Medical Devices; Reclassification of 38 Preamendments Class III

Devices into Class II

AGENCY: Food and Drug Administration, HHS.

ACTION: Proposed rule.

-----------------------------------------------------------------------

SUMMARY: The Food and Drug Administration (FDA) is proposing to

reclassify 38 preamendments class III devices into class II (special

controls). FDA is also identifying the proposed special controls that

the agency believes will reasonably ensure the safety and effectiveness

of the devices. This reclassification is being proposed on the agency's

own initiative based on new information. This action is being taken

under the Federal Food, Drug, and Cosmetic Act (the act), as amended by

the Safe Medical Devices Act of 1990 (the SMDA) and the FDA

Modernization Act of 1997 (FDAMA). The agency is also proposing that

the identification of six of the devices subject to this proposal be

modified to more accurately reflect the characteristics of devices

actually being marketed.

DATES: Written comments by June 14, 1999. See section X of this

document for the proposed effective date of a final rule based on this

document.

ADDRESSES: Submit written comments to the Documents Management Branch

(HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061,

Rockville, MD 20852.

FOR FURTHER INFORMATION CONTACT: Janet L. Scudiero, Center for Devices

and Radiological Health (HFZ-410), Food and Drug Administration, 9200

Corporate Blvd., Rockville, MD 20850 301-594-1184.

SUPPLEMENTARY INFORMATION:

I. Regulatory Authorities

The act, as amended by the 1976 Medical Device Amendments (the

amendments) (Pub. L. 94-295), the SMDA (Pub. L. 101-629), and FDAMA

(Pub. L. 105-115), establishes a comprehensive system for the

regulation of medical devices intended for human use. Section 513 of

the act (21 U.S.C. 360c) establishes three categories (classes) of

devices, depending on the regulatory controls needed to provide

reasonable assurance of their safety and effectiveness. The three

categories of devices are class I (general controls), class II (special

controls), and class III (premarket approval).

Under section 513 of the act, devices that were in commercial

distribution before May 28, 1976 (the date of enactment of the

amendments), generally referred to as preamendments devices, are

classified after FDA has: (1) Received a recommendation from a device

classification panel (an FDA advisory committee); (2) published the

panel's recommendation for comment, along with a proposed regulation

classifying the device; and (3) published a final regulation

classifying the device. FDA has classified most preamendments devices

under these procedures.

Devices that were not in commercial distribution prior to May 28,

1976, generally referred to as postamendment devices, are classified

automatically by statute (section 513(f) of the act (21 U.S.C. 360c(f))

into class III without any FDA rulemaking process. Those devices remain

in class III and require premarket approval, unless and until FDA

issues an order finding the device to be substantially equivalent,

under section 513(i) of the act (21 U.S.C. 360c(i)), to a predicate

device that does not require premarket approval, or reclassifies the

device under 513(f). The agency determines whether new devices are

substantially equivalent to previously offered devices by means of

premarket notification procedures in section 510(k) of the act (21

U.S.C. 360(k)) and part 807 of the regulations (21 CFR part 807).

A preamendments device that has been classified into class III may

be marketed, by means of premarket notification (510(k)) procedures,

without submission of a premarket approval application (PMA) until FDA

issues a final regulation under section 515(b) of the act (21 U.S.C.

360e(b)) requiring premarket approval.

The SMDA added section 515(i) (21 U.S.C. 360e(i)) to the act. This

section requires FDA to issue an order to manufacturers of preamendment

class III devices and substantially equivalent postamendments devices

for which no final regulation requiring the submission of PMA's has

been issued. This order requires such manufacturers to submit to the

agency a summary of, and a citation to, any information known or

otherwise available to them respecting such devices, including adverse

safety and effectiveness information that has not been submitted under

section 519 of the act (21 U.S.C. 360i), which requires manufacturers,

importers, distributors, and device user facilities to submit adverse

event reports of certain device-related events and reports of certain

corrective actions taken. Section 515(i) of the act (21 U.S.C. 360e)

also directs FDA to either revise the classification of the device into

class I or class II or require the device to remain in class III and

establish a schedule for the issuance of a rule requiring the

submission of PMA's for those devices remaining in class III.

In the Federal Register of May 6, 1994 (59 FR 23731), FDA announced

the availability of a document setting forth its strategy for

implementing section 515(i) of the act. Under this plan, the agency

divided preamendment class III devices into the following three groups:

Group 1 devices are devices that FDA believes raise significant

questions of safety and/or effectiveness, but are no longer used or are

in very limited use; Group 2 devices are devices that FDA believes have

a high potential for being reclassified; and Group 3 devices are

devices that FDA believes are currently in commercial distribution and

are not likely candidates for reclassification. FDA also announced its

intention to call for submission of PMA's for the 15 highest priority

devices in Group 3, and for all Group 1 devices. The agency also

announced its intention to issue an order under section 515(i) of the

act for the remaining Group 3 devices and for all Group 2 devices.

In the Federal Register of August 14, 1995 (60 FR 41984 and

41986), FDA published two orders for certain class III devices,

requiring the submission of safety and effectiveness information in

accordance with the preamendments class III strategy for implementing

[[Page 12775]]

section 515(i) of the act. FDA published two updated orders in the

Federal Register of June 13, 1997 (62 FR 32352 and 32355). The orders

described in detail the format for submitting the type of information

required by section 515(i) of the act so that the information submitted

would clearly support reclassification or indicate that a device should

be retained in class III. The orders also scheduled the required

submissions in groups, at 6-month intervals, beginning on August 14,

1996.

Reclassification of classified preamendments devices is governed

by section 513(e) of the act (21 U.S.C. 360c(e)). This section provides

that FDA may, by rulemaking, reclassify a device based upon ``new

information.'' The reclassification can be initiated by FDA or by the

petition of an interested person. The term ``new information,'' as used

in section 513(e) of the act (21 U.S.C. 360c(e)), includes information

developed as a result of a reevaluation of the data before the agency

when the device was originally classified, as well as information not

presented, not available, or not developed at that time. (See, e.g.,

Holland Rantos v. United States Department of Health, Education, and

Welfare, 587 F.2d 1173, 1174 n.1 (D.C. Cir. 1978); Upjohn v. Finch, 422

F.2d 944 (6th Cir. 1970); Bell v. Goddard, 366 F.2d 177 (7th Cir.

1966).)

Reevaluation of the data previously before the agency is an

appropriate basis for subsequent regulatory action where the

reevaluation is made in light of changes in ``medical science.,'' (see

Upjohn v. Finch, supra, 422 F.2d at 951), or in light of newly

available regulatory controls (cf. Ethicon, Inc. v. FDA, 762 F. Supp.

382, 388-389 (D.D.C. 1991)), such as special controls or design

controls. However, regardless of whether data before the agency are

past or new data, the ``new information'' on which any reclassification

is based is required to consist of ``valid scientific evidence,'' as

defined in section 513(a)(3) of the act (21 U.S.C. 360c(a)(3)) and

Sec. 860.7(c)(2) (21 CFR 860.7(c)(2)). FDA relies upon ``valid

scientific evidence'' in the classification process to determine the

level of regulation for devices.

II. Regulatory History of the Devices

The 38 devices subject to this proposal were classified by final

rules published in the Federal Register in parts 864, 866, 868, 870,

872, 874, 876, 878, 884, 886, and 888 (21 CFR parts 864, 866, 868, 870,

872, 874, 876, 878, 884, 886, and 888). In the proposed rules upon

which the final rules were based, FDA considered the recommendations of

the device classification advisory panels regarding the classification

of preamendments medical devices. Subsequently, FDA classified the

devices subject to this proposal into class III, because there was

insufficient information to determine that class I or class II controls

could provide reasonable assurance of the safety and effectiveness of

these devices. The Federal Register citations and publication dates for

the proposed and final rules classifying the devices subject to this

proposal are provided in Table 1. as follows:

Table 1.--Publication Dates for the Proposal and Final Rules Classifying

the Devices Subject to This Proposal

------------------------------------------------------------------------

21 CFR Part and Device Proposed Rule Final Rule

------------------------------------------------------------------------

Part 864, Hematology/ September 11, 1979, 44 September 12, 1980, 45

Pathology FR 52950 FR 60576

Part 866, Immunology/ April 22, 1980, 45 FR November 9, 1982, 47

Microbiology 27204 FR 50283

Part 868, November 2, 1979, 44 FR July 16, 1982, 47 FR

Anesthesiology 63292 31130

Part 870, March 9, 1979, 44 FR February 5, 1980, 45

Cardiovascular 13284 FR 7904

Part 872, Dental December 30, 1980, 45 August 12, 1987, 52 FR

FR 85962 30082

Part 874, Ear, Nose, January 22, 1982, 47 FR November 6, 1986, 51

and Throat 3280 FR 40389

Part 876, January 23, 1981, 46 FR November 23, 1983, 48

Gastroenterology/ 7562 FR 53012

Urology

Part 878, General and January 19, 1982, 47 FR June 24, 1988, 53 FR

Plastic Surgery 2810 23856

Part 884, Obstetrical April 3, 1979, 44 FR February 26, 1980, 45

and Gynecological 19894 FR 12682

Part 886, Ophthalmic January 26, 1982, 47 FR September 2, 1987, 52

3694 FR 33346

Part 888, Orthopedic July 2, 1982, 47 FR September 4, 1987, 52

29052 FR 33686

------------------------------------------------------------------------

In accordance with section 513(e) of the act and Sec. 860.130 (21

CFR 860.130), based upon new information received or otherwise

available to the agency with respect to the devices subject to this

proposal, FDA, on its own initiative, is proposing to reclassify 38

preamendments class III devices to class II. Consistent with the act

and the regulation, FDA did not refer the proposed reclassifications to

the panels for their recommendations on the requested change in

classifications.

III. Proposed Changes to Device Names and Identifications

Since initial classification of the 38 devices subject to this

proposal, there have been many advances in the medical device industry.

These advances have led to many changes, including the use of

alternative materials, and/or modifications of the intended uses for

some devices. Because the changes have been of sufficiently low impact

on safety and effectiveness, FDA determined that the modified devices

were substantially equivalent to the respective predicate devices. In

some cases, however, the substantially equivalent device differs

slightly from the device description found in the agency's regulations.

In order to more accurately reflect the characteristics of the actual

marketed devices subject to this proposal, the agency is proposing

certain technical amendments be made to six device identifications, as

listed in section III of this document. The agency stresses that these

amendments are not intended to impose any additional restrictions on

the marketed devices; rather, they are intended to accurately reflect

the characteristics of marketed devices. The following changes in

device identifications are being proposed.

A. Section 876.5860--High Permeability Hemodialysis System

A high permeability hemodialysis system is a device intended for

use as an artificial kidney system for the treatment of patients with

renal failure, fluid overload, or toxemic conditions by performing such

therapies as hemodialysis, hemofiltration, and hemoconcentration. Using

a hemodialyzer with a semipermeable membrane that is more permeable to

water than the semipermeable membrane of the conventional hemodialysis

system described in Sec. 876.5820, the high permeability hemodialysis

system removes toxins or excess fluid from the patient's blood

[[Page 12776]]

using the principles of convection (via a high ultrafiltration rate)

and/or diffusion (via a concentration gradient in dialysate). During

treatment, blood from the patient is circulated through the

hemodialyzer's blood compartment, while the dialysate solution flows

countercurrent through the dialysate compartment. In this process,

toxins and/or fluid are transferred across the membrane from the blood

to the dialysate compartment. The hemodialysis delivery machine

controls and monitors the parameters related to this processing,

including the rate at which blood and dialysate are pumped through the

system, and the rate at which fluid is removed from the patient. The

high permeability hemodialysis system consists of the following

devices:

(1) The hemodialyzer consists of a semipermeable membrane with an

in vitro ultrafiltration coefficient (Kuf) greater than 12 milliliters

per hour per millimeters of mercury (mL/h/mmHg), and is used with

either an automated ultrafiltration controller or another method of

ultrafiltration control to prevent fluid imbalance.

(2) The hemodialysis delivery machine is similar to the

extracorporeal blood system and dialysate delivery system of the

hemodialysis system and accessories (Sec. 876.5820), with the addition

of an ultrafiltration controller and mechanisms that monitor and/or

control such parameters as fluid balance, dialysate composition, and

patient treatment parameters (e.g., blood pressure, hematocrit, urea,

etc.).

(3) The high permeability hemodialysis system accessories include,

but are not limited to, tubing lines and various treatment related

monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood

recirculation monitors).

B. Section 878.3610--Esophageal Prosthesis

An esophageal prosthesis is a rigid, flexible, or expandable

tubular device constructed of a plastic, metal, or polymeric material

that is intended to be implanted to restore the structure and/or

function of the esophagus. The metal esophageal prosthesis may be

uncovered or covered with a polymeric material. This device may also

include a device delivery system.

C. Section 878.3720--Tracheal Prosthesis

A tracheal prosthesis is a rigid, flexible, or expandable tubular

device constructed of a silicone, metal, or polymeric material that is

intended to be implanted to restore the structure and/or function of

the trachea or tracheal-bronchial tree. It may be unbranched or contain

one or two branches. The metal tracheal prosthesis may be uncovered or

covered with a polymeric material. This device may also include a

device delivery system.

D. Section 886.3400--Keratoprosthesis

A keratoprosthesis is a device intended to provide a transparent

optical pathway through an opacified cornea, either intraoperatively or

permanently, in an eye which is not a reasonable candidate for a

corneal transplant.

This identification recognizes the temporary use of the device

intraoperatively, and removes the description of the device as being

made of only plastic material.

E. Section 886.3920--Aqueous Shunt (previously ``Eye valve implant'')

An aqueous shunt is a one-way, pressure sensitive device intended

to be implanted to normalize intraocular pressure. The device is

intended to treat neurovascular glaucoma or glaucomas where medical and

conventional surgical treatment have failed

The agency is proposing that the name of this device be ``aqueous

shunt'' rather than ``eye valve implant,'' because certain marketed

devices, which have been determined to be substantially equivalent to

the eye valve implant, do not contain a valve or a valve-like

component.

The agency is also proposing to modify the identification of this

device to more accurately reflect the device's actual use. Because the

identified use of ``treatment of glaucoma'' is unnecessarily broad, the

agency proposes that the identification state that the device may be

used for the treatment of neovascular glaucoma or glaucomas where

medical and conventional surgical treatment have failed.

F. Section 888.3150--Elbow Joint Metal/Polymer Constrained Cemented

Prosthesis

An elbow joint metal/polymer constrained cemented prosthesis is a

device intended to be implanted made of alloys such as cobalt-chromium-

molybdenum and of an ultra-high molecular weight polyethylene bushing,

and used to replace an elbow joint. The device presents dislocation in

more than one anatomic plane and consists of two components which are

linked together. This generic type of device is limited to those

prostheses intended for use with bone cement (Sec. 888.3027).

The agency is proposing that the name and identification of the

elbow joint metal/metal or metal/polymer constrained cemented

prosthesis be modified to remove reference to the metal/metal

prosthesis, because no metal/metal constrained cemented elbow

prosthesis has ever been marketed.

IV. Proposed Reclassification

FDA is proposing that the devices subject to this proposal be

reclassified from class III to class II. FDA believes that the

identified special controls would provide reasonable assurance of

safety and effectiveness. Therefore, in accordance with sections 513(e)

and 515(i) of the act and Sec. 860.130, based on new information with

respect to the devices, FDA, on its own initiative, is proposing to

reclassify these 38 preamendments class III devices into class II. The

agency has identified special controls that would provide reasonable

assurance of their safety and effectiveness. The agency does not intend

to exempt these proposed class II devices from premarket notification

(510(k)) submissions as provided for under section 510(m) of the act

(21 U.S.C. 360(m)).

V. Proposed Special Controls.

Because several of the special controls identified in this

proposal apply to 2 or more of the 38 devices addressed by this

proposal, the agency has determined that it would be inefficient and

redundant to individually identify, for each device, shared risks to

health and corresponding special controls to address the risks to

health. Instead, this document focuses on the special controls,

explains the types of risks to health addressed by the special

controls, and identifies the devices to which the special controls

apply. For ease of review, Table 1 is included in section VI of this

document following the discussion of special controls. The summary

table identifies each device by name and Code of Federal Regulations

(CFR) citation section number, the citation for the final rule which

classified the preamendments device into Class III, and the proposed

special controls applicable to the device. The special controls

identified in this proposal are of four general types: FDA guidance

documents, consensus standards, device specific labeling, and design

and performance testing.

A. FDA Guidance Documents

Based on its premarket and postmarket experience and the published

literature, the agency has developed the guidance documents in section

V.A of this document that are designed to inform manufacturers of

[[Page 12777]]

how the agency evaluates the safety and effectiveness of devices and

reaches determinations of substantial equivalence. The guidance

documents are also intended for use by FDA reviewers to ensure

consistency of premarket reviews. Some FDA guidance documents are

generic guidances applicable to many different devices, while others

are applicable to a few related devices, or a specific device. The

generic guidance documents may be referenced, and thereby incorporated

into, other guidances.

The agency has adopted Good Guidance Practices (GGP's), which set

forth the agency's policies and procedures for the development,

issuance, and use of guidance documents (62 FR 8961, February 27,

1997). When FDA issues a final rule based on this proposal, all of the

guidance documents identified as special controls will have been issued

in accordance with GGP's.

Persons interested in obtaining a copy of a guidance may do so

using the World Wide Web (WWW). The Center for Devices and Radiological

Health (CDRH) maintains an entry on the WWW for easy access to

information including text, graphics, and files that may be downloaded

to a personal computer with access to the WWW. The CDRH home page may

be accessed at http://www.fda.gov/cdrh. Guidance documents are also

available from the Division of Small Manufacturers' Assistance (HFZ-

220), Food and Drug Administration, Center for Devices and Radiological

Health, 1350 Piccard Dr., Rockville, MD 20850.

FDA guidances are periodically updated as new information becomes

available. When an FDA guidance that has been identified as a special

control is revised, a notice of availability of the revised guidance

will be published in the Federal Register, as well as a proposal to

amend the special control(s) for the relevant device(s) to include the

revised guidance. The following is a list and description of guidance

documents that FDA proposes to use as special controls:

1. Use of International Standard ISO 10993, ``Biological Evaluation of

Medical Devices Part I: Evaluation and Testing'' (biocompatibility

guidance)

During the classification of the preamendments devices, the device

classification panels (the panels) identified potential adverse tissue

reactions as a risk to health common to devices that contact the body.

These adverse tissue reactions were identified generally, or more

specifically according to the type of tissue reaction (e.g.,

sensitization, pyrogen reaction, hemolysis, etc.). The agency believes

that the information contained in this biocompatibility guidance is

adequate to control the risks to health related to adverse tissue

reaction.

Therefore, the agency is proposing that the biocompability

guidance be a special control applied to the following 27 devices:

Indwelling blood carbon dioxide partial pressure (PCO2)

analyzer (Sec. 868.1150), indwelling blood hydrogen ion concentration

(pH) analyzer (Sec. 868.1170), indwelling blood oxygen partial pressure

(PO2) analyzer (Sec. 868.1200), cardiovascular intravascular

filter (Sec. 870.3375), vascular graft prosthesis of less than 6-

millimeters diameter (Sec. 870.3450), pacemaker lead adaptor

(Sec. 870.3620), annuloplasty ring (Sec. 870.3800), cardiopulmonary

bypass defoamer (Sec. 870.4230), cardiopulmonary bypass arterial line

blood filter (Sec. 870.4260), cardiopulmonary bypass oxygenator

(Sec. 870.4350), OTC (over-the-counter) denture cushion or pad

(Sec. 872.3540), OTC denture reliner (Sec. 872.3560), OTC denture

repair kit (Sec. 872.3570), partially fabricated denture kit

(Sec. 872.3600), high permeability hemodialysis system (Sec. 876.5860),

peritoneo-venous shunt (Sec. 876.5955), endometrial aspirator

(Sec. 884.1060), endometrial brush (Sec. 884.1100), endometrial washer

(Sec. 884.1185), endoscopic electrocautery and accessories

(Sec. 884.4100), bipolar endoscopic coagulator-cutter and accessories

(Sec. 884.4150), keratoprosthesis (Sec. 886.3400), aqueous shunt

(Sec. 886.3920), elbow joint metal/polymer constrained cemented

prosthesis (Sec. 888.3150), knee joint patellofemoral polymer/metal

semi-constrained cemented prosthesis (Sec. 888.3540), shoulder joint

metal/polymer nonconstrained cemented prosthesis (Sec. 888.3650), and

shoulder joint metal/polymer semi-constrained cemented prosthesis

(Sec. 888.3660).

2. ``510(k) Sterility Review Guidance and Revision of 11/18/94 K90-1

(Sterility Guidance)''

During the classification of the preamendments devices, the panels

identified potential infection as a risk to health common to the use of

many devices. The potential risk of infection would be minimized if the

device were properly sterilized prior to use and appropriately labeled.

Since classification of the devices subject to this proposal, the

agency has developed the sterility guidance. It provides information

about the use and application of national and international sterility

consensus standards for devices to be labeled as ``sterile.'' The

agency believes that the information contained in this guidance

document is adequate to control for the potential risks to health

related to infection.

Therefore, the agency is proposing that the sterility guidance be

a special control for the following 23 devices: Indwelling blood carbon

dioxide partial pressure (PCO2) analyzer (Sec. 868.1150),

indwelling blood hydrogen ion concentration (pH) analyzer

(Sec. 868.1170), indwelling blood oxygen partial pressure

(PO2) analyzer (Sec. 868.1200), cardiovascular intravascular

filter (Sec. 870.3375), vascular graft prosthesis of less than 6-

millimeters diameter (Sec. 870.3450), pacemaker lead adaptor

(Sec. 870.3620), annuloplasty ring (Sec. 870.3800), cardiopulmonary

bypass defoamer (Sec. 870.4230), cardiopulmonary bypass arterial line

blood filter (Sec. 870.4260), cardiopulmonary bypass oxygenator

(Sec. 870.4350), electrohydraulic lithotriptor (Sec. 876.4480),

peritoneo-venous shunt (Sec. 876.5955), endometrial aspirator

(Sec. 884.1060), endometrial brush (Sec. 884.1100), endometrial washer

(Sec. 884.1185), endoscopic electrocautery and accessories

(Sec. 884.4100), bipolar endoscopic coagulator-cutter and accessories

(Sec. 884.4150), keratoprosthesis (Sec. 886.3400), eye valve implant

(Sec. 886.3920), elbow joint metal/polymer constrained cemented

prosthesis (Sec. 888.3150), knee joint patellofemoral polymer/metal

semi-constrained cemented prosthesis (Sec. 888.3540), shoulder joint

metal/polymer nonconstrained cemented prosthesis (Sec. 888.3650), and

shoulder joint metal/polymer semi-constrained cemented prosthesis

(Sec. 888.3660).

3. ``Guidance Document for the Submission of Erythropoietin Assay

Premarket Notification (510(k))''

During the classification of the preamendments devices, the

Hematology and Pathology Devices Classification Panel identified as

risks to health, complications associated with misdiagnosis of a

disease state. Since classification of this device, the agency has

developed a guidance document describing its present conclusions

regarding the materials, labeling, and testing controls for

erythropoietin assay devices. Because the agency believes that the

information contained in this guidance document is adequate to control

for the identified risks to health, the agency is proposing that the

``Guidance Document for the Submission of Erythropoietin Assay

Premarket Notification (510(k))'' be a special control for the

erythropoietin assay (Sec. 864.7250).

4. ``Guidance Document for the Submission of Fibrin Monomer

Paracoagulation Test Premarket Notification (510(k))''

[[Page 12778]]

During the classification of the preamendments devices, the

Hematology and Pathology Devices Classification Panel identified as

risks to health associated with the use of this device, complications

associated with misdiagnosis of a disease state. Since classification

of this device, the agency has developed a guidance document describing

its present conclusions regarding the materials, labeling, and testing

controls for fibrin monomer paracoagulation test devices. Because the

agency believes that the information contained in this guidance

document is adequate to control for the identified risks to health, the

agency is proposing that the ``Guidance Document for the Submission of

Fibrin Monomer Paracoagulation Test Premarket Notifications (510(k))''

be a special control for the fibrin monomer paracoagulation test

(Sec. 864.7300).

5. ``Reviewer Guidance for Clinical Studies and Labeling for Indwelling

Blood Gas Analyzers''

During the classification of the preamendments devices, the

Anesthesiology and Respiratory Therapy Device Classification Panel

identified as a risk to health common to indwelling blood gas

analyzers, the potential for inaccurate measurement which would lead to

inappropriate therapy. Since their classification, the agency has

developed a guidance document describing its present conclusions

regarding the appropriate clinical testing to ensure that indwelling

blood gas analyzers function properly, and labeling which would ensure

that the devices would be used properly, thus minimizing the risk of

inaccurate measurement of blood gasses. Because the agency believes

that the information contained in this guidance document, in

combination with the guidances described below, is adequate to address

the risks to health, the agency is proposing that the ``Reviewer

Guidance for Clinical Studies and Labeling for Indwelling Blood Gas

Analyzers'' be a special control for the following three devices:

Indwelling blood carbon dioxide partial pressure (PCO2)

analyzer (Sec. 868.1150), indwelling blood hydrogen ion concentration

(pH) analyzer (Sec. 868.1170), and indwelling blood oxygen partial

pressure (PO2) analyzer (Sec. 868.1200).

To further minimize the risk of inaccurate measurement by

indwelling blood gas analyzers, the agency is proposing that the

``Reviewer Guidance for Computer Controlled Medical Devices Undergoing

510(k) Review and the Reviewer Guidance for Format and Content of

Premarket Notifications (510(k) Submissions), Labeling, Performance and

Environmental Testing for Electronic Devices'' be special controls for

the following three devices: Indwelling blood carbon dioxide partial

pressure (PCO2) analyzer (Sec. 868.1150), indwelling blood

hydrogen ion concentration (pH) analyzer (Sec. 868.1170), and

indwelling blood oxygen partial pressure (PO2) analyzer

(Sec. 868.1200). These guidance documents provide more details about

the agency's present conclusions regarding testing, labeling, and

manufacturing information which would be required for premarket

notifications (510(k)) for these device. The agency believes that these

guidance documents are adequate to address the identified risk to

health.

6. ``Guidance Document for the Submission of 510(k) Premarket

Notifications for Cardiovascular Intravascular Filters''

During the classification of the preamendments devices, the

Circulatory System Devices Classification Panel identified as risks to

health the following potential complications associated with the use of

the cardiovascular intravascular filter: Pulmonary thromboembolism when

anticoagulants are contraindicated, failure of anticoagulant therapy in

thromboembolic diseases, chronic recurrent pulmonary embolism where

anticoagulant therapy has failed or is contraindicated. Since

classification of this device, the agency has developed a guidance

document describing its present conclusions regarding the labeling,

biocompatibility testing, mechanical testing, sterilization procedures

and labeling, and clinical data controls that would ensure the safety

and effectiveness of cardiovascular intravascular filters seeking

510(k) clearance. Because the agency believes that the information

contained in this guidance document is adequate to control for the

identified risks to health, the agency is proposing that the ``Guidance

Document for the Submission of 510(k) Premarket Notifications for

Cardiovascular Intravascular Filters'' be a special control for the

cardiovascular intravascular filter (Sec. 870.3375).

7. ``Document for Special Controls for Vascular Prostheses''

During the classification of the preamendments devices, the

Circulatory System Devices Classification Panel identified as risks to

health associated with the use of the vascular prosthesis, the

potentials for: Thrombosis, embolism, occlusion stenosis, leakage,

graft disruption, seroma, pseudoaneurisms, aneurisms, dilation,

infection, and device failure. Since classification of this device, the

agency has established certain labeling, testing, and manufacturing

controls to minimize the potential of the identified risks to health.

These controls are discussed in this guidance document. Because the

agency believes that the information contained in this guidance

document is adequate to address the identified risks to health, the

agency is proposing that the ``Document for Special Controls for

Vascular Prostheses'' be a special control for the vascular graft

prosthesis of less than 6-millimeters diameter (Sec. 870.3450).

8. ``Document for Special Controls for Annuloplasty Rings''

During the classification of the preamendments devices, the

Circulatory System Devices Classification Panel identified as a risk to

health associated with the use of the annuloplasty ring, the potentials

for: Stenosis, thrombosis, thromboembolism, regurgitation, ring

fracture, obstruction, low cardiac output, ring dehiscence,

endocarditis, bleeding, blockage, and suture injury. Since

classification of this device, the agency has established certain

labeling, testing, and manufacturing controls to minimize the potential

of the identified risks to health. These controls are discussed in this

guidance document. Because the agency believes that the information

contained in this guidance document is adequate to address the

identified risks to health, the agency is proposing that the ``Document

for Special Controls for Annuloplasty Rings'' be a special control for

the annuloplasty ring (Sec. 870.3800).

9. ``Document for Special Controls for the Cardiopulmonary Bypass

Defoamer''

During the classification of the preamendments devices, the

Circulatory System Devices Classification Panel identified as risks to

health associated with the use of the cardiopulmonary bypass defoamer,

the potentials for: Blood damage, gaseous embolism, thromboembolism,

blood incompatibility, and inadequate blood flow. Since classification

of this device, the agency has established labeling, testing, and

manufacturing controls to minimize the potential of the identified

risks to health. These controls are discussed in this guidance

document. Because the agency believes that the information contained in

this guidance document is adequate to control for the identified risks

to health, the agency is proposing that the ``Document for Special

Controls for the Cardiopulmonary Bypass Defoamer'' be a special control

for the cardiopulmonary bypass defoamer (Sec. 870.4230).

[[Page 12779]]

10. ``Document for Special Controls for the Cardiopulmonary Bypass

Arterial Filter''

During the classification of the preamendments devices, the

Circulatory System Devices Classification Panel identified as risks to

health associated with the use of the cardiopulmonary bypass arterial

filter, the potentials for: Gaseous embolism, thromboembolism, blood

incompatibility, and inadequate blood flow. Since classification of

this device, the agency has established certain labeling, testing, and

manufacturing controls to minimize the potential of the identified

risks to health. These controls are discussed in this guidance

document. Because the agency believes that the information contained in

this guidance document is adequate to control for the identified risks

to health, the agency is proposing that the ``Document for Special

Controls for the Cardiopulmonary Bypass Arterial Filter'' be a special

control for the cardiopulmonary bypass arterial line blood filter

(Sec. 870.4260).

11. ``Information for Manufacturers Seeking Marketing Clearance for

Blood-Gas Exchangers (Oxygenators) Used in Cardiopulmonary Bypass''

During the classification of the preamendments devices, the

Circulatory System Devices Classification Panel identified as risks to

health associated with the use of the cardiopulmonary bypass

oxygenator, the potentials for: Failure, improper gas transfer

function, hemolysis, destruction of platelets and white blood cells,

sludging, leaking, and emboli formation. Since classification of this

device, the agency has developed a guidance document describing its

present conclusions regarding the testing, labeling, and manufacturing

controls that would be necessary to ensure the safety and effectiveness

of the cardiopulmonary bypass oxygenator. The controls are discussed in

this guidance document. Because the agency believes that the

information contained in this guidance document is adequate to control

for the identified risks to health, the agency is proposing that the

``Information for Manufacturers Seeking Marketing Clearance for Blood-

Gas Exchangers'' be a special control for the cardiopulmonary bypass

oxygenator (Sec. 870.4350).

12. ``Guidance Document for the Submission of Research and Marketing

Applications for Permanent Pacemaker Leads''

During the classification of the preamendments devices, the

Circulatory System Devices Classification Panel identified as risks to

health associated with the use of the pacemaker lead adaptor, improper

pacing, failure to pace, and tissue damage. Since classification of

this device, the agency has developed a guidance document describing

its present conclusions regarding the research and marketing

information which should be submitted to the agency to support 510(k)

clearance for pacemaker lead adaptors. Because the agency believes that

the information contained in this guidance document is adequate to

control for the identified risk to health, the agency is proposing that

the ``Guidance Document for the Submission of Research and Marketing

Applications for Permanent Pacemaker Leads'' be a special control for

the pacemaker lead adaptor (Sec. 870.3620).

13. ``OTC Denture Reliners, Repair Kits, and Partially Fabricated

Denture Kits''

During the classification of the preamendments devices, the Dental

Product Classification Panel identified as risks to health common to

the use of certain denture accessories, complications resulting from an

alteration of the vertical dimension of a patient's jaw and irritation

of oral tissues. Since classification of these devices, the agency has

developed a guidance document describing its present conclusions

regarding procedures to minimize the risk of such complications.

Because the agency believes that the information contained in this

guidance document is adequate to control for the identified risks to

health, the agency is proposing that the ``OTC Denture Reliners, Repair

Kits, and Partially Fabricated Denture Kits'' be a special control for

the following four devices: OTC denture cushion or pad (Sec. 872.3540),

OTC denture reliner (Sec. 872.3560), OTC denture repair kit

(Sec. 872.3570), and partially fabricated denture kit (Sec. 872.3600).

14. ``Tympanostomy Tubes, Submission Guidance for a 510(k) Premarket

Notification''

During the classification of the preamendments devices, the Ear,

Nose, and Throat Devices Classification Panel identified as risks to

health associated with the use of the tympanostomy tube with

semipermeable membrane, the potentials for hearing loss or premature

extrusion. Since classification of this device, the agency has

developed a guidance document describing its present conclusions

regarding the labeling, testing, and manufacturing information which

should be submitted to the agency to support 510(k) clearance for

tympanostomy tubes. Because the agency believes that the information

contained in this guidance document is adequate to control for the

identified risks to health, the agency is proposing that the document

titled ``Tympanostomy Tubes, Submission Guidance for a 510(k) Premarket

Notification'' be a special control for the tympanostomy tube with

semipermeable membrane (Sec. 874.3930).

15. ``Guidance for the Content of Premarket Notifications for

Intracorporal Lithotripters''

During the classification of the preamendments devices, the

Gastroenterology-Urology Devices Classification Panel identified as

risks to health associated with the use of the electrohydraulic

lithotripter, potential: Infection, tissue damage, failure, breakage,

bleeding, pain, renal damage, and the formation of new stones. Since

classification of this device, the agency has developed a guidance

document describing its present conclusions regarding the labeling,

testing, and manufacturing information which should be submitted to the

agency to support 510(k) clearance for lithotripters. Because the

agency believes that the information contained in the guidance

document, when coupled with the guidance document described below, is

adequate to control for the identified risks to health, the agency is

proposing that the ``Guidance for the Content of Premarket

Notifications for Intracorporeal Lithotripters'' be a special control

for the electrohydraulic lithotripter (Sec. 876.4480).

To further minimize the potential risk of infection associated

with the reuse of electrohydraulic lithotripters, the agency believes

that certain labeling regarding the reuse of the device is appropriate.

The agency has developed a guidance document describing its present

conclusions regarding the labeling for certain reusable devices. FDA is

also proposing that the guidance document entitled ``Labeling Reusable

Medical Devices for Reprocessing in Health Care Facilities: FDA

Reviewer Guidance'' be a special control for the electrohydraulic

lithotripter (Sec. 876.4480).

16. ``Guidance for the Content of 510(k)s for Conventional and High

Permeability Hemodialyzers, Guidance for Industry and CDRH Reviewers on

the Content of Premarket Notifications for Hemodialysis Delivery

Systems, Guidance for the Content of Premarket Notifications for Water

Purification Components and Systems for Hemodialysis, and Guidance for

Hemodialyzer Reuse Labeling''

During the classification of the preamendments devices, the

Gastroenterology-Urology Devices Classification Panel identified as

risks to health associated with the use of the high permeability

hemodialysis system,

[[Page 12780]]

potential infection, electrical injury, adverse tissue reaction,

pyrogen reaction, hemolysis, electrolyte imbalance, hypovolemic shock,

air embolisms, loss of protein, and blood loss. Since classification of

this device, the agency has developed four guidance documents

describing its present conclusions regarding the labeling, testing, and

manufacturing information which should be submitted to the agency to

support 510(k) clearance for hemodialysis devices and accessories.

Because the agency believes that the information contained in the

guidance documents is adequate to control for the identified risks to

health, the agency is proposing that these four guidance documents be

applied as special controls for the high permeability hemodialysis

system (Sec. 876.5860).

17. ``Guidance for the Content of Premarket Notification Submissions

for Esophageal and Tracheal Prostheses''

During the classification of the preamendments devices, the

General and Plastic Surgery Devices Classification Panel and the Ear,

Nose, and Throat Devices Classification Panel identified as potential

risks to health common to the use of the esophageal prosthesis and the

tracheal prosthesis, certain complications resulting from migration,

obstruction, or placement of the devices, and potential gastric reflux

associated with the use of the esophageal prosthesis. Since

classification of these devices, the agency has developed a guidance

document describing its present conclusions regarding the labeling,

biocompatibility testing, mechanical testing, sterilization procedures

and labeling, and clinical data controls for esophageal or tracheal

prostheses seeking 510(k) clearance. Because the agency believes that

the information contained in this guidance document is adequate to

control for the identified risks to health, the agency is proposing

that the ``Guidance for the Content of Premarket Notification

Submissions for Esophageal and Tracheal Prostheses'' be a special

control for the following two devices: Esophageal prosthesis

(Sec. 878.3610) and tracheal prosthesis (Sec. 878.3720).

18. ``Guidance for Evaluation of Laproscopic Bipolar and Thermal

Coagulators (and Accessories)''

During the classification of the preamendments devices, the

Obstetrical and Gynecological Devices Classification Panel identified

potential complications from use in pregnant women as a risk to health

associated with the use of endoscopic electrocautery. Since

classification of these devices, the agency has developed a document

which provides information for the evaluation of laproscopic and

bipolar thermal coagulators. Among the information contained in this

document, is a discussion of the agency's present conclusions regarding

the labeling, testing, and manufacturing of such devices. Because the

agency believes that the information contained in the document is

adequate to control for the identified risk to health, the agency is

proposing that the ``Guidance for Evaluation of Laproscopic Bipolar and

Thermal Coagulators (and Accessories)'' be a special control for the

endoscopic electrocautery and accessories (Sec. 884.4100) and the

bipolar endoscopic coagulator-cutter and accessories (Sec. 884.4150).

19. ``Keratoprosthesis Guidance Document''

During the classification of the preamendments devices, the

Ophthalmic Devices Classification Panel identified as risks to health

associated with keratoprostheses, potentials for extrusion, infection,

adverse tissue reaction, glaucoma, retinal detachment, and development

of a retroprosthetic membrane. Since classification of this device, the

agency has developed a guidance document describing its present

conclusions regarding the labeling, testing, and manufacturing

information which should be submitted to the agency to support 510(k)

clearance for keratoprosthesis devices. Because the agency believes

that the information contained in the guidance document is adequate to

control for the identified risks to health, the agency is proposing

that the ``Keratoprosthesis Guidance Document'' be a special control

for the keratoprosthesis (Sec. 886.3400).

20. ``Aqueous Shunt-510(k) Submission''

During the classification of the preamendments devices, the

Ophthalmic Devices Classification Panel identified as risks to health

associated with aqueous shunts, the potentials for hypotony, extrusion,

infection, adverse tissue reaction, misplacement, migration, and

failure to filter. Since classification of these devices, the agency

has developed a guidance document describing its present conclusions

regarding the labeling, testing, and manufacturing information to be

submitted to the agency to support 510(k) clearance for aqueous shunts.

Because the agency believes that the information contained in the

guidance document is adequate to control for the identified risks to

health, the agency is proposing that the ``Aqueous Shunt-510(k)

Submission'' be a special control for the eye valve implant

(Sec. 886.3920).

21. ``Guidance Document for Testing Orthopedic Implants with Modified

Metallic Surfaces Apposing Bone or Bone Cement, Guidance Document for

Testing Non-articulating, Mechanically Locked' Modular Implant

Components and Guidance Document for the Preparation of Premarket

Notification (510(k) Applications for Orthopedic Devices''

During the classification of the preamendments devices, the

Orthopedic and Rehabilitation Devices Classification Panel identified

as risks to health common to the use of certain orthopedic implants,

the potential for: Pain, loss of joint function, adverse tissue

reaction, infection, and device failure. Since classification of these

devices, the agency has provided more information about the agency's

present conclusions regarding the labeling, testing, and manufacturing

information required for 510(k) clearance of orthopedic devices, the

agency has also developed the guidance document titled ``Guidance

Document for the Preparation of Premarket Notification (510(k))

Application for Orthopedic Devices.'' Because the information contained

in these guidance documents will help minimize the risks to health, the

agency is proposing that these guidances be applied as a special

control for the following four devices: Elbow joint metal/polymer

constrained cemented prosthesis (Sec. 888.3150), knee joint

patellofemoral polymer/metal semi-constrained cemented prosthesis

(Sec. 888.3540), shoulder joint metal/polymer nonconstrained cemented

prosthesis (Sec. 888.3650), and shoulder joint metal/polymer semi-

constrained cemented prosthesis (Sec. 888.3660).

B. Consensus Standards

FDA has a long history of participating in the development of

consensus standards relating to the safety and effectiveness of medical

devices. These consensus standards are developed by independent

standards organizations based upon discussions among experts from

industry, the agency, and other interested parties, and after a series

of ballots on draft and final documents. Consensus standards define

terminology, describe test methods, and set performance limits for a

given product or products. The agency believes that conformity with a

consensus standard helps to ensure acceptable quality and performance

of the device to which the standard is applied. The use of standards

helps to ensure the safety and effectiveness of the devices to which

the consensus standards apply, and it helps to

[[Page 12781]]

minimize the potential risks to health associated with the use of these

devices.

Section 204 of FDAMA amended section 514 of the act (21 U.S.C.

360d) to allow the agency to recognize consensus standards established

by international and national standards development organizations for

use in certain regulatory decision making concerning devices. On

February 25, 1998 (63 FR 9561), FDA issued a notice of availability of

a guidance entitled ``Guidance on the Recognition and Use of Consensus

Standards'' and also published in that document a list of the consensus

standards that FDA was recognizing for use in the premarket review

process. FDA will update this list at least annually.

Consensus standards are periodically updated as new information

becomes available. When a consensus standard that has been identified

as a special control is revised, the agency will publish in the Federal

Register a proposal to amend the special controls for the relevant

devices to include the revised consensus standard.

Accordingly, the agency is proposing that the following consensus

standards be adopted as special controls for the devices identified:

1. American Society for Testing and Materials (ASTM) Standards

These standards may be obtained from ASTM Customer Services, 100

Barr Harbor Dr., West Conshohocken, PA 19428 (Telephone 610-832-9585).

ASTM also maintains a site on the WWW at the address ``http://

www.astm.org''.

a. The following standard is proposed as a special control for the

cutaneous oxygen monitor (21 CFR 868.2500) ASTM F984-86:

``Specification for Cutaneous Gas Monitoring Devices for Oxygen and

Carbon Dioxide.''

b. The following seven standards are proposed as special controls

for the elbow joint metal/polymer constrained cemented prosthesis

(Sec. 888.3150), the knee joint patellofemoral polymer/metal semi-

constrained cemented prosthesis (Sec. 888.3540), the shoulder joint

metal/polymer nonconstrained cemented prosthesis (Sec. 888.3650), and

the shoulder joint metal/polymer semi-constrained cemented prosthesis

(Sec. 888.3660):

(1) ASTM F75-92 ``Specification for Cast Cobalt-Chromium-

Molybdenum Alloy for Surgical Implant Material,''

(2) STM F799-96 ``Specification for Cobalt-28 Chromium-6

Molybdenum Alloy Forgings for Surgical Implants,''

(3) ASTM F1108-97 ``Specification for Ti6Al4V Alloy Castings for

Surgical Implants,''

(4) ASTM F648-96 ``Specification for Ultra-High-Molecular-Weight

Polyethylene Powder and Fabricated Form for Surgical Implants,''

(5) ASTM F1537-94 ``Specification for Wrought Cobalt-Chromium-

Molybdenum Alloy for Surgical Implants,''

(6) ASTM 1044 ``Test Method for Shear Testing of Porous Metal

Coatings,'' and

(7) ASTM 1147 ``Test Method for Tension Testing of Porous Metal

Coatings.''

c. The following standards are proposed as special controls for

the knee joint patellofemoral polymer/metal semi-constrained cemented

prosthesis (Sec. 888.3540):

(1) ASTM F370-94 ``Specification for Proximal Femoral

Prosthesis,''

(2) ASTM F1672-95 ``Specification for Resurfacing Patellar

Prosthesis,'' and

(3) ASTM F1223-96 ``Test Method for Determination of Total Knee

Replacement Constraint.''

d. The following standard is proposed as a special control for the

elbow joint metal/polymer constrained cemented prosthesis

(Sec. 888.3150) ASTM 981: ``Practice for Assessment of Compatibility of

Biomaterials for Surgical Implant with Respect to Effect of Material on

Muscle and Bone.''

e. The following standard is proposed as a special control for the

shoulder joint metal/polymer nonconstrained cemented prosthesis

(Sec. 888.3650), and the shoulder joint metal/polymer semi-constrained

cemented prosthesis (Sec. 888.3660) ASTM 1378: ``Specification for

Shoulder Prosthesis.''

2. American National Standards Institute/American Association for

Medical Instrumentation (ANSI/AAMI)

These standards may be obtained from ANSI/AAMI, 11 West 42d St.,

New York, NY 10036. ANSI also maintains a site on the world wide web at

``http://www.ansi.org''. FDA proposes the following ANSI/AAMI standards

as special controls for the identified devices:

a. ANSI/AAMI DF2 ``Cardiac Defibrillator Devices'' as applied to

the external transcutaneous cardiac pacemaker (noninvasive) (21 CFR

870.5550);

b. ANSI/AAMI/ISO 11135 ``Medical Devices-Validation and Routine

Control of Ethylene Oxide Sterilization'' as applied to the peritoneo-

venous shunt (Sec. 876.5955); and

c. ANSI/AAMI HF-18 ``Electrosurgical Devices'' as applied to the

endoscopic electrocautery and accessories (Sec. 884.4100), the bipolar

endoscopic coagulator-cutter and accessories (Sec. 884.4150), and the

electrohydraulic lithotriptor (Sec. 876.4480).

3. International Standards Organization (ISO).

These standards may be obtained from International Organization

for Standardization, Case Postale, Geneva, Switzerland, CH-1121. ISO

also maintains a site on the world wide web at ``http://www.iso.org''.

a. FDA proposes the following ISO standards as special controls

for the elbow joint metal/polymer constrained cemented prosthesis

(Sec. 888.3150), the knee joint patellofemoral polymer/metal semi-

constrained cemented prosthesis (Sec. 888.3540), the shoulder joint

metal/polymer nonconstrained cemented prosthesis (Sec. 888.3650), and

the shoulder joint metal/polymer semi-constrained cemented prosthesis

(Sec. 888.3660):

(1) ISO 5832 ``Implants for Surgery-Metallic Materials;''

(2) ISO 5833 ``Implants for Surgery-Acrylic Resin Cements;'' and

(3) ISO 5834 ``Implants for Surgery-Ultra High Molecular Weight

Polyethylene;''

(4) ISO 9001 ``Quality Systems-Model for Quality Assurance in

Design/Development, Production, Installation, and Servicing;'' and

(5) ISO 6018 ``General Requirements for Marketing, Packaging, and

Labeling.''

b. The following ISO standard is proposed as a special control for

the elbow joint metal/polymer constrained cemented prosthesis

(Sec. 888.3150): ISO 14630 ``Non-active Surgical Implants-General

Requirements.''

c. The following ISO standard is proposed as a special control for

the knee joint patellofemoral polymer/metal semi-constrained cemented

prosthesis (Sec. 888.3540): ISO 7207 ``Implants for Surgery-Femoral and

Tibial Components for Partial and Total Knee Joint Prostheses.''

4. National Committee for Clinical Laboratory Standards (NCCLS)

Copies of these standards may be obtained from NCCLS Executive

Offices, 940 West Valley Rd., suite 1400, Wayne, PA 19087 (Telephone

610-688-0100). NCCLS also maintains a site on the WWW at ``http://

www.nccls.org''.

a. FDA proposes the following NCCLS standards as special controls

for the rubella virus serological reagents (Sec. 866.3510):

(1) NCCLS I/LA6 ``Evaluation and Performance Criteria for Multiple

Component Test Products Intended for the Detection and Quantitation of

Rubella IgG Antibody,''

[[Page 12782]]

(2) NCCLS D13 ``Agglutination Characteristics, Methodology,

Limitations, and Clinical Validation,''

(3) NCCLS I/LA18 ``Specifications for Immunological Testing for

Infectious Diseases,''

(4) NCCLS EP5 ``Evaluation of Precision Performance of Clinical

Chemistry Devices,'' and

(5) NCCLS EP10 ``Preliminary Evaluation of Quantitative Clinical

Laboratory Methods-Second edition, 1993.''

b. FDA proposes the following NCCLS standards as special controls

for the indwelling blood carbon dioxide partial pressure

(PCO2) analyzer (Sec. 868.1150), the indwelling blood

hydrogen ion concentration (pH) analyzer (Sec. 868.1170), and the

indwelling blood oxygen partial pressure (PO2) analyzer

(Sec. 868.1200):

(1) NCCLS EP5 ``Evaluation of Precision Performance of Clinical

Chemistry Devices,''

(2) NCCLS EP6 ``Evaluation of the Linearity of Quantitive

Analytical Methods,''

(3) NCCLS EP7 ``Interference Testing in Clinical

Chemistry''(PO2) analyzer (21 CFR 868.1200),

(4) NCCLS EP9 ``User Comparison of Quantitative Clinical

Laboratory Methods Using Patient Samples,'' and

(5) NCCLS EP10 ``Preliminary Evaluation of Quantitative Clinical

Laboratory Methods-Second edition, 1993.''

5. International Electrotechnical Commission (IEC)

Copies of these standards may be obtained from IEC, AT3, Rue de

Varembe, P.O. Box 131, Geneva, Switzerland, ch-1211. IEC also maintains

a site on the WWW at ``http://www.iec.ch''.

a. FDA proposes the following IEC standard as special controls for

the indwelling blood carbon dioxide partial pressure (PCO2)

analyzer (Sec. 868.1150), the indwelling blood hydrogen ion

concentration (pH) analyzer (Sec. 868.1170), the indwelling blood

oxygen partial pressure (PO2) analyzer (Sec. 868.1200), the

electrohydraulic lithotripter (Sec. 876.4480), the endoscopic

electrocautery and accessories (Sec. 884.4100), and the bipolar

endoscopic coagulator-cutter and accessories (Sec. 884.4150): IEC 60601

``Electrical Safety Standard.''

b. FDA also proposes the following IEC standard as a special

control for the cutaneous oxygen monitor (Sec. 868.2500), and the

airbrush (Sec. 872.6080): IEC 601 ``Medical Device Electrical

Standard.''

6. Underwriters Laboratory (UL)

These standards may be obtained from Underwriters Laboratories,

Inc., 333 Pfingsten Rd., Northbrook, IL 60062 (Telephone 847-272-8800).

UL also maintains a site on the WWW at ``http://www.ul.com''.

FDA proposes the following standard as a special control for the

cutaneous oxygen monitor (Sec. 868.2500): UL 2601-1 ``Standard for

Safety, Medical Electrical Equipment, Part 1: General Requirements for

Safety.''

7. The International Federation of Clinical Chemistry (IFCC)

These standards may be obtained from IFCC through their site on

the WWW at ``http://www.leeds.ac.uk/ifcc''.

FDA proposes the following standard as a special control for the

cutaneous oxygen monitor (Sec. 868.2500): ``IFCC Guidelines for

Transcutaneous PO2 and PCO2 Measurement.''

8. Centers for Disease Control and Prevention (CDC)

CDC has developed standards associated with the detection or

prevention of disease. These standards may be obtained from the Center

for Disease Control and Prevention, Mail Stop G18, 1600 Clifton Rd.,

NE., Atlanta, GA 30333.

FDA proposes the following CDC standards as special controls for

the rubella virus serological reagents (Sec. 866.3510):

(1) ``CDC Low Titer Rubella Standard'' as applied to

(2) ``CDC Reference Panel of Well Characterized Rubella Sera.''

9. World Health Organization International (WHO)

WHO has also developed standards associated with the detection or

prevention of disease. These standards may be obtained from the World

Health Organization International, Laboratory for Biological Standards,

Statens Seruminstitut, Center for Prevention and Control of Infectious

Diseases and Congenital Disorders, 5. Artillerivej, DK-2300 Copenhagen

S, Denmark. FDA proposes the following as a special control for the

identified device proposed for reclassification: ``WHO Rubella

Standard'' as applied to rubella virus serological reagents

(Sec. 866.3510).

C. Device-specific Labeling

When considering the preamendments devices, the panels identified

certain risks to health which would result from the improper use of a

device, or use in improper circumstances. The agency believes that

general labeling controls such as adequate directions for use, as

required by section 502(f) of the act (21 U.S.C. 352(f)), and the

labeling requirements for medical devices in 21 CFR part 801, and for

in vitro diagnostic products at 21 CFR 809.10 minimize the potential

for most identified risks to health.

However, the agency recognizes that, for certain devices, the

general labeling requirements are not sufficiently specific to

adequately address and minimize specifically identified risks to

health. These risks may be addressed by a more specific labeling

regulation (e.g., 21 CFR part 801, subpart H), by guidance, or by

issuing specific labeling as a special control. Indeed, several of the

FDA guidance documents, which have been identified in this proposal as

special controls, contain a section on device labeling. For other

devices, no device-specific labeling is addressed in regulations or FDA

guidance, although the agency believes that device-specific labeling

would be an appropriate special control. Labeling is being proposed as

a special control for the following devices:

1. Tinnitus masker (Sec. 874.3400)

The agency is proposing that the professional labeling of this

device contain patient information that describes the risks, benefits,

warnings for safe use, and technical specifications of the device in

terminology understandable to the average layman. Patient information

would also include recommending that the patient seek medical

consultation to determine the cause of tinnitus, fitting of the device,

and followup care by a hearing health care professional.

2. Tympanostomy tube with semipermeable membrane (Sec. 874.3930)

The agency is proposing that the labeling for this device describe

the risk of clogging, and state that the device is intended for use

only in ears that have been evacuated.

3. Endometrial aspirator (Sec. 884.1060), endometrial brush

(Sec. 884.1100), and endometrial washer (Sec. 884.1185)

The agency is proposing that the labeling for these devices state

that the device is only intended as an adjunctive tool to evaluate the

endometrium, and that it is contraindicated in cases of pregnancy,

history of uterine perforation, and recent cesarean section.

Furthermore, the agency is proposing that the labeling of the

endometrial washer (Sec. 884.1185) also contain a statement warning

that the device should not be attached to wall or any external suction.

4. Endoscopic electrocautery and accessories (Sec. 884.4100) and

bipolar endoscopic coagulator-cutter and accessories (Sec. 884.4150)

The agency is proposing that the labeling of these devices:

Contain an indication for use statement: ``for female tubal

sterilization,'' contain instructions for use that recommend

destruction of at least 2 cm of the fallopian tube, use of a ``cut''

(or undamped sinusoidal)

[[Page 12783]]

waveform, and use of minimum power of 25 watts. For devices that have

ammeters, the labeling must state that activation of electrode is

recommended for 4 to 5 seconds after the visual endpoint is reached or

current flow ceases, to achieve complete destruction of tissue.

D. Design and Performance Testing.

The agency has often relied upon consensus standards for the

establishment of design specifications for medical devices. For certain

devices for which neither consensus standards nor FDA guidances are

available to address critical design or performance criteria, the

agency believes it is appropriate to identify design specifications and

performance testing as a special control. Accordingly, design

specifications and performance testing are proposed as special controls

for the following devices:

1. External transcutaneous cardiac pacemaker (noninvasive)

(Sec. 870.5550)

The agency is proposing that this device shall not have the

capability of delivering pulses in excess of 200 microamperes with a

width less than or equal to 50 milliseconds.

2. Tympanostomy tube with semipermeable membrane (Sec. 874.3930)

The agency is proposing that the membrane material be

polytetrafluoroethylene (PTFE) sheeting with no more than a 1-micron

pore size and 0.003-inch thickness. Furthermore, the agency is

proposing to require functional testing of these devices to verify air

passage.

3. Peritoneo-venous shunt (Sec. 876.5955)

The agency is proposing that these devices provide a specification

for backflow that ensures against excessive reflux of blood into the

shunt. Furthermore, the agency is proposing that these devices undergo

pyrogenicity testing using either the U.S. Pharmacopeia (USP) Rabbit

Pyrogen Test or USP Bacterial Endotoxins Test.

4. Endometrial aspirator (Sec. 884.1060), endometrial brush

(Sec. 884.1100), and endometrial washer (Sec. 884.1185)

The agency is proposing that these devices be designed such that

the sampling part of the device is covered while entering and leaving

the vagina. Furthermore, the agency is proposing that the endometrial

brush (Sec. 884.1100) be tested to demonstrate adequate adherence of

bristles and brush head, and the endometrial washer (Sec. 884.1185)

undergo testing to demonstrate that maximum intrauterine pressure does

not exceed 50 millimeters of mercury.

VI. Summary of Special Controls

[[Page 12784]]

Table 2.--Summary of Special Controls Listed by Device1

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

FDA Sterility Review FDA Biocompatibility Design Controls,

CFR Section Device Name Guidance Guidance Other FDA Guidance1 Labeling Standards Performance Testing

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

864.7250............ Erythropoietin assay 1

864.7300............ Fibrin monomer 2

paracoagulation test

866.3510............ Rubella virus NCCLS\2\ 1/LA6; 1/

serological reagents LA18; D13; EP5; EP10;

CDC\3\ Low Titer

Rubella Standard;

WHO\4\ Rubella

Standard;

CDC Reference Panel of

Well Characterized

Rubella Sera

868.1150............ Indwelling blood carbon X X 3, 4, 5 NCCLS standards,

dioxide partial EP5,

pressure (PCO2) EP6,

analyzer EP7,

EP9,

EP10,

IEC\5\ 60601

868.1170............ Indwelling blood X X 3, 4, 5 NCCLS standards,

hydrogen ion EP5,

concentration (pH) EP6,

analyzer EP7,

EP9,

EP10,

IEC 60601

868.1200............ Indwelling blood oxygen X X 3, 4, 5 NCCLS standards,

partial pressure (PO2) EP5,

analyzer EP6,

EP7,

EP9,

EP10,

IEC 60601

868.2500(b)......... Cutaneous oxygen ASTM\6\ F984-86,

monitor IEC 601,

UL\7\ 2601-1,

IFCC\8\ Guidelines for

Transcutaneous PO2

and PCO2 Measurement

870.3375............ Cardiovascular X X 6

intravascular filter

870.3450............ Vascular graft X X 7

prosthesis of less

than 6 millimeters

diameter

870.3620............ Pacemaker lead adaptor X X 8

870.3800............ Annuloplasty ring X X 9

[[Page 12785]]

870.4230............ Cardiopulmonary bypass X X 10

defoamer

870.4260............ Cardiopulmonary bypass X X 11

arterial line blood

filter

870.4350............ Cardiopulmonary bypass X X 12

oxygenator

870.5550............ External transcutaneous ANSI/AAMI\9\ DF-2 Shall not have

cardiac pacemaker capability of

(noninvasive) delivering pulses in

excess of 200

milliamperes with a

width less than or

equal to 50

milliseconds

872.3540............ OTC denture cushion or X 13

pad

872.3560............ OTC denture reliner X 13

872.3570............ OTC denture repair kit X 13

872.3600............ Partially fabricated X 13

denture kit

872.6080............ Airbrush IEC-601

874.3400............ Tinnitus masker Patient labeling re:

medical consultation,

fitting and follow-up

care by a hearing

health care

professional, risks,

benefits, warnings

for safe use, and

technical

specifications

874.3930............ Tympanostomy tube with 14 Risk of clogging Functional testing to

semipermeable membrane described, verify air passage.

Only for use in ears Membrane material:

that have been PTFE\10\ sheeting;

evacuated 1.0 micron pore size;

0.003 inch thickness

876.4480............ Electrohydraulic X 15, 16 ANSI/AAMI HF-18

lithotriptor IEC 60601

876.5860............ High permeability X 17, 18, 19, 20

hemodialysis system

876.5955............ Peritoneo-venous shunt X X Sterilization Pyrogeni-city testing

validation per ANSI/ per USP Rabbit

AAMI/ISO\11\ 11135 Pyrogen Test or USP

Bacterial Endotoxins

Test,

Backflow

specifications that

ensure against

excessive reflux of

blood into the shunt

878.3610............ Esophageal prosthesis 21

878.3720............ Tracheal prosthesis 21

[[Page 12786]]

884.1060............ Endometrial aspirator X X Only for use as an Device design to

adjunctive tool to ensure that the

evaluate the sampling part is

endometrium; covered while

Contraindications: entering and leaving

pregnancy, history of vagina

uterine perforation,

and recent cesarean

section

884.1100............ Endometrial brush X X Only for use as an Device design so that

adjunctive tool to the sampling part is

evaluate the covered within the

endometrium; vagina,

Contraindication: Testing to demonstrate

pregnancy, history of adequate adherence of

uterine perforation, bristles and brush

and recent cesarean head

section

884.1185............ Endometrial washer X X Only for use as an Testing to demonstrate

adjunctive tool to that maximum

evaluate the intrauterine pressure

endometrium; should not exceed 50

Device should not be millimeters of

attached to wall or mercury;

any external suction; Device design so that

Contraindications: the sampling part is

Pregnancy, history of covered within vagina

uterine perforation,

and recent cesarean

section

884.4100............ Endoscopic X 22 Indication: female IEC 60601 ANSI/AAMI HF-

electrocautery and tubal sterilization; 18

accessories Treatment

instructions:

``destruction of at

least 2 cm of

tube,''; use of a

cut' (or undamped

sinusoidal) waveform,

and minimum power of

25 watts;

For devices with

ammeters: activation

of electrode for 4 to

5 seconds

[[Page 12787]]

884.4150............ Bipolar endoscopic X 22 Indication: female IEC 60601 or ANSI/AAMI

coagulator-cutter tubal sterilization; HF-18

Treatment

instructions:

``destruction of at

least 2 centimeters

of tube,''; use of a

cut' (or undamped

sinusoidal) waveform,

and minimum power of

25 watts;

For devices with

ammeters: activation

of electrode for 4 to

5 seconds

886.3400............ Keratoprosthesis X X 23

886.3920............ Eye valve implant X X 24

888.3150............ Elbow joint metal/ X X 25, 26, 27 ASTM F75-92, F799-96,

polymer constrained F1108-97, F648-96,

cemented prosthesis F1537-94, F981,

F1044, F1147;

ISO 5832, 5833, 5834,

14630, 10993, 9001,

6018

888.3540............ Knee joint X X 25, 26, 27 ASTM F75-92, F799-96,

patellofemoral polymer/ F1108-97, F648-96,

metal semi-constrained F1537-94, F1044,

cemented prosthesis F1147, F1223-96, F370-

94; F1672-95

ISO 5832, 5833, 5834,

6018, 7207, 9001

888.3650............ Shoulder joint metal/ X X 25, 26, 27 ASTM F75-92, F799-96,

polymer non- F1108-97, F648-96,

constrained cemented F1537-94, F1044,

prosthesis F1147, F1378;

ISO 5832, 5833, 5834,

6018, 9001

888.3660............ Shoulder joint metal/ X X 25, 26, 27 ASTM F75-92, F799-96,

polymer semi- F1108-97, F648-96,

constrained cemented F1537-94, F1044,

prosthesis F1147, F1378;

ISO 5832, 5833, 5834,

6018, 9001

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

\1\ The following is a list of guidances FDA has developed to inform manufactures of how the agency evaluates the safety and effectiveness of devices and reaches determination of substantial

equivalency:

(1) ``Guidance Document for the Submission of Erythropoietin Assay Premarket Notification (510(k)),''

(2) ``Guidance Document for the Submission of Fibrin Monomer Paracoagulation Test Premarket Notification (510(k)),''

(3)``Reviewer Guidance for Clinical Studies and Labeling for Indwelling Blood Gas Analyzers,''

(4) ``Reviewer Guidance for Computer Controlled Medical Devices Undergoing 510(k) Review,''

(5) ``Reviewer Guidance for Format and Content of Premarket Notifications (510(k) Submissions): Labeling, Performance and Environmental Testing for Electronic Devices,''

(6) ``Guidance Document for the Submission of 510(k) Premarket Notifications for Cardiovascular Intravascular Filters,''

(7) ``Document for Special Controls for Vascular Prostheses,''

[[Page 12788]]

(8) ``Guidance Document for the Submission of Research and Marketing Applications for Permanent Pacemaker Leads,''

(9) ``Document for Special Controls for the Cardiopulmonary Bypass Defoamer,''

(10) ``Document for Special Controls for the Cardiopulmonary Bypass Arterial Filter,''

(11) ``Information for Manufacturers Seeking Marketing Clearance for Blood-Gas Exchangers (Oxygenators) Used in Cardiopulmonary Bypass,''

(12) ``Document for Special Controls for Annuloplasty Rings,''

(13) ``OTC Denture Reliners, Repair Kits, and Partially Fabricated Denture Kits,''

(14) ``Tympanostomy Tubes, Submission Guidance for a 510(k) Premarket Notification,''

(15) ``Guidance for the Content of Premarket Notifications for Intracorporal Lithotripters,''

(16) ``Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: FDA Reviewer Guidance,''

(17) ``Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,''

(18) ``Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,''

(19) ``Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,''

(20) ``Guidance for Hemodialyzer Reuse Labeling,''

(21) ``Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses,''

(22) ``Guidance for Evaluation of Laproscopic Bipolar and Thermal Coagulators (and Accessories),''

(23) ``Keratoprosthesis Guidance Document,''

(24) ``Aqueous Shunt-510(k) Submission,''

(25) ``Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,''

(26) ``Guidance Document for Testing Non-articulating, Mechanically Locked' Modular Implant Components,'' and

(27) ``Guidance Document for the Preparation of Premarket Notification (510(k) Applications for Orthopedic Devices.''

\2\ National Committee for Clinical Laboratory Standards.

\3\ Centers for Disease Control and Prevention.

\4\ World Health Organization.

\5\ International Electrotechnical Commission.

\6\ American National Standards Institute.

\7\ Underwriters Laboratories.

\8\ International Federation of Clinical Chemistry.

\9\ Association for the Advancement of Medical Instrumentation.

\10\ Polytetrafluoroethylene.

\11\ International Standards Organization.

[[Page 12789]]

VII. Environmental Impact

The agency has determined under 21 CFR 25.34(b) that this proposed

classification action is of a type that does not individually or

cumulatively have a significant effect on the human environment.

Therefore, neither an environmental assessment nor an environmental

impact statement is required.

VIII. Analysis of Impacts

FDA has examined the impacts of the proposed rule under Executive

Order 12866, and the Regulatory Flexibility Act (5 U.S.C 601-612) (as

amended by subtitle D of the Small Business Regulatory Fairness Act of

1996 (Pub. L 104-121), and the Unfunded Mandates Reform Act of 1995

(Pub. L. 104-4)). Executive Order 12866 directs agencies to assess all

costs and benefits of available regulatory alternatives and, when

regulation is necessary, to select regulatory approaches that maximize

net benefits (including potential economic, environmental, public

health and safety and other advantages distributive impacts and

equity). The agency believes that this proposed rule is consistent with

the regulatory philosophy and principles identified in the Executive

Order. In addition, the proposed rule is not a significant regulatory

action as defined by the Executive Order and so is not subject to

review under the Executive Order.

The Regulatory Flexibility Act requires agencies to analyze

regulatory options that would minimize any significant impact of a rule

on small entities. Reclassification of these devices from class III

will relieve all manufacturers of these devices of the cost of

complying with the premarket approval requirements in section 515 of

the act. Moreover, compliance with special controls proposed for these

devices will not impose significant new costs on affected manufacturers

as most of these devices already comply with the proposed special

controls. Because reclassification will reduce regulatory costs with

respect to these devices, it will impose no significant economic impact

on any small entities, and it may permit small potential competitors to

enter the marketplace by lowering their costs. The agency therefore

certifies that this proposed rule, if issued, will not have a

significant economic impact on a substantial number of small entities.

In addition, this proposed rule will not impose costs of $100 million

or more on either the private sector or state, local, and tribal

governments in the aggregate, and therefore a summary statement of

analysis under section 202(a) of the Unfunded Mandates Reform Act of

1995 is not required.

IX. Paperwork Reduction Act of 1995

FDA tentatively concludes that this proposed rule contains no

collections of information. Therefore, clearance by the Office of

Management and Budget under the Paperwork Reduction Act of 1995 is not

required.

X. Submission of Comments and Proposed Effective Dates

Interested persons may, on or before June 14, 1999 submit to the

Dockets Management Branch (address above) written comments regarding

this proposal. Two copies of any comments are to be submitted except

that individuals may submit one copy. Comments are to be identified

with the docket number found in brackets in the heading of this

document. Received comments may be seen in the office above between 9

a.m. and 4 p.m., Monday through Friday.

FDA proposes that any final regulation that may issue based on

this proposal becomes effective 30 days after its date of publication

in the Federal Register.

List of Subjects

21 CFR Part 864

Blood, Medical devices, Packaging and containers.

21 CFR Part 866

Biologics, Laboratories, Medical devices.

21 CFR Parts 868, 870, 872, 874, 876, 878, 884, and 888

Medical devices.

21 CFR Part 886

Medical devices, Ophthalmic goods and services.

Therefore, under the Federal Food, Drug, and Cosmetic Act and under

authority delegated to the Commissioner of Food and Drugs, it is

proposed that 21 CFR parts 864, 866, 868, 870, 872, 874, 876, 878, 884,

886, and 888 be amended as follows:

PART 864--HEMATOLOGY AND PATHOLOGY DEVICES

1. The authority citation for 21 CFR part 864 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

2. Section 864.7250 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 864.7250 Erythropoietin assay.

* * * * *

(b) Classification. Class II. The special control for this device

is FDA's ``Guidance Document for Submission of Erythropoietin Assay

Premarket Notification (510(k)).''

3. Section 864.7300 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 864.7300 Fibrin monomer paracoagulation test.

* * * * *

(b) Classification. Class II. The special control for this device

is FDA's ``Guidance Document for Submission of Fibrin Monomer

Paracoagulation Test Premarket Notification (510(k)).''

PART 866--IMMUNOLOGY AND MICROBIOLOGY DEVICES

4. The authority citation for 21 CFR part 866 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

5. Section 866.3510 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 866.3510 Rubella virus serological reagents.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) National Committee for Clinical Laboratory Standards':

(i) I/LA6 ``Evaluation and Performance Criteria for Multiple

Component Test Products Intended for the Detection and Quantitation of

Rubella IgG Antibody,''

(ii) 1/LA18 ``Specifications for Immunological Testing for

Infectious Diseases,''

(iii) D13 ``Agglutination Characteristics, Methodology,

Limitations, and Clinical Validation,''

(iv) EP5 ``Evaluation of Precision Performance of Clinical

Chemistry Devices,'' and

(v) EP10 ``Evaluation of the Linearity of Quantitive Analytical

Methods,''

(2) Centers for Disease Control's:

(i) Low Titer Rubella Standard,

(ii) Reference Panel of Well Characterized Rubella Sera, and

(3) World Health Organization's International Rubella Standard.

PART 868--ANESTHESIOLOGY DEVICES

6. The authority citation for 21 CFR part 868 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

7. Section 868.1150 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

[[Page 12790]]

Sec. 868.1150 Indwelling blood carbon dioxide partial pressure

(PCO2) analyzer.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,''

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,''

(ii) ``Reviewer Guidance for Computer Controlled Medical Devices

Undergoing 510(k) Review,''

(iii) ``Reviewer Guidance for Clinical Studies for Indwelling Blood

Gas Analyzers,'' and

(iv) ``Reviewer Guidance for Format and Content of Premarket

Notifications (510(k) Submissions), Labeling, Performance and

Environmental Testing for Electronic Devices,''

(3) National Committee for Clinical Laboratory Standards':

(i) EP5 ``Evaluation of Precision Performance of Clinical Chemistry

Devices,''

(ii) EP6 ``Evaluation of the Linearity of Quantitive Analytical

Methods,''

(iii) EP7 ``Interference Testing in Clinical Chemistry,''

(iv) EP9 ``User Comparison of Quantitative Clinical Laboratory

Methods Using Patient Samples,'' and

(v) EP10 ``Preliminary Evaluation of Quantitative Clinical

Laboratory Methods--Second edition, 1993,'' and

(4) International Electrotechnical Commission's 60601 ``Electrical

Safety Standard.''

8. Section 868.1170 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 868.1170 Indwelling blood hydrogen ion concentration (pH)

analyzer.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,''

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,''

(ii) ``FDA Reviewer Guidance for Computer Controlled Medical

Devices Undergoing 510(k) Review,''

(iii) ``Reviewer Guidance for Clinical Studies for Indwelling Blood

Gas Analyzers,'' and

(iv) ``Reviewer Guidance for Format and Content of Premarket

Notifications (510(k) Submissions), Labeling, Performance and

Environmental Testing for Electronic Devices,''

(3) National Committee for Clinical Laboratory Standards':

(i) EP5 ``Evaluation of Precision Performance of Clinical Chemistry

Devices,''

(ii) EP6 ``Evaluation of the Linearity of Quantitive Analytical

Methods,''

(iii) EP7 ``Interference Testing in Clinical Chemistry,''

(iv) EP9 ``User Comparison of Quantitative Clinical Laboratory

Methods Using Patient Samples,'' and

(v) EP10 ``Preliminary Evaluation of Quantitative Clinical

Laboratory Methods--Second edition, 1993,'' and

(4) International Electrotechnical Commission's 60601 ``Electrical

Safety Standard.''

9. Section 868.1200 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 868.1200 Indwelling blood oxygen partial pressure

(PO2) analyzer.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) FDA's:

(i) ``Reviewer Guidance for Computer Controlled Medical Devices

Undergoing 510(k),''

(ii) ``Reviewer Guidance for Clinical Studies for Indwelling Blood

Gas Analyzers,''

(iii) ``Reviewer Guidance for Format and Content of Premarket

Notifications (510(k) Submissions), Labeling, Performance and

Environmental Testing for Electronic Devices,'' and

(iv) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-10,''

(2) National Committee for Clinical Laboratory Standards':

(i) EP5 ``Evaluation of Precision Performance of Clinical Chemistry

Devices,''

(ii) EP6 ``Evaluation of the Linearity of Quantitive Analytical

Methods,''

(iii) EP7 ``Interference Testing in Clinical Chemistry,''

(iv) EP9 ``User Comparison of Quantitative Clinical Laboratory

Methods Using Patient Samples,'' and

(v) EP10 ``Preliminary Evaluation of Quantitative Clinical

Laboratory Methods--Second edition, 1993,''

(3) International Electrotechnical Commission's 60601 ``Electrical

Safety Standard,'' and

(4) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing.''

10. Section 868.2500 is amended by revising paragraph (b)(2) and by

removing paragraph (c) to read as follows:

Sec. 868.2500 Cutaneous oxygen monitor.

* * * * *

(b) * * *

(2) Classification. Class II. The special controls for this device

are:

(i) American Society for Testing and Materials' F984-86

``Specification for Cutaneous Gas Monitoring Devices for Oxygen and

Carbon Dioxide,''

(ii) International Electrotechnical Commission's IEC 601 ``Medical

Device Electrical Standard,''

(iii) Underwriters Laboratory's ``Medical Electrical Equipment (UL

2601-1),'' and

(iv) The International Federation of Clinical Chemistry's

``Guidelines for Transcutaneous PO2 and PCO2

Measurement.''

PART 870--CARDIOVASCULAR DEVICES

11. The authority citation for 21 CFR part 870 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

12. Section 870.3375 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 870.3375 Cardiovascular intravascular filter.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-10,'' and

(ii) ``Guidance Document for the Submission of 510(k) Premarket

Notifications for Cardiovascular Intravascular Filters.''

13. Section 870.3450 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 870.3450 Vascular graft prosthesis of less than 6 millimeters

diameter.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-10,'' and

(ii) ``Document on Special Controls for Vascular Prostheses.''

14. Section 870.3620 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 870.3620 Pacemaker lead adaptor.

* * * * *

[[Page 12791]]

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,'' and

(ii) ``Guidance Document for the Submission of Research and

Marketing Applications for Permanent Pacemaker Leads.''

15. Section 870.3800 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 870.3800 Annuloplasty rings.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,'' and

(ii) ``Document for Special Controls for Annuloplasty Rings,'' and

(2) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing.''

16. Section 870.4230 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 870.4230 Cardiopulmonary bypass defoamer.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) FDA's:

(i) ``Bluebook Guidance for Sterility, K90-1,''

(ii) ``Document for Special Controls for Cardiopulmonary Bypass

Defoamer,'' and

(2) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing.''

17. Section 870.4260 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 870.4260 Cardiopulmonary bypass arterial line blood filter.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) FDA's:

(i) ``Bluebook Guidance for Sterility, K90-1,'' and

(ii) ``Document for Special Controls for Cardiopulmonary Bypass

Arterial Filters,'' and

(2) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing.''

18. Section 870.4350 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 870.4350 Cardiopulmonary bypass oxygenator.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,'' and

(ii) ``Information for Manufacturers Seeking Marketing Clearance

for Blood Gas Exchangers (Oxygenators) Used in Cardiopulmonary

Bypass.''

19. Section 870.5550 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 870.5550 External transcutaneous cardiac pacemaker (noninvasive).

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) ``American National Standards Institute/American Association

for Medical Instrumentation's DF-2,'' and

(2) The device shall not have capability of delivering pulses in

excess of 200 milliamperes with a width less than or equal to 50

milliseconds.

PART 872--DENTAL DEVICES

20. The authority citation for 21 CFR part 872 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

21. Section 872.3540 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 872.3540 OTC denture cushion or pad.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's ``OTC Denture Reliners, Repair Kits, and Partially

Fabricated Denture Kits.''

22. Section 872.3560 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 872.3560 OTC denture reliner.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's ``OTC Denture Reliners, Repair Kits, and Partially

Fabricated Denture Kits.''

23. Section 872.3570 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 872.3570 OTC denture repair kit.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's ``OTC Denture Reliners, Repair Kits, Partially Fabricated

Denture Kits.''

24. Section 872.3600 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 872.3600 Partially fabricated denture kit.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's ``OTC Denture Reliners, Repair Kits, Partially Fabricated

Denture Kits.''

25. Section 872.6080 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 872.6080 Airbrush.

* * * * *

(b) Classification. Class II. The special control for this device

is International Electrotechnical Commission's IEC-601 ``Medical Device

Electrical Standard.''

PART 874--EAR, NOSE, AND THROAT DEVICES

26. The authority citation for 21 CFR part 874 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

27. Section 874.3400 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 874.3400 Tinnitus masker.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) Patient labeling to include information about:

(i) Risks,

(ii) Benefits,

(iii) Warnings for safe use, and

(iv) Technical specifications, and

(2) Medical consultation for:

[[Page 12792]]

(i) Determination of the cause of tinnitus,

(ii) Fitting of the device, and

(iii) Followup care by a hearing health care professional.

28. Section 874.3930 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 874.3930 Tympanostomy tube with semipermeable membrane.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) FDA's ``Tympanostomy Tubes, Submission Guidance for a 510(k),''

(2) Functional testing to verify air passage,

(3) Use of polytetrafluoroethylene sheeting with 1.0 micron pore

size and 0.003 inch thickness as membrane material, and

(4) Labeling to:

(i) Describe risk of clogging, and

(ii) State that device is only for use in ears that have been

evacuated.

PART 876--GASTROENTEROLOGY-UROLOGY DEVICES

29. The authority citation for 21 CFR part 876 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

30. Section 876.4480 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 876.4480 Electrohydraulic lithotriptor.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,''

(ii) ``Guidance for the Content of Premarket Notifications for

Intracorporeal Lithotripters,'' and

(iii) ``Labeling Reusable Medical Devices for Reprocessing in

Health Care Facilities: FDA Reviewer Guidance,''

(2) American National Standards Institute/American Association for

Medical Instrumentation's HF-18 ``Electrosurgical Devices,'' and

(3) International Electrotechnical Commission's IEC 60601

``Electrical Safety Standard.''

31. Section 876.5860 is amended by revising paragraphs (a) and (b)

and by removing paragraph (c) to read as follows:

Sec. 876.5860 High permeability hemodialysis system.

(a) Identification. A high permeability hemodialysis system is a

device intended for use as an artificial kidney system for the

treatment of patients with renal failure, fluid overload, or toxemic

conditions by performing such therapies as hemodialysis,

hemofiltration, and hemoconcentration. Using a hemodialyzer with a

semipermeable membrane that is more permeable to water than the

semipermeable membrane of the conventional hemodialysis system

described in Sec. 876.5820, the high permeability hemodialysis system

removes toxins or excess fluid from the patient's blood using the

principles of convection (via a high ultrafiltration rate) and/or

diffusion (via a concentration gradient in dialysate). During

treatment, blood is circulated from the patient through the

hemodialyzer's blood compartment, while the dialysate solution flows

countercurrent through the dialysate compartment. In this process,

toxins and/or fluid are transferred across the membrane from the blood

to the dialysate compartment. The hemodialysis delivery machine

controls and monitors the parameters related to this processing,

including the rate at which blood and dialysate are pumped through the

system, and the rate at which fluid is removed from the patient. The

high permeability hemodialysis system consists of the following

devices:

(1) The hemodialyzer consists of a semipermeable membrane with an

in vitro ultrafiltration coefficient (Kuf) greater than 12

milliliters per hour per conventional millimeter of mercury, and is

used with either an automated ultrafiltration controller or another

method of ultrafiltration control to prevent fluid imbalance.

(2) The hemodialysis delivery machine is similar to the

extracorporeal blood system and dialysate delivery system of the

hemodialysis system and accessories (Sec. 876.5820), with the addition

of an ultrafiltration controller and mechanisms that monitor and/or

control such parameters as fluid balance, dialysate composition, and

patient treatment parameters (e.g., blood pressure, hematocrit, urea,

etc.).

(3) The high permeability hemodialysis system accessories include,

but are not limited to, tubing lines and various treatment related

monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood

recirculation monitors).

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's:

(i) ``Guidance for the Content of 510(k)s for Conventional and High

Permeability Hemodialyzers,''

(ii) ``Guidance for Industry and CDRH Reviewers on the Content of

Premarket Notifications for Hemodialysis Delivery Systems,''

(iii) ``Guidance for the Content of Premarket Notifications for

Water Purification Components and Systems for Hemodialysis,'' and

(iv) ``Guidance for Hemodialyzer Reuse Labeling.''

32. Section 876.5955 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 876.5955 Peritoneo-venous shunt.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,''

(2) FDA's ``510(k) Sterility Review Guidance and Revision of 11/18/

94 K90-1,''

(3) ``Pyrogenicity Testing per USP Rabbit Pyrogen Test or USP

Bacterial Endotoxins Test,''

(4) American National Standards Institute/American Association for

Medical Instrumentation's ANSI/AAMI/ISO 11135 ``Medical Devices--

Validation and Routine Control of Ethylene Oxide Sterilization,'' and

(5) Specification for backflow that ensures against excessive

reflux of blood into the shunt.

PART 878--GENERAL AND PLASTIC SURGERY DEVICES

33. The authority citation for 21 CFR part 878 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

34. Section 878.3610 is amended by revising paragraphs (a) and (b)

and by removing paragraph (c) to read as follows:

Sec. 878.3610 Esophageal prosthesis.

(a) Identification. An esophageal prosthesis is a rigid, flexible,

or expandable tubular device constructed of a plastic, metal, or

polymeric material that is intended to be implanted to restore the

structure and/or function of the esophagus. The metal esophageal

prosthesis may be uncovered or covered with a polymeric material. This

device may also include a device delivery system.

(b) Classification. Class II. The special control for this device

is FDA's ``Guidance for the Content of Premarket

[[Page 12793]]

Notification Submissions for Esophageal and Tracheal Prostheses.''

35. Section 878.3720 is amended by revising paragraphs (a) and (b)

and by removing paragraph (c) to read as follows:

Sec. 878.3720 Tracheal prosthesis.

(a) Identification. The tracheal prosthesis is a rigid, flexible,

or expandable tubular device constructed of a silicone, metal, or

polymeric material that is intended to be implanted to restore the

structure and/or function of the trachea or trachealbronchial tree. It

may be unbranched or contain one or two branches. The metal tracheal

prosthesis may be uncovered or covered with a polymeric material. This

device may also include a device delivery system.

(b) Classification. Class II. The special control for this device

is FDA's ``Guidance for the Content of Premarket Notification

Submissions for Esophageal and Tracheal Prostheses.''

PART 884--OBSTETRICAL AND GYNECOLOGICAL DEVICES

36. The authority citation for 21 CFR part 884 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

37. Section 884.1060 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 884.1060 Endometrial aspirator.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,''

(2) FDA's ``510(k) Sterility Review Guidance and Revision of 11/18/

94 K90-1,''

(3) Device design so that sampling part is covered while entering

or leaving vagina, and

(4) Labeling to state that the device is only an adjunctive tool to

evaluate the endometrium and to contraindicate use of device in

pregnant patients and patients with a history of uterus perforation or

recent cesarean section.

38. Section 884.1100 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 884.1100 Endometrial brush.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,''

(2) FDA's ``510(k) Sterility Review Guidance and Revision of 11/18/

94 K90-1,''

(3) Device design so that sampling part is covered while entering

or leaving vagina,

(4) Testing to demonstrate adequate adherence of bristles and brush

head, and

(5) Labeling to state that the device is only an adjunctive tool to

evaluate the endometrium and to contraindicate use of device in

pregnant patients and patients with a history of uterus perforation or

recent cesarean section.

39. Section 884.1185 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 884.1185 Endometrial washer.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,''

(2) FDA's ``510(k) Sterility Review Guidance and Revision of 11/18/

94 K90-1,''

(3) Device design so that sampling part is covered while entering

or leaving vagina,

(4) Intrauterine pressure not to exceed 50 conventional millimeters

of mercury, and

(5) Labeling to:

(i) Contraindicate use of the device in pregnant patients and

patients with a history of uterus perforation or recent cesarean

section,

(ii) Warn that the device should not be attached to wall or any

other external source of suction, and

(iii) State that the device is only an adjunctive tool to evaluate

the endometrium.

40. Section 884.4100 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 884.4100 Endoscopic electrocautery and accessories.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,''

(2) FDA's ``Guidelines for Evaluation of Laproscopic Bipolar and

Thermal Coagulators,''

(3) International Electrotechnical Commission's IEC 60601

``Electrical Safety Standard,''

(4) American National Standards Institute/American Association for

Medical Instrumentation's HF-18 ``Electrosurgical Devices,''

(5) Labeling: an indication for female tubal sterilization,

(6) Treatment instructions to:

(i) Destroy at least 2 centimeters of a tube,

(ii) Use a cut or undampened sinusoidal waveform, and

(iii) Use a minimum power of 25 watts, and

(7) Labeling of devices with ammeters to activate the electrode for

4 to 5 seconds after the visual endpoint is reached or current flow

ceases for complete destruction of tissue.

41. Section 884.4150 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 884.4150 Bipolar endoscopic coagulator-cutter and accessories.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,''

(2) FDA's ``Guidance for Evaluation of Laproscopic Bipolar and

Thermal Coagulators (and Accessories),''

(3) International Electrotechnical Commission's IEC 60601

``Electrical Safety Standard,''

(4) American National Standards Institute/American Association for

Medical Instrumentation's HF-18 ``Electrosurgical Devices,''

(5) Labeling: An indication for female tubal sterilization,

(6) Treatment instructions to:

(i) Destroy at least 2 centimeters of a tube,

(ii) Use a cut or undampened sinusoidal waveform, and

(iii) Use a minimum power of 25 watts, and

(7) Labeling of devices with ammeters to activate the electrode for

4 to 5 seconds after the visual endpoint is reached or current flow

ceases for complete destruction of tissue.

PART 886--OPHTHALMIC DEVICES

42. The authority citation for 21 CFR part 886 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

43. Section 886.3400 is amended by revising paragraphs (a) and (b)

and by removing paragraph (c) to read as follows:

Sec. 886.3400 Keratoprosthesis.

(a) Identification. A keratoprosthesis is a device intended to

provide a

[[Page 12794]]

transparent optical pathway through an opacified cornea, either

intraoperatively or permanently, in an eye which is not a reasonable

candidate for a corneal transplant.

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,'' and

(ii) ``Keratoprosthesis Guidance Document.''

44. Section 886.3920 is amended by revising the section heading and

paragraphs (a) and (b) and by removing paragraph (c) to read as

follows:

Sec. 886.3920 Aqueous shunt.

(a) Identification. An aqueous shunt is a one-way, pressure

sensitive device intended to be implanted to normalize intraocular

pressure. The device is intended to treat neovascular glaucoma or

glaucomas where medical and conventional surgical treatment have

failed.

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,'' and

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,'' and

(ii) ``Aqueous Shunts--510(k) Submissions.''

PART 888--ORTHOPEDIC DEVICES

45. The authority citation for 21 CFR part 888 continues to read as

follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

46. Section 888.3150 is revised to read as follows:

Sec. 888.3150 Elbow joint metal/polymer constrained cemented

prosthesis.

(a) Identification. An elbow joint metal/polymer constrained

cemented prosthesis is a device intended to be implanted to replace an

elbow joint. It is made exclusively of alloys, such as cobalt-chromium-

molybdenum, or of these alloys and of an ultra-high molecular weight

polyethylene bushing. The device prevents dislocation in more than one

anatomic plane and consists of two components which are linked

together. This generic type of device is limited to those prostheses

intended for use with bone cement (Sec. 888.3027).

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,''

(i) ISO 5832 ``Implants for Surgery--Metallic Materials,''

(ii) ISO 5833 ``Implants for Surgery--Acrylic Resin Cements,''

(iii) ISO 5834 ``Implants for Surgery--Ultra High Molecular Weight

Polyethylene,''

(iv) ISO 14630 ``Non-active Surgical Implants--General

Requirements,''

(v) ISO 10993 ``Biocompatibility Test Methods,''

(vi) ISO 9001 ``Quality Systems--Model for Quality Assurance in

Design/Development, Production, Installation, and Servicing,'' and

(vii) ISO 6018 ``General Requirements for Marketing, Packaging, and

Labeling.''

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,''

(ii) ``Guidance Document for Testing Orthopedic Implants with

Modified Metallic Surfaces Apposing Bone or Bone Cement,''

(iii) ``Guidance Document for the Preparation of Premarket

Notification (510(k)) Application for Orthopedic Devices,'' and

(iv) ``Guidance Document for Testing Non-articulating,

`Mechanically Locked' Modular Implant Components,'' and

(3) American Society for Testing and Materials':

(i) F75-92 ``Specification for Cast Cobalt-Chromium-Molybdenum

Alloy for Surgical Implant Material,''

(ii) F799-96 ``Specification for Cobalt-28 Chromium-6 Molybdenum

Alloy Forgings for Surgical Implants,''

(iii) F1108-97 ``Specification for Ti6Al4V Alloy Castings for

Surgical Implants,''

(iv) F648-96 ``Specification for Ultra-High-Molecular-Weight

Polyethylene Powder and Fabricated Form for Surgical Implants,''

(v) F1537-94 ``Specification for Wrought Cobalt-Chromium-Molybdenum

Alloy for Surgical Implants,''

(vi) F981 ``Practice for Assessment of Compatibility of

Biomaterials (Nonporous) for Surgical Implant with Respect to Effect of

Material on Muscle and Bone,''

(vii) F1044 ``Test Method for Shear Testing of Porous Metal

Coatings,'' and

(viii) F1147 ``Test Method for Tension Testing of Porous Metal

Coatings.''

47. Section 888.3540 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 888.3540 Knee joint patellofemoral polymer/metal semi-constrained

cemented prosthesis.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's ISO 10993 ``Biological

Evaluation of Medical Devices Part I: Evaluation and Testing,''

(i) ISO 5832 ``Implants for Surgery--Metallic Materials,''

(ii) ISO 5833 ``Implants for Surgery--Acrylic Resin Cements,''

(iii) ISO 5834 ``Implants for Surgery--Ultra High Molecular Weight

Polyethylene,''

(iv) ISO 9001 ``Quality Systems--Model for Quality Assurance in

Design/Development, Production, Installation, and Servicing,''

(v) ISO 7207 ``Implants for Surgery--Femoral and Tibial Components

for Partial and Total Knee Joint Prostheses,'' and

(vi) ISO 6018 ``General Requirements for Marketing, Packaging, and

Labeling.''

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,''

(ii) ``Guidance Document for Testing Orthopedic Implants with

Modified Metallic Surfaces Apposing Bone or Bone Cement,''

(iii) ``Guidance Document for the Preparation of Premarket

Notification (510(k)) Applications for Orthopedic Devices,''

(iv) ``Guidance Document for Testing Non-articulating,

`Mechanically Locked' Modular Implant Components,'' and

(3) American Society for Testing and Materials':

(i) F75-92 ``Specification for Cast Cobalt-Chromium-Molybdenum

Alloy for Surgical Implant Material,''

(ii) F799-96 ``Specification for Cobalt-28 Chromium-6 Molybdenum

Alloy Forgings for Surgical Implants,''

(iii) F1108-97 ``Ti6Al4V Alloy Castings for Surgical Implants,''

(iv) F648-96 ``Specification for Ultra-High-Molecular-Weight

Polyethylene Powder and Fabricated Form for Surgical Implants,''

(v) F1537-94 ``Specification for Wrought Cobalt-Chromium-Molybdenum

Alloy for Surgical Implants,''

(vi) F1044 ``Test Method for Shear Testing of Porous Metal

Coatings,''

(vii) F1147 ``Test Method for Tension Testing of Porous Metal

Coatings,''

(viii) F370-94 ``Specification for Proximal Femoral Prosthesis,''

and

(ix) F1672-95 ``Specification for Resurfacing Patellar

Prosthesis.''

[[Page 12795]]

48. Section 888.3650 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 888.3650 Shoulder joint metal/polymer non-constrained cemented

prosthesis.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's:

(i) ISO 10993 ``Biological Evaluation of Medical Devices Part I:

Evaluation and Testing,''

(ii) ISO 5832 ``Implants for Surgery--Metallic Materials,''

(iii) ISO 5833 ``Implants for Surgery--Acrylic Resin Cements,''

(iv) ISO 5834 ``Implants for Surgery--Ultra High Molecular Weight

Polyethylene,''

(v) ISO 9001 ``Quality Systems--Model for Quality Assurance in

Design/Development, Production, Installation, and Servicing,'' and

(vi) ISO 6018 ``General Requirements for Marketing, Packaging, and

Labeling.''

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,''

(ii) ``Guidance Document for Testing Orthopedic Implants with

Modified Metallic Surfaces Apposing Bone or Bone Cement,''

(iii) ``Guidance Document for the Preparation of Premarket

Notification (510(k)) Application for Orthopedic Devices,'' and

(iv) ``Guidance Document for Testing Non-articulating,

`Mechanically Locked' Modular Implant Components,''

(3) American Society for Testing and Materials':

(i) F75-92 ``Specification for Cast Cobalt-Chromium-Molybdenum

Alloy for Surgical Implant Material,''

(ii) F799-96 ``Specification for Cobalt-28 Chromium-6 Molybdenum

Alloy Forgings for Surgical Implants,''

(iii) F1108-97 ``Ti6Al4V Alloy Castings for Surgical Implants,''

(iv) F648-96 ``Specification for Ultra-High-Molecular-Weight

Polyethylene Powder and Fabricated Form for Surgical Implants,''

(v) F1537-94 ``Specification for Wrought Cobalt-Chromium-Molybdenum

Alloy for Surgical Implants,''

(vi) F1044 ``Test Method for Shear Testing of Porous Metal

Coatings,''

(vii) F1147 ``Test Method for Tension Testing of Porous Metal

Coatings,'' and

(viii) F1378 ``Specification for Shoulder Prosthesis.''

49. Section 888.3660 is amended by revising paragraph (b) and by

removing paragraph (c) to read as follows:

Sec. 888.3660 Shoulder joint metal/polymer semi-constrained cemented

prosthesis.

* * * * *

(b) Classification. Class II. The special controls for this device

are:

(1) International Standards Organization's:

(i) ISO 10993 ``Biological Evaluation of Medical Devices Part I:

Evaluation and Testing,''

(ii) ISO 5832 ``Implants for Surgery--Metallic Materials,''

(iii) ISO 5833 ``Implants for Surgery--Acrylic Resin Cements,''

(iv) ISO 5834 ``Implants for Surgery--Ultra High Molecular Weight

Polyethylene,''

(v) ISO 9001 ``Quality Systems--Model for Quality Assurance in

Design/Development, Production, Installation, and Servicing,'' and

(vi) ISO 6018 ``General Requirements for Marketing, Packaging, and

Labeling,''

(2) FDA's:

(i) ``510(k) Sterility Review Guidance and Revision of 11/18/94

K90-1,''

(ii) ``Guidance Document for Testing Orthopedic Implants with

Modified Metallic Surfaces Apposing Bone or Bone Cement,''

(iii) ``Guidance Document for the Preparation of Premarket

Notification (510(k)) Application for Orthopedic Devices,'' and

(iv) ``Guidance Document for Testing Non-articulating,

`Mechanically Locked' Modular Implant Components,'' and

(3) American Society for Testing and Materials':

(i) F75-92 ``Specification for Cast Cobalt-Chromium-Molybdenum

Alloy for Surgical Implant Material,''

(ii) F799-96 ``Specification for Cobalt-28 Chromium-6 Molybdenum

Alloy Forgings for Surgical Implants,''

(iii) F1108-97 ``Specification for Ti6Al4V Alloy Castings for

Surgical Implants,''

(iv) F648-96 ``Specification for Ultra-High-Molecular-Weight

Polyethylene Powder and Fabricated Form for Surgical Implants,''

(v) F1537-94 ``Specification for Wrought Cobalt-Chromium-Molybdenum

Alloy for Surgical Implants,''

(vi) F1044 ``Test Method for Shear Testing of Porous Metal

Coatings,''

(vii) F1147 ``Test Method for Tension Testing of Porous Metal

Coatings,'' and

(viii) F1378 ``Standard Specification for Shoulder Prosthesis.''

Dated: March 1, 1999.

William K. Hubbard,

Acting Deputy Commissioner for Policy.

[FR Doc. 99-6266 Filed 3-12-99; 8:45 am]