct to spinal fusion in the treatment of grades 3 or 4 severe

spondylolisthesis at the L5-S1 spinal level be

classified into class II.

In the Federal Register of October 4, 1995 (60 FR 51946), FDA

published a proposed rule to classify certain unclassified

preamendments pedicle screw spinal systems (for use in certain types of

severe spondylolisthesis ) into class II, to reclassify certain

postamendments pedicle screw spinal systems (for use in fracture and

other conditions) from class III to class II, and to retain in class

III other postamendments pedicle screw spinal systems. The proposed

rule reflected FDA's belief that the clinical outcomes and

complications described in the literature, clinical data, and MDR and

MedWatch surveillance data bases,\2\ described patient risks and

benefits of pedicle screw spinal systems comparable to other class II

spinal fixation devices and that special controls have been identified

which would provide a reasonable assurance of safety and effectiveness,

i.e., compliance with material standards, mechanical testing standards,

biocompatibility standards, and special labeling requirements.

Initially, FDA provided for interested persons to submit comments on

the proposal by January 2, 1996. Subsequently, in the Federal Register

of December 29, 1995 (60 FR 67345), FDA extended the comment period

until March 4, 1996, in response to several requests for extension of

the comment period.

---------------------------------------------------------------------------

\2\ MDR and MedWatch data bases are two reporting systems that

FDA uses to track adverse events, e.g., injuries, deaths, and device

malfunctions, related to medical devices. The information consists

of a combination of mandatory and/or voluntary adverse event reports

from manufacturers, distributors, user facilities, healthcare

professionals, as well as consumers.

---------------------------------------------------------------------------

FDA received 4,060 comments in response to the proposed rule. These

comments were submitted by physicians, patients, lawyers, device

manufacturers, trade associations, and other interested parties. The

overwhelming majority of these comments were in favor of the proposed

rule, although some comments were opposed to the proposed rule, and a

few were both in favor of some aspects of the proposed rule and opposed

to others.

In response to comments received on the proposed rule, FDA

reanalyzed the meta-analysis of the literature, the Cohort study, and

the publicly released IDE data for the indications of spinal fractures

and degenerative spondylolisthesis. The reanalysis of the meta-analysis

of the literature consisted of a review of the summary data and

conclusions from the original, published analysis. The review of the

Cohort study consisted of an audit (Ref. 1) of a structured sample of

all 377 patients enrolled by 21 of the 314 participating surgeons, a

reanalysis (Ref. 2) of all of the data from the audit, and a comparison

to the data from unaudited surgeons. The Division of Bioresearch

Monitoring (BIMO) in the Office of Compliance performed the data audit,

while the Office of Device Evaluation and the Office of Surveillance

and Biometrics performed the reanalyses. This audit found records were

incomplete and investigators had not followed the protocol. In review

of the audit, the agency concluded that the disparities and

irregularities were consistent, with respect to both type and scope,

with other audits of similar studies. After careful reanalysis of the

potential impact of the ``problem'' records, the agency concluded that

they could not account for the favorable results reported in this

study.

The review of the Cohort study in the context of the audit findings

yielded results that supported the safety and effectiveness of these

devices. For spinal fracture, pedicle screw spinal systems presented

risks and benefits that were comparable to those presented by

nonpedicle screw instrumented spinal fusion. The devices used in the

comparison group are class II medical devices. For spondylolisthesis,

the review in the context of the audit findings described an advantage

for pedicle screw spinal systems with regard to the clinical outcome

parameters of fusion and improvement in neurological status when

compared to noninstrumented spinal fusions. For the other parameters

that were analyzed, e.g., pain, function, and reoperation rate, pedicle

screw spinal systems did not always demonstrate an advantage compared

to noninstrumented spinal fusion. When compared to instrumented spinal

fusions, however, results among pedicle screw spinal system patients

for these parameters were not statistically equivalent and not worse.

Thus, FDA has concluded that the results from the review of the Cohort

study are consistent with those reported in the literature and the

publicly released IDE data.

The reanalysis of the meta-analysis of the literature describing

experience with pedicle screw spinal systems in treating spinal

fracture and degenerative spondylolisthesis found that pedicle screw

spinal systems present risks and benefits that are comparable to those

presented by nonpedicle screw spinal systems and noninstrumented spinal

fusions. For degenerative spondylolisthesis, the reanalysis found

[[Page 40027]]

that patient results with pedicle screw spinal systems were comparable

to those with noninstrumented spinal fusions; it did not find a

clinically significant improvement in results at followup obtained with

instrumented spinal fusions over noninstrumented spinal fusions.

The reanalysis of the publicly available IDE data supports the

Panel's recommendation for the classification and reclassification of

pedicle screw spinal systems intended to treat spinal fractures and

severe spondylolisthesis. It also supports the use of pedicle screw

spinal systems when intended to provide immobilization and

stabilization of spinal segments in skeletally mature patients as an

adjunct to fusion for the treatment of the following acute and chronic

instabilities or deformities of the thoracic, lumbar, and sacral spine:

degenerative spondylolisthesis with objective evidence of neurologic

impairment, fractures, dislocations, scoliosis, kyphosis, spinal

tumors, and failed previous fusion (pseudarthrosis).

When all of these data are viewed in conjunction with the medical

literature and the MDR and MedWatch surveillance data, no new issues

relating to the safety or effectiveness of pedicle screw spinal systems

are raised. Therefore, the agency has concluded that these data provide

valid scientific evidence that certain special controls in conjunction

with the general controls applicable to all devices, will provide a

reasonable assurance of the safety and effectiveness of pedicle screw

spinal systems for L5-S1 use and for use at other

levels for the treatment of degenerative spondylolisthesis with

objective evidence of neurologic impairment.

The agency also reviewed whether the Panel was properly

constituted. Investigation of alleged undisclosed and unwaived

conflicts of interest held by Panel members found either no omissions

of current interests or omissions of minor interests for all but one of

the Panel members. The agency has concluded that the minor omissions

are insignificant and do not constitute a financial conflict of

interest that would credibly influence the members' actions in forming

the Panel's recommendations.

The agency has found that one voting Panel member did have

significant undisclosed financial conflicts. However, because the

recommendation of the Panel, both in the July 23, 1994, meeting and on

the subsequent homework assignment, was unanimous and this individual

was not controlling, or unduly influential, of the votes of the other

Panel members and was not necessary to constitute a quorum, after

expunging the participation of this Panel member, FDA has concluded

that this Panel, both in the meeting and on the subsequent homework

assignment, was a valid scientific Panel to make recommendations to the

agency.

The agency's reanalysis of these data has confirmed its original

conclusion, reflected in the proposed rule, that the risks and benefits

of pedicle screw spinal systems are comparable to those of other class

II spinal fixation devices. FDA's decision to classify and reclassify

these devices into class II is based upon valid scientific evidence

establishing that the special controls described above, along with the

general controls applicable to all devices under the act, provide a

reasonable assurance of the safety and effectiveness of pedicle screw

spinal systems.

III. Summary of the Final Rule

In this final rule, FDA is classifying into class II the

unclassified preamendments pedicle screw spinal systems intended for

treatment of severe spondylolisthesis (grades 3 and 4) of the

L5-S1 vertebra in skeletally mature patients

receiving fusion by autogenous bone graft having implants attached to

the lumbar and sacral spine with removal of the implants after the

attainment of a solid fusion. In addition, FDA is reclassifying into

class II the postamendments class III pedicle screw spinal systems

intended to provide immobilization and stabilization of spinal segments

in skeletally mature patients as an adjunct to fusion in the treatment

of the following acute and chronic instabilities or deformities of the

thoracic, lumbar, and sacral spine: Degenerative spondylolisthesis with

objective evidence of neurologic impairment, fracture, dislocation,

scoliosis, kyphosis, spinal tumor, and failed previous fusion

(pseudarthrosis). Pedicle screw spinal systems intended for any other

uses are considered postamendments class III devices for which

premarket approval is required. The following four special controls

apply to the devices being classified and reclassified into class II:

(1) Compliance with materials standards, (2) compliance with mechanical

testing standards of performance, (3) compliance with biocompatibility

standards, and (4) adherence to labeling requirements.

IV. Proposed Rule Clarifications

FDA is taking this opportunity to clarify that neither well-

controlled investigations nor valid scientific evidence relating to

pedicle screw spinal systems intended for use in the cervical spine is

available and, therefore, the safety and effectiveness of these devices

for this intended use have not been demonstrated. As a result, pedicle

screw spinal systems intended for use in the cervical spine are

excluded from this classification and reclassification and are

considered postamendments class III devices for which premarket

approval is required.

In addition, although not specifically stated in the preamble to

the proposed rule, all valid scientific evidence reviewed by the Panel

and FDA were obtained from skeletally mature populations. To date, the

safety and effectiveness of pedicle screw spinal systems in pediatric

populations have not been demonstrated. Consequently, pedicle screw

spinal systems intended for use in pediatric populations are

postamendments class III devices for which premarket approval is

required.

V. Analysis of Comments and FDA's Response

A. Issues Relating to the Recommendations of the Panel, FDA's Tentative

Findings, and Summary of the Data Upon Which FDA's Findings Were Based

1. Several comments believed that valid scientific evidence was not

presented to the Panel or used in formulating the proposed rule. These

comments argued that only prospective, randomized, concurrently-

controlled clinical trials constitute valid scientific evidence and

that anything else is insufficient to support device reclassification.

FDA disagrees that only data from prospective, randomized,

concurrently-controlled clinical trials can constitute valid scientific

evidence. Although prospective, randomized, concurrently controlled

clinical trials have the potential to produce the most convincing and

reliable data, e.g., all sources of bias have been reduced to a

minimum, such clinical trials are not the only type of study that can

produce data adequate to support a determination that there is

reasonable assurance that a device is safe and effective for its

conditions for use. In fact, Sec. 860.7(c)(2) defines valid scientific

evidence as

* * * evidence from well-controlled investigations, partially

controlled studies, studies and objective trials without matched

controls, well-documented case histories conducted by qualified

experts, and reports of significant human experience with a marketed

device, from which it can fairly and responsibly be concluded by

qualified experts that there is reasonable assurance of the safety

and effectiveness of a device under

[[Page 40028]]

its conditions of use. The evidence may vary according to the

characteristics of the device, the existence and adequacy of

warnings and other restrictions, and the extent of experience with

its use * * *.

From this definition, it is clear that there is a hierarchy of data

that fits the definition of valid scientific and that, while well-

controlled, prospective clinical trials are at the top of the

hierarchy, they are not the only source of data that may support a

determination regarding reasonable assurance of safety and

effectiveness for purposes of classifying and/or reclassifying a

device.

FDA also disagrees that valid scientific evidence was not presented

to the Panel or used in support of the proposed rule. The three sources

of data presented to the Panel and used in support of the proposed rule

were: (1) Reformatted IDE data, (2) a meta-analysis of the literature,

and (3) the Cohort study. The reformatted IDE data represent data from

well-controlled investigations, while the meta-analysis of the

literature and the Cohort study represent studies and objective trials

without matched controls or well-documented case histories or reports

of significant human experience. All three sources of data used in

support of the classification and reclassification of pedicle screw

spinal systems clearly fall within the definition of valid scientific

evidence in Sec. 860.7(c)(2).

2. One comment objected that, in addition to not being valid

scientific evidence, the three sources of data, i.e., the meta-analysis

of the literature, the reformatted IDE data and the Cohort study, were

flawed.

The comment noted the following deficiencies with the meta-

analysis:

FDA previously determined that the available literature

on pedicle screw spinal systems could not be used to support device

reclassification.

FDA disagrees. FDA made that statement prior to January, 1993, when

no adequate analysis of pedicle screw literature had been provided to

the agency. FDA believes that, while individual literature articles

describing the use of pedicle screw spinal systems would be

insufficient to support reclassification of a device, group analysis of

relevant articles may be adequate, especially where, as here, the group

analysis is considered in conjunction with other supporting data.

Furthermore, after noting the limitations of the individual studies

reported in the literature, FDA concluded that the literature, taken as

a whole and used in conjunction with the other sources of data,

provided adequate information to support the reclassification of

pedicle screw spinal systems intended to treat degenerative

spondylolisthesis with objective evidence of neurologic impairment or

spinal trauma.

The meta-analysis is not an appropriate scientific

technique, as applied to retrospective studies, because different

studies have different parameters, biases, and strengths and

weaknesses, all of which invalidate the pooling of data.

FDA disagrees. Although meta-analysis of literature may be less

rigorous than other forms of scientific research, it still provides

useful information. As discussed in section V.A.1 of this document,

Sec. 860.7(c)(2) defines ``valid scientific evidence'' to include many

types of evidence of varying degrees of scientific rigor, including

meta-analysis of literature. FDA participated in the development of the

meta-analysis because the agency believed that this analysis could

produce data meeting the definition of valid scientific evidence.

Finally, the inherent limitations of a literature meta-analysis were

discussed during the presentation of this analysis at the July 23,

1994, Panel meeting and in the preamble to the proposed rule (60 FR

51946).

The meta-analysis actually lent support to the conclusion

that pedicle screw fixation is less effective than other methods of

treating degenerative spondylolisthesis and spinal fracture and that it

may present the patient with more risks.

FDA disagrees. With respect to degenerative spondylolisthesis,

there was no statistically significant difference in fusion rates

between the control and the pedicle screw spinal system treatment

groups. This is supportive data that clarifies the relative safety and

effectiveness of pedicle screw spinal systems for this use. With

respect to spinal fracture, significantly higher fusion rates were

achieved in the pedicle screw spinal system treatment group than in the

nonpedicle screw treatment groups. Thus, the meta-analysis confirmed

the comparability of pedicle screw spinal systems to other class II

devices used to treat spinal fracture in terms of safety and

effectiveness.

Fifty-five of the 58 studies in the meta-analysis were

nonexperimental case-series having no validity as scientific evidence.

FDA disagrees. As discussed in section V.A.1 of this document,

Sec. 860.7(c)(2) states that valid scientific evidence may include ``*

* * well-documented case histories conducted by qualified experts * *

*''. Moreover, these well-documented case studies, which were conducted

by qualified experts, were not the sole basis for the proposed

classification/reclassification, but rather were considered in

conjunction with data from various other sources.

The comment also noted the following deficiencies with the

reformatted IDE data:

The reformatted IDE data are not appropriate for

classifying and reclassifying pedicle screw spinal systems because FDA

previously had determined that these data could not support PMA's for

these devices.

FDA disagrees in part. Prior to the August 20, 1993, Panel meeting,

FDA had determined that data from individual IDE's were insufficient to

support PMA's for those devices. Nevertheless, FDA recognized that the

IDE data could still be valuable. In 1993, after receiving permission

from nine IDE sponsors to publicly release and use their combined data,

FDA determined that the data, reviewed as a whole, corroborated the

results of other available data sets demonstrating the safety and

effectiveness of pedicle screw spinal systems.

The reformatted IDE data are inherently suspect because

they (1) were reformatted by the sponsors and not by FDA, (2) were not

provided for public scrutiny during the Panel meeting or at any other

time, and (3) may have omitted poor results.

FDA disagrees that the reformatted data were suspect because they

were reformatted by the sponsors and not by FDA. If IDE data are not

properly formatted, FDA requests the sponsor to reformat its data for

proper presentation to the agency. Furthermore, data in all marketing

applications are formatted by the sponsor. Therefore, the simple fact

that the IDE data were reformatted by the sponsor, not by FDA, does not

make these data inherently suspect.

FDA also disagrees that the data were suspect because they were not

presented for public scrutiny. For reclassification purposes, the valid

scientific evidence upon which the agency relies must be publicly

available Sec. 860.5(e) (21 CFR 860.5(e)). Publicly available

information excludes trade secret and/or confidential commercial

information (21 CFR 20.61). IDE data typically contain trade secret

and/or confidential commercial information and, consequently,

ordinarily may not be publicly disclosed by the agency to support

reclassification of a device (49 FR 17523 at 17531 and 17532, April 24,

1984). In fact, under Sec. 812.38(a) and (b)(3) (21 CFR 812.38 (a) and

(b)(3)), FDA generally does not acknowledge the existence of an IDE or

disclose any

[[Page 40029]]

of the collected data. However, on August 13, 1993, after receiving

permission from nine IDE sponsors to publicly release and use their

combined data, the Commissioner of Food and Drugs (the Commissioner)

exercised his discretionary authority under Sec. 812.38(b)(2) and

publicly released the data from nine IDE's, redacted of the

identification of the IDE sponsors, institutional review boards,

investigators, and patients. Although FDA did not make publicly

available the unformatted data from the IDE studies or the

identification of the IDE sponsors, institutional review boards,

investigators or patients, the agency did provide the public with a

detailed report of the combined IDE data (60 FR 51946 at 51961, ref.

173). This information was publicly available for analysis for more

than 2 years before the publication of the proposed rule.

Finally, FDA disagrees that the data were suspect because they may

have omitted poor results. Nine of fourteen sponsors provided their

reformatted IDE data for analysis. There is no evidence that the five

sponsors who did not offer their data did so because the data reflected

adversely on the performance of their products. They may not have

provided their data for any number of reasons. For instance, the

sponsors may have believed that they had an inadequate amount of data

to contribute to the effort or that the data may not have been in a

readily accessible format. Regardless of the reason, the publicly

available reformatted IDE data corroborate the results of other studies

that demonstrate the safety and effectiveness of pedicle screw spinal

systems. Specifically, the fusion rates, complication rates, and

reoperation, revision, and removal rates attained under publicly

available IDE studies were consistent with what was observed in the

literature for such devices.

The 12-month followup time period was inadequate to

support any conclusions. Specifically, the comment stated that the

Panel was not supplied with any information on the safety and

effectiveness of these devices at more than 1 year following surgery.

The comment continued that, without a minimum followup period of 2

years, it is impossible to make appropriate conclusions with regard to

the longer-term safety and efficacy of these devices in accordance with

accepted scientific convention.

FDA agrees that a 12-month followup time period would be inadequate

and, therefore, selected a 24-month followup period for analysis. The

24-month followup period was also supported by the Panel and the

literature. Contrary to the comment's statement, the Panel was supplied

with information on the safety and effectiveness of pedicle screw

spinal systems at more than one year following surgery. Spinal fusion

generally occurs within 6 to 18 months after surgery. The majority of

post-operative complications occur by the 18th month time point. For

these reasons, FDA concluded that a 24-month followup period was

adequate. FDA recognizes that not all of the reformatted IDE data were

from a 24-month followup examination. However, a sufficient amount of

data from a 24 month followup evaluation was examined for the Panel to

make a recommendation about the reasonable assurance of safety and

effectiveness of pedicle screw spinal systems for their class II

intended uses.

The comment stated that the lost-to-followup rate was too

high.

FDA agrees that the lost-to-followup rate was high. FDA believes

that patients with poor results tend to either return to their surgeons

more frequently or go to other caregivers, attempting to receive the

pain relief and return of function that they were originally seeking.

It cannot be determined whether the patients who were lost-to-followup

had acceptable results or went to other caregivers. However, FDA does

not believe that this theoretical weakness in the data is of such a

magnitude as to justify rejecting the studies. Thus, both the Panel and

FDA believe that the lost-to-followup rate was not unacceptably high.

The comment noted the following deficiencies with the Cohort study:

The Cohort study did not constitute valid scientific

evidence.

FDA disagrees. As described above, valid scientific evidence

encompasses a wide variety of data. The Cohort study satisfies the

definition of valid scientific evidence because it consisted of data

from well-documented case histories conducted by qualified experts and

reports of significant human experience.

The sample size and statistical power used in designing

the Cohort study were inadequate and, therefore, no reliable

conclusions can be drawn from the study. Another comment attempted to

rebut this allegation.

FDA believes that the sample size and statistical power

calculations that were performed in the Cohort study were accurate and

appropriate and, consequently, that the conclusions drawn from the

study had a sound basis.

The Cohort study was biased and the data were not

independently audited.

FDA disagrees. While the potential for bias exists in any study, it

was of particular concern in the design of the Cohort study due to its

retrospective nature. As described at the July 22, 1994, Panel meeting

and in the preamble to the proposed rule (60 FR 51946 at 51954),

various steps were taken to minimize the potential effects of bias due

to the study design. In addition, contrary to the comment's assertion,

there was a review of the data by an independent auditor and a

subsequent FDA BIMO audit and review. The review by the independent

auditor was not extensive and no definitive conclusions can be drawn

from its analysis of the Cohort study data. Although both audits

uncovered instances of protocol departures, recordkeeping

inconsistencies, or a lack of clear understanding or unfamiliarity with

the protocol requirements on the part of a participating surgeon, these

inconsistencies and protocol departures did not affect the reliability

of the data. For example, one type of reported protocol recordkeeping

departure was that some data forms were incomplete. In some instances,

the data forms simply omitted the patient's weight, but not the

patient's fusion status. The absence of that piece of information,

while rendering the form incomplete, clearly did not affect the

clinical outcome analysis. A more significant protocol departure

related to the inclusion and analysis of data from patients whose

diagnosis did not meet patient eligibility criteria. However, no

obvious pattern that would improve overall patient outcomes was

identified because these departures included indications for surgery

both more and less severe than those targeted by the protocol.

The data retrieved from the BIMO audit were analyzed to determine

if the major outcomes of the Cohort study were significantly different

(statistically or clinically) with or without the presence of protocol

departures, with or without the presence of recordkeeping

inconsistencies, or at sites where the participating investigator,

based on the audit, was or was not familiar with the protocol

requirements. While some differences were noted between sites with and

without inconsistencies, in most cases, these were not statistically

significant and no consistent or clinically relevant patterns were

noted. The analysis of the audited data did not find systemic bias in

either the conduct of the study or its reported results. None of the

analyzed audit data contradicted the published results of the Cohort

study. Finally, the data audit analysis concluded that the audited data

were consistent with other publicly available

[[Page 40030]]

data and that the Cohort study data could be used as part of a larger

body of data to support the classification and reclassification of

pedicle screw spinal systems.

Documents relating to the Cohort study were destroyed.

FDA disagrees. All Cohort study data were maintained in a master

file. Only extra copies of information were destroyed in an effort to

maintain the confidentiality of the identities of the participating

surgeons and their patients. In addition, as a matter of course, FDA

routinely assists Panel members in destroying copies of documents

containing trade secret and/or confidential commercial information that

they have received from FDA as preparatory material for a Panel

meeting.

Certain FDA employees had inappropriate relationships

with pedicle screw manufacturers and others involved in the Cohort

study.

This allegation, which has two parts, is unfounded. FDA performed

an internal affairs investigation of the employees about whom

allegations were made. This investigation showed that their attendance

at a health professional meeting was properly paid for by the agency,

not subsidized by the regulated industry. Also in the case of one

employee, FDA's investigation showed that negotiations regarding

outside employment with the regulated industry had been properly

reported to the employee's supervisors and immediate colleagues in all

instances.

The Scientific Committee and the Spinal Implant

Manufacturers Group (SIMG) were not independent.

FDA disagrees. The preamble to the proposed rule and the subsequent

correction (60 FR 51946 and 60 FR 66227, December 21, 1995) described

the makeup of the Scientific Committee and SIMG. SIMG consisted of

representatives of manufacturers who provided funding to support a

nationwide analysis of clinical data relating to pedicle screw spinal

systems. SIMG did not participate in the design of the study. The study

was designed and implemented by the Scientific Committee with input

from FDA as to the feasibility of various clinical study design

parameters. The Scientific Committee was formed by five professional

medical societies. Although two SIMG representatives were part of the

Scientific Committee, they were nonvoting members. Furthermore, even if

there were not independence between the Scientific Committee and SIMG,

there is no requirement that clinical studies be performed by parties

independent of device manufacturers. In fact, FDA routinely receives

and relies upon studies performed by manufacturers.

3. Several comments contended that financial conflicts of interest

were present in the three sources of data relied on by FDA to support

the classification/reclassification of pedicle screw spinal systems.

The comments claimed that, in the meta-analysis of the literature, the

authors of the individual articles had financial conflicts of interest

due to their relationships with device manufacturers. With respect to

the analysis of the reformatted IDE data and the Cohort study, the

comment stated that the IDE investigators and Cohort study participants

had significant financial interests in the companies whose devices they

were using and, therefore, had a strong financial incentive to report

only successful results. Similar objections were raised about the

authors of the 206 articles cited as constituting the body of medical

literature bearing on pedicle screw fixation. The comments stated that

almost all of the surgeons who authored these articles failed to

disclose their financial connections to manufacturers. The comments

stated that such interests raise serious concerns about researchers'

motivation to perform the research, the propriety and importance of

research questions and research designs, the adequacy of protection of

human subjects, lack of bias, and veracity in collecting and analyzing

the data and reporting the results.

FDA recognizes that some of the clinical investigators involved in

the three sources of data, as well as some of the authors of the 206

literature articles used to support classification and reclassification

of pedicle screw spinal systems, had financial interests in the devices

they were studying. FDA disagrees, however, that these financial

interests resulted in biased or unreliable data. Regardless of the

source of the data, the meta-analysis, the reformatted IDE data, the

Cohort study, or the collection of cited literature, the conclusions

were similar, i.e., that pedicle screw spinal systems are safe and

effective for the uses examined. Because of this, even if financial

conflicts of interest were present, they did not affect the resulting

data and the conclusions. Moreover, the agency has concluded that,

despite the failure to disclose the financial interests of clinical

investigators, the sponsors of these investigations and/or articles

took reasonable steps to minimize potential bias.

Furthermore, the fact that some spine surgeons were compensated by

industry for research or consulting services, or were reimbursed for

expenses incurred in connection with continuing medical education

courses, did not affect the validity of any of the data. Moreover, many

of the grants to support research were made directly to university

accounts for general research and development, not directly to

individual investigators. Consequently, the existence of a financial

relationship between some surgeons and manufacturers did not

necessarily result in biased case selection or reporting. Finally, FDA

notes that research used to support a medical device marketing

application has always been supported by the sponsor of the device and

there is neither an expectation of nonsupport nor a requirement of

disclosure of such support.

4. Several comments stated that pedicle screw spinal systems

present different safety and effectiveness issues than do either class

II spinal devices using hooks and/or wires or noninstrumented spinal

fusions. One comment identified the following areas of concern as

having the potential of presenting unreasonable danger for patients:

(1) Difficulty in placing screws completely within the walls of the

pedicle;

(2) Inability to determine screw placement postoperatively using

radiographic techniques;

(3) Damage to nerve tissue as a result of transient contact with a

screw during screw placement;

(4) Nerve root damage (irritation or compression) as a result of

screw malposition;

(5) Device failure;

(6) Loss of bone density as a result of stress shielding;

(7) Foreign body tissue response;

(8) Crevice corrosion;

(9) Fretting corrosion;

(10) Fibrosis;

(11) Bone fracture, particularly that of the pedicles;

(12) Nerve root or spinal cord compression as a result of fibrosis

or foreign body tissue response;

(13) Chronic irritation;

(14) Spine destabilization possibly leading to nonunion;

(15) Increased venous pressure as a result of blocked venous

channels within the bone;

(16) Increased risk of infection;

(17) Loss or decrease of sensory and/or motor function;

(18) Loss of bowel or bladder control; and

(19) Loss of sexual function.

FDA agrees that pedicle screw spinal systems have some potential

risks that

[[Page 40031]]

are different from those of other class II spinal devices. However, the

majority of the potential risks presented by these devices, e.g., bone

fracture, foreign body tissue response, loss or decrease in sensory

and/or motor function, and device failure or corrosion, are also

associated with class II spinal devices which use hooks and/or wires

for the same intended uses. Similarly, potential risks such as nonunion

and instability are also associated with noninstrumented spinal

fusions. Moreover, as described in the proposed rule, the incidence of

these adverse outcomes is no greater when a pedicle screw spinal system

is used than when other types of spinal fusions, instrumented and

noninstrumented, are performed in appropriately selected patients (60

FR 51946 at 51957). Finally, FDA believes that the potential risks that

are unique to pedicle screw spinal systems, e.g., difficulty in placing

screws completely within the walls of the pedicle, inability to

determine screw placement postoperatively using radiographic

techniques, damage to nerve tissue as a result of transient contact

with a screw during screw placement, and nerve root damage (irritation

or compression) as a result of screw malposition, can be adequately

addressed by the identified special controls and proper surgeon

training and surgical technique.

5. One comment asserted that the supposed advantages of pedicle

screw spinal systems are largely theoretical. The comment stated that,

while some investigators have shown that instrumented fusions increase

the likelihood of obtaining a solid fusion, others have demonstrated

that there is no significant increase in fusion rates performed with

instrumentation as compared with noninstrumented fusions performed with

bone graft alone.

FDA agrees that the data do not always support the theoretical

advantages of using pedicle screw spinal systems compared to alternate

methods of achieving spinal fusion. However, in forming its

recommendations, neither FDA nor the Panel is required to analyze the

theoretical behavior of a given device. It is only required to

determine whether the data demonstrate that there is a reasonable

assurance of safety and effectiveness for its intended uses.

6. The same comment stated that spinal fusion surgery is usually

performed because of the belief that spinal instability results in

pain. The clinical indicators used to determine which patients have

spinal instability and, therefore, are candidates for spinal fusion

surgery, are not clearly defined and are often not measurable. Because

the results of spinal fusion surgery are also dependent on measurements

of instability, a determination of success is difficult, if not

impossible.

FDA agrees that the methods used to measure instability are not

perfect and that several working definitions of instability exist.

Nevertheless, instability is measurable. In addition, the same

instability definitions and measurement techniques are used in

determining how a patient is to be treated, i.e., with pedicle screw

spinal systems, class II spinal devices using hooks and/or wires, or

noninstrumented fusions. FDA agrees that the determination of success

of spinal fusion surgery is often difficult, but disagrees that it is

impossible to determine. In fact, the success rates from using the

three treatment methods described above have been determined and found

to be reasonably equivalent (60 FR 51946 at 51954).

7. Three comments questioned the most appropriate classification

for pedicle screw spinal systems. One comment proposed that pedicle

screw spinal systems be classified into class I and two comments

suggested placing them in class III.

FDA disagrees. Based on the available information, both the Panel

and FDA concluded that general controls alone are not sufficient to

provide a reasonable assurance of the safety and effectiveness of

pedicle screw spinal systems intended to provide immobilization and

stabilization of spinal segments as an adjunct to fusion in the

treatment of the following acute and chronic instabilities or

deformities of the thoracic, lumbar, and sacral spine: Degenerative

spondylolisthesis with objective evidence of neurologic impairment,

fractures, dislocations, scoliosis, kyphosis, spinal tumors, and failed

previous fusion attempts (pseudarthrosis). The Panel and FDA also

concluded that premarket approval was not necessary to provide such

assurance. Both the Panel and FDA identified the potential risks

associated with the use of these devices and concluded that sufficient

information exists to establish special controls to provide reasonable

assurance of their safety and effectiveness. As a result, FDA is

classifying and reclassifying these devices in class II.

8. A comment believed that classification and reclassification of

pedicle screw spinal systems into class II is inappropriate because FDA

was correct in its prior determination that basic principles of

physiology, anatomy, biology, and biomechanical engineering demonstrate

that pedicle screw spinal systems present a serious risk of injury to

the spinal nerves, nerve roots, and surrounding vascular structures,

and increase the risk of pseudarthrosis. According to this comment,

these risks are not posed by existing spine fusion technology and

pedicle screw spinal systems are of questionable efficacy in comparison

to existing methodologies of treatment.

FDA disagrees. FDA did not determine that basic principles of

physiology, anatomy, biology, and biomechanical engineering demonstrate

that pedicle screw spinal systems present a serious risk of injury.

Rather, in 1984, FDA determined that a multiple component device system

intended for attachment to the spine via the pedicles was not

substantially equivalent to any legally marketed predicate device, in

accordance with section 513(i)(1) of the act. FDA's decision was based

on the fact that: (1) The sponsor did not identify a legally marketed

preamendments device incorporating pedicle screw components and (2) the

device posed potential risks not exhibited by other legally marketed

predicate spinal fixation systems, such as a greater chance of

neurological deficit due to imprecise screw placement or the event of a

screw failure; pedicle fracture during placement of screws; soft tissue

damage or inadequate fusion due to bending or fracture of device

components; and greater risk of pseudarthrosis due to instability of

the device design (60 FR 51946 at 51947). As stated previously, FDA

believes that the risks to health presented by pedicle screw spinal

systems intended to provide immobilization and stabilization of spinal

segments as an adjunct to fusion in the treatment of the following

acute and chronic instabilities or deformities of the thoracic, lumbar,

and sacral spine: Degenerative spondylolisthesis with objective

evidence of neurologic impairment, fractures, dislocations, scoliosis,

kyphosis, spinal tumors, and failed previous fusion (pseudarthrosis)

can be adequately addressed by special controls. Consequently, FDA is

classifying and reclassifying these devices into class II.

9. One comment argued that manufacturers of pedicle screw spinal

systems are seeking to have FDA down classify the device into class II

because the manufacturers are unable to prove that pedicle screws are

safe and effective for posterior implantation into the spine.

FDA disagrees. First, contrary to the comment's statement, this

classification and reclassification proceeding was

[[Page 40032]]

initiated by FDA; it is not in response to a petition for

reclassification. Second, under section 513 of the act, devices are

classified/reclassified into one of three classes based on reasonable

assurance, not ``absolute proof,'' of their safety and effectiveness.

Contrary to the comment's statement, it was not pedicle screw spinal

system manufacturers, but rather the Panel and FDA, that concluded that

pedicle screw spinal systems should be classified and reclassified into

class II because they determined that premarket approval is not

necessary to provide reasonable assurance of the device's safety and

effectiveness, general controls alone are insufficient to provide such

assurance, and there is sufficient information to establish special

controls to provide such assurance.

10. According to another comment, by classifying and reclassifying

pedicle screw spinal systems into class II, FDA is acknowledging that

there is no need for the manufacturers of pedicle screw spinal systems

to prove that the devices are safe and effective.

FDA agrees. The agency has determined that sufficient information

exists to establish special controls to provide reasonable assurance of

the safety and effectiveness of pedicle screw spinal systems intended

to provide immobilization and stabilization of spinal segments in

skeletally mature patients as an adjunct to fusion in the treatment of

the following acute and chronic instabilities or deformities of the

thoracic, lumbar, and sacral spine: Degenerative spondylolisthesis with

objective evidence of neurologic impairment, fracture, dislocation,

scoliosis, kyphosis, spinal tumor, and failed previous fusion

(pseudarthrosis). FDA has determined that premarket approval is not

necessary to provide such assurance.

B. Issues Relating to Information Published in the 1994 Supplementary

Issue of the Journal Spine (vol. 20S, 1994)

11. One comment objected that the manner in which the Scientific

Committee communicated to the public the results of the Cohort study

and related meta-analyses of the literature lacked scientific

integrity. According to the comment, the articles were not peer

reviewed, but rather they were accepted for publication solely by the

Editor-in-Chief of the peer-reviewed journal Spine. The comment

contended that publication of the articles without peer review

prevented the studies from being submitted to the usual critical

scrutiny of any peer review in the future.

While the articles describing the Cohort study and related meta-

analysis were not peer-reviewed in the usual manner, they were

subjected to a review process and published in an October 19, 1994,

Special Supplement of Spine. The editorial at the beginning of the

supplement states that,

The members of the Scientific Committee and editors of Spine

felt it important that presentations from the (July 1994, Panel)

meeting be available to the readers of Spine in an expedited manner.

The articles have been reviewed by the Scientific Committee, but

have not gone through the normal review process of the Spine

Editorial Board. However, it has been prepared, written, re-written,

and critiqued by all members of the Scientific Committee and member

of the Spine Editorial Board, as well as presented in an open public

forum to the scientists who comprised the Orthopedic and

Rehabilitation Devices Advisory Panel to the FDA. Weaknesses and

strengths of the studies are readily apparent and have been

addressed by each author, as well as in my summation.

12. The comment also stated that the articles should not have been

accepted for publication because the editorial policy of the journal

requires that the recommended minimum followup period for studies

should be 24 months.

FDA disagrees. Under Spine policy, a sufficient length of time for

followup of articles is necessary for publication. While the

recommended time period for surgical procedures is 2 years, the policy

does not state that studies with less than 2-year followup will not be

published.

C. Issues Relating to the January 1995, 510(k) Substantial Equivalence

Determination for a Pedicle Screw Spinal System Intended for Severe

Spondylolisthesis

13. A comment stated that the circumstances surrounding the first

510(k) clearance of a pedicle screw spinal system in January 1995, were

highly suspect because, until that time, FDA consistently had found

bone screws for use in the pedicles to be not substantially equivalent

to the identified predicate device, the lag screws used by Dr.

Harrington. The comment also stated that the lag screws were

manufactured as a custom device and used under a funded research grant

and, therefore, were not in commercial distribution prior to 1976.

FDA disagrees. The 510(k) applicant provided new evidence

documenting, for the first time, that: (1) A medical device company had

manufactured and shipped in interstate commerce bone (lag) screws

intended for use in the pedicles of the spine prior to May 28, 1976;

(2) the devices were marketed to physicians, including, but not limited

to, Dr. Harrington; and (3) the devices were not used solely for

research purposes.

14. The same comment also argued that the two devices had different

technological characteristics because the lag screws attach to fixation

constructs by wires whereas the pedicle screws attach directly to

plates or rods. The comment concluded that the applicant could not

demonstrate that its device did not raise different questions of safety

and effectiveness compared to the predicate device because the lag

screws were used on an extremely limited basis and were abandoned

because of a lack of effectiveness.

FDA disagrees. The presence of technological differences does not

preclude a finding of substantial equivalence under section 513(i) of

the act. In accordance with section 513(i)(1)(A)) of the act and

Sec. 807.100(b)(2)(ii)(B), for purposes of determining substantial

equivalence, manufacturers have to demonstrate that their device (1)

Has the same intended use as a predicate device and (2) if it has

different technological characteristics than the predicate device, that

the device is as safe and as effective as a legally marketed device,

and it does not raise different questions of safety and effectiveness.

The relative extent of use of one device compared to another is not

relevant.

In making its decision, FDA analyzed all of the data provided by

the sponsor. This included reports describing the clinical and

mechanical behavior of the device, in addition to affidavits. From

these data, the Panel and FDA determined that the complications were

similar to those of a predicate device and that the technological

differences raised no new questions relating to safety or

effectiveness.

15. The comment also stated that FDA's reversal of its position

with regard to the preamendments status of pedicle fixation devices was

insupportable and a clear violation of its own regulations.

Specifically, the comment stated that the agency took the unprecedented

step of determining the existence of commercial distribution based

solely on the affidavit of a former employee of a pedicle screw

manufacturer. According to the comment, this was not sufficient

evidence to demonstrate that the device was in commercial distribution

prior to 1976.

The use of affidavits to document the preamendments status of a

predicate device is not unprecedented. In fact, FDA routinely allows

affidavits to be used to document the preamendments status of a device.

FDA recognizes that obtaining labeling, advertising, and

[[Page 40033]]

other records concerning the marketing status of a device dating back

more than 20 years is often difficult, if not impossible. Therefore,

FDA allows sponsors to rely on alternative methods to demonstrate

interstate commerce. Moreover, contrary to the comment's statement, the

preamendments status of the device was established by much more than a

single affidavit. In fact, the 510(k) submission contained several

affidavits from individuals other than the sponsor, correspondence, and

other documents, e.g., shipping documentation, that demonstrated the

preamendments status of the Harrington lag screws for use in a limited

area of the spine, i.e., L5-S1, and for a

particular indication, i.e., severe spondylolisthesis.

16. Finally, the comment alleged that FDA changed its regulatory

position regarding pedicle screw spinal systems after it made a

``deal'' with the affected industry on or about June 15, 1993. The

comment stated that, if manufacturers funded a retrospective study, FDA

provided assurances that it would (1) Refrain from taking criminal,

regulatory, or other legal actions against them; and (2) reclassify

pedicle screw spinal systems without requiring prospective studies and

without regard to the quality of any of the retrospective data.

FDA disagrees. Prior to its January 1995, 510(k) decision and the

publication of this classification and reclassification regulation, FDA

consistently maintained that pedicle screw spinal systems, except when

intended for a very limited use, were class III devices requiring

premarket approval. The purpose of FDA's meeting with the affected

industry and the orthopaedic professional societies was to request that

these groups submit to the agency all available clinical data on the

performance of pedicle screw spinal systems. FDA, at no time, agreed to

change the regulatory status of these devices without regard to the

quality of the data or to refrain from taking regulatory action if a

retrospective study were funded.

D. Issues Relating to Misstatements or False Statements Appearing in

the Proposed Rule

17. One comment alleged that the statement in the preamble to the

proposed rule regarding the conclusion of the August 20, 1993, Panel

meeting, i.e., that pedicle screw spinal systems appear to be safe and

effective when used as adjuncts to spinal fusion procedures, was

inaccurate.

FDA disagrees. The description of the August 20, 1993, Panel

meeting contained in the preamble to the proposed rule states that the

Panel concluded that mechanical testing data demonstrated that pedicle

screw spinal systems exhibit adequate mechanical strength, rigidity,

and fatigue resistance (60 FR 51946 at 51948).

18. The same comment alleged that neither the transcripts from the

two Panel meetings, nor the summary in the preamble to the proposed

rule accurately reflected the Panel's conclusions regarding potential

risks to health associated with the use of the pedicle screw spinal

system, special controls, development of performance standards,

mechanical performance of the device, and the Panel members' own

personal knowledge of, and clinical experience with, the device.

FDA disagrees that the transcripts of the two Panel meetings did

not accurately reflect the Panel's conclusions. The proceedings from

the two meetings were verbatim stenographic transcripts of oral

testimony prepared by an independent transcriptionist. FDA also

disagrees that the preamble to the proposed rule did not accurately

reflect the Panel's conclusions. The preamble to the proposed rule

mirrors the transcripts of the meetings.

19. The same comment alleged that the Panel members (voting members

and voting/nonvoting consultants), who met July 23, 1994, had

inappropriate relationships, e.g., financial arrangements and ex parte

communications, with pedicle screw spinal system manufacturers and had

participated substantially in the design of the Cohort study, thereby

compromising their impartiality.

FDA disagrees in part. While it is expected that Panel members, who

are experts in a given field, will often have some financial interests

related to that field (e.g., certain arrangements with a manufacturer

(designing a device sold by a particular manufacturer; serving as a

consultant to a manufacturer; or receiving funding, directly or

indirectly, for research), the required FDA conflict-of-interest

questionnaire (FDA From 2725a) enables FDA to identify conflicts-of-

interest with a device or manufacturer that all substantial and/or

material to the subject of a particular Panel meeting, and thereby

facilitates the disclosure and possible waived for the Panel member(s)

in order to permit their participation in Panel deliberations.

FDA performed an internal affairs investigation of the Panel

members regarding conflicts and ex parte communications . The agency

reviewed whether the Panel was properly constituted. Investigation of

alleged undisclosed and unwaived conflicts of interest held by Panel

members found minor disparities and reporting omissions for two voting

Panel members and one nonvoting consultant. The agency has concluded

these disparities and omissions were insignificant and did not

constitute financial conflicts of interest that would credibly

influence their recommendations.

The agency has found that one other voting Panel member had

significant undisclosed financial conflicts. However, because the

recommendation of the Panel, both in the July 23 meeting and on the

subsequent homework assignment, was unanimous and this individual was

not controlling, of or unduly influential of, the votes of the other

Panel members and was not necessary to constitute a quorum, after

expunging the participation of this Panel member, FDA has concluded

that this Panel, both in the meeting and on the subsequent homework

assignment, was a valid scientific Panel for purposes of making

recommendations regarding classification and reclassification.

E. Issues Relating to FDA's Issuance of Regulations

20. One comment argued that, in issuing a classification

regulation, FDA may not rely on a scientific study unless it makes

publicly available all study data, as well as the identities of the

persons who furnished the data. The comment cited 21 CFR 10.20(j),

20.63, and 860.5 as authority. In addition, the comment objected that

FDA refused to disclose the identities of the physician-investigators

who contributed data to the Cohort study, did not disclose the

reformatted IDE analysis, the IDE data, or internal information bearing

on the reliability of such data.

FDA disagrees. Although the agency did not disclose the raw IDE or

the Cohort study data, or the identities of the clinical investigators

who furnished such data to the agency, FDA did provide a detailed

analysis of the Cohort Study, the clinical data released by the IDE

sponsors, and the meta-analysis (60 FR 51946 at 51960-51962; refs. 51,

65, 66, 119, and 201). FDA believes these publicly available data not

only satisfy the requirements under the statute, but provide the public

with at least the level of detailed information as that usually

available from published reports regularly relied upon to support

classification and reclassification.

F. Response to Comments Which Contained Clinical Data

21. Several comments provided clinical information to support the

comment's position on the proposed

[[Page 40034]]

rule. The submitted clinical information consisted of literature

articles describing clinical trials and two questionnaires, a surgeon/

patient questionnaire and a lawyer/client questionnaire. The surgeon/

patient questionnaire provided mixed results, i.e., some patients were

satisfied with their clinical results and others were not satisfied,

whereas the lawyer/client questionnaire provided only negative results,

i.e., all clients were dissatisfied with their results.

The majority of the articles submitted or referenced in these

comments were already reviewed by the Panel and used as part of the

basis for their recommendation to classify and reclassify pedicle screw

spinal systems into class II. The remainder of these articles were not

reviewed by the Panel because they were published after the July 1994,

Panel meeting. As described in section V.M of this document, these

articles did not raise new issues or concerns relating to the safety or

effectiveness of pedicle screw spinal systems. Because of the inherent

bias present in the questionnaires, e.g., the total number of

questionnaires sent to patients/clients in relation to the number

returned and the number included as part of the comment are unknown,

the data cannot be used in analyzing the success rate of pedicle screw

spinal systems. These data can be used, however, as part of an analysis

of the complications. As such, the questionnaires did not describe any

complications or raise any issues that had not already been reviewed by

the Panel and FDA in making their determinations with respect to the

classification and reclassification of pedicle screw spinal systems.

G. Requests for Additional Pedicle Screw Clinical Trials and Data

Analyses

22. Ten comments requested that FDA require submission of

additional data before finalizing the classification and

reclassification of pedicle screw spinal systems. The comments

recommended that the following types of data be required: Studies to

analyze the long-term effects of the device, continuing evaluations,

collections of data using a recommended data report form for obtaining

data directly from patients rather than from their surgeons, studies

similar to the Cohort study but with larger sample sizes, comprehensive

reviews of the literature, and comprehensive reviews of all data. In

addition, one comment suggested that FDA was reclassifying these

devices without reviewing clinical trial data documenting their safety

and effectiveness.

FDA disagrees. As previously explained, under section 513 of the

act, devices are classified and reclassified into one of three classes

based on reasonable assurance, not absolute proof, of their safety and

effectiveness. The Panel recommended, and FDA concurred, that pedicle

screw spinal systems intended to provide immobilization and

stabilization of spinal segments in skeletally mature patients as an

adjunct to fusion in the treatment of degenerative spondylolisthesis

and spinal fractures be classified and reclassified into class II

because they determined that premarket approval is not necessary to

provide reasonable assurance of safety and effectiveness; general

controls alone are insufficient to provide reasonable assurance of the

device's safety and effectiveness; and there is sufficient information

to establish special controls to provide such assurance. FDA also

determined that, when intended to provide immobilization and

stabilization of spinal segments in skeletally mature patients as an

adjunct to fusion in the treatment of the following acute and chronic

instabilities or deformities of the thoracic, lumbar, and sacral spine:

Degenerative spondylolisthesis with objective evidence of neurologic

impairment, fractures, dislocations, scoliosis, kyphosis, spinal

tumors, and failed previous fusion attempts (pseudarthrosis), special

controls would provide a reasonable assurance of safety and

effectiveness. The Panel and FDA reached these conclusions only after

considering a substantial amount of valid scientific evidence. As

described previously, this valid scientific evidence consisted of

clinical data collected from three sources--data from IDE's (the

reformatted IDE data), data from the literature (the meta-analysis),

and data collected directly from surgeons (the Cohort study). The IDE

data was prospective clinical data collected under the protocols of

FDA-approved clinical trials. The meta-analysis was retrospective

clinical data published in peer-reviewed literature. The Cohort study

consisted of retrospective nationwide clinical data collected from

surgeons of various experience levels from a patient population that

was homogeneous in terms of diagnosis, but mixed in terms of severity

of disease. In addition to these sources of clinical data, MDR and

MedWatch reports were analyzed for device problems. FDA does not

believe that it is necessary to require submission of additional data,

to conduct additional studies, or to re-review the literature before

classifying and reclassifying these devices. FDA does agree, however,

that the longer-term performance of these devices is not fully

characterized. For this reason, postmarket surveillance (PMS) studies

will be required.

H. Issues Relating to Indications for Use

Over 200 comments addressed the various intended uses of pedicle

screw spinal systems.

23. Twenty-three comments questioned FDA's authority to regulate

the indications for use of medical devices. They believed that,

although restrictions on the use of pedicle screw spinal systems may be

appropriate, this aspect of medical device regulation is outside the

scope of FDA's authority and should be decided by professional

societies, peer review groups, credentialing organizations, and

hospitals. One comment stated that FDA should regulate the safety of

medical devices only for certain indications. Several other comments

stated that there should be no restrictions on the use of pedicle screw

spinal systems. All of these comments argued that FDA's actions

interfered with the practice of medicine.

FDA disagrees. In determining whether or not a device is safe and

effective, FDA first considers the intended uses for the device. Spinal

fusion is not a medical indication but a treatment option which can be

approached in a variety of ways. It is one of the desired outcomes from

using pedicle screw spinal systems. FDA recognizes, however, that

fusion in and of itself is not what patients with spinal disease are

seeking. They wish to be relieved of their symptoms, have their

objective impairment alleviated, and avoid more symptomatic or

functional impairment. Devices that share the same outcome for a given

condition do not necessarily share the same benefits and risks. One of

the aspects in determining if a device may be legally marketed is

deciding, based on the available data, what the appropriate indications

are. A device may be an appropriate treatment for one indication, but

not for another. In addition, to understand the evidence supporting a

device's safety and effectiveness, a distinct medical condition

requiring treatment must be identified. In reviewing the valid

scientific evidence, the Panel recommended and FDA found that the use

of pedicle screw spinal systems were safe and effective only for

certain indications. The valid scientific evidence did not support

unrestricted use of the device.

In determining the safety and effectiveness of a device for the

purpose

[[Page 40035]]

of classification or reclassification, both the Panel and the agency

are to consider the persons for whose use the device is represented or

intended, the conditions of use for the device, and the probable

benefit to health from the use of the device weighed against any

probable injury or illness from such use (Sec. 860.7(b)). The device is

to be considered, not in a vacuum, but rather in the context of the

patient population for whose use it is intended. Accordingly, there is

reasonable assurance that a device is safe when it can be determined

that the probable benefits to health from use of the device for its

intended uses and conditions of use outweigh any probable risks

(Sec. 860.7(d)(1)). The benefits and risks to health presented by a

device depend, in large part, on the specific use for which the device

is intended. There may be reasonable assurance that a device is safe

for some, but not other, uses. Similarly, there is reasonable assurance

that a device is effective when it can be determined that, ``in a

significant portion of the target population,'' the use of the device

for its intended uses and conditions of use will provide clinically

significant results (Sec. 860.7(e)(1) (emphasis added)). It is clear,

then, that when making determinations regarding the classification or

reclassification of a device, it is appropriate for the agency to

consider the specific intended uses of a device, including the specific

patient populations for which it is intended. Consequently, the agency

disagrees that it does not have authority to regulate the indications

for use for pedicle screws and that it is interfering with the practice

of medicine.

24. One comment objected that FDA's proposed reclassification

improperly exceeded the recommendations of the Panel.

The Panel determined that the evidence demonstrated a reasonable

assurance of safety and effectiveness of pedicle screw spinal systems

intended for two severe and diagnostically distinct indications--

fracture and degenerative spondylolisthesis. Accordingly, the Panel

recommended that the device be classified and reclassified into class

II only when intended for these uses. FDA proposed that the device also

be classified and reclassified into class II when intended for the

following acute and chronic mechanical instabilities or deformities of

the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis

with objective evidence of neurologic impairment, fractures,

dislocations, scoliosis, kyphosis, spinal tumors, and failed previous

fusion (pseudarthrosis). FDA disagrees that it exceeded its authority.

21 CFR 860.3(h) defines a classification panel as an advisory committee

established by the Commissioner for the purpose of making

``recommendations'' (emphasis added) to the Commissioner on the

classification/reclassification of devices. These recommendations are

designed to assist the Commissioner in the proper classification and/or

reclassification of a device. While FDA usually follows a Panel's

recommendations, it is not required to do so.

As stated in the preamble to the proposed rule, FDA believes that

sufficient clinical data exist to classify and reclassify into class II

pedicle screw spinal systems intended for degenerative

spondylolisthesis with objective evidence of neurologic impairment,

fractures, dislocations, scoliosis, kyphosis, spinal tumors, and failed

previous fusion (pseudarthrosis). The medical literature and data from

IDE clinical investigations provide adequate evidence that the device

can safely and effectively stabilize the spine and maintain spinal

alignment while fusion takes place. The risks associated with the use

of pedicle screw spinal systems intended to provide immobilization and

stabilization of spinal segments as an adjunct to fusion in the

treatment of these acute and chronic instabilities or deformities of

the thoracic, lumbar, and sacral spine are similar to those associated

with other class II spinal implant devices, such as those classified in

Sec. 888.3050 (21 CFR 888.3050) (60 FR 51946 at 51956).

25. Several comments advocated classifying and reclassifying into

class II pedicle screw spinal systems intended for additional uses,

including degenerative disc disease, degenerative deformities,

stenosis, iatrogenic instability and previous multiple laminectomies,

facet joint disease, pseudospondylolisthesis, low back pain, disc

herniation, arthritis, and osteomyelitis.

FDA believes that valid scientific evidence does not currently

exist to support classifying and reclassifying into class II pedicle

screw spinal systems when intended for the indications listed above.

Neither the literature nor the clinical data establish the safe and

effective use of pedicle screw spinal systems for degenerative disc

disease, degenerative deformities, stenosis, iatrogenic instability and

previous multiple laminectomies, facet joint disease,

pseudospondylolisthesis, low back pain, disc herniation, arthritis, or

osteomyelitis. As stated in the preamble to the proposed rule, FDA has

determined that, when intended for use in conditions not categorized as

acute or chronic instabilities or deformities of the thoracic, lumbar,

and sacral spine, premarket approval is necessary to ensure the safety

and effectiveness of the device (60 FR 51946 at 51957). FDA-approved

clinical trials for some of these indications are ongoing. When data

from these or other studies become available for any of the indications

described above, they may be submitted in either an application for

premarket approval or reclassification petition.

26. Eight comments advocated adding specific pediatric indications

and one comment advocated adding general pediatric use to the list of

indications. The specific indications included myelodysplasia, spina

bifida, cerebral palsy, muscular dystrophy, myelomeningocele, and

congenital subluxation.

FDA disagrees. As stated previously, all valid scientific evidence

reviewed by the Panel and FDA were obtained from skeletally mature

populations. To date, the safety and effectiveness of pedicle screw

spinal systems in pediatric populations have not been demonstrated.

Therefore, this patient population is excluded from this classification

and reclassification. When intended for use in pediatric populations,

pedicle screw spinal systems are considered postamendments class III

devices for which premarket approval is required.

27. Several comments addressed ways in which FDA should further

limit the indications for use of pedicle screw spinal systems, such as

by including specific patient evaluation criteria or by specifying the

severity of the condition.

FDA disagrees that these actions are necessary. FDA classifies

devices based upon, among other things, patient selection, not

individual patient management. FDA notes that it is the responsibility

of individual surgeons to determine the appropriateness of using a

specific medical device for a given patient.

28. Four comments stated that pedicle screw spinal systems should

not be allowed on the market for any use. Another comment requested

that an additional Panel meeting be convened to discuss further

restricting the intended uses of pedicle screw spinal systems.

FDA disagrees. After reviewing all available data and information,

FDA believes that there is reasonable assurance that pedicle screw

spinal systems are safe and effective for certain intended uses. FDA

does not believe that pedicle screw spinal systems present a

substantial deception or an

[[Page 40036]]

unreasonable and substantial risk of illness or injury. Consequently,

FDA does not believe it would be appropriate to ban them under section

516 of the act (21 U.S.C. 360f).

FDA also disagrees that an additional Panel meeting is necessary

because the relevant available data have been reviewed.

I. Issues Relating to Special Controls

29. One comment asserted that PMS studies cannot legally be

required for pedicle screw spinal systems because the devices are not

intended for use in supporting or sustaining life and pose risks no

different from those associated with the use of other preamendments

class II spinal fixation devices.

FDA disagrees. Under section 522 of the act (21 U.S.C. 360l),

postmarket surveillance is required for certain devices and may be

required for any device for which FDA determines that it is necessary

to protect the public health or to provide safety or effectiveness data

for the device. FDA has determined that PMS studies are necessary to

provide longer-term data on the safety and effectiveness of pedicle

screw spinal systems.

Although originally proposed as a special control, FDA has

determined that PMS studies are best imposed by order in the

substantial equivalence determination letter for each device. This will

preserve the discretionary nature of the PMS studies and will allow the

agency to more easily remove the requirement once it determines that

these studies are no longer necessary to assure the safety and

effectiveness of pedicle screw spinal systems. The final regulation has

been modified to reflect that PMS studies are no longer one of the

special controls for these devices.

30. One comment stated that PMS studies are appropriate only for

devices cleared for marketing with limited clinical performance data.

The comment noted that there now exists a vast amount of clinical

information gained from use of pedicle screw spinal systems in several

thousand patients. The comment also noted that, based on these data,

the Panel concluded that, with respect to safety and effectiveness,

these devices are comparable to, or better than, currently available

spinal systems. The comment concluded that this clinical information

and the conclusions drawn from this information provide sufficient

clinical data to adequately identify and characterize the performance

of pedicle screw spinal systems and the issues pertinent to safety and

effectiveness, thereby obviating the need to conduct PMS studies.

FDA disagrees that PMS studies are appropriate only for devices

cleared for marketing with limited clinical performance data. Section

522 of the act allows FDA to require PMS studies for any device for

which it determines such studies would protect the public health or

provide safety or effectiveness data for the device. As stated in the

preamble to the proposed rule, FDA will require PMS studies in order to

address issues related to device specific design differences, surgical

techniques, and device usage (60 FR 51946 at 51955). Although there is

ample short-term clinical performance data for these devices, there

does not now exist sufficient longer-term, i.e., more than 24-month

followup, safety and/or effectiveness data regarding device specific

design differences, surgical techniques, and device usage.

31. A second comment noted that components used to construct

pedicle screw spinal systems could be identical to those used to

construct either spinal interlaminal fixation orthoses (Sec. 888.3050)

or spinal intervertebral body fixation orthoses (21 CFR 888.3060).

Because PMS studies are not required for these devices, they should not

be required for pedicle screw spinal systems. A third comment believed

that PMS studies are inappropriate for well-established, standard of

care treatments involving medical devices that were in existence prior

to the 1976 amendments, including pedicle screw spinal systems.

FDA disagrees that PMS studies are inappropriate for devices that

were in existence prior to the 1976 amendments. Section 522(a)(2) of

the act specifically authorizes FDA to require a manufacturer to

conduct PMS studies for any device, regardless of when it was first

introduced or delivered for introduction into interstate commerce, for

which FDA determines that PMS studies are necessary to protect the

public health or to provide safety or effectiveness data for the

device. Although, as the comment states, certain devices have been used

as pedicle screw spinal systems for some time, except for the limited

severe spondylolisthesis intended use available since January 1995,

pedicle screw spinal systems have not been legally marketed. Collection

of the PMS study data will allow FDA to analyze information on the use

of devices specifically intended, and legally marketed, for use as

pedicle screw spinal systems.

32. Five comments believed that PMS studies are unnecessary and

will not further protect the public health because one or more of the

following current reporting systems already provides adequate

information on the performance of pedicle screw spinal systems: (1) The

MDR System, (2) Voluntary Reporting under MedWatch, (3) User Reporting,

and (4) Complaint Handling under the current good manufacturing

practices. One comment supported a requirement that labeling remind

surgeons they are required to report certain events under MDR. Two

comments suggested that a statement which encourages health care

professionals to submit MDR's under the Voluntary MedWatch System be

placed in the required package insert of the device. Two other comments

noted that no other class II spinal implant device is subject to PMS

studies. Three comments also stated that collecting additional

information will increase health care costs.

FDA disagrees in part. The purposes of PMS studies and current

reporting systems are different. PMS studies are active investigations

of device performance during actual use, whereas other reporting

systems, i.e., MedWatch, MDR, User Reporting, and Complaint Reporting,

are passive reporting mechanisms. As such, these current reporting

systems would not provide the agency with clinical monitoring

information on pedicle screw spinal systems other than unexpected

problems in the marketplace. The PMS studies, in contrast, will provide

longer-term safety and effectiveness data for pedicle screw spinal

systems once the devices are distributed in the general population

under actual conditions of use. Finally, FDA is aware that PMS studies

might have an impact on health care costs. Although this is

unfortunate, the agency believes that it is necessary to impose this

requirement and collect this information in order to assure the safety

and effectiveness of pedicle screw spinal systems.

33. A comment suggested that, due to the litigious climate

surrounding these devices, it may be very difficult for manufacturers

to recruit surgeons to participate in PMS studies.

FDA recognizes the concern that there may be conditions which would

make the collection of the data somewhat difficult. However, FDA

believes that it is important that the data be obtained and that it is

possible to recruit a sufficient number of surgeons to participate in

PMS studies.

34. One comment stated that the proposed identification for pedicle

screw spinal systems was inaccurate, or at least misleading. The

comment noted that, as proposed, a pedicle screw spinal system assembly

must contain all of the components listed as part of the pedicle screw

spinal system. The comment stated that, for any given assembly,

[[Page 40037]]

some or all of the system components could be used.

FDA agrees in part. As proposed, the identification could be

interpreted to require that all of the described components were

necessary to construct a pedicle screw spinal system assembly. FDA has

amended the identification of the device to clarify that not all of the

described components are required to be used in a pedicle screw spinal

system assembly.

35. In the preamble to the proposed rule, FDA proposed two labeling

special controls. These controls described the intended uses and

indications for pedicle screw spinal systems and cautioned the user

about potential risks to health if the devices were used under certain

conditions. Three comments stated that the two labeling special

controls were incorrectly categorized as ``warnings'' according to

FDA's General Program Memorandum No. G91-1, ``Device Labeling

Guidance.'' They believed that these labeling requirements are more

appropriately described as ``precautions'' or ``important notes''

because they describe a particular patient population and not specific

risks or hazards associated with the use of a device. Four comments

objected that: (1) Use of the phrase ``* * * with significant potential

risk for serious injury to patients * * *'' in the second labeling

statement did not accurately reflect the data reviewed by the Panel to

make its recommendation,(2) references to training and experience

should not be part of the second labeling special control, and (3) the

controls containing the language referred to in (1) and (2) should be

removed or modified.

FDA agrees with the comments that the two labeling special controls

should be rewritten, but disagrees with the specific reasons. General

Program Memorandum No. G91-1 states that ``A warning is appropriate

when the device is commonly used for a disease or condition for which

there is a lack of valid scientific evidence of effectiveness for that

disease or condition and such usage is associated with a serious risk

to health * * * Include an appropriate warning if there is reasonable

evidence of an association of a serious hazard with the use of the

device. A causal relationship need not have been proved.'' This is the

case when pedicle screw spinal systems are used for indications other

than significant mechanical instabilities or deformities of the

thoracic, lumbar, and sacral spine. Because valid scientific evidence

is not available to support a determination that a reasonable assurance

exists that pedicle screw spinal systems are safe and effective for

other indications, categorizing the first labeling special control as a

``warning'' is the appropriate mechanism to alert users to the

potential for injury to a patient.

The second labeling special control does not warrant being

described as a ``warning'' because it does not meet the definition of

this term. It does not describe known serious adverse reactions or

known potential safety hazards; it does not provide specific steps to

be taken; it does not concern a use for which there is reasonable

evidence of association with a serious hazard. It does, however,

provide information on special care to be exercised by a practitioner,

although the need for special care is implied, not explicitly stated.

Accordingly, FDA concludes that it is more appropriately categorized as

a ``precaution''.

After reviewing the proposed special controls regarding labeling,

FDA has concluded that the information should be stated more clearly.

FDA believes that the labeling special controls reflect the data

reviewed by the Panel. FDA also believes that the labeling special

controls are necessary to provide reasonable assurance of the safety

and effectiveness of the devices. Finally, as described in the next

section, the intent of the second control was not to specify the type

of training that should be available or to suggest that FDA would

provide or approve any training. Rather, it was intended to alert

surgeons to the necessity of receiving appropriate training in the use

of specific pedicle screw spinal systems. Because of concerns with the

proposed wording, the labeling special controls have been modified to

read as follows:

``Warning: The safety and effectiveness of pedicle screw spinal

systems have been established only for spinal conditions with

significant mechanical instability or deformity requiring fusion

with instrumentation. These conditions are significant mechanical

instability secondary to degenerative spondylolisthesis with

objective evidence of neurologic impairment, fracture, dislocation,

scoliosis, kyphosis, spinal tumor, and failed previous fusion

(pseudarthrosis). The safety and effectiveness of these devices for

any other conditions are unknown.''

``Precaution: The implantation of pedicle screw spinal systems

should be performed only by experienced surgeons with specific

training in the use of this pedicle screw spinal system because this

is a technically demanding procedure presenting a risk of serious

injury to the patient.''

36. A number of comments stated that appropriate surgeon training

should be required prior to use of pedicle screw spinal systems and

that classification/reclassification into class II would make access to

training and device information easier. In addition, several comments

believed that professional societies and hospitals, not FDA or the

manufacturers, should determine what constitutes adequate training for

surgeons implanting pedicle screw spinal systems.

FDA agrees that it is important that surgeons who use pedicle screw

spinal systems have proper training prior to using the device. FDA does

not believe, however, that it should identify who is most qualified to

provide such training or determine what constitutes adequate training.

The precaution statement is intended to inform surgeons (and patients)

of the possible effect the device could have on the patient if the

surgeon implanting the device is not trained or experienced in the

proper use of pedicle screw spinal systems. This includes knowledge of

the indications, patient selection criteria, and appropriate surgical

techniques.

37. A comment questioned the proposed warning label because, in the

past, FDA has prohibited pedicle screw spinal system manufacturers from

supporting courses that described surgical techniques of ``off label''

uses demonstrating such uses or providing hands-on workshops to learn

such uses.

FDA disagrees. Previously, the agency issued several warning

letters to pedicle screw spinal system manufacturers for participating

in or supporting the training of practitioners in the use of long bone

screw, pedicle fixation because, at that time, no long bone screw

devices had received FDA clearance for use in the pedicles of the

spine. As a result, FDA considered such use ``off label.'' Because the

association with these training programs was considered the promotion

of an ``off-label'' use, the agency stated that the manufacturers had

misbranded and adulterated the long bone screws in accordance with

sections 501(f)(1)(B) and 502(o) of the act (21 U.S.C. 351(f)(1)(B) and

352(o)) and promotion of this use was considered a major modification

of the intended use, requiring a new premarket notification (510(k))

submission under Sec. 807.81(a)(3)(ii). The regulations and the act are

clear that manufacturers must have clearance for the intended use for

which their device(s) are promoted, advertised, or held for sale.

With the issuance of this final regulation, the agency now

encourages pedicle screw spinal system manufacturers to support

training for the class II intended uses. Such training, however, should

not be provided before FDA clearance is received. The above referenced

warning label will appear

[[Page 40038]]

only on devices that have been cleared for pedicle screw spinal

fixation.

38. The comment also claimed that the right to free speech

guaranteed by the First Amendment to the U.S. Constitution should not

be restricted by FDA's suppression of training for ``off label'' use.

FDA disagrees that its limitations on promotional training

conducted or sponsored by manufacturers for ``off label'' uses for

pedicle screw spinal systems violate the First Amendment. As described

above, the act requires that FDA regulate devices based on their

intended use. The term ``intended use'' is broadly defined and

encompasses the manner in which a company characterizes its product in

the marketplace. The intended use of a device refers to the objective

intent of the persons legally responsible for its labeling (Sec. 801.4

(21 CFR 801.4)). ``The intent is determined by such persons'

expressions or may be shown by the circumstances surrounding the

distribution of the article. This objective intent may, for example, be

shown by labeling claims, advertising matter, or oral or written

statements by such persons or their representatives.'' ( Sec. 801.4

(emphasis added)); (see e.g., Coyne Beahm, Inc. et al. v. United States

Food and Drug Administration, et al., 958 F. Supp. 1060 (M.D.N.C.

1997).) Consequently, oral statements and materials presented at

industry-supported training programs may provide evidence of a device's

intended use. If these statements or materials promote a use that has

not been approved by the agency, the device is misbranded under section

502(f)(1) of the act for failure to bear labeling with adequate

directions for all intended uses, and under section 502(o) of the act

because premarket notification was not provided as required under

section 510(k) of the act. The device is also adulterated under section

501(f) of the act for failure to have FDA approval. Thus, the various

means by which manufacturers and their representatives provide

information about their products to healthcare professionals and

consumers, including statements and materials presented at industry-

supported scientific and educational activities, directly bear on

whether a device is improperly promoted and, therefore, adulterated or

misbranded.

Because the regulation of devices is an area of extensive Federal

regulation, the agency may regulate the communications at industry-

supported scientific and educational activities without violating the

First Amendment. (Cf. SEC v. Wall Street Publishing Institute, Inc.,

851 F.2d 365 (D.C.Cir. 1988), cert. denied, 109 S.Ct. 1342 (1989).)

Moreover, to the extent that such communications constitute protected

speech, they are commercial speech and FDA's regulation of such

activities does not violate the First Amendment. (See Bolger v. Youngs

Drug Products, 103 S.Ct. 2875 (1983); S.U.N.Y. v. Fox, 109 S.Ct. 3028

(1989); Cincinnati v. Discovery Network, 113 S.Ct. 1505 (1993).)

Industry-supported scientific educational activities refer to a

specific product, are economically motivated, and propose a commercial

transaction. These programs are intended to convince the audience to

prescribe, purchase, or otherwise use the particular product.

The Supreme Court has afforded commercial speech limited

constitutional protection. (See, e.g., Virginia State Board of Pharmacy

v. Virginia Citizens Consumer Council, Inc., 96 S.Ct. 1817 (1976);

Central Hudson Gas & Electric Corp. v. Public Service Commission, 100

S. Ct. 2343 (1990).) In Central Hudson, the Supreme Court established a

four-prong test to determine whether limitations on commercial speech

are constitutional. The four prongs are: (1) Whether the speech

concerns lawful activity and is not misleading, (2) whether the

asserted government interest is substantial, (3) whether the limitation

directly advances the governmental interest asserted, and (4) whether

the limitation is not more extensive than is necessary to serve that

interest. The Court has clarified that the fourth prong requires that

the restriction be ``narrowly tailored'' to serve the asserted

government interest. Narrow tailoring means a fit that is reasonable.

(See S.U.N.Y. v. Fox, 109 S.Ct. 3028. 3035 (1989).)

FDA's regulation of industry-supported scientific and educational

activities satisfies all four prongs. First, as previously discussed,

industry-supported scientific and educational activities that promote

an unapproved device, or promote an approved device for an unapproved

use, create an unlawful product--a misbranded or adulterated device.

Accordingly, industry-supported activities that promote unlawful

products concern illegal activity and may be prohibited. Second, FDA's

limitations on promotional activities with respect to off label uses

serve the substantial government interest of protecting the public

health and safety by helping to ensure the dissemination of truthful

and nonmisleading information about devices. The Supreme Court has

repeatedly held that the government's ``interest in the health, safety,

and welfare of its citizens constitutes a substantial interest.''

(Posadas de Puerto Rico Associates v. Tourism Co., 106 S.Ct. 2968, 2977

(1986); Rubin v. Coors, 115 S.Ct. 1585, 1591 (1995).) The limitations

also serve the second substantial government interest of protecting the

public health by preserving the integrity of the premarket approval

process under which manufacturers are required to establish that their

devices are safe and effective for each of their intended uses before

they may be marketed and promoted for those uses. Third, FDA's

limitations on promotional activities with respect to off-label uses

directly advance the government's substantial interests in protecting

the public health and safety by helping to ensure the dissemination of

truthful and nonmisleading information about devices and by preserving

the integrity of the premarket approval process by dissuading

manufacturers from using such activities as a means to promote

unapproved products and unapproved uses, thereby encouraging scientific

research and avoiding unnecessary harm to patients. Finally, FDA's

limitations on industry-sponsored training sessions are narrowly

tailored and are a reasonable approach to protect the public health and

safety by discouraging the dissemination of misleading or biased

information, and by maintaining the integrity of the premarket approval

process. FDA's limitations apply only to industry-supported activities

that relate to the supporting company's device or to competing devices.

They are directed to the regulated sponsors of such activities, and do

not apply to participating professionals or independent scientists and

organizations.

39. Several comments believed that the device should be available

for use only by neurosurgeons or orthopaedic surgeons supervised by

neurosurgeons.

FDA disagrees. According to section 520(e)(1)(B), FDA may not

restrict access to medical devices based on specialty or board

certification.

J. Other Issues

40. Several comments objected that publication of the proposed rule

in the Federal Register was not appropriate because the general public

is not aware of the Federal Register. The comments noted that another

vehicle for disseminating the information would have been more

appropriate.

FDA disagrees. The act (sec. 513(d)(1) and 513(e)(1)) requires that

a proposed rule be published in the Federal Register as the formal

mechanism to provide all interested parties an opportunity to submit

comments when

[[Page 40039]]

an advisory panel recommends an initial classification or change in

classification for a medical device. Comments are invited from anyone.

FDA recognizes that other mechanisms for distribution of this type of

information is also appropriate. One of the alternate mechanisms

currently being tested is electronic publication on the World Wide Web.

41. Several comments objected to FDA's consideration of public

comments, which may contain only anecdotal information, in determining

the appropriate class for these devices.

FDA agrees that comments provided by the public may contain

anecdotal information that does not meet the definition of valid

scientific evidence. However, FDA considers this information along with

the information provided in other comments. These anecdotal comments

did not raise any issues or comments that were not already addressed by

the information that the Panel reviewed in making its determination

that safety and effectiveness of pedicle screw spinal systems could be

assured by special controls.

42. Six comments disapproved of the release of the PIN's which

identified the surgeons participating in the Cohort study.

FDA regrets any problems that may have been caused by this

inadvertent release of information. However, release of this

information did not affect the quality, integrity, or value of the data

upon which the Panel's recommendation was based.

43. A comment noted that there is no consensus among spine surgeons

that pedicle screw fixation has become the standard of care or the gold

standard for treatment of spinal instability so as to justify the

conclusion that the devices are safe and effective and to justify

abandonment of the randomized control trial in making such an

assessment.

FDA agrees that there is no consensus among spine surgeons

regarding pedicle screw spinal systems. However, a medical device does

not need to be viewed as the ``gold standard'' in order for the agency

to determine that there is reasonable assurance of its safety and

effectiveness. Nor is it a requirement for the classification and

reclassification process that all members of a medical specialty agree

that a particular device should be used under all conditions. It is

recognized that certain devices provide their best outcome when used

for specific indications. This is one of the reasons why degenerative

disc disease is not included as one of the intended uses in the

classification and reclassification of pedicle screw spinal systems.

Finally, as described above, randomized clinical trials are only one of

the types of valid scientific evidence upon which FDA may rely in

support of a classification/reclassification determination. Many IDE

studies from which the reformatted IDE data came are still being

actively pursued by their sponsors and the patients are being actively

followed.

K. Labeling of Bone Screws

44. A comment requested FDA to formally rescind its April 8, 1994,

and June 15, 1994, letters to manufacturers of bone screws and devices

classified under Secs. 888.3030 and 888.3040 (21 CFR 888.3030 and

888.3040), directing them to amend their labeling by including the

following: ``Warning: This device is not approved for screw attachment

or fixation to the posterior elements (pedicles) of the cervical,

thoracic, or lumbar spine.'' According to the comment, this labeling

requirement will become unnecessary when pedicle screw spinal systems

are classified into class II.

FDA disagrees. In this final rule, FDA is classifying and

reclassifying only pedicle screw spinal systems intended for screw

attachment or fixation to the pedicles of the thoracic, lumbar, or

sacral spine for immobilization and stabilization of spinal segments

for the treatment of significant medical instability or deformity

requiring fusion with instrumentation. This classification and

reclassification in no way affects devices classified as single/

multiple component metallic bone fixation appliances and accessories

(Sec. 888.3030) or smooth or threaded metallic bone fixation fasteners

(Sec. 888.3040). Those devices are still not approved for screw

attachment or fixation to the posterior elements (pedicles) of the

cervical, thoracic, or lumbar spine. Hence, the 1994 amended labeling

remains appropriate for these devices.

45. One comment noted that in January 1995, FDA began clearing

510(k)'s intended to treat grades 3 and 4 spondylolisthesis at the

L5-S1 junction. The comment concluded that, by

default, grades 1 and 2 spondylolisthesis, less severe conditions, are

considered to be postamendments intended uses resulting in the device

being automatically classified into class III. According to this

comment, this means that FDA, through required manufacturer labeling,

is instructing physicians to wait until grades 1 and 2

spondylolisthesis develop into grades 3 and 4 spondylolisthesis before

employing treatments utilizing pedicle screw spinal systems, which is

not in the patient's best interests.

FDA disagrees. FDA is not limiting physicians, through required

manufacturer labeling, to wait until grades 1 or 2 spondylolisthesis

develop into grades 3 or 4 spondylolisthesis. FDA is stating that the

preamendments documentation in the 510(k) described marketing of the

device only for the treatment of grades 3 and 4 spondylolisthesis at

L5-S1. Treatment of grades 1 or 2

spondylolisthesis does not have to wait until it progresses to grades 3

or 4. Legally marketed devices which do not utilize pedicle screws are

available for this purpose.

L. Review of New Pedicle Screw Spinal System 510(k)'s

46. A comment pointed out that since FDA's January 1995,

determination regarding the preamendments status of pedicle screw

spinal systems in the treatment of severe spondylolisthesis, many

510(k) submissions have been cleared for this use. FDA's proposed rule

for pedicle screw spinal systems, once final, will essentially

represent a labeling change for these devices, requiring new 501(k)

submissions. The comment suggested that the new 510(k)'s should provide

a draft copy of the revised labeling and a statement that the

previously-cleared device has not been modified in any way that may

affect its safety or effectiveness. According to the comment, this

limited type of review would facilitate and expedite the review process

and would not unnecessarily burden FDA's device evaluation staff.

FDA agrees with this approach and intends to apply it in its review

of 510(k)'s for pedicle screw spinal systems that were cleared

previously for use in severe spondylolisthesis. Pedicle screw spinal

systems which have not been previously reviewed, or that represent

significant modifications compared to the previously cleared device(s),

will require a complete 510(k) submission, including the device

labeling.

M. Review of New Information Published and Submitted After Publication

of the Proposed Rule: Pedicle Screw and Related Literature and MedWatch

and MDR System Reports

FDA performed a comprehensive search of the English-language

medical literature published between 1994 and the present. Thirty-five

articles pertained to the clinical performance of pedicle screw spinal

systems. The clinical performance results, e.g., fusion rate and

complication types and rates,

[[Page 40040]]

from these peer-reviewed articles did not differ from those previously

reported in the preamble to the proposed rule for either pedicle screw

spinal systems or the group of class II spinal devices using hooks and/

or wires or noninstrumented fusions.

FDA also performed a review of the MedWatch and MDR databases from

1994 to the present. The complications associated with pedicle screw

spinal systems during this period were comparable to those reported in

the preamble to the proposed rule for pedicle screw spinal systems and

the group of class II spinal devices using hooks and/or wires and

noninstrumented fusions.

VI. References

The following references have been placed on display in the

Dockets Management Branch (HFA-305), Food and Drug Administration, 5630

Fishers Lane, rm. 1061, Rockville, MD 20852. These references may be

seen by interested persons between 9 a.m. and 4 p.m., Monday through

Friday.

1. Rodgers, A. E., ``FDA Pedicle Screw Cohort Study: Audit

Findings,'' July 30, 1996.

2. Richter, K. C., ``Assessment of the Impact of BIMO Audit

Findings for the Pedicle Screw Cohort Study on Study Results,'' August

29, 1997.

VII. Environmental Impact

The agency has determined under 21 CFR 25.24(e)(2) that this action

is of a type that does not individually or cumulatively have a

significant effect on the human environment. Therefore, neither an

environmental assessment nor an environmental impact statement is

required.

VIII. Analysis of Impacts

FDA has examined the impact of the final rule under Executive Order

12866 and the Regulatory Flexibility Act (5 U.S.C. 601-612) (as amended

by Subtitle D of the Small Business Regulatory Fairness Enforcement Act

of 1996 (Pub. L. 104-121), and the Unfunded Mandates Reform Act of 1995

(Pub. L. 104-4)). Executive Order 12866 directs agencies to assess all

costs and benefits of available regulatory alternatives and, when

regulation is necessary, to select regulatory approaches that maximize

net benefits (including potential economic, environmental, public

health and safety, and other advantages; distributive impacts; and

equity). The agency believes that this final rule is consistent with

the regulatory philosophy and principles identified in the Executive

Order. In addition, the final rule has been determined to be a

significant regulatory action as defined by the Executive Order and so

is not subject to review under the Executive Order.

The Regulatory Flexibility Act requires agencies to analyze

regulatory options that would minimize any significant impact of a rule

on small entities. Classification and reclassification of the device

from class III to class II when the device is intended to provide

immobilization and stabilization of spinal segments as an adjunct to

fusion in the treatment of the following acute and chronic

instabilities or deformities of the thoracic, lumbar, and sacral spine:

Degenerative spondylolisthesis with objective evidence of neurologic

impairment, fractures, dislocations, scoliosis, kyphosis, spinal

tumors, and failed previous fusion attempts (pseudarthrosis) will

relieve all manufacturers of the device of the cost of complying with

the premarket approval requirements in section 515(b) of the act.

Because classification and reclassification will reduce regulatory

costs with respect to this device, it will not impose significant

economic impact on any small entities, and it may permit small

potential competitors to enter the marketplace by lowering their costs.

The Commissioner of Food and Drugs, therefore, certifies that this

final rule will not have a significant economic impact on a substantial

number of small entities. In addition, this final rule will not impose

costs of $100 million or more on either the private sector or State,

local, and tribal governments in any one year, therefore, a summary

statement of analysis under section 202(a) of the Unfunded Mandates

Reform Act of 1995 is not required.

List of Subjects in 21 CFR Part 888

Medical devices.

Therefore, under the Federal Food, Drug, and Cosmetic Act and under

authority delegated to the Commissioner of Food and Drugs, 21 CFR part

888 is amended as follows:

PART 888--ORTHOPEDIC DEVICES

1. The authority citation for 21 CFR part 888 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

2. Section 888.3070 is added to subpart D to read as follows:

Sec. 888.3070 Pedicle screw spinal system.

(a) Pedicle screw spinal systems--(1) Identification. Pedicle screw

spinal systems are multiple component devices, made from a variety of

materials, including alloys such as 316L stainless steel, 316LVM

stainless steel, 22Cr-13Ni-5Mn stainless steel, Ti-6Al-4V, and

unalloyed titanium, that allows the surgeon to build an implant system

to fit the patient's anatomical and physiological requirements. Such a

spinal implant assembly consists of a combination of anchors (e.g.,

bolts, hooks, and/or screws); interconnection mechanisms incorporating

nuts, screws, sleeves, or bolts; longitudinal members (e.g., plates,

rods, and/or plate/rod combinations); and/or transverse connectors. The

devices are intended to provide immobilization and stabilization of

spinal segments in skeletally mature patients as an adjunct to fusion

in the treatment of the following acute and chronic instabilities or

deformities of the thoracic, lumbar, and sacral spine: degenerative

spondylolisthesis with objective evidence of neurologic impairment,

fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed

previous fusion (pseudarthrosis).

(2) Classification. Class II (special controls). Pedicle screw

spinal systems must comply with the following special controls:

(i) Compliance with material standards,

(ii) Compliance with mechanical testing standards,

(iii) Compliance with biocompatibility standards, and

(iv) Labeling which contains these two statements in addition to

other appropriate labeling information:

``Warning: The safety and effectiveness of pedicle screw spinal

systems have been established only for spinal conditions with

significant mechanical instability or deformity requiring fusion

with instrumentation. These conditions are significant mechanical

instability or deformity of the thoracic, lumbar, and sacral spine

secondary to degenerative spondylolisthesis with objective evidence

of neurologic impairment, fracture, dislocation, scoliosis,

kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

The safety and effectiveness of these devices for any other

conditions are unknown.''

``Precaution: The implantation of pedicle screw spinal systems

should be performed only by experienced spinal surgeons with

specific training in the use of this pedicle screw spinal system

because this is a technically demanding procedure presenting a risk

of serious injury to the patient.''

(b) Pedicle screw spinal systems for all other uses--(1)

Identification. Pedicle screw spinal systems for all other uses are

multiple component devices, made from a variety of materials, including

alloys such as 316L stainless steel, 316LVM stainless steel, 22Cr-13Ni-

5Mn stainless steel, Ti-6Al-4V, and unalloyed titanium, that allow

[[Page 40041]]

the surgeon to build an implant system to fit the patient's anatomical

and physiological requirements. Such an spinal implant assembly

consists of a combination of anchors (e.g., bolts, hooks, and/or

screws); interconnection mechanisms incorporating nuts, screws,

sleeves, or bolts; longitudinal members (e.g., plates, rods, and/or

plate/rod combinations); and/or transverse connectors.

(2) Classification. Class III (premarket approval).

(c) Date PMA or notice of completion of a PDP is required. An

approved PMA or a declared completed PDP must be in effect before

placing the device in commercial distribution. See Sec. 888.3.

Dated: April 22, 1998,

Michael A. Friedman,

Lead Deputy Commissioner for the Food and Drug Administration.

Donna E. Shalala,

Secretary of Health and Human Services.

[FR Doc. 98-19944 Filed 7-23-98; 8:45 am]

BILLING CODE 4160-01-F