Surgeon's and Patient Examination Gloves; Reclassification

Federal RegisterJul 30, 1999

Ask Donna

What actually matters in this document.

Text

SUMMARY: The Food and Drug Administration (FDA) is proposing

regulations to reclassify all surgeon's and patient examination gloves

as class II medical devices because it believes that general controls

are insufficient to provide a reasonable assurance of safety and

effectiveness. The reclassified gloves, including those made of natural

rubber latex (NRL) or synthetic material, will be regulated in four

categories: Powdered surgeon's gloves, powder-free surgeon's gloves,

powdered patient examination gloves, and powder-free patient

examination gloves. The proposed special controls are in the form of a

proposed guidance document entitled ``Medical Glove Guidance Manual,''

which includes recommended protein and glove powder limits, and new

label caution statements including protein and powder labeling

requirements. FDA is also proposing to require expiration dating. This

proposed rule is intended to reduce the adverse health effects from

allergic and foreign body reactions caused by the natural latex (NL)

protein allergens and glove powder found on surgeon's and patient

examination gloves and to reduce the adverse health effects from

defects in the barrier integrity and quality of surgeon's and patient

examination gloves.

DATES: Written comments by October 28, 1999. Written comments on the

information collection requirements should be submitted by August 30,

1999.

ADDRESSES: Submit written comments to the Dockets Management Branch

(HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061,

Rockville, MD 20852. Submit written comments on the information

collection requirements to the Office of Information and Regulatory

Affairs, Office of Management and Budget (OMB), New Executive Office

Bldg., 725 17th St. NW., rm. 10235, Washington, DC 20503, Attn.: Wendy

Taylor, Desk Officer for FDA.

FOR FURTHER INFORMATION CONTACT: Donald E. Marlowe, Center for Devices

and Radiological Health (HFZ-100), Food and Drug Administration, 5600

Fishers Lane, Rockville, MD 20857, 301-827-4777.

SUPPLEMENTARY INFORMATION:

I. Background

Surgeon's and patient examination gloves are intended to provide an

effective barrier against potentially infectious materials and other

contaminants. However, the use of surgeon's and patient examination

gloves has been associated with a number of adverse health effects in

patients and users, including allergic reactions, foreign body

reactions, and irritation.

NL is a milky fluid that consists of extremely small particles of

rubber obtained from plants, principally from the Heavea brasiliensis

(rubber) tree, dispersed in an aqueous medium. NL contains a variety of

naturally occurring substances, including plant proteins, which are

believed to be the primary allergens associated with NL allergy. NL is

employed in the natural rubber latex manufacturing process. Products

made by the natural rubber latex manufacturing process, such as medical

gloves, are referred to as containing or made of NRL. For a more

complete description of the NRL manufacturing process and further

definition of related terms, see the final rule entitled ``Natural

Rubber-Containing Medical Devices; User Labeling,'' published on

September 30, 1997 (62 FR 51021), and codified in part 801 (21 CFR part

801) at Sec. 801.437.

Glove powder is defined as the total particulate matter on a

finished glove, including donning and dusting powder, as well as

former-release (or mold-release) compounds and manufacturing debris.

The main component of donning and dusting powder is most commonly

cornstarch.

Health care workers, comprised of physicians, dentists,

pharmacists, nurses, technologists, technicians, and phlebotomists, use

millions of NRL gloves during procedures involving millions of

patients; this makes NRL gloves a significant source of exposure to NL

allergens (Ref. 1).

Studies of health care workers, blood donors, and ambulatory

surgical patients have demonstrated an appreciable prevalence of NL

sensitivity (Refs. 2 to 8). FDA has received 330 reports of adverse

events attributed to NL allergy occurring in patients and health care

workers, which suggests that allergic reaction to NRL products in

health care settings manifests itself in a variety of symptoms ranging

from dermatitis to anaphylaxis (Ref. 9). The general population is

directly exposed to NRL from a variety of sources, including consumer

products such as industrial gloves and NRL balloons, as well as medical

devices such as barrier contraceptives and NRL gloves.

FDA has significant concerns about the role of glove powder as a

carrier of airborne allergens, because NL allergens have been shown to

bind to cornstarch. A number of published clinical and experimental

studies support this conclusion (Refs. 10 to 14). In addition to the

role of glove powder as a carrier of airborne allergens, FDA is also

aware that glove powder contributes to a number of other adverse health

effects. As particulate matter, it can cause foreign body reactions,

resulting in inflammation, granulomas and adhesions of peritoneal

tissues after surgery (Refs. 15 to 19). Glove powder may serve as an

absorbent or adsorbent for unbound chemicals that may be irritants or

chemical contact sensitizers. In addition, glove powder from nonsterile

patient examination gloves may also support microbial growth and act as

a carrier for endotoxins (Ref. 20). These multiple concerns of adverse

health effects associated with particulate matter from the surface of

medical gloves constitute compelling reasons for FDA to reduce the

amount of powder on all gloves, as well as to ensure that both powdered

gloves and powder-free alternatives are clearly labeled so users and

consumers may make informed choices. Although data is not currently

available to quantify a maximum allowable level of glove powder,

decreased exposure to glove powder will decrease the prevalence of

adverse health effects. Therefore, FDA is recommending a powder level

it believes is achievable by industry.

In June 1997, the National Institute of Occupational Safety and

Health (NIOSH) issued a safety alert recommending the use of powder-

free, reduced protein content NRL or synthetic gloves as a means to

reduce exposure to NL allergens, specifically via the airborne route of

exposure (Ref. 21). While FDA agrees with the goal of reducing exposure

to airborne allergens, FDA is concerned that efforts to produce powder-

free gloves with satisfactory donning properties may require additional

manufacturing processes that, if not appropriately controlled, have

deleterious effects on physical properties, performance, and shelf-life

of the gloves (Refs. 22 and 23).

[[Page 41711]]

II. Statutory Authority

The Federal Food, Drug, and Cosmetic Act (the act), as amended by

the Medical Device Amendments of 1976 (the 1976 amendments) (Public Law

94-295), the Safe Medical Devices Act of 1990 (SMDA) (Public Law 101-

629), and the Food and Drug Administration Modernization Act of 1997

(FDAMA) (Public Law 105-115), established a comprehensive system for

the regulation of medical devices intended for human use. Section 513

of the act (21 U.S.C. 360c) established three categories (classes) of

devices, depending on the regulatory controls needed to provide a

reasonable assurance of their safety and effectiveness.

The three categories of devices are class I (general controls),

class II (special controls), and class III (premarket approval). The

effect of classifying a device into class I is to require that the

device meet only the general controls that are applicable to all

devices. The effect of classifying a device into class II is to require

the device to meet special controls as well as general controls, which

together provide reasonable assurance of the safety and effectiveness

of the device. Class II devices are devices which cannot be classified

in class I because general controls by themselves are insufficient to

provide reasonable assurance of safety and effectiveness and for which

there is sufficient information to establish special controls to

provide such assurance, including the issuance of performance

standards, postmarket surveillance, patient registries, and guidelines

(see section 513(a)(1)(B) of the act). The effect of classifying a

device into class III is to require each manufacturer of the device to

submit to FDA a premarket approval application (PMA) that includes

information concerning safety and effectiveness of the device.

Under section 513 of the act, devices that were in commercial

distribution before May 28, 1976 (the date of enactment of the

amendments), generally referred to as preamendments devices, are

classified after FDA has: (1) Received a recommendation from a device

classification panel (an FDA advisory committee); (2) published the

panel's recommendation for comment, along with a proposed regulation

classifying the device; and (3) published a final regulation

classifying the device. FDA has classified most preamendments devices

under these procedures.

A device that is first offered in commercial distribution after May

28, 1976, generally referred to as a postamendments device, and which

FDA determines to be substantially equivalent to a device classified

under this scheme, is classified into the same class as the device to

which it is substantially equivalent. The agency determines whether new

devices are substantially equivalent to previously offered devices by

means of premarket notification procedures in section 510(k) of the act

(21 U.S.C. 360(k)) and part 807 of the regulations (21 CFR part 807). A

device that was not in commercial distribution prior to May 28, 1976,

and that has not been found by FDA to be substantially equivalent to a

legally marketed device, is classified automatically by statute

(section 513(f) of the act) into class III, without any FDA rulemaking

proceeding.

Reclassification of classified preamendments devices is governed by

section 513(e) of the act (21 U.S.C. 360c(e)). This section provides

that FDA may, by rulemaking, reclassify a device (in a proceeding that

parallels the initial classification proceeding) based upon ``new

information.'' The reclassification can be initiated by FDA or by the

petition of an interested person. The term ``new information,'' as used

in section 513(e) of the act, includes information developed as a

result of the reevaluation of the data before the agency when the

device was originally classified, as well as information not presented,

not available, or not developed at that time. (See, e.g., Holland

Rantos v. United States Department of Health, Education, and Welfare,

587 F.2d 1173, 1174 n.1 (D.C. Cir. 1978); Upjohn v. Finch, 422 F.2d 944

(6th Cir. 1970); Bell v. Goddard, 366 F.2d 177 (7th Cir. 1966).)

Reevaluation of the data previously before the agency is an

appropriate basis for subsequent regulatory action where the

reevaluation is made in light of changes in ``medical science.'' (See

Upjohn v. Finch, supra, 422 F.2d at 951.) However, regardless of

whether data before the agency are past or new data, the ``new

information'' on which any reclassification is based is required to

consist of ``valid scientific evidence,'' as defined in section

513(a)(3) of the act (21 U.S.C. 360c(a)(3)) and 21 CFR 860.7(c)(2). FDA

relies upon ``valid scientific evidence'' in the classification process

to determine the level of regulation for devices. For the purpose of

reclassification, the valid scientific evidence upon which the agency

relies must be publicly available. Publicly available information

excludes trade secret and/or confidential commercial information.

On November 21, 1997, the President signed into law FDAMA. Section

206 of FDAMA added a new section 510(m) (21 U.S.C. 360(m)) to the act.

Section 510(m)(2) of the act provides that FDA may, on its own

initiative or upon petition of an interested person, exempt a class II

device from the requirement of premarket notification in section 510(k)

of the act, if FDA determines that a 510(k) submission is not necessary

to provide reasonable assurance of the safety and effectiveness of the

device. Such an exemption would permit manufacturers to introduce the

generic type of device into commercial distribution without first

submitting a premarket notification to FDA.

Section 701(a) of the act (21 U.S.C. 371(a)) authorizes FDA to

issue substantive binding regulations for the efficient enforcement of

the act. (Weinberger v. Hynson, Westcott & Dunning, Inc., 412 U.S. 609

(1973); see also Weinberger v. Bentex Pharmaceuticals, Inc., 412 U.S.

645, 653 (1973); National Ass'n of Pharmaceutical Manufacturers v. FDA,

637 F.2d 877 (2d Cir. 1981); National Confectioners Ass'n v. Califano,

569 F.2d 690 (D.C. Cir. 1978); National Nutritional Foods Ass'n v.

Weinberger, 512 F.2d 688 (2d Cir.), cert. denied, 423 U.S. 825 (1975).)

Section 502(a) of the act (21 U.S.C. 352(a)) provides that a device

is misbranded ``[I]f its labeling is false or misleading in any

particular.'' Section 201(n) of the act (21 U.S.C. 321 (n)) provides

that, in determining whether labeling of a regulated article (such as a

device) is misleading

* * * there shall be taken into account * * * not only

representations made or suggested by statement, word, design,

device, or any combination thereof, but also the extent to which the

labeling * * * fails to reveal facts material in light of such

representations * * * with respect to consequences which may result

from the use of the article to which the labeling * * * relates

under the conditions of use prescribed in the labeling or

advertising thereof or under such conditions of use as are customary

or usual.

The courts have upheld FDA's authority to prevent false or

misleading labeling by issuing regulations requiring label warnings and

other affirmative disclosures (See, e.g., Cosmetic, Toiletry, and

Fragrance Association v. Schmidt, 409 F. Supp. 57 (D.D.C. 1976), aff'd

without opinion, Civil No. 75-1715 (D.C. Cir. August 19, 1977), even in

the absence of a proven cause-and-effect relationship between product

usage and harm (Council for Responsible Nutrition v. Goyan, Civil No.

80-1124 (D. D. C. August 1, 1980)).

FDA may impose testing requirements in a labeling regulation issued

under its general rulemaking authority. (See, e.g.,

[[Page 41712]]

American Frozen Food Inst. v. Mathews, 413 F. Supp. 548 (D.D.C. 1976),

aff'd per curiam sub nom. American Frozen Food Inst. v. Califano, 555

F.2d 1059 (D.C. Cir. 1977); see also National Nutritional Foods Ass'n

v. Weinberger, supra.) Thus, FDA may require that all manufacturers use

the same conditions to test aging to ensure that the expiration date

reflects the period of time a product can be used safely. Similar

requirements are imposed in Sec. 801.430(f) for absorbency testing for

menstrual tampons, and in Sec. 801.420(c)(4) on hearing aid

manufacturers and distributors who must determine and state technical

data values for hearing aid labeling in accordance with specified test

procedures. The hearing aid regulation has been upheld. (American

Speech and Hearing Ass'n v. Califano, Medical Devices Report (CCH) No.

77-1327 Secs. 15004, 15007 (D.D.C. August 23, 1977) aff'd No. 77-1327

(D.C. Cir. Dec. 19, 1977).) Food regulations issued under section

701(a) of the act also impose many such specific testing requirements

(see e.g., 21 CFR 113.40 (tests for low-acid canned foods); 21 CFR

155.190(b)(2)(i) (test for determining drained weight of canned

tomatos); 21 CFR 161.190 (method for determining color designation of

tuna).

III. Powder and Protein Concerns

Although FDA has been concerned about airborne NL allergens

associated with the use of powdered medical gloves and has undertaken

continued efforts to address these concerns, recent heightened

awareness within the health care community and State and Federal

Government agencies of adverse health effects has prompted this

proposed action.

Over the past 3 years, FDA has received requests to ban the use of

all glove powders. These requests have been based on a number of

clinical and experimental studies reporting that cornstarch on surgical

gloves can reduce tissue resistance to infection, enhance the

development of infection, cause formation of granulomas and adhesions,

act as a carrier of NL protein from NRL products, and serve as a

potential source of occupational asthma. Although a ban of all powdered

medical gloves has been requested by petitioners and would reduce the

problem of airborne powder, it would not completely address the problem

of NL allergy and would potentially leave a significant and important

need for high quality barrier products unmet.

One of the concerns regarding glove powder, in general, is its

capability, as particulate material, to cause foreign body reaction,

resulting in inflammation, granulomas and adhesions of peritoneal

tissues after surgery (Refs. 15 to 19). Although cornstarch was

considered to be absorbable by United States Pharmacopeia (USP),

changes in the sterilization processes have reduced absorbability

significantly (Ref. 15). Cornstarch represents a growth source for

bacteria, and it is also a carrier of endotoxin, which can play a role

in enhancing both delayed and immediate hypersensitivity (Ref. 20).

Clinical experience suggests that powder on NRL gloves, in addition to

its role in Type I allergy, may also be a contributing factor in the

development of irritant dermatitis and Type IV allergy. Irritant skin

reactions have been observed in association with frequent exposure to

glove powder. Compromised skin barrier properties resulting from such

reactions may permit penetration of allergens and other substances into

the skin, thereby increasing chances for the development of both Type I

and Type IV allergy (Ref. 24).

In addition, a significant concern, specific to NL gloves, exists

regarding the role of glove powder as a carrier of airborne NL

allergens. A number of respiratory problems and episodes of bronchial

spasms in hospital employees and patients, reported since the mid

1980's, were ascribed to inhalation of airborne NL allergens in

settings with heavy use of powdered gloves (Refs. 25 to 30). The

implication of glove powder in the previous clinical reports was based

on medical histories of individuals presenting with symptoms, on

positive skin tests, positive tests for the presence of antibodies to

NL allergens in blood and, in some cases, on positive inhalation

challenge tests. A number of published clinical and experimental

studies support this conclusion.

Binding of NL proteins to cornstarch was demonstrated in recent

laboratory studies, which support a causal relationship between

asthmatic reactions in individuals with NL allergy and the exposure to

airborne particles from NL products (Refs. 10 and 11). The level of

exposure and the severity of the reactions depend on both the amount of

powder and the amount of NL protein allergens on the finished products.

Measurements of airborne particle levels in environments where NL

gloves were used frequently demonstrated that the level of airborne

allergen is directly related to the frequency of powdered NL glove

usage in particular areas and to the level of allergen and/or powder on

the gloves used (Refs. 12 and 14).

Direct evidence that NL protein allergens, bound to the glove

powder particles, provoke respiratory allergic reactions and asthma-

like attacks has been documented by the bronchial provocation tests

with powders on NL gloves. The bronchial provocation tests were

performed by having allergic individuals inhale the extracts from

powder-free surgeon's gloves, from powdered surgeon's gloves, and from

cornstarch powder not exposed to NL. The studies indicated that

cornstarch powder not exposed to NL did not cause any reaction in

sensitized subjects, while nebulized powdered NL surgeon's glove

extract, and to some extent, nebulized powder-free glove extract

induced bronchoconstriction in tested subjects (Ref. 31).

However, the scientific data to define the quantitative

relationship between respiratory allergic reactions and powder level on

NL gloves are not available at this time. Such data and the specific

dose-response relationship would be difficult to establish, because

allergenicity of the airborne glove powder depends on the amount of

powder and also on the amount of powder-bound allergenic proteins.

Standardized methods for measuring the amount of powder-bound proteins

or allergens and the amount of inhaled powder are not available.

NL protein has been widely reported as a cause of Type I

sensitivity in individuals who have been exposed to NL devices (Refs. 2

to 8). Repeated exposure to NL protein is considered to increase the

probability that an individual will become sensitized. Total water-

extractable protein on the finished NL product is considered an

indirect measure of the potential allergenicity. Because several NL

proteins have already been identified as allergenic and others may be

identified in the future, exclusion of any proteins from the evaluation

may result in an inaccurate determination of potential allergenicity.

The total water-extractable protein level measured using the standard

American Society for Testing and Materials (ASTM) D 5712 method was

found to correlate well with currently used allergen measurement

methods. Most importantly, a total water-extractable protein level

correlates also with the skin prick test, which is a direct measure of

allergic response in sensitized individuals (Ref. 32). Since May 1991,

FDA has advised manufacturers of NL devices to reduce the water-

extractable protein on their NL devices. This reduction is now

addressed in the Quality System (QS) Regulation at 21 CFR 820.3(p) and

820.70(h).

[[Page 41713]]

Initially, a labeling claim for a protein level was not accepted in

a 510(k) submission because a standard test method for measuring water-

extractable protein in NL did not exist. In 1995, with the help of

industry and FDA, ASTM published the ``ASTM Standard Test Method for

Analysis of Protein in Natural Rubber and its Products, D 5712-95.''

FDA subsequently issued a document entitled ``Interim Guidance On

Protein Content Labeling Claim For Latex Medical Gloves,'' which is

based on this test method. Manufacturers were allowed to use this

guidance to submit a 510(k) submission for NL gloves identifying the

level of water-extractable protein for the device. FDA is now proposing

that a recommended limit on water-extractable protein per glove and the

actual protein level appear on the label.

The amount of powder required for satisfactory donning of gloves

has not been quantified, and the level of glove powder used varies

greatly. Limited laboratory data from measurements of a number of

surgeon's and patient examination gloves demonstrated that powder

levels ranged from 70 to 375 milligrams (mg) per glove for surgeon's

gloves and from 50 to 426 mg per glove for patient examination gloves

(Ref. 31). Because of the multiple concerns regarding adverse health

effects associated with particulate matter from the surface of

surgeon's and patient examination gloves, FDA is now proposing that a

recommended limit on glove powder and the actual level of glove powder

appear on the label. FDA recognizes there is a correlation between

powder level and ease of glove donning and that powder level is

correlated with adverse health effects. For this reason, FDA is

encouraging industry to find the balance between donning requirements

and reducing the risks of adverse health effects.

Lowering the powder level and the amount of protein on surgeon's

and patient examination gloves will reduce exposure to NL allergens and

benefit both allergic individuals and those at risk to develop allergy.

In addition, the reduction of glove powder levels will help reduce

exposure to particulate materials responsible for foreign body

reactions. However, the reduction of powder and protein levels must be

accomplished by methods that do not compromise the availability of or

barrier properties of surgeon's and patient examination gloves.

IV. Barrier and Other Quality Issues

In the Federal Register of October 21, 1980 (45 FR 69723), FDA

issued a final rule classifying the patient examination glove into

class I and exempting manufacturers of the device from compliance with

premarket notification procedures under section 510(k) of the act and

certain requirements of the current good manufacturing practice (CGMP)

regulation. FDA granted the exemptions in the 1980 regulation because,

at that time, no adverse experiences had been related to patient

examination gloves. Furthermore, the role of the gloves as a protective

barrier against human immunodeficiency virus (HIV) transmission was not

recognized and the concomitant risks associated with glove failure were

not well understood.

In the Federal Register of January 19, 1982 (47 FR 2810 at 2852),

FDA proposed that the surgeon's glove be classified into class II

because of concerns about tissue compatibility and the risk of

infection if the devices were not properly sterilized. Comments offered

in response to the proposed classification stated that those problems

could be addressed through general controls, including labeling and

CGMP adherence, and recommended that the device be classified into

class I because of the history of its safe and effective use. In the

Federal Register of June 24, 1988 (53 FR 23856), FDA issued a final

rule classifying the surgeon's glove into class I without exemptions.

Manufacturers and importers of surgeon's gloves have been required to

comply with the premarket notification and CGMP regulations since the

initial classification of the device.

Over the years, many issues regarding surgeon's and patient

examination gloves have been brought to the attention of FDA. The

acquired immune deficiency syndrome (AIDS) epidemic resulted in an

elevated reliance on medical gloves as a barrier against blood-borne

viral transmission. The increased demand for gloves soon outstripped

the domestic supply. Foreign glove manufacturers began to meet the

demand for additional gloves. Many manufacturers with little or no

medical glove manufacturing experience began operations, resulting in

large quantities of gloves of uncertain quality entering the U.S.

market.

Following the advent of AIDS as a major public health concern and

recommendations from the Centers for Disease Control and Prevention

(CDC) that health care workers use appropriate barrier precautions to

prevent exposure to the HIV virus, FDA recognized the need for greater

assurance that cross-contamination between patients and health care

workers be prevented. Accordingly, in the Federal Register of January

13, 1989 (54 FR 1602), FDA revoked the exemption for patient

examination gloves from certain CGMP requirements in order to assure

that manufacturers provide an acceptable manufacturing quality level.

FDA similarly revoked the exemption from premarket notification

requirements for patient examination gloves. On December 12, 1990 (55

FR 51254), FDA published regulations describing certain circumstances

under which surgeon's and patient examination gloves would be

considered adulterated, and establishing the sampling plans and test

methods the agency would use to determine whether gloves were

adulterated (Sec. 800.20 (21 CFR 800.20)). Subsequently, FDA initiated

inspections of glove manufacturers to assure conformance with the

acceptable quality levels (AQL) identified in that regulation.

FDA has sought to address many concerns regarding the quality and

barrier integrity of medical gloves. Certain processes or conditions

can often contribute to degradation of the barrier. NL degrades if it

is not correctly formulated and processed. Proper formulation includes

the use of stabilizers, antiozonants, and antioxidants to reduce

degradation. Improper curing can also cause thin spots on the glove

surface, which may lead to early barrier failure.

Gloves composed of synthetic polymer, such as nitrile, are produced

by essentially the same processes as NL. The same accelerators,

antioxidants, and stabilizers are used to reduce degradation. Thus,

improper formulation and processing may also lead to rapid degradation

of synthetic gloves.

Storage conditions can also cause degradation of the polymers,

whether natural or synthetic. These storage conditions include the

temperature at which the material is held, the humidity of their

environment, and any radiation (for example, sunlight or fluorescent

lights) to which the material may be exposed.

Additionally, chlorination is widely used to reduce the tackiness

of NL gloves and thus eliminate the need for donning powder.

Chlorination works by degrading the surface of the gloves. Therefore,

chlorination must be carefully controlled in order to prevent

destruction of the glove barrier. Improperly chlorinated gloves rapidly

degrade, and breaks in the latex film may occur in a matter of months.

Another concern has been the presence of minute defects known as

pinholes, which directly affect the barrier integrity of the gloves.

FDA studies of micro-photographs of

[[Page 41714]]

defective NL devices have shown that dust, dirt, rust, paint chips,

charred starch, insect parts, and other debris may cause pinholes.

Therefore, appropriate environmental and processing controls, as

required by the QS regulation, are needed. Manufacturers also need to

control other causes of pinholes such as former vibration, air bubbles

in the dipping tanks, dirty formers, incorrect formulation, and

excessive curing temperatures.

If gloves have pinholes, breaks or tears, viruses can potentially

penetrate the glove wall, eliminating or reducing the gloves'

effectiveness as a barrier. On April 6 and 7, 1989, the University of

Maryland, in conjunction with FDA, held a conference entitled ``Latex

as a Barrier Material,'' which reiterated the value of NL as a barrier

film and generated continued support towards more research in this area

by industry and FDA.

Although manufacturers have data to show that their gloves meet

their company AQL for defects when the gloves are shipped, for some

manufacturers, the same gloves which passed the manufacturer's tests

are sometimes rejected at the port of entry in the United States

because the gloves fail the FDA water leak test at that point. This

test result disparity, whether due to degradation or for other reasons,

is a primary reason why, upon importation, the gloves of some

manufacturers have been detained without physical examination.

Manufacturers should assure, by means of stability testing, that their

surgeon's and patient examination gloves will continue to meet the

manufacturers' specifications over the expected life of the gloves.

FDA is aware that microbial growth on gloves also can be a problem.

The QS regulation requires manufacturers to control processing,

shipping and storage environment, and contamination when these can

adversely affect the product. Therefore, processing controls should

include: Using only cornstarch with an acceptable bioburden, properly

storing the cornstarch until it is used, applying cornstarch by

established procedures, cooling the cornstarch slurry and/or using an

antimicrobial in the cornstarch slurry tanks, checking finished gloves

on a sampling basis to assure that excessive cornstarch is not applied,

keeping the finished gloves clean, establishing and meeting a dryness

specification for finished gloves, and protecting finished gloves from

adverse environmental conditions.

Although synthetic materials have improved in recent years, NL

gloves may be superior to some synthetic gloves in regard to barrier

properties (Ref. 34). Both NL and synthetic surgeon's and patient

examination gloves provide protection against microorganisms; however,

it has been demonstrated that compared to vinyl, NL has more effective

and durable barrier qualities (Refs. 35 and 36).

There are other safety and performance issues related to gloves and

other barrier devices that are currently being considered by industry

and FDA. These issues include puncture resistance, tear resistance,

reliability, and biocidal claims.

V. The Proposed Rule

Based upon new information that was not presented, not available,

or not developed when FDA originally classified surgeon's and patient

examination gloves, FDA has reevaluated its classification in light of

changes in the medical science discussed in sections III and IV of this

document. The new, publicly available, valid scientific evidence

demonstrates that these gloves should not remain as class I devices

because of: (1) Barrier integrity concerns; (2) degradation of quality

during storage; (3) contamination concerns; and (4) concerns about

exposure to NL allergens and the role of glove powder as a carrier of

airborne NL allergens, and the inability of general controls to address

these concerns. The agency believes that general controls are no longer

sufficient to provide reasonable assurance of the gloves' safety and

effectiveness and, therefore, FDA is proposing that these gloves be

reclassified into class II.

Surgeon's and patient examination gloves are intended for use as an

effective barrier against potentially infectious materials and other

contaminants. Risk to the user or patient may result from lack of

barrier integrity from degradation, pinholes, breaks, tears, or loss of

quality during storage, potentially causing penetration of the glove

wall by viruses or other infectious materials. When glove powder comes

into contact with compromised human tissue, risk to the user or patient

may result from foreign body reactions caused by NL allergens bound to

the glove powder. Allergic reactions may also be caused by inhalation

of NL allergens bound to the glove powder. Reducing the degree of risk

to acceptable levels depends on effective maintenance of the barrier

properties of the gloves and on reducing exposure to NL allergens,

particularly exposure to airborne NL allergens. The highest risk

products are those with large amounts of glove powder and NL protein

and those products with poor barrier properties.

In order to enable users to distinguish between powdered and

powder-free gloves and to choose the glove type appropriate for their

needs, FDA proposes to reclassify surgeon's gloves into two separate

classifications, based on powder level: Powdered surgeon's gloves, and

powder-free surgeon's gloves. FDA similarly proposes to reclassify

patient examination gloves into two categories: Powdered patient

examination gloves, and powder-free patient examination gloves.

FDA is proposing that these gloves be subject to two special

controls: A guidance document entitled, ``Medical Glove Guidance

Manual,'' and new user labeling requirements. FDA believes that the

proposed guidance document and user labeling requirements are necessary

to provide reasonable assurance of the safe and effective use of the

devices. The guidance is currently being issued in draft as a Level 1

guidance consistent with the good guidance practices (GGP's) FDA

adopted for the development, issuance, and use of guidance documents

(62 FR 8961, February 27, 1997). Elsewhere in this issue of the Federal

Register, FDA is announcing the availability of the guidance in draft

form, to provide an opportunity for comment.

The proposed guidance document recommends that manufacturers of

powdered surgeon's and patient examination gloves limit the amount of

powder to no more than 120 mg of powder per glove, regardless of glove

size. In order to limit total exposure to the user, a ``per glove''

measurement (mg per glove) is used instead of the ``per unit'' dose (mg

per gram (g) of glove material). Under the proposed labeling

requirements, manufacturers of all powdered gloves would be required to

include the actual level of glove powder on the label. FDA believes

that the recommended limit should be sufficient for proper donning of

gloves, but would reduce exposure to airborne glove powder particles.

In addition to the role of glove powder as a carrier of airborne

allergens, FDA is also aware that glove powder contributes to a number

of other adverse health effects. As particulate matter, it can cause

foreign body reactions, resulting in inflammation, granulomas and

adhesions of peritoneal tissues after surgery (Refs. 15 to 19). Glove

powder may serve as an absorbent or adsorbent for unbound chemicals

that may be irritants or chemical contact sensitizers.

The proposed guidance document further recommends that

manufacturers of powder-free surgeon's and patient

[[Page 41715]]

examination gloves limit the amount of total trace (residual) powder to

no more than 2-mg particulate weight (based on the ASTM test standard D

6124-97) per glove, regardless of glove size. Previously, this limit

was recommended to manufacturers who wanted to market gloves with a

powder-free labeling claim. A number of premarket notification

submissions based on this claim already have been cleared for market.

The proposed guidance document also recommends that manufacturers

of NL surgeon's and patient examination gloves limit the amount of

water-extractable protein on the gloves to no more than 1,200

micrograms (g) of protein per glove, regardless of glove size.

In order to limit total exposure to the user, a ``per glove

measurement'' (mg per glove) is used instead of a ``per unit'' dose (mg

per g of glove material). Under the proposed labeling requirements,

labeling on all NL gloves would be required to include the level of

water-extractable protein measured, as recommended in the guidance, by

the currently recognized ASTM D 5712 modified Lowry method. The lowest

acceptable amount of water-extractable protein that may be stated in

the labeling will be limited by the sensitivity of the current ASTM D

5712 test method to 50 g of protein per g of natural rubber

product (which translates to 300 g per glove for a 6 g glove,

i.e., 6 x 50 = 300). FDA believes that without a more sensitive

standard method, lower claims would be misleading.

The proposed labeling requirements are a special control intended

to provide guidance to users of surgeon's and patient examination

gloves. They would require manufacturers to provide new caution

statements, which would include both the FDA recommended limit for

glove powder and protein levels, as well as the actual glove powder and

protein levels present in the manufacturer's gloves. The labeling

special control provides essential decisionmaking information for

health professionals, patients, and lay users. The information required

under the proposed regulations would assist health care professionals,

patients and lay users to select a lower risk device by providing

information about protein and glove powder levels.

The proposed caution statements would be required to appear on all

device labels and other labeling, including the principal display panel

of the device packaging, the outside package, container or wrapper, and

the immediate device package, container or wrapper. The proposed

caution statements for powdered and powder-free NL gloves (surgeon's

and patient examination) would supersede the caution statements in

Sec. 801.437(d) for devices containing NRL currently required in the

regulation published in the Federal Register of September 30, 1997

(effective September 30, 1998).

Labeling for powdered surgeon's and patient examination gloves

containing NL that contacts humans would be required to bear the

following statement:

``Caution: This product contains natural rubber latex which may

cause allergic reactions. FDA recommends that this product contain

no more than 120 mg powder and 1,200 g extractable protein

per glove. This product contains no more than [insert level] mg

powder and no more than [insert level] g extractable

protein per glove.''

Labeling for powder-free surgeon's and patient examination gloves

containing NL that contacts humans would be required to bear the

following statement:

``Caution: This product contains natural rubber latex which may

cause allergic reactions. FDA recommends that this product contain

no more than 1,200 g extractable protein per glove. This

product contains no more than [insert level] g extractable

protein per glove.''

FDA is also proposing new labeling requirements for powdered gloves

made of synthetic material. FDA proposes that labeling for those gloves

bear the following statement:

``Caution: Glove powder is associated with adverse reactions.

FDA recommends that this product contain no more than 120 mg powder

per glove. This product contains no more than [insert level] mg

powder per glove.''

FDA is proposing no new labeling for powder-free surgeon's gloves and

patient examination gloves made of synthetic materials.

FDA is also proposing to require expiration dating on the labeling

of all powdered surgeon's and patient examination gloves and powder-

free surgeon's and patient examination gloves. Previously, expiration

dating has not been required for surgeon's or patient examination

gloves, although it is customary for surgeon's gloves to bear an

expiration date for sterility. A few glove manufacturers have

voluntarily used expiration dates based on real-time data to support

the integrity of the gloves throughout the shelf-life period.

In view of the quality concerns discussed in section IV of this

document, especially those relating to degradation of barrier integrity

over time, FDA believes that expiration dating is necessary to allow

users to correctly store and use stock of gloves, and to allow users to

avoid gloves that may have degraded. Users must be aware of the

potential for degradation of gloves in order to safely use such

products to provide a barrier from infectious agents. Accordingly, FDA

believes that shelf life is a fact material to the consequences of use

of surgeon's and patient examination gloves. Therefore, FDA is now

proposing that all surgeon's and patient examination gloves be required

to bear an expiration date on their primary and retail packaging and

shipping carton. The expiration date should consist of the month and

year for which data exists to support the shelf-life of the gloves. The

time period upon which the expiration date is based starts with the

date of manufacture.

This expiration date must be based on testing conducted according

to a validated stability study protocol to determine the shelf-life of

the gloves. The stability study protocol should employ tests commonly

used by industry to demonstrate the physical and mechanical integrity

of the gloves over their claimed shelf-life.

Manufacturers will not be required to provide new section 510(k) of

the act submissions to demonstrate the shelf-life of gloves. However,

for each distinct glove design, the records of study protocols and test

data must be retained for a period equivalent to the design and

expected life of the gloves, and must be made available for inspection

by FDA personnel.

Expiration dates for sterile surgeon's or patient examination

gloves should either be based on the shelf-life determined by stability

studies as outlined in the proposed rule, or on the sterility shelf-

life, whichever is shorter. Only one expiration date should appear on

each product.

FDA does not intend to require a new submission under section

510(k) of the act based upon labeling changes or reductions in glove

powder or NL protein made to comply with any final regulation based

upon this proposed regulation, provided that no other changes requiring

a new 510(k) submission under Sec. 807.81 are made to the device.

Section 510(m) of the act allows FDA to exempt a class II device

from the requirement of premarket notification in section 510(k) of the

act. FDA does not intend to exempt powdered or powder-free surgeon's or

patient examination gloves from premarket notification because of FDA's

concerns regarding the effective maintenance of barrier properties and

adverse health effects associated with NL allergens, glove powder and

residual chemical sensitizers and irritants.

This proposed rule would not impose requirements on glove users or

user facilities. Therefore, it would not affect

[[Page 41716]]

the authority of the Secretary of Labor, under the Occupational Safety

and Health Act (OSH act), to enforce regulations, standards, or other

directives issued under the OSH act.

VI. Specific Request for Comments

FDA recognizes that this regulation affects surgeon's and patient

examination gloves in different ways, depending on glove powder level.

FDA also recognizes that manufacturing processes for powdered and

powder-free gloves vary. FDA welcomes comments on all aspects of the

proposed regulation, but particularly invites comments on the following

issues:

1. FDA requests comments on the timeframe for implementation of the

proposed rule considering the need for changes in production,

technology, and labeling, as well as the immediate need to address

adverse health concerns associated with medical gloves. Although FDA

prefers a 1-year effective date, FDA is proposing a 2-year effective

date based on indications from industry that the necessary changes

could not be made in 1 year and that a shortage of medical gloves could

result.

2. In the proposed guidance document, FDA recommends a limit of no

more than 120 mg powder per powdered glove, regardless of size, as the

maximum level in order to reduce exposure to particulates and airborne

allergens. FDA requests comments on the recommended limit with regard

to the minimum level of powder needed for adequate donning of gloves.

3. FDA requests comments on the feasibility and desirability of

additional labeling requiring manufacturers to state the primary

ingredients in glove powder in the product labeling.

4. In the proposed guidance document, FDA is recommending no more

than 2 mg powder per glove, regardless of size, as the recommended

powder level for those surgeon's and patient examination gloves labeled

``powder-free.'' FDA requests comments on the proposed limit. FDA is

also seeking comments on the possible impact of this powder limit on

barrier properties and shelf-life of NL gloves.

5. FDA is also considering a future requirement that all surgeon's

and patient examination gloves marketed in the United States be powder-

free. FDA requests comments as to whether a continued need for powdered

gloves exists, and, if so, the reason for this need.

6. FDA considered restrictions on the sale (advertising),

distribution, and use of powdered surgeon's and patient examination

gloves. FDA is seeking comments on the feasibility of such

restrictions.

7. In the proposed guidance document, FDA is recommending an upper

limit of no more than 1,200 g protein per NL glove, regardless

of size, as the maximum level for NL surgeon's and patient examination

gloves. FDA is seeking comments on the proposed recommended limit.

8. FDA's objectives in this proposed rulemaking are to reduce

adverse health effects from allergic reactions and foreign body

reactions by controlling the levels of water-extractable protein and

glove powder on NL gloves. FDA requests comments as to whether there

are feasible alternative approaches to achieve these objectives. If

other alternatives or data submitted present feasible methods to

protect the public health or suggest that different powder or protein

levels are adequate to protect the public health, FDA may incorporate

such data or approaches in a final rule.

9. FDA also invites comments on the issue of whether the

recommended limits on powder and protein proposed in this rule should

be recommended limits or required limits.

10. FDA considered allowing manufacturers to establish an initial

tentative shelf-life up to a certain duration based on accelerated

aging data, provided that manufacturers initiate concurrent real-time

shelf-life studies to confirm and extend the tentative shelf-life. FDA

has been unable, however, to determine whether any validated stability

study protocols exist employing accelerated aging methodologies. The

agency invites comments or information on the availability of

accelerated aging stability study protocols which are predictive of

glove shelf-life. If convincing information concerning such protocols

is available, FDA may incorporate such an approach in a final rule.

11. FDA considered requiring the use of a special air handling

system at the point of use for those facilities using powdered

surgeon's and patient examination gloves with powder levels over 120 mg

per glove, regardless of glove size. FDA is seeking comments on the

appropriateness of this restriction.

12. FDA seeks comments as to whether a provision permitting

affected persons to request exemptions or variances from the labeling

requirements or restrictions on distribution and use proposed in this

rule should be added.

VII. General Request for Comments

Interested persons may submit written comments regarding this

proposed rule by October 28, 1999, to the Dockets Management Branch

(address above). Comments regarding the information collection

provisions should be submitted by August 30, 1999, to the Office of

Information and Regulatory Affairs, Office of Management and Budget

(address above). Two copies of any comments are to be submitted, except

that individuals may submit one copy. Comments are to be identified

with the docket number found in brackets in the heading of this

document. Received comments may be seen in the office above between 9

a.m. and 4 p.m., Monday through Friday.

VIII. Access to Special Control

The availability of the special control entitled ``Medical Glove

Guidance Manual'' is being announced elsewhere in this issue of the

Federal Register. A copy of the ``Medical Glove Guidance Manual'' may

be seen by interested persons in the Dockets Management Branch (address

above) between 9 a.m. and 4 p.m., Monday through Friday.

Persons interested in obtaining a copy of the guidance may also do

so using the World Wide Web (WWW). FDA's Center for Devices and

Radiological Health (CDRH), maintains an entry on the WWW for easy

access to information including text, graphics, and files that may be

downloaded to a PC with access to the Web. The CDRH home page is

updated on a regular basis and includes the draft ``Medical Glove

Guidance Manual;'' device safety alerts; Federal Register reprints;

information on premarket submissions (including lists of approved

applications and manufacturers' addresses); small manufacturers'

assistance; and information on video conferencing and electronic

submissions, mammography matters, and other device-oriented

information. The CDRH home page may be accessed at ``http://

www.fda.gov/cdrh''.

Submit written requests for single copies of the draft guidance to

the Division of Small Manufacturers Assistance, Center for Devices and

Radiological Health (HFZ-220), Food and Drug Administration, 1350

Piccard Dr., Rockville, MD 20850. Send two self-addressed adhesive

labels to assist that office in processing your request, or fax your

request to 301-443-8818.

To receive the directions via fax machine on receiving the proposed

guidance document, call CDRH Facts-on-Demand system at 800-399-0381, or

301-827-0111 from a touch-tone telephone. At the first voice prompt,

press 1 to access the Division of Small Manufacturers Assistance (DSMA)

Fax, at the second voice prompt, press 2, and then enter the document

number 852

[[Page 41717]]

followed by the pound sign (#). Then follow the remaining voice prompts

to complete your request.

IX. Analysis of Impacts

FDA has examined the impacts of the proposed rule under Executive

Order 12866, under the Regulatory Flexibility Act (5 U.S.C. 601-612),

and under the Unfunded Mandates Reform Act (Public Law 104-4).

Executive Order 12866 directs agencies to assess all costs and benefits

of available regulatory alternatives and, when regulation is necessary,

to select regulatory approaches that maximize net benefits (including

potential economic, environmental, public health and safety, and other

advantages; distributive impacts; and equity). Unless the agency

certifies that the rule is not expected to have a significant economic

impact on a substantial number of small entities, the Regulatory

Flexibility Act requires agencies to analyze regulatory options that

would minimize any significant economic impact of a rule on small

entities. Section 202 of the Unfunded Mandates Reform Act requires that

agencies prepare an assessment of anticipated costs and benefits before

proposing any rule that may result in an expenditure by State, local,

and tribal governments, in the aggregate, or by the private sector, of

$100 million in any one year (adjusted annually for inflation).

The agency believes that this proposed rule is consistent with the

principles set out in the Executive Order and in these two statutes.

The rule is an economically significant regulatory action as defined by

the Executive Order. With respect to the Regulatory Flexibility Act,

FDA does not believe that this proposal will have a significant effect

on a substantial number of small entities, but recognizes the

uncertainty of its estimates. Therefore, the agency has prepared an

IRFA. FDA is not required to conduct a cost-benefit analysis according

to the Unfunded Mandates Reform Act, because the rule will not impose

any mandates on State, local, or tribal governments, or the private

sector, that will result in an annual expenditure of $100 million or

more.

Furthermore, in accordance with the Small Business Regulatory

Enforcement Fairness Act of 1995 (Public Law 104-121), it has been

determined that this proposed rule would be a major rule for the

purpose of congressional review.

A. Objectives of the Proposed Regulations

The objectives of this proposed regulation are to reduce the

adverse health effects from allergic and foreign body reactions caused

by the NL protein allergens and glove powder found on surgeon's and

patient examination gloves, and from defects in the barrier integrity

and quality of surgeon's and patient examination gloves. The rule will

accomplish these objectives by encouraging manufacturers to limit both

the level of water-extractable protein allowed on gloves and the level

of powder packaged with the gloves, and by requiring the inclusion of

caution statements and the actual level of protein and powder in the

labeling of the gloves. In addition, labeling will include expiration

dates to ensure that the gloves provide adequate barrier protection and

that all medical gloves meet quality standards specified in the special

control guidance referenced elsewhere in this preamble. FDA believes

that by reducing the amount of powder dispersed, these special controls

will reduce the incidence and severity of the allergic reactions caused

by NL proteins without compromising the barrier performance of these

products.

B. Risks of NL Protein Allergic Reactions

FDA recognizes that no systemic epidemiological data exist to

identify the risk of airborne NL protein allergens. However, several

sources indicate that a proportion of the U.S. population have

developed NL sensitivity (Refs. 1 to 8) due to increased exposure to NL

proteins. The increased use of NL gloves with unlimited powder and

protein levels in recent years is believed to contribute to these

adverse events.

FDA's Adverse Experience Reporting System received a total of 330

NL allergy Medical Device Reports (MDR's) associated with medical

gloves for the 12-month period of August 15, 1996, through August 15,

1997 (Ref. 9). These reports included reactions of 435 affected

persons. Despite the lack of representative sampling and the

unconfirmed nature of these reports, FDA believes these data may

provide a reasonable measure of the magnitude of existing risk. Table 1

classifies these reports by type and severity of reaction and shows the

results by number of affected patients.

Table 1.--Number of Patients Reporting to FDA Natural Rubber Latex Allergies Reactions Associated with Medical Gloves Between August 15, 1996, and

August 15, 1997

--------------------------------------------------------------------------------------------------------------------------------------------------------

Type of Allergic Reaction1

-------------------------------------------------------------------------------------------------------

Respiratory Requiring

Systemic Topical (i.e., Systemic Respiratory Aggressive Treatment

Local Topical rash not in area with (e.g., wheezing, (e.g., anaphylaxis,

direct contact) shortness of breath) hospitalization)

--------------------------------------------------------------------------------------------------------------------------------------------------------

Number of Patients Reporting Reaction 20 21 294 100

--------------------------------------------------------------------------------------------------------------------------------------------------------

\1\Includes 40 patients with unclassified reactions that were distributed by proportion of reported reactions.

FDA has long been aware that MDR's received by the agency may

account for as little as one percent of the actual events (Ref. 37). If

true, the reports received for allergic reactions associated with

medical gloves could represent as many as 43,500 allergic incidents

during the 12-month period. Because patients may often fail to connect

an allergic incident to use of gloves, FDA believes that this estimate

better reflects the true number of incidents associated with medical

gloves. Given that approximately 22.0 billion gloves (Ref. 38) were

used and 2.16 billion patient visits occurred during that period (Ref.

39), the projected baseline rate of annual allergic reaction incidents

to the total population (0.0001626) at current protein/powder levels

does not seem unreasonable.

Despite the widespread under-reporting cited in the General

Accounting Office (GAO) report, FDA believes that those allergic

reactions that require the most aggressive treatment would be subject

to less under-reporting. For this analysis, FDA has assumed that MDR's

for patients with severe allergic reactions are under-reported by 33

percent, and the other three categories are proportionally

[[Page 41718]]

increased to account for the total under-reporting (Table 2).

Specifically, FDA believes that the 100 reported incidents of

respiratory allergic reactions requiring aggressive treatment (from

Table 1) represent only 150 actual such incidents; not 1,000 as would

be indicated by MDR underreporting. The difference of 850 expected

incidents were distributed to the remaining three categories to result

in 43,500 total incidents. Table 2 also shows the proportion of each

category of reactions reporting long-term and short-term effects, based

on reported lost work-time due to recovery. As expected, only 6 percent

of all topical local reactions were considered long-term, while almost

half of the serious systemic reports were long-term. As discussed in

the benefits section (section IX.F of this document), FDA has assumed,

based on discussions with clinicians, that short-term impacts have a

duration of 1 day and long-term impacts a duration of 2 months.

Table 2 also presents FDA's estimated annual number of each type of

allergic reaction. Although no mortalities were reported in the MDR's

for this period, anaphylaxis carries a risk of mortality that FDA

statisticians place at up to 2 percent, even in health care settings.

Because not all reported serious systemic respiratory reactions were

anaphylaxis, FDA assigned a probability of 0.002 to the adjusted

reports to account for potential fatalities due to anaphylactic shock

caused by NL allergens. (This assumes that only 10 percent of all

respiratory reactions that require aggressive treatment were due to

anaphylaxis.) Given the estimated under-reporting rate, this implies an

annual risk of 0.3 mortalities. FDA expects that by encouraging lower

protein and powder levels for medical gloves, the proportion of

allergic reactions to NL protein allergens will be reduced.

Table 2.--Estimated Number of Patients Experiencing Natural Rubber Latex Allergic Reaction Associated With Medical Gloves From August 15, 1996, to

August 15, 1997, and Proportion Experiencing Short- and Long-Term Effects

--------------------------------------------------------------------------------------------------------------------------------------------------------

Type of Allergic Reaction

---------------------------------------------------------------------------------------------------

Respiratory Requiring Aggressive

Systemic Treatment

Local Topical Systemic Topical Respiratory ---------------------------------------

Other Reactions Mortality

--------------------------------------------------------------------------------------------------------------------------------------------------------

Estimated Number of Patients Experiencing Reaction 2,588 2,717 38,045 149.7 0.3

Proportion Exhibiting Short-Term Effects (duration 94% 74% 73% 51% NA

of 1 day)

Proportion Exhibiting Long-Term Effects (duration of 6% 26% 27% 49% NA

2 months)

--------------------------------------------------------------------------------------------------------------------------------------------------------

C. Costs of the Proposed Regulation

This section develops estimates of the costs of compliance with the

proposed rule by comparing the expected costs of using surgeon's and

patient examination gloves after the rule is in effect to the costs

that would have been incurred in the absence of the rule. Regulatory

costs occur in four categories. First, the proposed regulation is

expected to accelerate the trend of the glove market towards more

costly, powder-free products. Second, higher average glove purchase

prices will result from the increased cost of gloves with recommended

levels of powder and NL proteins compared to the cost of gloves with

unregulated levels of powder and NL proteins. Third manufacturers will

be required to conduct shelf-life testing on gloves in order to support

expiration dates. Fourth, increased labeling costs will result from the

addition of protein and powder levels and/or expiration dating to each

package of surgeon's and patient examination gloves. Because many of

the estimates are derived from uncertain projections based on limited

data, sensitivity analyses are presented for the most critical

variables and assumptions.

D. Baseline Conditions

1. Annual Number of Gloves

To measure the incremental costs of the regulation against a

baseline of nonregulation, FDA first projected future glove sales. An

estimated 22.0 billion surgeon's and patient examination gloves were

used in the United States in 1997, more than an 11-fold increase from

the approximately 2.0 billion gloves used in 1987 (Ref. 38). The major

contributors to this growth were the recognition of the potential risk

from AIDS infection and the publication of Occupational Safety and

Health Administration (OSHA) regulations requiring barrier protection

for patients and employees exposed to blood borne pathogens (Ref. 40).

FDA assumed that the demand for surgeon's and patient examination

gloves will continue to grow as a result of expected increases in

employment within the health services industry (Standard Industrial

Classification (SIC) 80). The Bureau of Labor Statistics has suggested

that employment within this industry may continue to grow at an annual

rate of 3.9 percent (Ref. 41). Assuming that annual glove use per

employee remains at current levels of approximately 10 pairs per day,

the agency projected that the annual demand for gloves will increase

over the next 10 years at an approximate rate of 3.9 percent per year

(see Table 3). As expected growth in employment or patient health

service visits may also predict future glove use. FDA tested this

assumption by forecasting alternative rates of growth in the

sensitivity analyses presented in section IX.G of this document.

About 65 percent of the current glove market consists of powdered

gloves (Ref. 38), but both health service facilities and glove

manufacturers agree that the market share of powdered gloves is

decreasing rapidly as facilities gain awareness of the potential

adverse health effects associated with NL protein allergens.

Manufacturers, however, explain that powdered gloves will not soon

disappear, because new chlorinators and production lines associated

with powder-free glove production take at least 18 months to

[[Page 41719]]

install and because powdered gloves are still desired by a proportion

of customers. However, manufacturers have estimated that even in the

absence of this regulation, the market share of powder-free gloves

could reach as high as 60 percent within 18 months (Ref. 38). For this

analysis, FDA assumed that, even in the absence of regulation, the

market share for powdered gloves would decrease from the current 65

percent down to 20 percent within 4 years. Concurrently, the market

share for powder-free gloves would increase from 35 percent up to 80

percent over the same period (see Table 3).

Next, FDA estimated that gloves manufactured with synthetic

materials (referred to as synthetic gloves), which are available in

both powdered and powder-free varieties, account for approximately 10

percent of the current market. Most synthetic gloves are manufactured

of vinyl, but other polymers are also used. Synthetic gloves are

generally believed to provide less acceptable barrier protection after

extended use and reduced tactile sensitivity compared to NL. FDA

assumed that, in the absence of regulation, this market share would

increase slightly each year, accounting for 20 percent of the market

within 5 years. Table 3 includes the projected market shares for each

glove type.

Because these projections contain considerable uncertainty, FDA

analyzed several alternative assumptions in the sensitivity analysis

section presented in section IX.G of this document. These scenarios

assume that, in the absence of this rule, the anticipated baseline

market adjustments would take either 10 years, or would not occur at

all.

[[Page 41720]]

Table 3.--Surgeon's and Patient Examination Glove Market Shares--Baseline Estimate

--------------------------------------------------------------------------------------------------------------------------------------------------------

All Surgeon's and Synthetic Gloves Natural Rubber Latex Gloves

Patient -----------------------------------------------------------------------------------------------------------------

Examination

Year Gloves Number of Number of

------------------- Number for Powder- Powdered Total (billion) Number of Powder- Powdered Total (billion)

Number of Gloves Free (billion) (billion) Free (billion) (billion)

(billion)

--------------------------------------------------------------------------------------------------------------------------------------------------------

Current 22.00 0.77 1.43 2.20 6.93 12.87 19.80

1 22.86 1.37 1.37 2.74 10.06 10.06 20.12

2 23.75 1.99 1.33 3.32 12.25 8.17 20.42

3 24.68 2.76 1.18 3.95 14.51 6.22 20.73

4 25.64 3.69 0.92 4.61 16.82 4.20 21.02

5 26.64 4.26 1.07 5.33 17.05 4.26 21.31

6 27.68 4.43 1.11 5.54 17.71 4.43 22.14

7 28.76 4.60 1.15 5.75 18.40 4.60 23.00

8 29.88 4.78 1.20 5.98 19.12 4.78 23.90

9 31.04 4.97 1.24 6.21 19.87 4.97 24.83

10 32.25 5.16 1.29 6.45 20.64 5.16 25.80

--------------------------------------------------------------------------------------------------------------------------------------------------------

[[Page 41721]]

2. Baseline Glove Cost

There are an estimated 198 current marketers of surgeon's and

patient examination gloves in the United States, 10 of which are

domestic manufacturers. Approximately 95 percent of all gloves

purchased in the United States are imported. Although individual

marketers of surgeon's and patient examination gloves operate in a

highly competitive industry and face highly elastic demand curves, the

aggregate market demand for the gloves is assumed to be price

inelastic, because of workplace regulations that require gloves as

barrier protection (Ref. 42). Demand is inelastic if the percentage

increase in price exceeds the percentage decrease in quantity sold.

Consequently, most glove manufacturing cost increases would be passed

on to health care facilities in the form of industry wide price

increases. Although over 95 percent of the manufacturing facilities are

located overseas and the world wide demand for gloves is high, the

United States market dominates global sales. According to Malaysian

manufacturers (Ref. 38), about 80 percent of their gloves are for U.S.

customers.

Current prices of powdered NL gloves average $3.90 per 100, while

powder-free NL gloves average $5.80 per 100 (Ref. 38). Prices were

reported as averages of both surgeon's and patient examination gloves.

The price difference of $1.90 per 100, or almost $.02 per pair, is

attributable to a number of factors, but the predominant reason is the

increased cost of removing former-release powder and/or applying other

lubricants to produce powder-free gloves. The estimated cost for

synthetic gloves is $4.15 per 100 for powdered and $5.03 per 100 for

powder-free. Vinyl gloves account for 90 percent of the synthetic glove

market, with the remaining gloves manufactured from polymers and other

materials.

The nation's annual expenditures for surgeon's and patient

examination gloves are currently estimated at over $1.0 billion. Even

in the absence of regulation, FDA expects that these outlays would

increase to $1.1 billion within 1 year and $1.7 billion within 10

years.

E. Estimation of Compliance Costs

The net costs of compliance with the proposed regulation is the

difference between glove-related costs with and without the regulation.

As noted earlier, industry comments suggest that even in the absence of

this regulation, the market share of powder-free gloves is expected to

increase from 35 percent to about 80 percent over a 4-year period. With

regulation, this trend will be accelerated. Although the market effects

of the rule cannot be known with certainty, FDA estimates that powder-

free gloves will achieve the 80 percent market share 2 years earlier,

or within 2 years of the rule's implementation. In addition,

manufacturers would experience increased costs due to the

recommendation to limit the level of protein to 1,200 g per glove and

the level of powder on NL and synthetic powdered gloves to 120 mg per

glove. These costs would be passed through to health care facilities in

the form of higher prices. Finally, each package of NL gloves must

include labeling that includes protein and powder levels and expiration

dating, and shelf-life testing must support this labeling.

1. Accelerated Market Share for Powder-Free and Synthetic Gloves

Figure 1 illustrates FDA's forecast that powder-free gloves would

gain 80 percent of the surgeon's and patient examination glove market

share within 4 years without regulation and within 2 years with

regulation. Manufacturers have indicated (Ref. 38) that if U.S.

facilities are willing to bear the market price for powder-free gloves,

the powder-free supply to other parts of the world could be shifted to

meet U.S. demand and powder-free market shares could reach as high as

60 percent within 18 months. FDA forecasts that the proposed

regulations will accelerate this trend by reinforcing incentives for

facilities to use powder-free gloves. The shaded area of the chart

measures the expected substitution of powder-free for powdered gloves

caused by facilities choosing to increase use of powder-free gloves in

response to regulatory controls. In addition, FDA projects that the

synthetic market share will rise from 10 to 20 percent within 5 years

without regulation, but within 2 years with regulation. The expected

market shares with the proposed regulation in place are shown in Table

4.

FDA also examined the potential of this regulation to result in

domestic shortages of latex gloves and concluded that there would be

minimal disruption to the U.S. market, as it constitutes such a major

proportion of global sales (up to 80 percent (Ref. 38)). If other

countries do not restrict glove powder, it is possible that the number

of powder-free gloves sold in those markets would fall in the short-

term, while producers adjusted to the demand shift. FDA solicits public

comment on how manufacturers would respond to these altered market

forces.

[[Page 41722]]

Table 4.--Surgeon's and Patient Examination Glove Market Shares with Regulation

--------------------------------------------------------------------------------------------------------------------------------------------------------

Surgeon's and Synthetic Gloves Natural Rubber Latex Gloves

Patient -----------------------------------------------------------------------------------------------------------------

Examination

Year Gloves Number of Number of

------------------- Number of Powder- Powdered Total (billion) Number of Powder- Powdered Total (billion)

Total Number Free (billion) (billion) Free (billion) (billion)

(billion)

--------------------------------------------------------------------------------------------------------------------------------------------------------

Current 22.00 0.77 1.43 2.20 6.93 12.87 19.80

1 22.86 1.37 1.37 2.74 10.06 10.06 20.12

2 23.75 2.47 1.33 3.80 12.97 6.98 19.95

3 24.68 3.95 0.99 4.94 15.79 3.95 19.74

4 25.64 4.10 1.03 5.13 16.41 4.10 20.51

5 26.64 4.26 1.07 5.33 17.05 4.26 21.31

6 27.68 4.43 1.11 5.54 17.71 4.43 22.14

7 28.76 4.60 1.15 5.75 18.40 4.60 23.00

8 29.88 4.78 1.20 5.98 19.12 4.78 23.90

9 31.04 4.97 1.24 6.21 19.87 4.97 24.83

10 32.25 5.16 1.29 6.45 20.64 5.16 25.80

--------------------------------------------------------------------------------------------------------------------------------------------------------

[[Page 41723]]

Because the regulation would not be implemented until 2 years after

publication of the final rule (as shown in Figure 1), no costs would be

incurred in the first year. Moreover, there would be no market share-

associated costs expected after the fourth year, because, by that time,

there would be no difference in the respective market shares of

powdered and powder-free gloves. Based on these assumptions, the

accelerated increase in the powder-free market share results in

increased regulatory costs of $18.9 million in the second year and

$37.3 million in the third year. In the fourth year following

implementation of the rule, costs would fall by $2.9 million due to the

increased use of lower cost synthetic gloves. As shown in Table 5, the

average annualized costs (at a 7 percent discount rate over a 10-year

period) attributable to the accelerated market share for powder-free

gloves are calculated at $6.4 million.

BILLING CODE 4160-01-F

[[Page 41724]]

[GRAPHIC] [TIFF OMITTED] TP30JY99.004

BILLING CODE 4160-01-C

[[Page 41725]]

2. Increased Costs for Powdered Gloves

Limiting the amount of protein and powder permitted on gloves would

increase the production cost and therefore raise the purchase price of

gloves to health care facilities. Although the limits are only

recommended, both the actual and recommended levels of protein and

powder must be included on the product label. Thus, FDA believes it

likely that most glove manufacturers will meet the recommended levels.

According to tests conducted by FDA, current powder levels on powdered

gloves vary between 50 mg and 426 mg per glove. For this analysis, FDA

assumed that a typical powdered NL or synthetic glove contains 260 mg

of powder (based on the observed distribution). Current glove protein

levels vary widely.

Several manufacturers indicated that even minimal recommendations

on powder and protein would result in cost increases of as much as five

percent. These increases would be due to the increased testing and

validation required to ensure that gloves did not exceed limits, the

slower production times resulting from more controlled processes, the

increased inventory damage when stripping gloves from molds, the

increased controls for slurry mixtures, the increased time spent

cleaning or replacing filters and other equipment, and the other costs

associated with more careful controls for the entire manufacturing

process. Manufacturers stated that limiting powder is more a question

of adding controls in the production process than adding new production

lines or facilities. Equipment such as slurries and tumblers are

currently in place, and controls are likely to consist of simply

weighing finished gloves or weighing the slurry filters. However, these

costs are expected to result in increased contract prices for U.S.

health facilities, because there are no substitute products for medical

gloves.

To calculate the costs of alternative permissible powder limits,

FDA estimated an average cost function where the cost of reducing each

mg of powder increases as the proportion of powder remaining on the

manufactured glove decreases. Because current powdered NL gloves cost

$3.90 per 100 and powder-free gloves cost $5.80 per 100, FDA calculated

that the $1.90 cost of removing the average 260 mg of powder per 100

gloves is about $0.0073 per mg ($1.90/260 mg). If the cost function

were linear, the incremental cost of reducing powder levels by 140 mg

(i.e., from the current average 260 mg of powder to the recommended

level of 120 mg) would be calculated as $0.0073 times 140, or $1.022

per 100 gloves. However, FDA believes that the relationship is unlikely

to be linear as several manufacturers indicated that significant

control costs would be needed to achieve even modest reductions in

powder levels, after which average costs would rise slowly and then

more steeply as powder concentrations approach zero. Such a functional

form is typical of many manufacturing processes and illustrated by the

solid sigmoid curve shown in Figure 2 (Refs. 44 and 45). A cost

equation fitting this illustrated functional form is:

Y = 0.00365 + 0.0292(X - 0.5)3

Where:

Y equals the cost per mg removed per 100 gloves, and

X equals the proportion of powder removed.

Figure 2 includes the estimated cost function for removing powder

from synthetic gloves as the hashed line. The expected costs per mg

removed are less than for NL gloves because the current price

difference between powder-free and powdered synthetic gloves ($0.88 per

100) is less than the difference for NL gloves ($1.90 per 100).

On the assumption that these equations approximate the actual

relationships, FDA estimates that the cost of limiting powder to 120 mg

per 100 NL gloves is about $0.003652 per mg removed, or about $0.511

per 100 NL gloves. For synthetic gloves, the estimated costs are

$0.001693 per mg removed, or about $0.237 per 100 synthetic gloves. As

shown in Figure 3, the control costs rise sharply for limits below 120

mg. For example, a proposed powder limit of 100 mg per NL and synthetic

glove would result in costs over 15 percent greater than the proposed

120 mg limit. Because of the control processes required, FDA assumes

that the previous estimates would also account for the cost of limiting

protein levels for NL gloves.

Table 5 shows these estimated costs over a 10-year period. Because

the regulation is expected to be implemented 2 years after publication

of the final rule, no increased powdered glove costs are incurred in

the first year. In year 2, the higher prices for powdered NL gloves

result in increased costs of $35.7 million. In year 3, these costs fall

to $20.2 million. Thereafter, the yearly incremental compliance costs

associated with NL glove powder and protein limits vary between $21.0

and $26.4 million. The average annualized contribution of this cost

category (at a 7 percent discount rate over 10 years) equals $21.4

million.

Within 2 years, higher costs for powdered synthetic gloves will

equal $3.1 million. The yearly incremental compliance cost for powdered

synthetic gloves is expected to decrease to $2.3 million in year 3, and

then increase slightly each year throughout the evaluation period. The

average annualized contribution of this cost category (at a 7 percent

discount rate over 10 years) equals $2.4 million.

BILLING CODE 4160-01-F

[[Page 41726]]

[GRAPHIC] [TIFF OMITTED] TP30JY99.005

[[Page 41727]]

[GRAPHIC] [TIFF OMITTED] TP30JY99.006

BILLING CODE 4160-01-C

[[Page 41728]]

Table 5.--Compliance Costs Over 10-Year Period

--------------------------------------------------------------------------------------------------------------------------------------------------------

Cost of Increased Cost of Cost of Shelf-Life Testing

Accelerated Powdered NRL Synthetic -------------------------------- Labeling Cost Total Cost ($

Year Market Share Gloves ($ Gloves ($ Test Cost ($ Lost Inventory ($ million) million)

($ million) million) million) million) ($ million)

--------------------------------------------------------------------------------------------------------------------------------------------------------

1 0.0 0.0 0.0 0.0 0.0 0.0 0.0

2 18.9 35.7 3.1 1.6 3.0 1.6 63.9

3 37.3 20.2 2.3 1.2 1.3 0.6 62.9

4 2.9 21.0 2.4 1.2 1.1 0.5 23.3

5 0.0 21.8 2.5 1.3 1.4 0.7 27.7

6 0.0 22.6 2.6 1.3 1.2 0.5 28.2

7 0.0 23.5 2.7 1.5 1.1 0.6 29.4

8 0.0 24.4 2.8 1.5 1.6 0.8 31.1

9 0.0 25.5 2.9 1.4 1.3 0.6 31.7

10 0.0 26.4 43.1 1.5 1.3 0.6 32.9

Total 53.3 221.0 24.6 12.5 13.3 6.5 331.2

Average Annualized (7 percent discount 6.4 21.4 2.4 1.2 1.3 0.7 33.4

rate)

--------------------------------------------------------------------------------------------------------------------------------------------------------

3. Costs of Shelf-life Testing and Inventory Loss

The proposed regulation will require manufacturers of patient

examination and surgical gloves to develop and affix labeling to their

products that will include expiration dating. To ensure that medical

gloves will maintain adequate barrier protection for the entire stated

period, manufacturers will likely conduct real-time shelf-life testing

of gloves. The compliance costs of this testing includes both the

actual cost of conducting laboratory tests, and the lost revenues of

inventory lost due to sampling.

a. Shelf-life testing. FDA contracted with the Eastern Research

Group (ERG), an economic consulting firm, to contact domestic and

foreign glove manufacturers and research laboratories to determine the

expected unit costs of shelf-life testing, and to determine current

levels of industry compliance. ERG developed a cost model that

estimated compliance costs according to the size of the manufacturer

(Ref. 45a).

ERG estimated that the expected marketing life for each glove model

is approximately 3 years. During this period, stability testing is

likely to occur at 6-month, 1-year, 2-year, and 3-year intervals. The

actual tests were assumed to consist of a combination of real-time and

accelerated tests. Overall, the estimated costs of a shelf-life test

was found to approximate $265 for foreign tests and $865 for domestic

tests. (The difference in testing costs are attributable to the lower

purchasing power parity per capita in foreign countries that produce

medical gloves.)

As explained in Ref. 45a, almost 3,000 separate glove models are

currently produced by 198 separate manufacturers. Only 160 models are

marketed by the 10 domestic manufacturers. Given the expected growth in

the demand for gloves, and the shift to powder-free and synthetic glove

models, the estimated costs of shelf-life testing varies with FDA's

projected number of future glove models. It was assumed that new models

would have two shelf-life tests during the year of introduction while

models already marketed would have one annual shelf-life test. Finally,

ERG and industry sources estimated the current level of shelf-life

testing based on both domestic/foreign and size characteristics.

Based on these assumptions, the greatest increase in shelf-life

testing is expected during year 2, with over 6,000 additional tests due

to this proposed regulation. The total cost of conducting these tests

equals $1.6 million, of which $0.1 million is incurred by domestic

glove manufacturers. Amortizing the annual testing costs by 7 percent

over 10 years, the average annualized costs of conducting the required

shelf-life tests equals $1.2 million.

b. Inventory losses. As part of these tests, manufacturers will be

required to set inventory aside from which test samples will be

selected. ERG, with discussions with laboratories and manufacturers,

has determined that small glove manufacturers would be likely to set

10,000 gloves per model aside for shelf-life testing while large

manufacturers would set 30,000 gloves per model. Given the industry

characteristics as discussed in Ref. 45a, this implies that over 115

million gloves would be set aside in year 2. In addition, the relative

market shares of synthetic, NL, powdered and powder-free gloves is

expected to change over time which will affect the average lost revenue

per sample. FDA analyzed the impact of this future inventory loss and

found that during year 2 of the evaluation period, the value of lost

inventory for testing is expected to equal over $3.0 million for the

entire industry. The average annualized cost of this lost inventory (as

shown in Table 5) at 7 percent over 10 years equals $1.3 million.

4. Costs of Labeling. ERG also developed estimates of the costs of

developing the proposed enhanced labeling for gloves. These estimates

included the costs of artwork, design, regulatory review, production

and application, as shown in Ref. 45a. Overall, the average cost of

developing a label for a foreign medical glove model was estimated to

equal $411, while a domestic model would cost $1,444. The number of

domestic and foreign glove models expected to be introduced throughout

the 10-year evaluation period and the market characteristics as

discussed in Ref. 45a, indicate that the costs of labeling will equal

$1.4 million in year 2. These yearly costs will then decrease to as low

as $0.3 million by the 10th year. The average annualized cost of

developing and producing labeling for medical gloves attributable to

this proposed regulation is estimated to equal $0.7 million, as shown

in Table 5.

5. Total Incremental Costs

Figure 4 presents the estimated annual expenditures imposed by the

proposed rule. Overall, costs of $63.9 million are expected in year 2.

These costs decreased to $62.9 million in year 3, and then decrease to

$23.3 million in the third year. Costs are expected to

[[Page 41729]]

increase slightly for each subsequent year. Most of the incremental

costs, as shown in Table 5, are due to increases in glove costs

(powdered NL and synthetic gloves with limited powder levels). The

estimated average annualized cost over a 10-year period (at a 7 percent

discount rate) is $33.4 million.

BILLING CODE 4160-01-F

[[Page 41730]]

[GRAPHIC] [TIFF OMITTED] TP30JY99.007

[[Page 41731]]

[GRAPHIC] [TIFF OMITTED] TP30JY99.008

BILLING CODE 4160-01-C

[[Page 41732]]

F. Benefits of the Proposed Regulations

1. Expected Risk Reduction

As discussed previously, the estimated annual proportion of the

population (0.0001626) that experiences allergic reactions associated

with medical gloves is assumed to be related to the prevalence of

environmental protein and powder. Consequently, reducing protein and

powder levels would reduce the proportion of the population expected to

experience an allergic reaction. Decreases would be expected in NL

sensitization as well as allergic reactions.

To estimate this relationship, FDA assumed that the proportion of

the population affected would vary directly with the total quantity of

environmental protein/powder. The annual level of environmental

protein/powder was calculated from the expected annual number of

powdered NL gloves multiplied by the average level of powder per glove.

The current market share of powdered NL gloves (Table 3) and the

current average level of glove powder (260 mg) yield an aggregate

estimate of 3.346 billion g of protein/powder. This quantity of

protein/powder is associated with allergic reactions in 0.0001626 of

the population, or 0.000049 reactions per billion g. If the

relationship between the number of reactions and the quantity of

protein/powder were linear, the model implies a 30 percent reduction in

allergic prevalence for each billion g of powder reduction.

Alternatively, the function relationship may take other forms, and FDA

suspects that the increasing number of reports of allergic reactions to

NL in recent years likely indicates a nonlinear relationship. Figure 5

presents a polynomial projection that FDA tentatively adopts as a

plausible estimate for this analysis. The equation of the function

illustrated in Figure 5 is:

Y = (0.0000143)X2

Where:

Y equals the proportion of the population with NL allergic

reactions, and

X equals the level of environmental protein/powder (in billions of

g).

Although the exact relationship is speculative, FDA believes that

an exponential relationship as shown in Figure 5 is most likely. As

shown in section IX.G of this document, the agency's sensitivity

analysis indicates that due to the rising baseline projection, this

polynomial projection yields smaller benefits than a linear model.

Table 6 shows the expected number of allergic reactions associated

with protein/powder levels with and without the proposed regulation.

The protein/powder amounts are derived from the expected numbers of

powdered NL gloves shown in Tables 3 and 4, the current average glove

powder level (260 mg per glove), and the new recommended glove powder

level (120 mg per glove). Powdered synthetic gloves do not affect this

relationship because no NL proteins are associated with those products.

Table 6 shows that in the absence of the proposed regulation, the

expected increased market share of powder-free gloves would reduce the

number of annual allergic reactions attributable to medical gloves from

43,500 to only 4,800 within 4 years. With the proposed regulation in

place, the expected number of allergic reactions would decrease to only

900 within 3 years, and consistently remain several thousand fewer than

those expected without regulations.

2. Benefits

To estimate the potential benefits of the proposed rule, the number

of reduced expected allergic reactions shown in Table 6 were

distributed in proportion to the categories shown in Table 2. Assuming

that the decreased number of reactions would not modify the severity

distribution as reported in the MDR's (as adjusted to account for

under-reporting), the proposed regulation would reduce annual allergic

reactions by 15,100 within 2 years. The characteristics of these second

year avoided reactions are shown in the first four columns of Table 7.

[[Page 41733]]

Table 6.--Expected Number of Allergic Reactions

--------------------------------------------------------------------------------------------------------------------------------------------------------

In the Absence of Regulation1 With Regulation2

------------------------------------------------------------------------------------------------------------------ Difference in

Number of Number of Allergic

Year Powdered Natural Level of Powder Estimated Number Powdered Natural Level of Powder Estimated Number Reactions with

Rubber Latex (billion g) of Allergic Rubber Latex (billion g) of Allergic Regulation (000)

Gloves (billion) Reactions (000) Gloves (billion) Reactions (000)

--------------------------------------------------------------------------------------------------------------------------------------------------------

Current 12.87 3.35 43.5 - NA - - NA - - NA - - NA -

1 10.06 2.62 26.9 10.06 2.62 26.9 0.0

2 8.17 2.13 17.9 6.98 0.84 2.8 (15.1)

3 6.22 1.62 10.4 3.95 0.47 0.9 (9.5)

4 4.20 1.09 4.8 4.10 0.49 1.0 (3.8)

5 4.26 1.11 5.0 4.26 0.51 1.1 (3.9)

6 4.43 1.15 5.4 4.43 0.53 1.2 (4.3)

7 4.60 1.20 5.9 4.60 0.55 1.3 (4.7)

8 4.78 1.24 6.4 4.78 0.57 1.4 (5.1)

9 4.97 1.29 7.0 4.97 0.60 1.5 (5.5)

10 5.16 1.34 7.6 5.16 0.62 1.6 (6.0)

--------------------------------------------------------------------------------------------------------------------------------------------------------

\1\ Powder level of 0.260 g per glove

\2\ Powder level of 0.120 g per glove

[[Page 41734]]

There is no methodology that permits a precise assignment of

monetary values to regulatory health benefits. However, one approach

recently described in the health economics literature (Refs. 47 and 48)

combines relative states of well-being with observed willingness to pay

for risk avoidance. FDA adopted this methodology and used the Kaplan-

Bush Indices of Well-Being (Refs. 49 and 50) to estimate the value of

reducing the number of allergic reactions.

The first step was to assign to each category of reaction a

functional index based on mobility/physical/sociability scales. The

index of relative well-being (as described in Refs. 49 and 50) utilize

functionality levels as a basis for estimating well-being. The

functionality scales are described in Table 8. Baseline levels of well-

being were defined for 43 distinct combinations of mobility, physical

activity, and sociability. For example, if a hypothetical patient could

drive a car and use transportation without help (mobility equals 5),

could walk without a physical problem (physical activity equals 4), and

had no morbidity symptoms or problem, then this patient would have an

assigned well-being of 1.0000. However, if this hypothetical patient

could perform all of these activities, but suffered from any morbidity

(including requiring eyeglasses), the assigned baseline level of well-

being was found to equal 0.7433. The baseline levels of well-being are

then adjusted, either up or down, based on the predominant symptom or

problem that is on-going. This methodology is described in detail in

Refs. 49 and 50. For example, a local topical reaction is unlikely to

interfere with normal activities, such as driving a car or performing

housework. A patient suffering from a local topical reaction is

expected to continue to be able to interact with others in a normal

manner. This functional state is assigned a relative well-being rate of

0.7433, or roughly 74 percent of optimum well-being. This baseline

functional index is based on the prevailing medical problem. In this

case, the problem/symptom is identified as ``burning and/or itching of

skin'' and the 0.0171 value for this problem/symptom (from Refs. 49 and

50) is added to the basic functional state. Thus, by combining these

indices, a person suffering a local, topical allergic reaction is

expected to have a relative well-being of 0.7604. Each of the

categories of reactions have been assigned values, as included in Table

7. Mortalities are valued as 0.0000.

Next, optimum values of well-being were derived for both short-term

durations (1 day) and long-term durations (2 months). The economic

literature includes many attempts to quantify society's willingness-to-

pay (WTP) to avoid risks. Various methodologies have resulted in an

average value of approximately $5.0 million as a measure of the WTP to

avoid a statistical death (Refs. 51, 52, and 53). By amortizing this

value to account for life expectancy and expected disability-days

(Refs. 54 and 55), FDA estimates that a quality-adjusted life-year

(QALY) has an approximate value of $373,000. Using this estimate, the

expected value of a quality-adjusted life-day is approximately $1,022

and the expected value of two quality-adjusted life-months is $62,166.

The relative wellness values for each category shown in Table 7

represent the proportion of wellness relative to an optimum level. The

willingness of society to pay for avoiding each incident were reflected

as the difference between the wellness state and an optimum level

multiplied by the duration of the event. For example, a local topical

allergic reaction has an expected wellness value of 0.7604, or 0.2396

below optimum. This difference is used to calculate the amount that

society is willing to pay to avoid a reaction of this type.

[[Page 41735]]

Table 7.--Characteristics of Reductions in Second Year Allergic Reactions

--------------------------------------------------------------------------------------------------------------------------------------------------------

Number of Number of Value per Value per

Category Number of Avoided Avoided Problem/ Value per Value per Short-Term Long-Term

of Avoided Short-Term Long-Term Functional Symptom Relative Short-Term Long-Term Reaction Reaction Total Value

Reaction Reactions Reactions Reactions State\1\ Weight\2\ Wellness Reaction Reaction ($ Avoided ($ Avoided ($ ($ 000)

(000) (000) (000) 000) 000) 000)

--------------------------------------------------------------------------------------------------------------------------------------------------------

Local 0.89 0.84 0.05 0.7433 0.0171 0.7604 245 14,900 205 794 999

Topical

Systemic 0.94 0.69 0.24 0.6065 0.0171 0.6236 385 23,400 267 5,709 5,976

Topical

Systemic 13.24 9.66 3.57 0.525 -0.0075 0.5175 493 30,000 4,764 107,194 111,958

Respira

tory

Respirat 0.05 0.03 0.03 0.5284 -0.1507 0.3777 636 38,700 17 972 989

ory

Requiri

ng

Aggress

ive

Treatme

nt

Mortalit 0.00 NA NA 0 0 0 NA NA NA NA 514\3\

y

Totals 15.11 11.22 3.90 120,436

--------------------------------------------------------------------------------------------------------------------------------------------------------

\1\ Functional states: Mobility - 5; Physical activity - 4; Social - 5 = 0.7433

Mobility - 4; Physical activity - 4; Social - .4 = 0.6065

Mobility - 4; Physical activity - 3; Social - 3 = 0.525

Mobility - 4; Physical activity - 1; Social - 2 = 0.5284

Mortality = 0.0000

\2\ Problem/Symptom Adjustments:

Burning or itching rash on body = +0.0171

Wheezing or shortness of breath = -0.0075

Loss of consciousness, fainting = -0.1507

\3\ Value per mortality is $5 million. May not add due to rounding.

[[Page 41736]]

For 1 day, this value is $245 ($1,022 x .2396) and for 2 months, the

estimated value per reaction is $14,895. The derived values for each of

the reaction categories and terms are shown in Table 7.

The values for each category, when multiplied by the number of

decreased reactions expected due to this regulation, result in the

expected annual benefit. Table 7 includes this estimate for only the

second evaluation year. It indicates that society would be willing to

pay a value of approximately $120.4 million to avoid 15,100 allergic

reactions to NL protein.

Taking these steps for each year in the evaluation period yields

estimates of the willingness to pay to avoid these reactions as shown

in Table 9. The undiscounted benefits equal $120.4 million in year 2,

then decrease to $30.4 million in year 4. Between years 4 and 10, the

estimated annual benefit increases to a value of $47.5 million. The

estimated annualized benefit of avoiding these reactions is $46.9

million.

FDA notes that other potential benefits, such as the avoidance of

third-party payments as a result of treating fewer allergic reactions,

the value of reduced anxiety due to lowering NL sensitization, the

reduction in defects in glove barrier integrity, and the reduction in

other foreign body reactions caused by glove powder have not been

quantified at this time. FDA recognizes the considerable uncertainty of

all of these estimates, however, and requests comment on all of the

data and assumptions.

Table 8.--Description of Inputs to Functionality Levels

----------------------------------------------------------------------------------------------------------------

Mobility Physical Activity Social Activity

----------------------------------------------------------------------------------------------------------------

5-Drove car and used transportation 5-Did work, school, or housework and

without help other activities

4-Did not drive, needed help with 4-Walked without physical problem 4-Did work, school, or housework,

transportation but other activities limited

3-In house 3-Walked with physical problem 3-Limited in work, school or

housework

2-In hospital 2-Moved own wheelchair without help 2-Performed self-care

1-In special unit 1-In bed or chair 1-Had help in self-care

----------------------------------------------------------------------------------------------------------------

Source: Kaplan, Bush, et. al. (Refs. 49 and 50)

Table 9.--Expected Benefit of Decreased NRL Allergic Reactions Due to Regulation

----------------------------------------------------------------------------------------------------------------

Net Present Value of

Year Decreased Reactions (000) Value of Decreased Reactions Decreased Reactions ($

($ million) millions)

----------------------------------------------------------------------------------------------------------------

Current NA NA NA

1 0.0 NA NA

2 (15.1) 120.4 105.2

3 (9.5) 76.0 62.0

4 (3.8) 30.4 23.2

5 (3.9) 31.1 22.2

6 (4.3) 34.0 22.7

7 (4.7) 37.1 23.1

8 (5.1) 40.3 23.4

9 (5.5) 43.7 23.8

10 (6.0) 47.5 24.2

Average Annual 46.9

Benefit ($

million)

----------------------------------------------------------------------------------------------------------------

G. Sensitivity Analyses

FDA examined the impact of various assumptions that affect future

conditions. These analyses are as follows:

1. Growth Rate of the Demand for Surgical and Patient Examination

Gloves

FDA used 1992 to 1994 rates of employment growth within the health

services industry (SIC 80) to project a 3.9 percent annual growth in

the future demand for surgical gloves (Ref. 41). However, more recent

data obtained for the period up to 1998 suggest the more modest growth

rate of 2.7 percent for this industry (Ref. 55a). Examining the

expected costs and benefits after lowering the expected growth for

surgical and patient examination gloves to 2.7 percent indicates that

average annual costs decrease from $33.4 to $31.5 million and average

annual benefits decrease from $46.9 to $42.1 million. If the forecast

relied instead on the growth of total employment hours in the health

service industry (Ref. 55b), the rate in recent years has been

approximately 2.0 percent. Using this rate as the expected growth rate

for surgical and patient examination gloves results in average annual

costs of $30.4 million and average annual benefits of $39.6 million.

FDA notes that under the alternative assumptions, both costs and

benefits are lower than under the scenario presented earlier, but the

regulation would still be justified.

2. Market Shares of Powder-Free and Synthetic Gloves

FDA has estimated that in the absence of regulation, within 4

years, 80 percent of the glove market would consist of powder-free

gloves; and within 5 years, 20 percent of all gloves would be

manufactured of synthetic material. The proposed regulation is expected

to accelerate these trends to within 2 years of implementation.

To examine the sensitivity of these assumptions, FDA calculated the

costs and benefits of the rule assuming that, in the absence of

regulation, it would take 10 years rather than 4 years for powder-free

gloves to account for 80 percent of the market and 10 years rather than

5 years for synthetic gloves to account for 20 percent of the market.

The expected average annual costs in this scenario equal $72.7 million,

and the average annual benefits equal $112.1 million. FDA also examined

the impact of assuming no expected change in baseline market share from

the first implementation year, in the absence of regulation. In this

case, the average

[[Page 41737]]

annual costs equal $135.7 million, and the average annual benefits

equal $283.2 million.

3. Linear Relationship between Environmental Protein/Powder and

Allergic Reactions

FDA expects that an exponential relationship exists between

protein/powder levels and allergic reactions, but the agency also

examined the effect of a linear relationship. The linear model

increased the expected average annual benefit of reducing exposure from

$46.9 million to $75.7 million, by increasing the number of avoided

incidents as protein/powder levels were decreased. Table 9 indicates

the magnitude of the expected decrease in NL reactions using the

expected exponential relationship. A total of 57,900 avoided reactions

were forecast. If the actual relationship were linear, the rule would

be expected to result in the avoidance of 88,100 incidents over the

same period.

4. Conclusion

FDA has tested several key assumptions used in the analysis of

impacts. Each simulation resulted in estimated benefits exceeding

costs. Nonetheless, FDA recognizes the significant uncertainty in this

analysis and requests any additional information that would improve the

projections.

H. Small Business Impact

1. Initial Regulatory Flexibility Analysis

FDA believes that the proposed regulation will not have a

significant impact on a substantial number of small entities, but

conducted an initial regulatory flexibility analysis (IRFA) to ensure

that impacts on small entities were assessed and to alert any

potentially impacted entities to the opportunity to submit comments to

the agency.

2. Description of Impact

The objectives of the proposed regulation are to reduce the adverse

health effects attributable to allergic and foreign body reactions from

NL allergens and glove powder and to defects in barrier protection and

quality of surgeon's and patient examination gloves. The proposed

regulation will accomplish these objectives by reclassifying surgeon's

and patient examination gloves into class II products, and requiring

product labeling. In addition, the proposed regulation recommends

protein and powder levels for surgeon's and patient examination gloves.

FDA's statutory authority for the proposed rulemaking under the act is

discussed in section II of this document.

Two separate industries will be affected by the proposed

regulation: Manufacturers of surgeon's and patient examination gloves

(found in Standard Industrial Classification 3842, Medical Equipment

and Supplies) and Health Facilities (found in SIC 80).

Table 10.--Non-Health Care Industries That Use Gloves as Protection

------------------------------------------------------------------------

Number of Number of

Industry Sector SIC Code Establishments Employees

------------------------------------------------------------------------

Government 9,461 10,893 56,345

Residential Care 836 2,423 NA

Personal Services 7,362 1,348 163,477

Funeral Services 726 19,890 57,013

Health Units in NA 202,540 178,732

Industry

Non-Health 8,221 1,453 89,159

Research

Laboratories

Linen Services 7,218 1,250 50,000

Medical Equipment 384 1,076 6,185

Repair

Law Enforcement 9,221 4,946 341,546

Fire and Rescue 9,224 3,174 252,048

Lifesaving 9,229 100 5,000

Schools 9,411 6,321 4,132

Waste Removal 4,953 940 13,300

------------------------------------------------------------------------

Source: OSHA (Ref. 40)

FDA considered the potential impact of the proposed regulation on a

number of nonhealth industries, but found that any impact would be

insignificant. When OSHA issued its final regulations on blood-borne

pathogens (Ref. 40), it considered a wide-range of establishments

including: Law enforcement agencies, schools, linen services, and

funeral parlors (see Table 10). While a substantial number of these

establishments are small under the Small Business Administration

definition, this proposed regulation does not require the use of FDA-

regulated medical gloves at these sites. OSHA assumed that many of

these industries would use utility gloves or consumer-grade gloves to

provide barrier protection. For example, janitorial services and waste

removal establishments were assumed to use utility work gloves, while

law enforcement agencies were expected to use consumer-grade vinyl

gloves. Few industries or establishments were expected to use FDA-

regulated medical gloves in nonmedical settings. However, even in

settings where medical gloves may be used, the frequency of glove usage

was much less in these sectors. OSHA estimated that an average school

would use approximately eight pairs of gloves per day. In contrast, a

small physician/dental office would be expected to use 30 pairs of

gloves per day. Both the relative frequency of glove use and the

concentration of FDA-regulated medical gloves convinced FDA to focus on

the Health Services Industry (Table 11) as the area of largest

potential impact.

Table 11.--Establishment and Employment in the Health Services1

----------------------------------------------------------------------------------------------------------------

Number of Number of Average Number of

Establishments and (Standard Industrial Classification Establishments Employees Employees per

Codes) (thousand) (thousand) Establishment

----------------------------------------------------------------------------------------------------------------

Total Health Services (80) 1,030.0 11,000.0 10.7

[[Page 41738]]

Clinics and Offices of MD's (801) 328.9 1,908.4 5.8

Clinics and Offices of Dentists (802) 138.5 709.4 5.1

Clinics and Offices of Osteopathy (803) 18.4 60.6 3.3

Other Health Practitioners (804) 243.0 483.6 2.0

Nursing Facilities (805) 57.7 2,011.8 34.9

Hospitals (806) 7.1 4,496.5 633.3

Medical/Dental Laboratories (807) 29.4 229.3 7.8

Home Health Services (808) 99.9 743.9 7.4

Other Allied Services (809) 107.7 356.5 3.3

----------------------------------------------------------------------------------------------------------------

\1\ 1992 Census of Service Industries and Bureau of Labor Statistics projections of employment trends in the

health services industries.

Glove manufacturers will be affected by labeling that requires

additional warnings and statements concerning recommended protein and

powder limits, testing and validation measures that are necessary to

ensure the accuracy of this information, and limitations on the use of

powder for mold release. Health facilities will face increased

expenditures for surgeon's and patient examination gloves by either

shifting from powdered gloves to more expensive powder-free products or

continuing to use powdered gloves that cost more due to production cost

increases.

Manufacturers classified within the four-digit SIC code 3842 are

typically small. Only 38 percent of all establishments had 20 or more

employees in 1992 (Ref. 56), and companies had an average of 1.12

separate establishments. The manufacturers are highly specialized, with

over 92 percent of their products considered within the medical

equipment and supplies industry, and 94 percent of all medical

equipment and supplies manufactured by these firms. The Small Business

Administration classifies as small any entity within this industry with

500 or fewer employees (Ref. 57), capturing the majority of

establishments. However, the affected manufacturers of surgeon's and

patient examination gloves have some product-specific characteristics

that distinguish them from the average establishment in this industry.

FDA's registration system for medical devices shows 198

manufacturers of surgeon's and patient examination gloves, the vast

majority of which are located outside the United States and operate in

a world-wide market, although the U.S. constitutes the most significant

regional market. FDA examined the records of current manufacturers and

identified 10 domestic manufacturers of surgeon's and patient

examination gloves out of the total 198 marketers. Only 1 of these 10

domestic manufacturers reported employment of fewer than 1,200

employees. However, FDA acknowledges that additional small domestic

manufacturers could enter the market in the future.

The main impacts of the proposed regulations on small manufacturers

would occur if the manufacturer had to conduct additional validation

tests to ensure the accuracy of protein and powder levels displayed on

the product labeling and if increased inventory loss or slower

production times occurred due to limited uses of powder as a mold

release. Although FDA does not stipulate the acceptable validation test

method in the regulation, and is soliciting comments on this issue in

order to minimize its impact, it is possible that a chemist would be

required on a contract basis to ensure that the actual levels of

protein and powder matched the levels on the label. FDA is working with

industry groups to ensure that an acceptable and reliable test method

is chosen. Despite this outreach, the selected test method could impose

additional and disparate costs to a small manufacturer. Similarly,

increased inventory loss because of tearing in the production process

due to limited powder would affect small production runs to a greater

degree than large production runs. Discussions with manufacturers have

indicated that any additional validation testing or negative impacts on

production capability could increase the production costs of medical

gloves by 5 percent or more.

As discussed earlier in the analysis of impacts section (section

IX.D of this document), the demand for medical gloves is highly price

inelastic due to the regulatory requirement for health facilities (SIC

80) to provide barrier protection (Ref. 40) and the lack of substitute

products (Ref. 42). The characteristics of the medical glove market

therefore indicate that production cost increases resulting from the

proposed rule are likely to be passed through in the form of higher

contract prices. In addition, many facilities are currently accepting

increased glove prices by establishing powder-free environments in the

absence of any rule-making. Thus, production cost increases by glove

manufacturers are likely to be offset by revenue gains for these same

manufacturers, with the result of shifting the cost impact to the

health facilities.

Small health facilities therefore will also bear some regulatory

impact. The Small Business Administration has defined as small any

``for-profit'' health facility with annual revenues of $5 million or

less (Ref. 57). Most hospitals and nursing facilities would be

considered large under this definition. However, nonprofit facilities

not dominant in their field are also considered small entities.

Industry characteristics of the health facility industry are shown in

Table 11. Approximately 95 percent of the hospitals and nursing

facilities are considered as small entities (6,700 hospitals and 54,800

nursing facilities).

FDA examined the potential impact of the proposed regulations on

two types of health care user facilities: Small physician/dental

facilities and small hospitals. A small physician or dental facility

may use as many as 25,000 (based on 120 patient visits per week) gloves

each year. If the facility substitutes powder-free for powdered gloves

as a result of this regulation, costs would increase by $475 per year

((25,000/100) x $1.90).

Similarly, a small hospital is also likely to experience increased

annual costs of acquiring gloves. An extremely small hospital with only

6 beds and a staff of 11 might use about 22,000 gloves annually. If the

facility faced increased glove costs, the total increase in costs could

amount to about $950.

FDA wishes to collect additional information on the nature of the

impacts on small entities in order to ensure that all such impacts are

noted. In addition,

[[Page 41739]]

other public facilities such as prisons, and police or fire departments

may face higher glove prices due to this regulation. FDA does not

expect these costs to be significant, but solicits comments on this

potential burden.

3. Analysis of Alternatives

FDA has examined and rejected the following alternatives to the

proposed rule:(1) Banning powdered gloves; (2) mandating protein and

powder levels on medical gloves; (3) requiring all users of powdered

gloves to comply with restrictions on distribution and use; (4)

retaining the class I classification for all (or some) of the medical

gloves; and (5) excluding powdered synthetic gloves from this

rulemaking; and (6) providing for a shorter or longer compliance

period. FDA has rejected the alternatives at this time for the

following reasons:

Alternative 1: A ban of all powdered medical gloves has been

requested in a citizen petition submitted to FDA. FDA considered

banning powdered gloves because that action would meet the stated

objective of eliminating airborne powder and greatly reducing exposure

to airborne allergens associated with the use of medical gloves.

However, FDA did not select this alternative because a ban would not

address exposure to NL allergens from medical gloves with high levels

of NL proteins. Moreover, such a ban of powdered gloves might

compromise the availability of high quality medical gloves and greatly

increase the annual costs by almost as much as $64 million over the

selected alternative.

Alternative 2: FDA also considered mandating powder and protein

levels for medical gloves because this alternative would accomplish the

stated objectives more completely than banning. FDA rejected mandating

powder and protein levels for medical gloves because the agency

believes that the increased regulatory flexibility of the proposed rule

may reduce the costs of compliance by allowing for more efficient

methods of reaching the goal. Inventories could be lowered and industry

capacity could be assured. Mandating specific protein and powder

levels, as well as the acceptable test method, may preclude all parties

from developing a more efficient system. In addition, FDA inspectional

and compliance costs are minimized by relying on recommended levels of

powder and protein. By ensuring user access to relevant information,

the agency believes that users will move the market to a more efficient

level.

Alternative 3: FDA considered restricting the distribution and use

of powdered NRL or synthetic material medical gloves by requiring that

establishments using powdered gloves establish and maintain written

procedures for selecting, purchasing and distributing gloves. FDA

further considered restricting the distribution and use of powdered NRL

or synthetic material medical gloves by requiring establishments using

powdered gloves with more than the recommended powder levels to

establish and maintain written procedures to evaluate, monitor and

control airborne particulate matter at the point of use, through the

use of an externally exhausted air handling system, HEPA filtration, or

other system. FDA believes that these restrictions would reduce the

risk of adverse foreign body and allergic reactions associated with

powdered glove use. However, the extent of the expected reduction is

uncertain. The expected costs of complying with these restrictions was

estimated to be over $21 million. Furthermore, any such workplace

restrictions may impede or preempt the authority of OSHA to regulate

gloves and glove powder in the workplace.

Before rejecting this alternative, the agency had examined the

feasibility of exempting small facilities from the requirements of

developing written procedures and air quality measures. Based on the

expectation that small establishments with 10 or fewer employees would

be able to communicate and control risks associated with powdered

medical easier than larger institutions. Exempting small medical

facilities from these controls lowers the added costs to $6.6 million.

However, FDA rejected this alternative because the expected benefits of

restricting glove use remained uncertain, and the potential overlap of

authority with OSHA would still exist.

Alternative 4: FDA considered retaining the class I classification

for all or some of the medical gloves. This alternative was rejected

because it did not meet the stated objectives. In light of new

information concerning barrier integrity, degradation of quality during

storage, contamination concerns and concerns about exposure to foreign

bodies and allergens, FDA found that general controls are no longer

sufficient to provide reasonable assurances of the safety and

effectiveness of medical gloves. Moreover, such concerns were not

limited to only powdered gloves. To require a device to meet special

controls as well as general controls, a device must be classified (or

reclassified) into class II. Consequently, although compliance costs

would have been reduced by this alternative, retaining some or all

gloves as class I devices was rejected.

Alternative 5: Alternative 5 (excluding powdered synthetic gloves

for this rulemaking) was considered in order to reduce cost by as much

as $2.4 million per year. FDA rejected this alternative because it

would not meet the stated objective of the applicable statutes. While

synthetic gloves do not contain NL proteins, FDA is concerned about

foreign body reactions caused by glove powder. These reactions occur

whether the powder is present on a NRL or synthetic glove.

Consequently, FDA is rejecting exempting powdered synthetic gloves from

this regulation.

Alternative 6: FDA considered providing a shorter compliance period

for implementation of the regulation. A compliance period of 90 days or

1 year would significantly increase the expected benefits of the rule

by decreasing the number of annual allergic reactions. FDA estimates

that a 90-day or 1-year implementation period would result in between

3,300 and 3,600 fewer annual allergic reactions to NL proteins than the

number expected with the selected 2-year compliance period. However,

FDA is concerned that the lead times necessary to manufacture limited

powder gloves would make compliance difficult. As stated earlier,

manufacturing equipment used to control glove powder levels is

currently backordered as much as 18 months, and short compliance

periods may result in inadequate supplies of medical gloves. Not

including the potential of shortages, FDA has estimated that average

annualized costs of shorter compliance periods could equal $10 million

to $16 million more than the selected alternative. The 2-year

compliance period allows firms to combine recommended changes with any

other market driven changes, and will allow firms to deplete their

supply of existing labels. As set forth above, however, FDA is

soliciting comment on the timeframe for implementation to determine

whether a 2-year compliance period is really needed. FDA also rejected

providing a longer compliance period. FDA has tentatively determined

that the decrease in costs is outweighed by the decrease in benefits if

the compliance period is lengthened to as many as 3 years. While annual

costs would decrease by almost $9 million, allowing such a long

compliance period would result in about 1,800 additional average annual

allergic reactions as compared to the selected alternative and benefits

would be reduced to $32.0 million. Since glove manufacturers would have

ample opportunity to comply within the

[[Page 41740]]

selected 2-year period, FDA does not believe that additional time is

justified.

FDA solicits comments on other alternatives that meet the stated

objectives.

4. Assuring Small Entity Participation in Rulemaking

At this time, FDA does not believe that the proposed regulation

will have a significant economic impact on a substantial number of

small entities. However, the agency recognizes that many facilities

will be affected. The impact may range from increased glove

manufacturing costs due to validation testing and control of mold

powder to increased contract prices of powdered gloves used by health

facilities. FDA solicits comments from affected entities to ensure that

this impact is analyzed.

FDA plans to provide for access to the Federal Register analysis

through FDA's website on the Internet. Notice of the availability of

this proposed rule and request for comment will be communicated to all

glove-related associations and include a request for comments.

FDA is currently preparing an article for publication in latex-

related trade publications that will highlight the proposed

requirements. In addition, notice of the proposed rulemaking and

request for comments will be available in health-related publications

and sent to trade organizations. FDA actively seeks input into this

proposal and requests comments on all aspects of the analysis of

impacts and the regulatory flexibility analysis.

X. Conclusion

FDA has examined the impacts of the proposed regulation of protein

and powder levels of NL gloves. Based on these estimates, the average

annual quantifiable benefits ($46.9 million) exceed the average annual

quantifiable costs ($32.5 million). Given the high level of uncertainty

and the existence of unquantified benefits, FDA solicits comment on

this analysis and all of its assumptions and projections.

XI. Environmental Impact

FDA has determined under 21 CFR 25.30(k) and 25.34(b) that this

action is of the type that does not individually or cumulatively have a

significant effect on the human environment. Therefore, neither an

environmental assessment nor an environmental impact statement is

required.

XII. Paperwork Reduction Act of 1995

This proposed rule contains information collections provisions that

are subject to review by the Office of Management and Budget (OMB)

under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). A

description of these provisions is given below with an estimate of the

annual reporting burden. Included in the estimate is the time for

reviewing instructions, searching existing data sources, gathering and

maintaining the data needed, and completing and reviewing each

collection of information.

FDA invites comments on: (1) Whether the proposed collection of

information is necessary for the proper performance of FDA's functions,

including whether the information will have practical utility; (2) the

accuracy of FDA's estimate of the burden of the proposed collection of

information, including the validity of the methodology and assumptions

used; (3) ways to enhance the quality, utility, and clarity of the

information to be collected; and (4) ways to minimize the burden of the

collection of information on respondents, including through the use of

automated collection techniques, when appropriate, and other forms of

information technology.

Title: Labeling and Written Procedures Requirements for Powdered

and Powder-Free Patient Examination Gloves, and Powdered and Powder-

Free Surgeon's Gloves.

Description: The proposed rule intends to provide users with

material information to safely use patient examination and surgeon's

gloves. The proposed rule expands the labeling for medical gloves to

include: (1) Caution statements including the actual levels of protein

and powder on the gloves, and (2) expiration dating.

The proposed labeling requirements would require manufacturers to

conduct tests to support the protein and glove powder levels and

expiration date. The proposed special control, a guidance document

entitled ``Medical Glove Guidance Manual,'' recommends that protein

levels be measured by the ASTM D 5712 modified Lorry method and that

glove powder levels be measured by the ASTM D6124-97 method. The

labeling requirements also require stability testing to support the

expiration date. The special control recommends that stability testing

include tensile strength, elongation and water leak tests.

The labeling is intended to communicate useful information to users

about FDA's guidance recommending the use of gloves with no more than

1,200 g of protein and 120 mg of glove powder (or 2 mg of

powder, for powder-free gloves) and to ensure that the labeling

contains adequate directions for use. The labeling would require

manufacturers to indicate the actual levels of protein and powder on

the gloves so that the user can ascertain if the gloves meet the

recommended limits on protein and powder, which are intended to reduce

exposure to particulates and airborne allergens. The expiration date

labeling is intended to ensure that medical glove users have

appropriate information regarding shelf life to enable them to use

medical gloves safely by avoiding products that may have degraded.

Description of Respondents: Businesses or other for profit

organizations.

Table 12.--Estimated Annual Reporting Burden1

----------------------------------------------------------------------------------------------------------------

Annual

21 CFR Section No. of Frequency per Total Annual Hours per Total Hours Total Capital

Respondents Response Responses Response Costs

----------------------------------------------------------------------------------------------------------------

801.440(a) 180 1 180 22 3,960 $985,248

801.440(b) 18 1 18 14 252

801.440(c) 178 1 178 16 2,848

801.440(d) 376 42 1,504 72 108,288

Total 115,348 $985,248

----------------------------------------------------------------------------------------------------------------

\1\ There are no operating and maintenance costs associated with this collection of information.

\2\ The annual burden reported here represents the first year in which a manufacturer would have conducted

testing at 0 days, 3 months, 6 months, and 1 year. FDA expects in any succeeding years, testing would only be

done at 6-month intervals.

[[Page 41741]]

For the proposed labeling requirements, the hours per response

included the hours estimated, based upon communications with industry,

to run the tests to support the powder and protein levels and the

expiration date, as well as the hours estimated to change the

respondent's labeling. The total capital costs were derived from

multiplying the total annual responses for protein testing and

multiplying it by the estimated costs of buying a spectrometer and a

plate reader, instruments that are necessary to conduct the protein

testing. That cost was then annualized over a 5-year period.

Based on communication with industry, FDA estimates that a

respondent would take approximately 8 hours to run the protein tests

necessary to obtain a protein level to add to the labeling. FDA bases

its estimate on the ASTM D 6124-97 protein test.

Based on communication with industry, FDA estimates that a

respondent would take approximately 6 hours to run the powder tests

necessary to obtain a powder level to add to the labeling. FDA bases

its estimate on the ASTM D 5712 modified Lowry method powder test.

Based on communication with industry, FDA estimates that a

respondent would take approximately 16 hours to run the elongation,

tensile strength, and waterleak tests recommended to support the

expiration date. In the first year, FDA estimates that the tests would

be run 4 times, at 0 days, 3 months, 6 months, and 1 year (16 X 4 =

64). In the second, or succeeding years, FDA expects the tests to be

run twice a year.

FDA estimates that a respondent would take approximately 8 hours to

change the labeling and approximately 8 hours to change the promotional

materials to include the appropriate caution statement and the

expiration date. This 16 hours is divided between the labeling changes

proposed in Sec. 801.440(a) and (d) resulting in 8 hours being assessed

for the caution statement and 8 hours being assessed for the expiration

date.

FDA estimates the number of burden hours per response for

Sec. 801.440(a) is 22. That burden comes from the sum of the hours for

running the powder and protein tests (8 hours plus 6 hours) and the

hours for changing the labeling (8 hours).

FDA estimates the number of burden hours per response for

Sec. 801.440(b) is 14. That burden comes from the sum of the hours for

running the powder tests (6 hours) and the hours for changing the

labeling (8 hours).

FDA estimates the number of burden hours per response for

Sec. 801.440(c) is 16. That burden comes from the sum of the hours for

running the protein tests (8 hours) and the hours for changing the

labeling (8 hours).

FDA estimates the number of burden hours per response for

Sec. 801.440(d) is 72. That burden comes from the sum of the hours for

running the elongation, tensile strength, and waterleak tests four

times in the first year (64 hours) and the hours for changing the

labeling (8 hours).

FDA believes that manufacturers already have the equipment

necessary to do the tests to support the powder levels and expiration

dating because such equipment is currently being used to test the

gloves. In order to do the protein tests recommended by FDA, FDA

believes a manufacturer would need to obtain a spectrometer and a plate

reader. FDA estimates that buying this equipment would cost

approximately $22,000 (approximately $10,000 for the spectrometer and

$12,000 for the plate reader). In addition, FDA assumed a 7 percent

discount on the price of the equipment and that the equipment would be

annualized over a 5-year period. In order to obtain a per annualized

year estimate, FDA multiplied the cost by the discount ($22,000 x

.244). FDA added the discounted amount ($5,368) to the cost of the

equipment ($22,000) for a total equipment cost of $27,368. That cost

annualized over a 5-year period is $5,473.60. FDA multiplied that cost

by the number of respondents testing for protein levels (180) for a

total capital cost of $985,248.

In compliance with the Paperwork Reduction Act of 1995 (44 U.S.C.

3507(d)), FDA has submitted the information collection provisions of

this proposed rule to OMB for review. Interested persons are requested

to send comments regarding information collection by August 30, 1999,

to the Office of Information and Regulatory Affairs, OMB, New Executive

Office Bldg., 725 17th St. NW., rm. 10235, Washington, DC 20503, Attn.:

Wendy Taylor, Desk Officer for FDA.

XIII. References

The following references have been placed on display in the Dockets

Management Branch (address above) and may be seen by interested persons

between 9 a.m. and 4 p.m., Monday through Friday.

1. Kaczmarek, R., R. Moore, J. McCrohan, et al., ``Glove Use by

Health Care Workers: Results of a Tri-state Investigation,''

American Journal of Infection Control, 19:228-232, 1991.

2. Kibby, T., and M. Akl, ``Prevalence of Latex Sensitization in

a Hospital Employee Population,'' Annals of Allergy, Asthma &

Immunology, 78:41-44, 1997.

3. Kaczmarek, R., B. Silverman, T. Gross, et al., ``Prevalence

of Latex-specific IgE Antibodies in Hospital Personnel,'' Annals of

Allergy, Asthma & Immunology, 76:51-56, 1996.

4. Arellano, R., J. Bradley, and G. Sussman, `` Prevalence of

Latex Sensitization Among Hospital Physicians Occupationally Exposed

to Latex Gloves,'' Anesthesiology, 77:905-908, 1992.

5. Lagier, F., D. Vervioet, I. Lhermet, et al., `` Prevalence of

Latex Allergy in Operating Room Nurses,'' Journal of Allergy and

Clinical Immunology, 90:319-322, 1992.

6. Yassin, M., M. Lierl, T. Fischer, et al., ``Latex Allergy in

Hospital Employees,'' Annals of Allergy, 72:245-249, 1994.

7. Ownby, D., H. Ownby, J. McCullough, and A. Shafer, ``The

Prevalence of Anti-latex IgE Antibodies in 1000 Volunteer Blood

Donors [Abstract],'' Journal of Allergy and Clinical Immunology,

97:1188-1192, 1996.

8. Lebenbom-Mansour, M., J. Oesterle, et al., ``The Incidence of

Latex Sensitivity in Ambulatory Surgical Patients: A Correlation of

Historical Factors with Positive Serum Immunoglobin E Levels,''

Anesthesia and Analgesia, 85:44-49, July 1997.

9. FDA, Medical device reporting databases of adverse event

reports, Rockville, MD, 1996-1997 (World Wide Web access: http://

www.fda.gov/cdrh/mdr.html).

10. Beezhold, D. and W. Beck, ``Surgical Glove Powders Bind

Latex Antigens,'' Archives of Surgery, 127:1354-1357, 1992.

11. Tomazic, V., E. Shampaine, A. Lamanna, T. Withrow, N.

Adkinson, Jr., and R. Hamilton, ``Cornstarch Powder on Latex

Products Is an Allergen Carrier,'' Journal of Allergy and Clinical

Immunology, 93:751-758, 1994.

12. Tarlo, S., G. Sussman, A. Contala, and M. Swanson, ``Control

of Airborne Latex by Use of Powder-free Latex Gloves,'' Journal of

Allergy and Clinical Immunology, 93:985-989, 1994.

13. Swanson, M., M. Bubak, L. Hunt, J. Yunginger, M. Warner, and

C. Reed, ``Clinical Aspects of Allergic Disease: Quantification of

Occupational Latex Aeroallergens in a Medical Center,'' Journal of

Allergy and Clinical Immunology, 94:445-451, 1994.

14. Heilman, D., R. Jones, M. Swanson, and J. Yunginger, ``A

Prospective, Controlled Study Showing that Rubber Gloves Are the

Major Contributor to Latex Aeroallergen Levels in the Operating

Room,'' Journal of Allergy and Clinical Immunology, 98:325-330,

1996.

15. Ellis, H., ``The Hazards of Surgical Glove Dusting

Powders,'' Surgery, Gynecology & Obstetrics, 171: 521-527, 1990.

16. Edlich, R., ``A Plea for Powder-free Surgical Gloves,'' The

Journal of Emergency Medicine, 12:69-71, 1994.

17. Hunt, T., J. Slavin, and W. Goodson, ``Starch Powder

Contamination of Surgical Wounds,'' Archives of Surgery, 129: 825-

828, 1994.

18. Luijendijk, R., D. deLange, C. Wauters, W. Hop, et al.,

``Foreign Material in Postoperative Adhesions,'' Annals of Surgery,

223: 242-248, 1996.

[[Page 41742]]

19. Holmdahl, L., B. Risberg, D. Beck, et al., ``Adhesions:

Pathogenesis and Prevention-Panel Discussion and Summary,'' European

Journal of Surgery, Supplement, 163 (Suppl. 577), 56-62, 1997.

20. Williams, P., and J. Halsey, ``Endotoxin as a Factor in

Adverse Reactions to Latex Gloves,'' Annals of Allergy, Asthma, and

Immunology, 79:303-310, October 1997.

21. Department of Health and Human Services (NIOSH), ``National

Institute of Occupational Safety and Health (NIOSH) Alert:

Preventing Allergic Reactions to Natural Rubber Latex in the

Workplace,'' Publ. No. 97-135, June 1997.

22. Aziz, N., ``Chlorination of Gloves,'' Paper No. 5 of the

Latex Protein Workshop of the International Rubber Technology

Conference, June 1993, Kuala Lumpur, Malaysia.

23. FDA, CDRH, ``Environmental Degradation of Latex Gloves: The

Effects of Elevated Temperature on Tensile Strength,'' Division of

Mechanics and Materials Science Report # 96-05, D. Walsh, D.

Chwirut, R. Kotz, and J. Dawson, Rockville, MD, 1997.

24. McLelland, J., S. Shuster, and J. Matthews, ``Irritants

Increase the Response to an Allergen in Allergic Contact

Dermatitis,'' Archives of Dermatology, 127:1016-1019, 1991.

25. van der Meeren, H., and P. van Erp, ``Life-threatening

Contact Urticaria from Glove Powder,'' Contact Dermatitis, 14:190-

191, 1986.

26. Seggev, J., T. Mawhinney, J. Yunginger, and S. Braun,

``Anaphylaxis Due to Cornstarch Surgical Glove Powder,'' Annals of

Allergy, 65:152-155, 1990.

27. Assalve D., C. Cicioni, P. Pernio, and P. List, ``Contact

Urticaria and Anaphylactoid Reaction from Cornstarch Surgical Glove

Powder,'' Contact Dermatitis, 19:61, 1988.

28. Ruff, F., P. Thomas, and B. Przybilla, ``Natural Rubber

Latex as an Aeroallergen in the General Environment,'' Contact

Dermatitis, 35:46-47, 1996.

29. Vandenplas, O., J. P. Delwiche, and Y. Sibille,

``Occupational Asthma Due to Latex in a Hospital Administrative

Employee,'' Thorax, 51:452-453, 1996.

30. Kujala, V., and K. Reijula, ``Glove-related Rhinopathy Among

Hospital Personnel,'' American Journal of Industrial Medicine,

30:164-170, 1996.

31. Pisati, G., A. Baruffini, Bernabeo, and R. Stanizzi,

``Bronchial Provocation Testing in the Diagnosis of Occupational

Asthma Due to Latex Surgical Gloves,'' European Respiratory Journal,

7:332-336, 1994.

32. Palosuo, T., S. Makinen-Kiljunen, H. Alenius, et al.,

``Measurement of Natural Rubber Latex Allergen Levels in Medical

Gloves by Allergen-specific IgE-ELISA Inhibition, RAST Inhibition

and Skin Prick Test,'' Allergy, 53:59-67, 1998.

33. FDA, CDRH, ``Glove Powder Content on Surgical and

Examination Gloves,'' V. Tomazic, Division of Life Sciences Progress

Report, 1998.

34. Rabussay, D., and D. Korniewicz, ``Improving Glove Barrier

Effectiveness,'' AORN Journal, 66:1043-1063, 1997.

35. Korniewicz, D., ``Barrier Protection of Latex,'' Immunology

and Allergy Clinics of North America, 15/1:123-137, 1995.

36. Korniewicz, D., M. Kirwin, K. Cresci, et al., ``In-use

Comparison of Latex Gloves in Two High-Risk Units: Surgical

Intensive Care and Acquired Immunodeficiency Syndrome,'' Heart &

Lung, 21:81-84, 1992.

37. U. S. General Accounting Office, ``Medical Devices: Early

Warning of Problems Is Hampered by Severe Underreporting,'' GAO/T-

PEMD-87-1, 1987.

38. U.S. FDA, CDRH, ``Medical Glove Powder Report,'' 1997.

39. U.S. National Center for Health Statistics, ``National

Health Interview Survey,'' 1996.

40. U.S. Occupational Health and Safety Administration, final

rule on ``occupational exposure to bloodborne pathogens'' (29 CFR

1910.1030) (56 FR 64004, December 6, 1991).

41. U.S. Department of Commerce, U.S. Industrial Outlook, 1994.

42. Katz, M. L., and H. Rosen, Microeconomics, 2d Ed., Irwin

Press, 1998.

43. U.S. Bureau of Labor Statistics, Census of Service

Industries, 1992.

44. Ramey, V., ``Nonconvex Costs and the Behavior of

Inventories,'' Journal of Political Economy, 99:306-334, April 1991.

45. Hall, G., ``Non-Convex Costs and Capital Utilization: A

Study of Production Scheduling at Automobile Assembly Plants,'' Yale

University Press, November 1997.

45a. Eastern Research Group; ``Preliminary Estimates: Labeling

and Related Testing Costs for Medical Glove Manufacturers,''

Memorandum, January 18, 1999.

46. U.S. Bureau of Labor Statistics, Monthly Labor Review, 120,

No. 11. 1997.

47. French M. T., J. A. Mauskopg, et al., ``Estimating the

Dollar Value of Health Outcomes from Drug Abuse Interventions,''

Medical Care, 34(9):890-910, 1996.

48. Johnson, F., E. Fries, et al., ``Valuing Morbidity: An

Integration of Willingness-to-Pay and Health-status Index

Literatures,'' Journal of Economic Literature, December 1996.

49. Kaplan, R., J. Bush, et al., ``Health status: Types of

Validity and the Index of Well-being,'' Health Services Research,

Winter 1976: 478-507.

50. Kaplan, R., and J. Bush, ``Health-Related Quality of Life

Measurement for Evaluation Research and Policy Analysis,'' Health

Psychology, 1(1):61-80, 1982.

51. Viscusi, K., ``Fatal Tradeoffs: Public and Private

Responsibilities for Risk,'' Oxford University Press, 1992.

52. Fisher, A., L. Chestnut, et al., ``The Value of Reducing

Risks of Death: A Note on New Evidence,'' Journal of Policy,

Analysis and Management, 8(1):88-100. 1989.

53. Mudarri, D., EPA, ``The Costs and Benefits of Smoking

Restrictions: An Assessment of the Smoke-free Environment Act of

1993 (H.R. 3434),'' 1994.

54. Chen, M., and J. Bush, `` Social Indicators for Health

Planning and Policy Analysis,'' Policy Sciences, 6:71-89, 1975.

55. FDA, Office of Planning and Evaluation and Eastern Research

Group, Economic Impact Analysis of Regulations Under the Mammography

Quality Standards Act of 1992, 1997.

55a. U.S. Bureau of Labor Statistics, ``Monthly Labor Review,''

1995.

55b. U.S. Bureau of Labor Statistics, ``Employment and

Earnings,'' September 1998.

56. U.S. Bureau of Census, ``Census of Manufacturers,'' 1992.

57. U.S. Small Business Administration, ``Table of Size

Standards,'' 1996.

List of Subjects

21 CFR Part 801

Labeling, Medical devices, Reporting and recordkeeping

requirements.

21 CFR Parts 878 and 880

Medical devices.

Therefore, under the Federal Food, Drug, and Cosmetic Act and under

authority delegated to the Commissioner of Food and Drugs, it is

proposed that 21 CFR parts 801, 878, and 880 be amended as follows:

PART 801--LABELING

1. The authority citation for 21 CFR part 801 is revised to read as

follows:

Authority: 21 U.S.C. 321, 331, 351, 352, 360c, 360i, 360j, 371,

374.

2. Section 801.437 is amended by revising paragraph (d) to read as

follows:

Sec. 801.437 User labeling for devices that contain natural rubber.

* * * * *

(d)(1) As described in paragraph (b) of this section, devices

containing natural rubber latex that contacts humans, except natural

rubber latex surgeon's and patient examination gloves shall bear the

following statement in bold print on the device labeling:

``Caution: This Product Contains Natural Rubber Latex Which May

Cause Allergic Reactions.''

This statement shall appear on all device labels, and other labeling,

and shall appear on the principal display panel of the device

packaging, the outside package, container, or wrapper, and the

immediate device package, container, or wrapper.

(2) Natural rubber latex surgeon's and patient examination gloves

shall bear the appropriate caution statement delineated in

Sec. 801.440(a) or (c). This statement shall appear on all device

labels, and other labeling, and shall appear on the principal display

panel of the device packaging, the outside package, container, or

wrapper, and the immediate device package, container, or wrapper.

* * * * *

3. Section 801.440 is added to subpart H to read as follows:

[[Page 41743]]

Sec. 801.440 User labeling for powdered and powder-free surgeon's and

patient examination gloves.

The caution statements required in this section shall appear on all

device labels, and other labeling, and shall appear on the principal

display panel of the device packaging, the outside package, container,

or wrapper, and the immediate device package, container, or wrapper.

(a) Natural rubber latex powdered surgeon's gloves and powdered

patient examination gloves shall bear the following statement:

``Caution: This product contains natural rubber latex which may cause

allergic reactions. FDA recommends that this product contain no more

than 120 mg powder and 1,200 g extractable protein per glove.

This product contains no more than [insert level] mg powder and no more

than [insert level] g extractable protein per glove.''

(b) Synthetic material powdered surgeon's or powdered patient

examination gloves shall bear the following statement: ``Caution: Glove

powder is associated with adverse reactions. FDA recommends that this

product contain no more than 120 milligrams powder per glove. This

product contains no more than [insert level] mg powder per glove.''

(c) Natural rubber latex powder-free surgeon's gloves and powder-

free patient examination gloves shall bear the following statement:

``Caution: This product contains natural rubber latex which may cause

allergic reactions. FDA recommends that this product contain no more

than 1,200 g extractable protein per glove. This product

contai