Making Permanent Regulatory Flexibilities Provided During the COVID-19 Public Health Emergency by Exempting Certain Medical Devices From Premarket Notification Requirements; Request for Information, Research, Analysis, and Public Comment on Opportunities for Further Science and Evidence-Based Reform of Section 510(k) Program

Federal RegisterJan 15, 2021

Ask Donna

What actually matters in this document.

Text

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

RIN 0991-ZA52

Making Permanent Regulatory Flexibilities Provided During the COVID-19 Public Health Emergency by Exempting Certain Medical Devices From Premarket Notification Requirements; Request for Information, Research, Analysis, and Public Comment on Opportunities for Further Science and Evidence-Based Reform of Section 510(k) Program

AGENCY:

Food and Drug Administration, Department of Health and Human Services (HHS).

ACTION:

Notice; request for information.

SUMMARY:

To provide Americans with expanded access to certain medical devices to respond to the COVID-19 Public Health Emergency, FDA issued guidance documents providing numerous regulatory flexibilities, including a temporary waiver of premarket notification requirements under section 510(k) of the Food, Drug, and Cosmetic Act. For seven class I devices for which 510(k) premarket review as temporarily waived during the PHE, the Department of Health and Human Services is permanently exempting those seven (7) class I devices from the 510(k) requirement and is also proposing to exempt an additional 83 class II devices and 1 unclassified device class from the 510(k) requirement, for which premarket review had also been waived during the PHE. The Department is soliciting the public's views on whether premarket review should be permanently waived for some or all of these 83 devices and views on ways to improve the 510(k) premarket notification program.

DATES:

Part III.A of this Notice shall be effective immediately on publication in the

Federal Register

. To be considered, responses and comments related to Part III.B of this Notice must be received electronically, within sixty days of publication in the

Federal Register

as provided below. The Department will consider information submitted by the public in response to Part IV of this Notice on a rolling basis, and until further notice.

ADDRESSES:

You may submit comments through the Federal eRulemaking Portal:

http://www.regulations.gov.

Follow the instructions for submitting comments.

Instructions:

All submissions received must include the agency name and docket number or Regulatory Information Number (RIN) for this rulemaking. All comments received will be posted to

http://regulations.gov,

including any personal information provided. For access to the docket to read background documents or comments received, go to

http://www.regulations.gov.

Comments must be identified by 0991-ZA52. Because of staff and resource limitations, all comments must be submitted electronically to

www.regulations.gov.

Follow the “Submit a comment” instructions.

Warning:

Do not include any personally identifiable information (such as name, address, or other contact information) or confidential business information that you do not want publicly disclosed. All comments may be posted on the internet and can be retrieved by most internet search engines. No deletions, modifications, or redactions will be made to comments received.

Inspection of Public Comments:

All comments received before the close of the comment period are available for viewing by the public, including personally identifiable or confidential business information that is included in a comment. You may wish to consider limiting the amount of personal information that you provide in any voluntary public comment submission you make. HHS may withhold information provided in comments from public viewing that it determines may impact the privacy of an individual or is offensive. For additional information, please read the Privacy Act notice that is available via the link in the footer of

http://www.regulations.gov.

Follow the search instructions on that website to view the public comments.

FOR FURTHER INFORMATION CONTACT:

Dan Barry, 200 Independence Ave. SW, Washington, DC 20201; or by email at

daniel.barry@hhs.gov;

or by telephone at 1-877-696-6775.

SUPPLEMENTARY INFORMATION:

The Administration is committed to creating a data-based regulatory process that appropriately balances benefits and costs. Consistent with the President's executive order on COVID-19 regulatory flexibilities, and Congress' direction in the 21st Century Cures Act, the Department is issuing this Notice to permanently exempt or proposing to permanently exempt certain class I and class II medical devices from the premarket notification requirement in section 510(k) of the Food, Drug, and Cosmetic Act, 21 U.S.C. 360(k). Under this notice, the Department is immediately making permanent the exemption of 7 class I device classes from the section 510(k) requirement and proposes to exempt an additional 84 class II and unclassified device classes from the same requirement on a permanent basis. These 91 devices were all subject a 510(k) waiving during the PHE.

I. Background

A. Statutory Framework

Under the Food, Drug, and Cosmetic Act (FD&C Act), medical devices are placed “in three categories based on the risk that they pose to the public.”

1

Class I devices, products “that present no unreasonable risk of illness or injury,”

2

are subject to general controls. FD&C Act 513(a)(1)(A), 21 U.S.C.

360c(a)(1)(A). Class II devices are “potentially more harmful” than class I devices, and “must comply with federal performance regulations known as `special controls.'”

3

Class III devices carry the highest risk, in that they are for “use in supporting or sustaining human life or for a use which is of substantial importance in preventing impairment of human health, or present[ ] a potential unreasonable risk of illness or injury.” FD&C Act 513(a)(1)(C)(ii)(I)-(II), 21 U.S.C. 360c(a)(1)(C)(ii)(I)-(II).

1

Medtronic, Inc.

v.

Lohr,

518 U.S. 470, 476 (1996).

2

Id.

at 476-77.

3

Lohr,

518 U.S. at 477;

see also

FD&C Act 513(a)(1)(B), 21 U.S.C. 360c(a)(1)(B).

Medical devices are generally subject to FDA premarket review in one of two forms. The first is premarket approval (PMA) review under section 515 of the FD&C Act, 21 U.S.C. 360e. This form of “rigorous” review, analogous to FDA review of a New Drug Application for a “new drug,” requires manufacturers to “submit detailed information regarding the safety and efficacy of their devices, which the FDA then reviews.”

4

During the mid-1990s, FDA reported spending “an average of 1,200 hours on each [PMA] submission,”

5

though the time for review has likely increased since

Lohr

was decided.

4

Lohr,

518 U.S. at 477.

5

Id.

The second form of premarket review is the premarket notification process, which is commonly referred to as the 510(k) process after section 510(k) of the FD&C Act, 21 U.S.C. 360(k). Generally, under the 510(k) process, a device that is “substantially equivalent” to another legally marketed predicate device is “cleared” (as opposed to “approved”)

6

by FDA for legal marketing in the United States.

See

FD&C Act 510(k), 513(i), 21 U.S.C. 360(k), 360c(i). FDA regulations specify the required contents of 510(k) notifications, including labeling, intended use, and clinical and performance data requirements. 21 CFR 807.92. FDA previously reported requiring “an average of only 20 hours” to complete a 510(k) review,

7

which would be around 60 times less than the time required for PMA review.

6

See

21 CFR 807.97 (providing that “determination by the Commissioner that the device intended for introduction into commercial distribution is substantially equivalent” to a predicate device “does not in any way denote official approval of the device”).

7

Id.

Obtaining either a PMA approval or a 510(k) clearance to legally market a medical device is expensive and time-consuming. According to a 2010 survey of medical device companies, “the average total cost from concept to approval [of a PMA device] was approximately $94 million, with $75 million spent on stages linked to the FDA.”

8

For PMAs, survey respondents reported “that it actually took them an average of 54 months to work with the FDA from first communication to approval.”

9

8

Josh Makower, Aabed Meer & Lyn Denend,

FDA Impact on U.S. Medical Technology Innovation: A Survey of Over 200 Medical Device Companies,

at 7 (Nov. 2010),

https://www.medtecheurope.org/wp-content/uploads/2015/07/01112010_FDA-impact-on-US-medical-technology-innovation_Backgrounder.pdf.

During a 2011 hearing before a House subcommittee, the Director for the Center for Devices and Radiological Health (CDRH) raised concerns regarding the methodology used in this study.

FDA Medical Device Approval: Is There a Better Way?, Hearing Before the H. Subcomm. on Health Care, District of Columbia, Census and the National Archives,

112th Cong. 29 (2011) (hereinafter the “2011 Hearing”). The CDRH Director's criticisms largely focused on the report's comparison of FDA's regulation of medical devices to the European Union's regulatory system. The CDRH Director otherwise acknowledged that FDA does not “do cost analyses for what the manufacturers are doing” and that the agency “would not know of the total cost to a particular company.”

Id.

at 32. Here, the Department is citing this study for 510(k) cost and time estimates, not for purposes of comparing the U.S. and E.U. medical device regulatory systems.

9

Id.

at 22.

While 510(k) devices trod a swifter, less expensive path to market than PMA devices do, the same survey found that “the average total cost for participants to bring a low-to-moderate-risk 510(k) product from concept to clearance was approximately $31 million, with $24 million spent on FDA dependent and/or related activities.”

10

Respondents also reported “an average of 10 months from first filing to clearance” for a 510(k) device.

11

The survey authors acknowledged that respondents “were most likely those companies working on innovative, new medical technologies that required clinical data to get through the FDA rather those seeking relatively simple extensions to low-risk, ubiquitous product lines already in existence.”

12

Nevertheless, the survey found the average total cost connected to the “Process of Obtaining [a] 510(k) [clearance]” to be more than $4 million per product.

13

Even if these estimates overstate costs by a factor of ten, a firm could still spend $2.4 million “on FDA dependent and/or related activities,” to include an estimated $400,000 on the 510(k) clearance process itself. Similarly, even if the survey respondents overstated delays, and the actual time were much closer to FDA's goal date of 90 days for review, it is undisputed that the 510(k) clearance process delays a device's introduction to the market.

10

Id.

at 7.

11

Id.

at 22.

12

Id.

at 29, fig. 10.

13

Id.

A 2014 report on antibacterial products produced for the Assistant Secretary for Planning and Evaluation (ASPE) contains similar findings. The report noted that, to conduct a pivotal clinical study to support a 510(k) submission for a MRSA point-of-care diagnostic, a manufacturer could expect to spend “from a low of $250,000 to as high as $4.0 million.”

14

The report estimated the cost to prepare and submit a 510(k) application “at $100,000” while acknowledging the amount “could be highly variable depending on device characteristics.”

15

14

Aylin Sertkaya et al.,

Analytical Framework for Examining the Value of Antibacterial Products,

at 5-3 (Apr. 15, 2014).

15

Id.

at 5-4.

These costs are barriers to new market entrants. To the extent imposing the section 510(k) premarket notification on a device does not create corresponding safety and efficacy benefits for Americans, those barriers are unjustified. Such barriers warrant scrutiny, particularly when market incumbents have an interest in retaining them. As FDA acknowledged in a 1975 proposed rule in the analogous context of drug approvals, “the manufacturer who holds the `pioneer' NDA for a drug may well have an economic interest in retaining the new drug status of that drug” because “[a]s long as either a full or an abbreviated NDA is required, entry into the market place, and thus increased competition is impeded.”

16

FDA noted its belief “that it was not the intention of Congress that section 505 of the [FD&C Act] would be used as an economic trade barrier.”

17

16

40 FR 26142, 26148 (June 20, 1975).

17

Id.

Congress has taken action to ensure that the section 510(k) premarket notification process does not create undue economic barriers for new medical devices. Under the FD&C Act, the Secretary is authorized to exempt class I and II medical devices from the 510(k) requirement if the Secretary finds those devices “no longer require[ ] a report under section [510](k) to provide reasonable assurance of safety and effectiveness.” FD&C Act 510(

l

)(2), 510(m)(1)(A)(i), 21 U.S.C. 360(

l

)(2), 360(m)(1)(A)(i). Congress did this in part to “allow the Secretary to expend limited premarket review resources on potentially risky and technologically advanced devices” so that “the public continues to be adequately protected and will still benefit from the earlier availability of new products.”

18

In section 3054 of the 21st Century Cures Act, Public Law 114-255, 130 Stat. 1033, 1126-27 (Dec. 13, 2016), Congress imposed additional requirements on the Secretary to take action to affirmatively review class I and II devices to determine whether they are exempt from the 510(k) requirement. This Notice is responsive to these previous mandates.

18

H.R. Conf. Rep. 105-399, at 96 (1997).

B. Waiver of Premarket Notification Requirement During COVID-19 PHE

Beginning in March 2020, in response to the COVID-19 Public Health Emergency (PHE), FDA issued a series of guidance documents designed to provide the private sector with regulatory flexibility to meet the sudden, increased need for personal protective equipment, disinfectant products, and other devices to combat the pandemic. The table below presents the various guidance documents issued in April 2020 to assist in the response to the PHE.

Table 1—List of FDA Medical Device Enforcement Policies Responsive to PHE

Title of guidance

Date

Enforcement Policy for Clinical Electronic Thermometers During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency

April 2020.

Enforcement Policy for Imaging Systems During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency

April 2020.

Enforcement Policy for Face Masks and Respirators During the Coronavirus Disease (COVID-19) Public Health Emergency (Revised)

May 2020.

Enforcement Policy for Gowns, Other Apparel, and Gloves During the Coronavirus Disease (COVID-19) Public Health Emergency

March 2020.

Enforcement Policy for Sterilizers, Disinfectant Devices, and Air Purifiers During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency

March 2020.

Enforcement Policy for Digital Health Devices For Treating Psychiatric Disorders During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency

April 2020.

Enforcement Policy for Extracorporeal Membrane Oxygenation and Cardiopulmonary Bypass Devices During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency

April 2020.

Enforcement Policy for Infusion Pumps and Accessories During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency

April 2020.

Enforcement Policy for Non-Invasive Fetal and Maternal Monitoring Devices Used to Support Patient Monitoring During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency

April 2020.

Enforcement Policy for Non-Invasive Remote Monitoring Devices Used to Support Patient Monitoring During the Coronavirus Disease-2019 (COVID-19) Public Health Emergency (Revised)

March 2020 (original).

June 2020 (revised).

October 2020 (revised).

Enforcement Policy for Remote Digital Pathology Devices During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency

April 2020.

Enforcement Policy for Remote Ophthalmic Assessment and Monitoring Devices During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency

April 2020.

Enforcement Policy for Ventilators and Accessories and Other Respiratory Devices During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency

March 2020.

As FDA explained in its clinical thermometer guidance, FDA provided these flexibilities to “ensure the availability of equipment that might offer some benefit to health care providers and the general public during the public health emergency.”

19

To that end, among other things, FDA announced that the agency “does not intend to object to the distribution and use of clinical thermometers that are not currently 510(k) cleared.”

20

Some of the flexibilities, such as those extended to remote patient monitoring, have helped facilitate telemedicine during the PHE. FDA extended similar flexibility to additional devices in other guidance documents shown in Table 1.

19

FDA, Enforcement Policy for Clinical Electronic Thermometers During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency, at 3 (Apr. 2020),

https://www.fda.gov/media/136698/download.

20

Id.

II. HHS' Review of 510(k) Premarket Notification Flexibilities

On May 19, 2020, the President issued Executive Order No. 13924, instructing “[t]he heads of all agencies” to “review any regulatory standards that they have temporarily rescinded, suspended, modified, or waived during the public health emergency,” in order to “determine which, if any, would promote economic recovery if made permanent.”

21

Further, Congress already instructed the Secretary to consider whether to exempt class I and II devices from the section 510(k) requirement “at least once every 5 years.” FDCA 510(

l

)(2), 510(m)(1)(A), 21 U.S.C. 360(

l

)(2), 360(m)(1)(A).

21

85 FR 31353, 31356 (May 22, 2020).

Consistent with the President's executive order, and Congress' direction in the 21st Century Cures Act, the Department conducted a data-driven review to determine whether temporary waiver of the section 510(k) premarket notification requirement for some devices during the PHE should be made permanent. The flexibilities given by FDA during the PHE presented the Department with a unique opportunity to analyze the adverse event records of

devices in periods of time with and without the premarket notification requirement. In view of this, the overarching question for HHS was whether premarket notification provided corresponding safety and efficacy benefits. Below the Department describes the methodology for its review and the results of the same.

A. Methodology

HHS first reviewed the thirteen FDA guidance documents listed in Table 1 to determine which device types are subject to those enforcement policies. The Department identified 221 unique device types. HHS analyzed those device types using FDA's Product Code Database

22

to determine how many of those devices require premarket review. Of those 221 device types, the Department determined that 5 require a PMA, 29 are exempt from the 510(k) requirement, 3 are marketed subject to FDA's enforcement discretion, and 184 require 510(k) clearance prior to marketing. Of the 184 devices types that would require 510(k) clearance without the guidance documents list in Table 1, 10 are class I devices, 173 are class II devices, and 1 is unclassified. These 184 devices are referred to collectively in this Notice as the “Review Devices.”

22

FDA,

Product Classification Database, https://www.fda.gov/medical-devices/classify-your-medical-device/product-code-classification-database.

FDA maintains a publicly available adverse event reporting database called the Manufacturer and User Facility Device Experience database or MAUDE. MAUDE “houses [medical device reports or] MDRs submitted to the FDA by mandatory reports (manufacturers, importers and device user facilities) and voluntary reporters such as health care professionals, patients and consumers.”

23

Like any “passive surveillance system,” MAUDE has “limitations, including the potential submission of incomplete, inaccurate, untimely, or biased data,” which means “incidence or prevalence of an event cannot be determined from this reporting system alone due to under-reporting of events, inaccuracies in reports, lack of verification that the device caused the reported event, and lack of information about frequency of device use.”

24

Even with the system's limitations, MAUDE is an important source of data. FDA has previously used data from MAUDE to inform the agency's decision making.

25

Products liability plaintiffs also make use of the database.

26

As the CDRH Director previously explained to Congress, systems like MAUDE set the United States apart from jurisdictions like the European Union that “do not have publicly available centralized data base[s] for that kind of information.”

27

23

FDA,

MAUDE—Manufacturer and User Facility Device Experience, https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/search.cfm.

24

Id.

25

E.g.,

85 FR 70003, 70006 (Dec. 20, 2019).

26

See

Patrick J. McGrather,

The FDA's MAUDE: Useful Insights for Medical Devices

(Oct. 31, 2017),

https://www.americanbar.org/groups/litigation/committees/mass-torts/practice/2017/manufacture-and-user-facility-device-experience/.

27

2011 Hearing at 30.

In this review, the Department performed searches of the Review Devices in FDA's online searchable MAUDE database. Updated on a monthly basis, the online “searchable database data contains the last 10 year's data” of reports “on medical devices which may have malfunctioned or caused a death or serious injury.”

28

HHS entered the three-letter product code for each Review Device into the online MAUDE database. HHS then collected data on the number of reports for each Review Device from November 1, 2010 to November 30, 2020, tabulating the reports from November 1, 2010 to the beginning of the PHE, and for the time period subsequent the beginning of the PHE to November 30, 2020.

28

FDA,

Manufacturer and User Facility Device Experience Database—(MAUDE), https://www.fda.gov/medical-devices/mandatory-reporting-requirements-manufacturers-importers-and-device-user-facilities/manufacturer-and-user-facility-device-experience-database-maude.

B. Results

Of the 184 Review Devices, HHS found there were 120 or more MAUDE reports for 74 devices and less than 100 MAUDE reports for the 110 other devices during the last ten years. This means roughly 60% of the Review Devices have less than 100 MAUDE reports over the last ten years. Of those 110 devices, 35 devices had no MAUDE reports from November 1, 2010 to November 30, 2020. Those 35 devices are shown in Table 2 below.

Table 2—Review Devices With Zero Adverse Event Reports in MAUDE for the Time Period November 1, 2010 to November 30, 2020

Device description

Device class

Product code

Section in

21 CFR

Powder-Free Polychloroprene Patient Examination Glove

I

OPC

880.6250

Ventilator, Continuous, Minimal Ventilatory Support, Home Use

II

NQY

868.5895

Airway Monitoring System

II

OQU

868.5730

Impedance Measuring Device Utilizing Oscillation Techniques

II

PNV

868.1840

Gauge, Pressure, Coronary, Cardiopulmonary Bypass

II

DXS

870.4310

Valve, Pressure Relief, Cardiopulmonary Bypass

II

MNJ

870.4400

Oximeter, Tissue Saturation, Reprocessed

II

NMD

870.2700

Multivariate Vital Signs Index

II

PLB

870.2300

Electrocardiograph Software For Over-The-Counter Use

II

QDA

870.2345

Sterilizer, Dry Heat

II

KMH

880.6870

Check Valve, Retrograde Flow (In-Line)

II

MJF

880.5440

Intravascular Administration Set, Automated Air Removal System

II

OKL

880.5445

Neuraxial Administration Set—Intrathecal Delivery

II

PYR

880.5440

High Level Disinfection Reprocessing Instrument For Ultrasonic Transducers, Liquid

II

PSW

892.1570

Pediatric/Child Facemask

II

OXZ

878.4040

Normalizing Quantitative Electroencephalograph Software

II

OLU

882.1400

Computerized Cognitive Assessment Aid

II

PKQ

882.1470

Physiological Signal Based Seizure Monitoring System

II

POS

882.1580

Computerized Behavioral Therapy Device For Psychiatric Disorders

II

PWE

882.5801

Monitor, Phonocardiographic, Fetal

II

HFP

884.2640

Monitor, Cardiac, Fetal

II

KXN

884.2600

Digital Pathology Display

II

PZZ

864.3700

Digital Pathology Image Viewing And Management Software

II

QKQ

864.3700

System, Imaging, Holography, Acoustic

II

NCS

892.1550

Lung Computed Tomography System, Computer-Aided Detection

II

OEB

892.2050

Chest X-Ray Computer Aided Detection

II

OMJ

892.2050

Computer-Assisted Diagnostic Software For Lesions Suspicious For Cancer

II

POK

892.2060

Radiological Computer-Assisted Triage And Notification Software

II

QAS

892.2080

Radiological Computer Assisted Detection/Diagnosis Software For Fracture

II

QBS

892.2090

Radiological Computer Assisted Detection/Diagnosis Software For Lesions Suspicious For Cancer

II

QDQ

892.2090

Radiological Computer-Assisted Prioritization Software For Lesions

II

QFM

892.2080

X-Ray Angiographic Imaging Based Coronary Vascular Simulation Software Device

II

QHA

892.1600

Automated Radiological Image Processing Software

II

QIH

892.2050

Image Acquisition And/Or Optimization Guided By Artificial Intelligence

II

QJU

892.2100

Apparatus, Vestibular Analysis

Unclassified

LXV

N/A

Another 43 devices had no reports in MAUDE following declaration of the PHE, and the waiver of the 510(k) premarket notification requirement, with anywhere from 1 to 86 reports in MAUDE prior to the PHE for those same devices. For the ten-year period spanning November 1, 2010 to November 30, 2020, there were a total of 637 reports in MAUDE associated with theses 43 devices listed in Table 3.1. This equates to about 1.5 MAUDE reports per year per device. Table 3.1 below shows each device with the corresponding number of adverse events before and after the PHE.

Table 3.1—Review Devices with Zero Adverse Events Post-PHE and 86 or Fewer Adverse Events Pre-PHE in MAUDE

Device description

Device class

Product code

Section in

21 CFR

MAUDE

events

November 1,

2010 to PHE

MAUDE

events

Post-PHE

to November 30, 2020

Patient Examination Glove, Specialty

I

LZC

880.6250

46

0

Radiation Attenuating Medical Glove

I

OPH

880.6250

1

0

Powder-Free Non-Natural Rubber Latex Surgeon's Gloves

I

OPA

878.4460

1

0

Powder-Free Guayle Rubber Examination Glove

I

OIG

880.6250

2

0

Latex Patient Examination Glove

I

LYY

880.6250

48

0

Meter, Peak Flow, Spirometry

II

BZH

868.1860

27

0

Monitor, Apnea, Facility Use

II

FLS

868.2377

86

0

Monitor, Apnea, Home Use

II

NPF

868.2377

41

0

Oximeter, Reprocessed

II

NLF

870.2700

65

0

Stethoscope, Electronic

II

DQD

870.1875

2

0

Defoamer, Cardiopulmonary Bypass

II

DTP

870.4230

4

0

Filter, Blood, Cardiotomy Suction Line, Cardiopulmonary Bypass

II

JOD

870.4270

2

0

Detector, Bubble, Cardiopulmonary Bypass

II

KRL

870.4205

44

0

Cpb Check Valve, Retrograde Flow, In-Line

II

MJJ

870.4400

12

0

Sterilizer, Ethylene-Oxide Gas

II

FLF

880.6860

29

0

Cabinet, Ethylene-Oxide Gas Aerator

II

FLI

880.6100

2

0

Purifier, Air, Ultraviolet, Medical

II

FRA

880.6500

1

0

Cleaner, Air, Medical Recirculating

II

FRF

880.5045

7

0

Controller, Infusion, Intravascular, Electronic

II

LDR

880.5725

27

0

Cleaners, Medical Devices

II

MDZ

880.6992

5

0

Percutaneous, Implanted, Long-Term Intravascular Catheter Accessory For Catheter Position

II

OMF

880.5970

9

0

N95 Respirator With Antimicrobial/Antiviral Agent For Use By The General Public In Public Health Medical Emergencies

II

ORW

880.6260

1

0

Two Or More Sterilant Sterilizer

II

PJJ

880.6860

6

0

High Level Disinfection Reprocessing Instrument For Ultrasonic Transducers, Mist

II

OUJ

892.1570

3

0

Gown, Patient

II

FYB

878.4040

1

0

Surgical Mask With Antimicrobial/Antiviral Agent

II

OUK

878.4040

1

0

Cerebral Oximeter

II

QEM

870.2700

2

0

Device, Sleep Assessment

II

LEL

882.5050

4

0

Standard Polysomnograph With Electroencephalograph

II

OLV

882.1400

9

0

Source Localization Software For Electroencephalograph Or Magnetoencephalograph

II

OLX

882.1400

2

0

Automatic Event Detection Software For Polysomnograph With Electroencephalograph

II

OLZ

882.1400

1

0

Amplitude-Integrated Electroencephalograph

II

OMA

882.1400

1

0

Automatic Event Detection Software For Full-Montage Electroencephalograph

II

OMB

882.1400

4

0

Burst Suppression Detection Software For Electroencephalograph

II

ORT

882.1400

1

0

Monitor, Heart Rate, Fetal, Ultrasonic

II

HEL

884.2660

12

0

Transducer, Ultrasonic, Obstetric

II

HGL

884.2960

6

0

Uterine Electromyographic Monitor

II

OSP

884.2720

3

0

Tonometer, Ac-Powered

II

HKX

886.1930

1

0

Tonometer, Manual

II

HKY

886.1930

8

0

Automated Digital Image Manual Interpretation Microscope

II

OEO

864.1860

1

0

System, X-Ray, Tomographic

II

IZF

892.1740

35

0

Analyzer, Medical Image

II

MYN

892.2070

1

0

C-Arm Fluoroscopic X-Ray System

II

RCC

892.1650

73

0

The Department further analyzed the details of the MAUDE reports listed in Table 3.1. For the 5 class I glove devices listed, there were 98 reports. As shown in Table 3.2 below, after review of the detailed narratives for those 98 reports, they can be broken down into eight categories.

Table 3.2—MAUDE Report Breakdown for 5 Class I Devices in Table 3.1

Device description (Product Code)

MAUDE report category

Rip/tear/

hole

Discolor/

debris

Allergy/skin

issue

Not device

related

Improper

use

Mislabeled

Odor

Total

Patient Examination Glove, Specialty (LZC)

22

19

4

1

0

0

0

46

Radiation Attenuating Medical Glove (OPH)

0

0

0

0

1

0

0

1

Powder-Free Non-Natural Rubber Latex Surgeon's Gloves (OPA)

0

0

1

0

0

0

0

1

Powder-Free Guayle Rubber Examination Glove (OIG)

0

0

2

0

0

0

0

2

Latex Patient Examination Glove (LYY)

6

7

29

4

0

1

1

48

Total

28

26

36

5

1

1

1

98

More than half of the reports (55%) related to material flaws such as tears, discoloration, or foreign debris in the gloves. For the 36 allergic reaction reports, there was only one report connected with a hospital visit for which the patient was ultimately monitored and discharged. There are 5 MAUDE events from which the report narrative does not provide a basis to infer that the device itself caused the harm.

29

None of the 98 reports involved a death.

29

For example, there was one MAUDE incident where a user reported suffering a third-degree burn after “pouring boiling nitric acid into a beaker and without warning a chemical reaction occurred causing acid to spill onto the wrist.” Nothing in the report indicates the gloves themselves caused the burn or otherwise exacerbated the burn.

The 38 class II devices listed in Table 3.1 were connected to another 539 MAUDE reports. Of those reports, 322 (59.7%) involved device malfunctions, 71 (13.2%) involved injuries, 22 (4.1%) involved deaths, and 124 (23%) have the event type listed as “other” or “NA.” Table 3.3 below provides the breakdown of the 539 MAUDE reports by device type.

Table 3.3—MAUDE Report Breakdown for 38 Class II Devices in Table 3.1

Device description

Product

code

MAUDE report category

Malfunction

Injury

Death

Other

NA

Total

Meter, Peak Flow, Spirometry

BZH

26

1

0

0

0

27

Monitor, Apnea, Facility Use

FLS

52

9

16

9

0

86

Monitor, Apnea, Home Use

NPF

34

5

2

0

0

41

Oximeter, Reprocessed

NLF

49

14

0

0

2

65

Stethoscope, Electronic

DQD

1

0

0

1

0

2

Defoamer, Cardiopulmonary Bypass

DTP

3

1

0

0

0

4

Filter, Blood, Cardiotomy Suction Line, Cardiopulmonary Bypass

JOD

1

0

0

1

0

2

Detector, Bubble, Cardiopulmonary Bypass

KRL

33

0

0

1

10

44

Cpb Check Valve, Retrograde Flow, In-Line

MJJ

11

1

0

0

0

12

Sterilizer, Ethylene-Oxide Gas

FLF

7

9

0

1

12

29

Cabinet, Ethylene-Oxide Gas Aerator

FLI

0

2

0

0

0

2

Purifier, Air, Ultraviolet, Medical

FRA

1

0

0

0

0

1

Cleaner, Air, Medical Recirculating

FRF

6

0

0

0

1

7

Controller, Infusion, Intravascular, Electronic

LDR

27

0

0

0

0

27

Cleaners, Medical Devices

MDZ

4

0

0

1

0

5

Percutaneous, Implanted, Long-Term Intravascular Catheter Accessory For Catheter Position

OMF

3

2

0

4

0

9

N95 Respirator With Antimicrobial/Antiviral Agent For Use By The General Public In Public Health Medical Emergencies

ORW

1

0

0

0

0

1

Two Or More Sterilant Sterilizer

PJJ

0

6

0

0

0

6

High Level Disinfection Reprocessing Instrument For Ultrasonic Transducers, Mist

OUJ

1

1

0

1

0

3

Gown, Patient

FYB

0

0

1

0

0

1

Surgical Mask With Antimicrobial/Antiviral Agent

OUK

0

1

0

0

0

1

Cerebral Oximeter

QEM

1

1

0

0

0

2

Device, Sleep Assessment

LEL

4

0

0

0

0

4

Standard Polysomnograph With Electroencephalograph

OLV

7

2

0

0

0

9

Source Localization Software For Electroencephalograph Or Magnetoencephalograph

OLX

1

1

0

0

0

2

Automatic Event Detection Software For Polysomnograph With Electroencephalograph

OLZ

0

1

0

0

0

1

Amplitude-Integrated Electroencephalograph

OMA

1

0

0

0

0

1

Automatic Event Detection Software For Full-Montage Electroencephalograph

OMB

1

0

0

1

2

4

Burst Suppression Detection Software For Electroencephalograph

ORT

1

0

0

0

0

1

Monitor, Heart Rate, Fetal, Ultrasonic

HEL

9

2

1

0

0

12

Transducer, Ultrasonic, Obstetric

HGL

0

6

0

0

0

6

Uterine Electromyographic Monitor

OSP

1

1

1

0

0

3

Tonometer, Ac-Powered

HKX

0

0

0

0

1

1

Tonometer, Manual

HKY

3

4

0

1

0

8

Automated Digital Image Manual Interpretation Microscope

OEO

1

0

0

0

0

1

System, X-Ray, Tomographic

IZF

31

1

1

1

1

35

Analyzer, Medical Image

MYN

1

0

0

0

0

1

C-Arm Fluoroscopic X-Ray System

RCC

0

0

0

1

72

73

Total

322

71

22

23

101

539

An additional 32 devices had from 1 to 32 reports in MAUDE after the PHE began and anywhere from 1 to 78 reports in MAUDE from November 1, 2010 to the start of the PHE. These devices are shown in Table 4.1 below.

Table 4.1—Review Devices with MAUDE Reports Before and After PHE

Device description

Device

class

Product

code

Section in 21 CFR

MAUDE events

November 1,

2010 to PHE

MAUDE events

Post-PHE to November 30,

2020

Vinyl Patient Examination Glove

I

LYZ

880.6250

40

1

Mechanical Ventilator

II

ONZ

868.5895

2

1

Cannula, Arterial, Cardiopulmonary Bypass (Cpb), Embolism Protection

II

NCP

870.4210

6

1

Dual Lumen Ecmo Cannula

II

PZS

870.4100

2

4

Respirator, N95, For Use By The General Public In Public Health Medical Emergencies

II

NZJ

880.6260

1

1

Sterilizer Automated Loading System

II

PEC

880.6880

8

1

Infusion Safety Management Software

II

PHC

880.5725

6

1

Gown, Isolation, Surgical

II

FYC

878.4040

12

1

Non-Normalizing Quantitative Electroencephalograph Software

II

OLT

882.1400

12

1

Monitor, Ultrasonic, Fetal

II

KNG

884.2660

16

2

Whole Slide Imaging System

II

PSY

864.3700

2

1

Oxygenator, Long Term Support Greater Than 6 Hours

II

BZG

868.1840

10

1

Transmitters And Receivers, Electrocardiograph, Telephone

II

BZQ

868.2375

38

8

Extracorporeal System For Long-Term Respiratory/Cardiopulmonary Failure

II

NFB

868.5905

24

1

Catheter, Percutaneous, Intraspinal, Short Term

II

NHJ

868.5905

18

2

Implanted Subcutaneous Securement Catheter

II

NHK

868.5905

78

1

Subcutaneous Implanted Apheresis Port

II

QAV

868.5454

0

1

Non-Coring (Huber) Needle

II

BYS

870.4100

0

1

Administrations Sets With Neuraxial Connectors

II

DXH

870.2920

18

5

Port & Catheter, Implanted, Subcutaneous, Intraventricular

II

QJZ

870.4100

0

12

Hood, Surgical

II

MAJ

868.5120

17

1

N95 Respirator With Antimicrobial/Antiviral Agent

II

OKC

880.5970

16

1

Reduced- Montage Standard Electroencephalograph

II

PTD

880.5965

40

32

Monitor, Uterine Contraction, External (For Use In Clinic)

II

PTI

880.5570

36

10

Coil, Magnetic Resonance, Specialty

II

PWH

880.5440

0

5

Solid State Fluoroscopic X-Ray Imager

II

LKG

882.5550

20

1

Oxygenator, Long Term Support Greater Than 6 Hours

II

FXY

878.4040

20

1

Transmitters And Receivers, Electrocardiograph, Telephone

II

ONT

878.4040

0

1

Extracorporeal System For Long-Term Respiratory/Cardiopulmonary Failure

II

OMC

882.1400

0

1

Catheter, Percutaneous, Intraspinal, Short Term

II

HFM

884.2720

13

1

Implanted Subcutaneous Securement Catheter

II

MOS

892.1000

72

1

Subcutaneous Implanted Apheresis Port

II

QHY

892.1650

0

1

Table 4.2 presents the devices in Table 4.1 broken down by type of MAUDE. Of the 630 MAUDE reports analyzed, the majority (383 or 60.7%) involved product malfunctions with a limited number connected to death (24 or 3.8%).

Table 4.2—Review Devices in Table 4.1 by MAUDE Report

Device descriptions

Device class

Product code

MAUDE Reports

Malfunction

Death

Injury

Other

NA

Total

Vinyl Patient Examination Glove

I

LYZ

20

0

19

1

1

41

Mechanical Ventilator

II

ONZ

1

1

1

0

0

3

Cannula, Arterial, Cardiopulmonary Bypass (Cpb), Embolism Protection

II

NCP

2

0

5

0

0

7

Dual Lumen Ecmo Cannula

II

PZS

1

1

4

0

0

6

Respirator, N95, For Use By The General Public In Public Health Medical Emergencies

II

NZJ

1

0

1

0

0

2

Sterilizer Automated Loading System

II

PEC

7

0

2

0

0

9

Infusion Safety Management Software

II

PHC

7

0

0

0

0

7

Gown, Isolation, Surgical

II

FYC

12

0

1

0

0

13

Non-Normalizing Quantitative Electroencephalograph Software

II

OLT

11

2

0

0

0

13

Monitor, Ultrasonic, Fetal

II

KNG

2

1

15

0

0

18

Whole Slide Imaging System

II

PSY

2

0

1

0

0

3

Oxygenator, Long Term Support Greater Than 6 Hours

II

BZG

7

0

2

1

1

11

Transmitters And Receivers, Electrocardiograph, Telephone

II

BZQ

38

5

3

0

0

46

Extracorporeal System For Long-Term Respiratory/Cardiopulmonary Failure

II

NFB

11

4

9

1

0

25

Catheter, Percutaneous, Intraspinal, Short Term

II

NHJ

10

1

9

0

0

20

Implanted Subcutaneous Securement Catheter

II

NHK

68

4

6

0

1

79

Subcutaneous Implanted Apheresis Port

II

QAV

1

0

0

0

0

1

Non-Coring (Huber) Needle

II

BYS

1

0

0

0

0

1

Administrations Sets With Neuraxial Connectors

II

DXH

9

1

9

0

4

23

Port & Catheter, Implanted, Subcutaneous, Intraventricular

II

QJZ

9

1

2

0

0

12

Hood, Surgical

II

MAJ

12

0

6

0

0

18

N95 Respirator With Antimicrobial/Antiviral Agent

II

OKC

12

0

3

1

1

17

Reduced- Montage Standard Electroencephalograph

II

PTD

52

1

19

0

0

72

Monitor, Uterine Contraction, External (For Use In Clinic)

II

PTI

22

0

24

0

0

46

Coil, Magnetic Resonance, Specialty

II

PWH

5

0

0

0

0

5

Solid State Fluoroscopic X-Ray Imager

II

LKG

16

1

4

0

0

21

Oxygenator, Long Term Support Greater Than 6 Hours

II

FXY

19

0

2

0

0

21

Transmitters And Receivers, Electrocardiograph, Telephone

II

ONT

0

0

1

0

0

1

Extracorporeal System For Long-Term Respiratory/Cardiopulmonary Failure

II

OMC

1

0

0

0

0

1

Catheter, Percutaneous, Intraspinal, Short Term

II

HFM

10

1

3

0

0

14

Implanted Subcutaneous Securement Catheter

II

MOS

13

0

52

7

1

73

Subcutaneous Implanted Apheresis Port

II

QHY

1

0

0

0

0

1

Total

383

24

203

11

9

630

III. Exemption from 510(k) Premarket Notification Requirement

A. Class I Devices

Section 510(

l

)(2)(A)-(B) of the FD&C Act, 21 U.S.C. 360(

l

)(2)(A)-(B), provides that “the Secretary shall identify through publication in the

Federal Register

, any type of class I device that the Secretary determines no longer requires a report under subsection (k) to provide reasonable assurance of safety and effectiveness” and that “[u]pon such publication—each type of class I device so identified shall be exempt from the requirement for a report under subsection (k); and the classification regulation applicable to each such type of device shall be deemed amended to incorporate such exemption.”

In view of the complete lack of or de minimis number of adverse events in MAUDE following FDA's waiver of the premarket notification requirement for the class I devices listed in Tables 2, 3.1, and 4.1, the Department has concluded that the premarket notification requirement is no longer required to provide reasonable assurance of the safety and efficacy of those devices. As such, as of this Notice, the 7 class I devices listed in Table 5 below shall be exempt from the 510(k) premarket notification requirement.

Table 5—Class I Devices Immediately Exempt from 510

(k)

Notification Requirement

Device

description

Device

class

Product

code

Section in 21 CFR

Powder-Free Polychloroprene Patient Examination Glove

I

OPC

880.6250

Patient Examination Glove, Specialty

I

LZC

880.6250

Radiation Attenuating Medical Glove

I

OPH

880.6250

Powder-Free Non-Natural Rubber Latex Surgeon”s Gloves

I

OPA

878.4460

Powder-Free Guayle Rubber Examination Glove

I

OIG

880.6250

Latex Patient Examination Glove

I

LYY

880.6250

Vinyl Patient Examination Glove

I

LYZ

880.6250

B. Class II Devices

Section 510(m)(2) of the FD&C Act, 21 U.S.C. 360(m)(2), provides that, after a 60-calendar-day-notice comment period, “the Secretary may exempt a class II device from the requirement to submit a report under subsection (k) . . . if the Secretary determines that such report is not necessary to assure the safety and effectiveness of the device.” Within 120 days of publication, “the Secretary shall publish an order in the

Federal Register

that sets forth the final determination of the Secretary regarding the exemption of the device that was the subject of the notice.” Given the lack of any adverse event reports in MAUDE for class II and the unclassified medical devices listed in Table 2, and the lack of non-death-related adverse event reports for class II devices in Tables 3.3 and 4.2, the Department has determined that 510(k) premarket notification for the 84 class II devices and the unclassified device listed in Table 6 below is no longer necessary to assure the safety and effectiveness of those devices.

Table 6—Class II Devices and Unclassified Devices Proposed Exempt from 510

(k)

Requirement

Device

description

Device

class

Product

code

Section in

21 CFR

Ventilator, Continuous, Minimal Ventilatory Support, Home Use

II

NQY

868.5895

Airway Monitoring System

II

OQU

868.5730

Impedance Measuring Device Utilizing Oscillation Techniques

II

PNV

868.1840

Gauge, Pressure, Coronary, Cardiopulmonary Bypass

II

DXS

870.4310

Valve, Pressure Relief, Cardiopulmonary Bypass

II

MNJ

870.4400

Oximeter, Tissue Saturation, Reprocessed

II

NMD

870.2700

Multivariate Vital Signs Index

II

PLB

870.2300

Electrocardiograph Software For Over-The-Counter Use

II

QDA

870.2345

Sterilizer, Dry Heat

II

KMH

880.6870

Check Valve, Retrograde Flow (In-Line)

II

MJF

880.5440

Intravascular Administration Set, Automated Air Removal System

II

OKL

880.5445

Neuraxial Administration Set—Intrathecal Delivery

II

PYR

880.5440

High Level Disinfection Reprocessing Instrument For Ultrasonic Transducers, Liquid

II

PSW

892.1570

Pediatric/Child Facemask

II

OXZ

878.4040

Normalizing Quantitative Electroencephalograph Software

II

OLU

882.1400

Computerized Cognitive Assessment Aid

II

PKQ

882.1470

Physiological Signal Based Seizure Monitoring System

II

POS

882.1580

Computerized Behavioral Therapy Device For Psychiatric Disorders

II

PWE

882.5801

Monitor, Phonocardiographic, Fetal

II

HFP

884.2640

Monitor, Cardiac, Fetal

II

KXN

884.2600

Digital Pathology Display

II

PZZ

864.3700

Digital Pathology Image Viewing And Management Software

II

QKQ

864.3700

System, Imaging, Holography, Acoustic

II

NCS

892.1550

Lung Computed Tomography System, Computer-Aided Detection

II

OEB

892.2050

Chest X-Ray Computer Aided Detection

II

OMJ

892.2050

Computer-Assisted Diagnostic Software For Lesions Suspicious For Cancer

II

POK

892.2060

Radiological Computer-Assisted Triage And Notification Software

II

QAS

892.2080

Radiological Computer Assisted Detection/Diagnosis Software For Fracture

II

QBS

892.2090

Radiological Computer Assisted Detection/Diagnosis Software For Lesions Suspicious For Cancer

II

QDQ

892.2090

Radiological Computer-Assisted Prioritization Software For Lesions

II

QFM

892.2080

X-Ray Angiographic Imaging Based Coronary Vascular Simulation Software Device

II

QHA

892.1600

Automated Radiological Image Processing Software

II

QIH

892.2050

Image Acquisition And/Or Optimization Guided By Artificial Intelligence

II

QJU

892.2100

Apparatus, Vestibular Analysis

Unclassified

LXV

N/A

Meter, Peak Flow, Spirometry

II

BZH

868.1860

Oximeter, Reprocessed

II

NLF

870.2700

Stethoscope, Electronic

II

DQD

870.1875

Defoamer, Cardiopulmonary Bypass

II

DTP

870.4230

Filter, Blood, Cardiotomy Suction Line, Cardiopulmonary Bypass

II

JOD

870.4270

Detector, Bubble, Cardiopulmonary Bypass

II

KRL

870.4205

Cpb Check Valve, Retrograde Flow, In-Line

II

MJJ

870.4400

Sterilizer, Ethylene-Oxide Gas

II

FLF

880.6860

Cabinet, Ethylene-Oxide Gas Aerator

II

FLI

880.6100

Purifier, Air, Ultraviolet, Medical

II

FRA

880.6500

Cleaner, Air, Medical Recirculating

II

FRF

880.5045

Controller, Infusion, Intravascular, Electronic

II

LDR

880.5725

Cleaners, Medical Devices

II

MDZ

880.6992

Percutaneous, Implanted, Long-Term Intravascular Catheter Accessory For Catheter Position

II

OMF

880.5970

N95 Respirator With Antimicrobial/Antiviral Agent For Use By The General Public In Public Health Medical Emergencies

II

ORW

880.6260

Two Or More Sterilant Sterilizer

II

PJJ

880.6860

High Level Disinfection Reprocessing Instrument For Ultrasonic Transducers, Mist

II

OUJ

892.1570

Surgical Mask With Antimicrobial/Antiviral Agent

II

OUK

878.4040

Cerebral Oximeter

II

QEM

870.2700

Device, Sleep Assessment

II

LEL

882.5050

Standard Polysomnograph With Electroencephalograph

II

OLV

882.1400

Source Localization Software For Electroencephalograph Or Magnetoencephalograph

II

OLX

882.1400

Automatic Event Detection Software For Polysomnograph With Electroencephalograph

II

OLZ

882.1400

Amplitude-Integrated Electroencephalograph

II

OMA

882.1400

Automatic Event Detection Software For Full-Montage Electroencephalograph

II

OMB

882.1400

Burst Suppression Detection Software For Electroencephalograph

II

ORT

882.1400

Transducer, Ultrasonic, Obstetric

II

HGL

884.2960

Tonometer, Ac-Powered

II

HKX

886.1930

Tonometer, Manual

II

HKY

886.1930

Automated Digital Image Manual Interpretation Microscope

II

OEO

864.1860

Analyzer, Medical Image

II

MYN

892.2070

C-Arm Fluoroscopic X-Ray System

II

RCC

892.1650

Cannula, Arterial, Cardiopulmonary Bypass (Cpb), Embolism Protection

II

NCP

870.4210

Respirator, N95, For Use By The General Public In Public Health Medical Emergencies

II

NZJ

880.6260

Sterilizer Automated Loading System

II

PEC

880.6880

Infusion Safety Management Software

II

PHC

880.5725

Gown, Isolation, Surgical

II

FYC

878.4040

Whole Slide Imaging System

II

PSY

864.3700

Oxygenator, Long Term Support Greater Than 6 Hours

II

BZG

868.1840

Subcutaneous Implanted Apheresis Port

II

QAV

868.5454

Non-Coring (Huber) Needle

II

BYS

870.4100

Hood, Surgical

II

MAJ

868.5120

N95 Respirator With Antimicrobial/Antiviral Agent

II

OKC

880.5970

Monitor, Uterine Contraction, External (For Use In Clinic)

II

PTI

880.5570

Coil, Magnetic Resonance, Specialty

II

PWH

880.5440

Oxygenator, Long Term Support Greater Than 6 Hours

II

FXY

878.4040

Transmitters And Receivers, Electrocardiograph, Telephone

II

ONT

878.4040

Extracorporeal System For Long-Term Respiratory/Cardiopulmonary Failure

II

OMC

882.1400

Implanted Subcutaneous Securement Catheter

II

MOS

892.1000

Subcutaneous Implanted Apheresis Port

II

QHY

892.1650

C. Impact of Exemptions on Patient Access to Medical Devices

With this Notice, the Department is immediately exempting 7 devices from the premarket notification requirement, and proposes to exempt an additional 84 devices from the requirement after public comment is closed. As noted above in Part I.A, estimates on the cost of preparing a 510(k) submission range from $100,000 to $4 million. The exemptions provided for and proposed under this Notice for these 91 device classes could eliminate anywhere from $9.1 to $364 million in startup costs if there were one new entrant into each device market. Savings could further accrue based on each new market entrant. Instead of being costs passed along to patients and taxpayers, these savings could be invested in other areas such as research and development and manufacturing.

At the same time, should these waivers go into effect as proposed, patients stand to gain more immediate access to new products that would otherwise be required to obtain a 510(k) clearance prior to marketing.

The exemptions provided for in this Notice also conserve FDA's scarce review resources. The COVID-19 PHE stretched FDA's review capacity. Under this Notice, FDA's review resources can be redeployed to review other innovative technology, to include devices designed to mitigate the impact of COVID-19.

IV. Request for Information, Data, and Further Study

HHS' review in this Notice warrants expansion and further study. FDA's medical device Product Code database contains 6,651 unique codes (to include those discussed in this Notice). Of those unique codes, 157 are for class I devices that require 510(k) clearance, and 2,662 are for class II devices that require 510(k) clearance. Applying the $100,000 to $4 million in estimated costs for 510(k) preparation and submission to these 2,819 devices yields approximately $281.9 million to $11.276 billion in startup costs, assuming one new market entrant in each of the 2,819 device classes. Further, again assuming a 90-day review period and one new device entrant in each of the 2,819 device classes that require 510(k) notification, FDA's current approach creates 253,710 review days or 695.1 review years between Americans and new devices. The question of whether the 510(k) notice is justified in view of safety and efficacy concerns merits comprehensive analysis for the benefit of Americans. The Department seeks public comment, research, and analysis on whether other devices should be exempt from the premarket notification requirement.

At a more detailed level, the Department observed internal inconsistencies in FDA's regulation of some device classes that merit discussion. Manual stethoscopes are exempt from the premarket notification requirement. 21 CFR 870.1875(a)(2). Electronic stethoscopes are also exempt, but only if the device “is a lung sound monitor.” 21 CFR 870.1875(b)(2). Similarly, FDA exempts “clinical mercury thermometer . . . device[s] used to measure oral, rectal, or axillary (armpit) body temperature using the thermal expansion of mercury” from the 510(k) premarket notification requirement. 21 CFR 880.2920. By contrast, clinical electronic thermometers which never enter into any body orifice require 510(k) premarket notification. 21 CFR 880.2910. These apparent inconsistencies merit scientific scrutiny. To that end, the Department seeks public comment as to whether other inconsistencies in the medical device regulatory framework exist.

Dated: January 8, 2021.

Alex M. Azar II,

Secretary, Department of Health and Human Services.

[FR Doc. 2021-00787 Filed 1-14-21; 8:45 am]

BILLING CODE 4150-26-P

This is a copy of a public record, reproduced as it was published. It is not legal advice, and it may not be the version a court would rely on. Check the official source before you cite it.

A word about cookies

We need a few to keep you signed in and the library working. The rest help us see which pages people use and where they get stuck. They stay off unless you say yes.

Making Permanent Regulatory Flexibilities Provided During the COVID-19 Public Health Emergency by Exempting Certain Medical Devices From Premarket Notification Requirements; Request for Information, Research, Analysis, and Public Comment on Opportunities for Further Science and Evidence-Based Reform of Section 510(k) Program · 86 FR 4088 | Frix