## Text

DEVICES NOT USED IN THE HEALING ARTS
6 CCR 1007-1 Part 08
[Editor’s Notes follow the text of the rules at the end of this CCR Document.]
_________________________________________________________________________
Adopted by the Board of Health June 16, 2021; effective August 14, 2021
PART 8:
RADIATION SAFETY REQUIREMENTS FOR RADIATION GENERATING DEVICES
NOT USED IN THE HEALING ARTS
8.1
Purpose and Scope.
8.1.1
Authority.
A.
Rules and regulations set forth herein are adopted pursuant to the provisions of sections
25-1-108, 25-1.5-101(1)(l), and 25-11-104, CRS.
8.1.2
Basis and Purpose.
B.
A statement of basis and purpose accompanies this part and changes to this part. A copy
may be obtained from the Department.
8.1.3
Scope.
A.
This part provides special requirements for non-healing-arts radiation generating devices
(RGDs) operating between 5 kiloelectron volts (keV) and 1 million electron volts (MeV).
For machines operating at energies greater than 1 MeV, see Part 9.
B.
The following machines and equipment are exempt from these regulations:
1.
Domestic television receivers.
2.
Cold-cathode gas discharge tubes, providing the exposure rates shall not exceed
10 mrem (0.1 mSv) per hour at a distance of thirty (30) centimeters from any
point on the external surface of the tube.
3.
Other electrical equipment that produces radiation incidental to its operation for
other purposes, providing the dose rate to the whole body at the point of nearest
approach to such equipment when any external shielding not integral to the
equipment is removed, does not exceed 25 mrem (0.25 mSv) per year. The
production testing or factory servicing for such equipment shall not be exempt.
4.
Equipment described in 8.1.3.B shall not be exempt if it is used or handled in
such a manner that any individual might receive a dose of radiation in excess of
the limits specified in Part 4 of these regulations.
C.
In addition to the requirements of this Part, all registrants are subject to the requirements
of Parts 1, 2, 4, 10 and 13 of these regulations. This Part does not pertain to radiation
safety requirements for x-ray equipment that is explicitly covered in other sections of
these regulations, such as Part 5 (Radiation safety requirements for industrial
radiographic operations), Part 6 (X-ray imaging in the healing arts), and Part 9 (Radiation
safety requirements for particle accelerators not used in the healing arts).
D.
Radiography that meets the definition of “cabinet radiography” as defined in 8.2 shall be
regulated under this Part. This includes certified x-ray systems.
E.
Radiography (excluding industrial radiography) that occurs in a shielded room as defined
in 8.2 shall be regulated under this Part.
F.
Industrial radiography that is open-beam, and not in a shielded room and not otherwise
listed in this section, shall be regulated under Part 5 of these regulations.
8.1.4
Applicability.
A.
The requirements and provisions of these regulations apply to applicants, and registrants
within the scope of by Part 8 unless specifically exempted by Part 8.
B.
The applicable special requirements of Part 5 also apply if an image receptor is used to
transform incident x-ray photons either into a visible image or into another form that can
be made into a visible image by further transformation.
C.
The requirements of Part 8 are in addition to, and not in substitution for, applicable
requirements in other parts of these regulations.
8.1.5
Intent.
RGDs are a broad class of equipment that generate x-rays or particle radiation having energies
between 5 keV and 1 MeV, and not intended for medical use on humans. If applicable, all RGDs
shall comply with FDA performance standards as defined in Title 21 Code of Federal Regulations,
parts 1010 thru 1050.
Examples of RGDs include, but are not limited to:
A.
Open and closed analytical x-ray equipment (table top and hand-held);
B.
X-ray gauges;
C.
Cabinet x-ray radiography;
D.
Security screening units;
E.
Quality or process control devices;
F.
Ion implantation devices;
G.
Electron beam welders; and
H.
Non-human use x-ray fluoroscopy.
8.1.6
Published Material Incorporated by Reference.
A.
Throughout this Part 8, federal regulations, state regulations, and standards or guidelines
of outside organizations have been adopted and incorporated by reference. Unless a
prior version of the incorporated material is otherwise specifically indicated, the materials
incorporated by reference cited herein include only those versions that were in effect as
of the most recent effective date of this Part 8 (August 2021), and not later amendments
or editions of the incorporated material.
B.
Materials incorporated by reference are available for public inspection, and copies
(including certified copies) can be obtained at reasonable cost, during normal business
hours from the Colorado Department of Public Health and Environment, Hazardous
Materials and Waste Management Division, 4300 Cherry Creek Drive South, Denver,
Colorado 80246. Additionally, https://www.colorado.gov/cdphe/radregs identifies where
the incorporated materials are available to the public on the internet at no cost. Due to
copyright restrictions certain materials incorporated in this Part are available for public
inspection at the state publications depository and distribution center.
C.
Availability from Source Agencies or Organizations.
1.
All federal agency regulations incorporated by reference herein are available at
no cost in the online edition of the Code of Federal Regulations (CFR) hosted by
the U.S. Government Printing Office, online at www.govinfo.gov.
2.
All state regulations incorporated by reference herein are available at no cost in
the online edition of the Code of Colorado Regulations (CCR) hosted by the
Colorado Secretary of State’s Office, online at
https://www.sos.state.co.us/CCR/RegisterHome.do.
8.2
Definitions.
8.2.1
Definitions of general applicability to these regulations are in Part 1, Section 1.2.
8.2.2
As used in Part 8, each term below has the definition set forth.
“Accessible surface” means the external or outside surface of the enclosure or housing provided
by the manufacturer. This includes high-voltage generator, doors, access panels, latches, control
knobs, and other permanently mounted hardware and including the plane across the exterior
edge of any opening.
“Analytical x-ray equipment” means equipment that generates (by electronic means) and uses
ionizing radiation for the purpose of examining the microstructure of materials, i.e., diffraction and
spectroscopy (including fluorescence).
“Baggage unit”. See “Security Screening Unit”.
“Beam-port” means an opening on the x-ray apparatus designed to emit a primary beam. This
does not include openings on baggage units.
“Cabinet radiography” means industrial radiography using radiation machines not subject to FDA
performance standard for cabinet x-ray systems, in an enclosed, interlocked cabinet in which the
portion of a material being irradiated is contained, and in which:
A.
The radiation machine will not operate unless all openings are closed with
interlocks activated;
B.
The cabinet is shielded such that every location on the exterior meets the
conditions for an unrestricted area as defined in Part 4 of these regulations; and
C.
The cabinet is constructed or arranged as to exclude the entrance of any part of
the body of an individual during irradiation.
“Cabinet x-ray system” means an x-ray system with the x-ray tube installed in an enclosure
which, independently of existing architectural structures except the floor on which it may be
placed, is intended to contain at least that portion of a material being irradiated, provide radiation
attenuation, and exclude personnel from its interior during generation of x radiation. Included are
all x-ray systems designed primarily for the inspection of bags, packages, and personal items at
airline, railroad, and bus terminals, and at other facilities for similar purposes. An x-ray tube used
within a shielded part of a building, or x-ray equipment which may temporarily or occasionally
incorporate portable shielding is not a cabinet x-ray system.
“Cathode ray tube” means any device used to accelerate electrons for demonstration or research
purposes, except where such cathode ray tube is incorporated into a television or display monitor
that is subject to, and has met applicable federal radiation safety performance standards in 21
CFR 1010 and 1020.10.
“Certified cabinet x-ray system” means a RGD certified by the manufacturer in accordance with
21 CFR 1010.2 as being manufactured and assembled pursuant to the provisions of applicable
federal radiation safety performance standards 21 CFR 1010 and 1020.40.
“Certifiable cabinet x-ray system” means an existing uncertified RGD that has been modified to
meet the certification requirements specified in 21 CFR 1020.40.
“Closed-beam x-ray equipment” means a system in which the beam path cannot be entered by
any part of the body during normal operation.
“Cold-cathode gas discharge tube” means an electronic device in which electron flow is produced
and sustained by ionization of contained gas atoms and ion bombardment of the cathode.
“Collimator” means a device for restricting the useful radiation in one or more directions.
“Control panel” means a device containing means for regulation and activation of a RGD or for
the preselection and indications of operating factors.
“Emergency procedure” means the written pre-planned steps to be taken in the event of actual or
suspected exposure of an individual in excess of administrative or regulatory limits. This
procedure shall include the names and telephone numbers of individuals to be contacted as well
as directives for processing the film badge or other personnel monitoring devices.
“Fail-safe design” means a design in which all realistically anticipated failures of indicators or
safety components result in a condition in which individuals are safe from exposure to radiation.
For example, if a light indicating “X-RAY ON” fails, the production of x-rays shall be prevented, or
if a shutter status indicator fails, the shutter shall close.
“General-use system” means a human body screening system that delivers an effective dose
equal to or less than 25 μrem (0.25 μSv) per screening. Given proper justification and certain
restrictions, general-use systems may be operated without specific controls that would limit the
number of individuals scanned or the number of scans per individual in a year.
“Hand-held x-ray system” means a portable instrument that is designed to operate when held in
the hand, such as a hand-held x-ray fluorescence (XRF) analytical device.
“Human body security screening system” means any x-ray equipment used on humans for
security evaluation purposes.
“Industrial radiography” means an examination of the structure of materials by nondestructive
methods utilizing ionizing radiation to make radiographic images for the purpose of detecting
structural flaws in objects. Industrial radiography does not include such imaging for education or
research purposes at a fixed location.
“Interlock” means a device or engineered system that precludes access to an area of radiation
hazard either by preventing entry or by automatically removing the hazard.
“Leakage radiation” means all radiation coming from within the source housing, except the useful
beam.
“Limited-use system” means a human body screening system that is capable of delivering an
effective dose greater than 25 µrem (0.25 µSv) per screening but cannot exceed an effective
dose of 1 mrem (10 µSv) per screening. Limited-use systems require additional controls and
documentation to ensure that annual individual dose limits required by 8.11.5 are not exceeded.
“Local components” means parts of a RGD x-ray system and include areas that are exposed to x-
rays, such as radiation source housings, beam port and shutter assemblies, collimators, sample
holders, cameras, goniometers, detectors, and shielding, but do not include power supplies,
transformers, amplifiers, readout devices, and control panels.
“Mobile equipment”. See “Radiation generating device.”
“Normal operating procedures” means a set of step-by-step instructions necessary to accomplish
the task. These procedures may include sample insertion and manipulation, equipment
alignment, routine maintenance by the registrant, and data recording procedures, which are
related to radiation safety.
“Open-beam x-ray equipment” means an open-beam x-ray system in which the beam path could
be entered by any part of the body at any time.
“Portable equipment”. See “Radiation generating device.”
“Primary beam” means the ionizing radiation coming directly from the radiation source through a
beam port into the volume defined by the collimation system.
“Qualified inspector” means an individual as defined in Part 1 of these regulations.
“Radiation generating device (RGD)” for purposes of Part 8, means any system, device,
subsystem, or component thereof, which may generate x-rays or particle radiation between 5 keV
and 1 MeV, and not intended for healing arts use for humans or animals. A RGD is x-ray
equipment (as defined in Part 1 of the regulations) that may be mobile, portable, or stationary.
“Radiation Safety Officer (RSO)” means an individual as defined in Part 1 of these regulations.
“Radiation source (or x-ray tube) housing” means that portion of an x-ray system which contains
the x-ray tube and/or secondary target. Often the housing contains radiation shielding material or
inherently provides shielding.
“Radiograph” means a permanent film or digital image produced on a sensitive surface by a form
of radiation other than direct visible light.
“Radiography” is the process of creating radiographic images.
“Safety device” means a device, interlock or system that prevents the entry of any portion of an
individual’s body into the primary x-ray beam or that causes the beam to shut off upon entry into
its path.
“Scattered radiation” means radiation that has been deviated in direction and / or energy by
passing through matter.
“Security screening unit” means a non-human use open-beam or cabinet x-ray system with
accessible openings designed for the detection of weapons, bombs, or contraband concealed in
baggage, mail, packages or other commodities or structure.
“Shielded room” means a room housing a RGD where, with the RGD at maximum exposure
setting, the exterior room environs meets the unrestricted area limits of 2 mrem (0.02 mSv) in any
one hour and 100 mrem (1 mSv) in a year at 30 centimeters from the barrier. A shielded room
does not include a RGD which meets the definition of cabinet x-ray systems.
“Shutter” means a moveable device used to block the useful (or primary) beam emitted from an x-
ray tube assembly.
“Source” means the point of origin of the radiation, for example, the focal spot of an x-ray tube.
“Stationary equipment”. See “Radiation generating device.”
“Stray radiation” means the sum of leakage and scatter radiation.
“Warning device” means a visible or audible signal that warns individuals of a potential radiation
hazard.
“X-ray generator” means that portion of an x-ray system which provides the accelerating high
voltage and current for the x-ray tube.
“X-ray gauge” means an x-ray producing device designed and manufactured for the purpose of
detecting, measuring, gauging, or controlling thickness, density, level, or interface location.
8.3
Administrative Requirements.
8.3.1
Each non-healing-arts radiation machine in the State of Colorado shall be registered with the
Department as required by Part 2, Section 2.4.
A.
In accordance with Part 2, such registration shall require:
1.
Designating a RSO;
2.
Maintaining documentation that a written shielding design has been completed,
as applicable.
Certified cabinet RGDs, hand-held XRF RGDs, and other systems as designated
in writing by the Department are not required to have a shielding design; and
3.
Inspection by a Qualified Inspector as required by Part 2, Section 2.5.
8.3.2
The registrant shall be responsible for directing the operation of the RGD(s) under their
administrative control and shall assure that the requirements of Parts 1, 2, 4, and 10 are met in
the operation of the system.
8.3.3
As provided in Part 2, Section 2.6.1.15, for any analytical, industrial or other non-healing-arts
RGD, “adequately trained” shall mean that the individual operator has met the requirements of
Part 2, Appendix 2N and any additional training requirements of Part 8.
8.3.5
Application for Exemptions.
Any RGD registrant that cannot meet the applicable requirements of this Part shall submit to the
Department a request for an exemption to the specific regulation or requirement in question. The
exemption request shall demonstrate to the Department’s satisfaction:
A.
That the use of the RGD will not result in undue hazard to public health and safety or
property;
B.
That compliance would require replacement or substantial modification of the RGD;
C.
That the registrant will achieve, through other means, radiation protection equivalent to
that required by the regulation; and
D.
Why the regulatory standard or requirement could not be met.
8.4
General regulatory requirements.
Unless otherwise provided in this Part, this Section 8.4 applies to all RGDs. Certified and
Certifiable Cabinet X-ray Systems as defined in this Part shall also meet the requirements of 21
CFR 1020.40.
8.4.1
Warning Devices.
A.
Warning devices shall be labeled so that their purpose is easily identified.
B.
An easily visible warning light labeled with the words “X-RAY ON”, or words/displays
having a similar intent, shall be located near any switch or control system that energizes
an x-ray tube and shall be illuminated only when the tube is energized. This warning
light/display shall be of a fail-safe design.
8.4.2
Labeling.
A.
All RGD equipment shall be labeled with a readily visible and discernible sign or signs
bearing the radiation symbol (in accordance with Part 4, Section 4.27) and the words:
“CAUTION - RADIATION - THIS EQUIPMENT PRODUCES RADIATION WHEN
ENERGIZED”, or words having a similar intent, near any switch that energizes an x-ray
tube.
B.
For RGDs with designed openings, for object entries (such as baggage units), the
following shall be posted at or near each opening: “CAUTION – X-RAY HAZARD: DO
NOT INSERT ANY PART OF THE BODY WHEN SYSTEM IS ENERGIZED”, or words
having similar intent.
8.4.3
Radiation Source Housing.
A.
Each radiation source housing shall be subject to the following requirements:
1.
Interlock.
When the x-ray tube housing is the primary shielding for the x-ray tube, and is
intended to be opened for normal use or maintenance, the housing shall be
equipped with an interlock that shuts off the high voltage to the x-ray tube if the
housing is opened; and
2.
Radiation emission limit.
Each x-ray tube housing shall be so constructed that, with all shutters closed, the
radiation measured at a distance of 5 centimeters from the x-ray tube housing
surface does not exceed 2.5 mrem (0.025 mSv) per hour This limit shall be met
at the maximum tube rating.
a.
For closed-beam RGDs, this requirement can be met by complying with
8.6.4, radiation emission limit (0.5 mrem or 0.005 mSv in one hour at five
centimeters outside any accessible surface).
b.
For a RGD in a shielded room, this limit can be met by measuring from
any accessible surface outside the room housing the RGD.
c.
For open-beam RGDs designed to be hand-held during operation, this
requirement can be met by complying with the limits in 8.8.3, radiation
emission limit (2.5 mrem or 0.025 mSv per hour at 5 centimeters.
8.4.4
Generator Cabinet or High Voltage Source Radiation Emission Limits
A.
Each x-ray generator or high-voltage source associated with an RGD shall be supplied
with a protective cabinet which limits leakage radiation to 0.25 mrem (2.5 µSv) per hour
at a distance of 5 centimeters measured at the nearest accessible surface.
1.
For closed beam systems, this requirement can be met by complying with
Section 8.6.4, radiation emission limit (0.5 mrem or 0.005 mSv in one hour at five
centimeters outside any accessible surface).
2.
For a RGD in a shielded room with the high-voltage generator also inside the
shielded room, this limit can be met by measuring from any accessible surface
outside the room housing the RGD. 3. For hand-held, open-beam RGDs, this
requirement can be met by complying with the limits in 8.8.3, radiation emission
limit (2.5 mrem or 0.025 mSv per hour at 5 centimeters).
8.4.5
Surveys.
A.
The registrant shall document performance of radiation surveys, as required by Part 4,
Section 4.17 of these regulations and shall be sufficient to show compliance with
radiation emission requirements of this Part, and as required by Part 4, Section 4.6 and
Part 4, Section 4.14 of these regulations. The radiation surveys shall be sufficient to
evaluate the magnitude and extent of radiation emissions and the potential radiological
hazards that could be present.
At a minimum, surveys shall be performed:
1.
Upon installation of the equipment, and at least once during each routine
certification evaluation thereafter;
2.
Following any change in the initial arrangement, number, or type of local
components in the system;
3.
Following any maintenance requiring the disassembly, removal, or repair of a
local component in the system;
4.
During the performance of maintenance, calibration and other procedures if the
procedures require the activation of a primary x-ray beam while any local
component in the system is disassembled or removed;
5.
Following the temporary bypass and restoration of a safety device or interlock
required by 8.4.7.B.
6.
Any time a visual inspection of the local components in the system reveals an
abnormal condition; and
7.
Whenever a personnel monitoring device shows a significant increase over the
previous monitoring period, or the readings are approaching the limits specified
in 4.6 of these regulations.
B.
The registrant shall have access to sufficiently calibrated, appropriate and operable
radiation survey instruments to make physical radiation surveys as required by this Part.
The instruments shall be capable of detecting and measuring the types and levels of
radiation involved (including primary, scattered, and leakage radiation).
C.
The registrant shall assure the maintenance and calibration of all monitoring and survey
instruments per Part 4, Section 4.17 of the regulations.
D.
Radiation survey measurements shall not be required if a registrant can otherwise
demonstrate compliance with the requirements of this Part to the satisfaction of the
Department.
8.4.6
Posting.
A.
Each area or room containing an RGD where an individual may receive 2 mrem (0.02
mSv) in any one hour or 100 mrem (1 mSv) per year shall be conspicuously posted with
a sign or signs bearing the radiation symbol (which meets the requirements of Part 4,
Section 4.27) and the words “CAUTION – X-RAY EQUIPMENT,” “CAUTION –
RADIATION GENERATING DEVICE” or words having a similar intent.
B.
Unless used in a dedicated location, hand-held RGDs are exempt from the requirements
of 8.4.6.A.
8.4.7
Security.
A.
When not in operation, RGDs shall be secured in such a way as to prevent access or
operation by unauthorized personnel.
8.4.8
Operating Requirements.
A.
Procedures.
1.
Normal operating procedures shall be written and available to all RGD operators.
2.
No individual shall be permitted to operate a RGD in any manner other than that
specified in the procedures unless such individual has obtained written approval
of the RSO and Department.
3.
Except as authorized under Part 6 for authorized medical purposes, or Section
8.11 for security screening purposes, intentional exposure of a living human to a
RGD for any purpose is strictly prohibited.
B.
Bypassing safety equipment.
1.
No individual shall bypass a safety device or interlock, unless such individual has
obtained the written approval of the RSO and Department. Such approval shall
be for a specified period of time.
2.
When a safety device or interlock has been bypassed, a readily discernible sign
bearing the words “SAFETY DEVICE NOT WORKING”, or words having a similar
intent, shall be placed on the radiation source housing and at the control switch.
3.
A record of any bypass of a safety device or interlock shall be maintained; the
record shall contain such information as the date the alteration was made, type of
alteration, length of time the unit remained in the altered condition, post bypass
survey and signed by the RSO, the individual who made the alteration, and the
individual who restored the unit to original condition.
4.
Control panel.
a.
The RGD can only be activated from a control panel.
b.
All indicators and controls that control the primary beam shall be
identifiable and discernible through the use of labels, symbols, software
displays or the equivalent.
C.
Interlocks.
1.
An interlock shall not be used to de-activate the x-ray tube or RGD, except in an
emergency or during testing of the interlock system.
2.
After triggering any interlock, it shall be possible to reset the RGD to full
operation only from a control panel.
3.
All interlocks shall be of a fail-safe design.
4.
When the system is in normal use, interlocks, including emergency shut off
controls, shall not be defeated, or otherwise made inoperable or inaccessible
such that the operability or primary purpose is impacted.
D.
Multiple sources.
If more than one x-ray tube assembly(s) or focal spot can be operated sequentially or
simultaneously from a control panel, visual indicators shall identify which tube
assembly(s) or focal spot has been selected. The selectors shall be identified as to their
function. If a letter or number is used, a reference card or table explaining the code shall
be affixed to the control panel.
8.4.9.
Repair or Modification of X-ray Tube or RGD Systems.
A.
Only registered service companies shall be permitted to install, repair, calibrate, or make
modifications to an RGD and only Qualified Inspectors shall perform machine
inspections.
B.
No operation involving removal of covers, shielding materials or tube housings or
modifications to shutters, collimators, or beam stops shall be performed without
ascertaining that the tube is off and will remain off until safe conditions have been
restored.
C.
The main power switch with a lock-out / tag-out, rather than interlocks, shall be used for
routine shutdown in preparation for repairs.
8.4.10. Testing of Safety Devices.
A.
Tests of all safety devices, such as interlocks, shutters, warning lights, and required
emergency shut-off switches shall be conducted at intervals not to exceed 6 months on
all operable RGDs. Cabinet integrity, where applicable, shall be evaluated at the same
time and frequency as safety device tests.
B.
If any safety device fails during testing, the RGD shall be removed from service until the
safety device failure is corrected or proper temporary administrative controls established
and approved in writing by the RSO.
C.
Records of safety device tests, check dates, findings and corrective actions shall be
available for inspection and maintained for 3 years.
D.
Records shall include the date of the test, a list of the safety devices tested, survey
instrument information, calibration date, the results of the test, the name of the person
performing the tests and corrective actions taken for safety devices that fail the required
test.
E.
Testing of safety devices may be deferred if the unit and/or installation is clearly marked
and kept out of service; units and/or installations brought back into service after
exceeding the 6 month interval shall be tested prior to use.
F.
If testing of a safety device cannot be performed due to manufacturer design, the
registrant shall document that the safety device will not be tested and specifically why the
safety device cannot be tested.
8.4.11 Instruction and Training.
The registrant shall document the scope of training required for the RGD they possess in
accordance with this section. No individual shall be permitted to operate or maintain an RGD, or
enter a shielded room without appropriate instruction and training. Records of all required training
and instruction shall be maintained onsite for a minimum of 3 years after the individual terminates
employment. Records shall be made available for review by the Department.
Each individual shall receive instruction in and demonstrated competence as to:
A.
Types of radiation and identification of radiation hazards associated with the use of the
RGD and associated equipment and precautions or measures to take to minimize
radiation exposure;
B.
Significance of the various radiation warning, safety devices, and interlocks incorporated
into the equipment, or the reasons they have not been installed on certain pieces of
equipment and the extra precautions required in such cases;
C.
Commensurate with potential hazards of use, biological effects of radiation, radiation
risks, and recognition of symptoms of an acute localized exposure;
D.
Normal operating procedures for each type of RGD and associated equipment, including
having received hands-on training, and procedures to prevent unauthorized use;
E.
Procedures for reporting an actual or suspected accidental exposure or other radiation
safety concerns, such as any unusual occurrence or malfunction that may involve
exposure to radiation; and
F.
Performing surveys where applicable.
8.4.12. Radiation Protection Responsibility.
A.
The registrant's senior management shall make the ultimate decision to use any RGD
and ultimately be responsible for radiation safety.
B.
The registrant's senior management shall designate an individual responsible for
radiation safety, or a RSO. This individual shall have direct access to senior management
for radiation safety issues. This individual shall have training and experience
commensurate with the scope of the radiation safety program to carry out the
responsibilities as indicated below.
1.
Ensuring that all RGDs are operated within the limitations of the established
radiation safety program and operating procedures.
2.
Instructing personnel with regard to safe working practices and ensuring all
personnel are trained in radiation safety commensurate with the hazards of the
job.
3.
Investigating any incident of abnormal operation or exposure or suspected
overexposure of personnel to determine the cause, take remedial action, and
report the incident to the proper authority.
4.
Ensuring that safety devices, interlocks, warning signals, labels, postings, and
signs are functioning and located where required.
5.
Maintain all radiation safety records.
8.5
Radiation levels and dosimetry monitoring.
8.5.1
Radiation Levels.
A.
X-ray equipment shall be located and arranged with sufficient shielding and area access
control to ensure compliance with Part 4.
B.
The registrant shall assure that no radiation levels exist in any area surrounding the local
component group which could result in a dose to an individual present therein in excess
of the dose limits in Part 4, Sections 4.14 or 4.15.
8.5.2
Personnel Monitoring Requirements.
A.
Each individual who is associated with the operation of a non-healing-arts radiation
generating device shall meet the requirements of Part 4, Sections 4.6, 4.10, 4.12, 4.13,
4.14, and 4.18.
1.
When personnel dosimetric monitoring devices are required, they shall be worn
in accordance with Part 4, Section 4.6.3.
2.
Each operator of portable hand-held x-ray equipment shall wear separate whole
body and extremity personnel dosimetric monitoring devices, unless otherwise
exempted by the Department or by regulation.
3.
Deliberate exposure of a personnel dosimetric monitoring device to deceptively
indicate a dose delivered to an individual is strictly prohibited.
B.
In addition to the requirements of Part 4, Section 4.6, extremity dosimetry shall be
provided to and shall be used by:
1.
Personnel working with or routinely working near and having potential for
extremity exposure to, the primary beam of an open-beam RGD; and
2.
Personnel maintaining RGDs if the maintenance procedures require the
activation of a primary x-ray beam when any local component in the x-ray system
is disassembled or removed.
8.6
Additional Requirements for Closed-Beam RGDs.
In addition to the requirements of 8.3, 8.4, and 8.5, the following applies to all closed-beam x-ray
RGDs:
8.6.1
System Enclosure.
The radiation source, sample or object, detector, and analyzing crystal (if used) shall be enclosed
in a chamber or coupled chambers that cannot be entered by any part of the body during normal
operation.
8.6.2
Interlocks.
All doors and panels accessing the RGDs shall be interlocked. The interlocks required by this
section shall be of a fail-safe design.
8.6.3
Interlock Functions.
The system enclosure, sample chamber, etc. closure shall be interlocked with the x–ray tube high
voltage supply and/or a shutter in the primary beam so that no x-ray beam can enter the sample
or object chamber while it is open unless the interlock has been conspicuously and deliberately
defeated. The interlock required by this section shall be of fail–safe design or adequate
administrative controls shall be exercised to ensure operations will not continue without a proper
functioning interlock.
8.6.4
Radiation Emission Limit.
The radiation emission for all closed beam RGDs shall not exceed a dose rate of 0.5 mrem
(0.005 mSv) in one hour at five centimeters outside any accessible surface.
8.6.5
Security Screening Units.
During operation and generation of radiation, security screening units shall require operator
presence at the control panel and shall permit surveillance of openings and doors.
A.
During an exposure or preset succession of exposures of one-half second or longer,
means shall be provided to terminate the exposure or preset succession of exposures at
any time.
B.
During an exposure or preset succession of exposures of less than one-half second,
means shall be provided to permit completion of the exposure in progress, but shall allow
the operator to prevent or terminate additional exposures.
8.7
Additional Requirements for Open Beam RGDs.
In addition to the requirements in 8.3, 8.4, and 8.5, the following requirements apply to all open
beam RGDs not otherwise addressed in this Part.
8.7.1
Safety Devices.
A.
The registrant shall document their justification of the use of open-beam instead of
closed-beam systems.
B.
If the registrant needs to use an open-beam system, the registrant shall require a safety
device which prevents the entry of any portion of the operator’s body into the path of the
primary beam or which causes the primary beam to be shut off upon entry into its path.
C.
If the registrant’s use of the open-beam RGD does not permit the use of a safety device
to prevent direct body exposure, the registrant shall maintain a written record of a
description of the various safety devices that have been evaluated and document why
these devices cannot be used. These records shall be available onsite for inspection.
D.
In lieu of the safety device described in 8.7.1.B, the registrant shall employ alternative
methods (such as policies and procedures) to minimize the possibility of unnecessary
exposure. These alternative methods shall be documented. The documentation shall
include information about the absence of safety devices. This documentation shall be
available for inspection as long as these methods are employed, plus an additional 5
years.
E.
For portable open-beam RGDs that are manufactured to be used hand-held, or
potentially used as a hand-held, without such safety devices, the requirements of 8.7.1
may be met by complying with all the requirements in 8.8.
8.7.2
X-ray On Status.
For open beam equipment, RGDs shall be provided with a readily discernible and active
indication of:
A.
X-ray tube “on-off” status located near the radiation source housing. The warning lights
as required by 8.4.1.B can meet this requirement if the warning lights are readily
discernible and viewable by anyone near the primary beam; or
B.
Shutter “open-closed” status located at the control panel and near each beam port on the
radiation source housing, if the primary beam is controlled with a shutter. The shutter
status device shall be clearly labeled as to the meaning of the status device (i.e., whether
the shutter is open or closed). The status light at the control panel can meet the
requirement for the status light at the beam port if the status light at the control panel is
readily discernible and viewable by anyone near the primary beam; and
C.
The x-ray tube “on-off” status indicator and the shutter “open-closed” status indicators
shall be of a fail-safe design.
8.7.3
Labeling.
Each unit will be labeled at or near the x-ray exit beam port to identify the location of the beam
with the words, “CAUTION - X-RAY BEAM”, “CAUTION - HIGH INTENSITY X-RAY BEAM”, or
words having a similar intent.
8.7.4
Beam Ports.
Unused beam ports on radiation source housings shall be secured in the closed position in a
manner which will prevent unintentional opening.
8.7.5
Shutters.
On open-beam RGD configurations that are designed to accommodate interchangeable
components, each beam port on the radiation source housing shall be equipped with a shutter
that cannot be opened unless a collimator or a component coupling has been connected to the
beam port.
8.7.6
Radiation Emission Limits.
The local components of an open-beam RGD shall be located and arranged and shall include
sufficient shielding or access control such that no radiation emissions exist (exclusive of the
primary beam) in any area surrounding the local component group which could result in a dose to
an individual present therein in excess of the dose limits as outlined in Part 4, Section 4.14 of
these regulations. These emissions shall be met at any specified tube rating.
8.7.7
Primary Beam Attenuation.
In cases where the primary x-ray beam is not intercepted by the detector device under all
conditions of operation, protective measures shall be provided, such as auxiliary shielding or
administrative procedures, to avoid exposure to any individual from the transmitted primary x-ray
beam.
8.7.8
Operator Attendance.
The operator shall be in immediate attendance at all times when the equipment is in operation
except when the area is locked or the equipment is secured to protect against unauthorized or
accidental entry.
8.7.9
Controlling Access.
A.
If the RGD is not in a restricted area as defined in Part 1 of these regulations, the
operator shall be able to control access to the RGD at all times during operation.
B.
If the RGD is not in a restricted area and the RGD is capable of creating a radiation area
or a high radiation area as defined in Part 1 of these regulations, the operator shall be
able to control access to the RGD at all times during operation, and:
1.
Radiation areas shall be conspicuously identified. The radiation source shall be
within a conspicuous perimeter (e.g., rope, tape, or other barrier) that identifies
the area in which the dose equivalent rate exceeds 5 mrem (0.05 mSv) per hour.
The area described by the temporary barricade shall be suitably posted with
“CAUTION - RADIATION AREA” signs. The operator shall ensure that no one is
inside or enters the radiation area during operation of the RGD;
2.
High radiation areas shall be conspicuously identified. The radiation source shall
be within a conspicuous perimeter (e.g., rope, tape, or other barrier) that
identifies the area in which the dose equivalent rate exceeds 1 mSv (100 mrem)
per hour. The area described by the temporary barricade shall be suitably posted
with “CAUTION - HIGH RADIATION AREA” signs. The operator shall ensure that
no one is inside or enters the high radiation area during operation of the RGD;
3.
The operator shall perform a visual check of the controlled area to ensure it is
free of all unauthorized personnel immediately prior to activating or exposing the
radiation source;
4.
Surveillance of the exposure area shall be maintained during operation, either by
visual or by other reliable means to ensure that no person enters the area;
5.
Excluding hand-held x-ray systems, when approaching the radiation source,
following the conclusion of an exposure, the operator shall use a suitable
calibrated and operable radiation detection instrument to verify that the x-ray tube
has been de-energized;
6.
A personal alarming dose rate meter may be worn to approach the work area if
the device is appropriately designed and calibrated for the type of x-ray emitted
(i.e., pulse or continuous), set at an appropriate level to detect the presence of
the source, for example 2 mrem (0.02 mSv) per hour, and has been source-
checked prior to use. The radiation in the work area must be reasonably uniform
so that the device responds to radiation exposure to any part of the body. It may
not be used to measure radiation levels, nor may it be used to indicate the
presence of the source for potential non-uniform exposure, such as may occur
during machine maintenance or work in a RGD target area;
7.
Measurement of radiation levels for a radiation survey shall be performed using
an appropriate calibrated radiation survey meter. A radiation survey meter shall
also be used when there is potential for non-uniform exposure to personnel, such
as may occur during machine maintenance or work in a RGD target area;
8.
During the initial exposure, the radiation levels shall be measured around the
perimeter of the controlled area. The perimeter shall be adjusted accordingly to
meet the access control requirement for radiation areas or high radiation areas;
and;
9.
The survey around the perimeter shall be made for each new operating condition
and the perimeter adjusted accordingly. The area of operation shall be monitored
periodically if radiation levels are variable.
8.7.10 Instruction and Training.
In addition to the requirements in 8.4.11, no individual shall be permitted to operate or maintain
an open-beam RGD unless such individual has received specific and detailed instruction in and
demonstrated competence as to:
A.
Sources and magnitude of common radiation exposure;
B.
Units of radiation measurement;
C.
Radiation protection concepts of time, distance, shielding, and ALARA;
D.
Procedures and rights of a declared pregnancy;
E.
Regulatory requirements and area postings;
F.
Worker, embryo/fetus, and public dose limits;
G.
Proper use of survey instruments and dosimetry; and
H.
The policies and procedures required by 8.7.1.
8.8
Additional Requirements for Open-beam, Hand-held RGDs.
In addition to the applicable requirements in 8.3, 8.4, 8.5 and 8.7, the following requirements in
this Section apply to open-beam, RGDs intended or designed to be hand-held during operation.
8.8.1
Procedures.
All registrants possessing open-beam, hand-held RGDs shall have available for review by the
Department, operating policies and procedures that contain measures to insure that:
A.
Radiation protection is provided equivalent to that afforded in Part 4, Section 4.14 of
these regulations (Dose Limits for Individual Members of the Public);
B.
Radiation protection is provided equivalent to that afforded in 8.7.7 (Primary Beam
Attenuation);
C.
The operator will not hold the sample during operation of the RGD and that the operator’s
hands will not approach the primary beam;
D.
The operator will not aim the primary beam at him/herself or at any individual during
operation of the RGD; and
E.
Operator radiation exposure is as low as reasonably achievable (ALARA), for example,
by use of ancillary equipment that will reduce exposure.
8.8.2
Training.
In addition to the training requirements of 8.4.11 and 8.7.10 above, the registrant shall provide
training for all users and operators on the subjects in section 8.8.1. Records shall be maintained
of all user and operator training.
8.8.3
Radiation Emission Limit.
For hand-held RGDs, the limits of 8.4.3.A and 8.4.4, excluding the primary beam, shall be met if
the radiation emission at any accessible surface of the RGD does not exceed 2.5 mrem (0.025
mSv) per hour at 5 centimeters.
8.8.4
Extremity Monitoring.
For the purposes of the requirements in 8.5.3 (extremity monitoring), operators of hand-held
RGDs shall be considered as working near the primary beam and shall be required to wear
extremity dosimetry, unless explicitly exempted by the Department.
8.8.5
Excluding hand-held XRF units, and other units exempted in writing by the Department, the
operator of a hand-held RGD shall be protected from scatter radiation by material of not less than
0.25 millimeter lead equivalent unless the RSO and Department determine that no added
protection is needed for the device use and model.
8.9
Shielded Room RGDs.
For RGDs that do not meet the limits of Part 4, Section 4.14, the RGD can be maintained inside a
shielded room such that the exterior of the room meets the limits of Part 4, Section 4.14 of these
regulations when the RGD is activated. RGDs in a shielded room shall be required to meet only
the requirements of 8.3, 8.4, and 8.5 and the following:
8.9.1
Posting.
The door to the room containing the RGD shall be posted “CAUTION – RADIATION AREA”, or
“CAUTION – HIGH RADIATION AREA”, or “GRAVE DANGER – VERY HIGH RADIATION
AREA”, as required by Part 4 of these regulations.
8.9.2
Entrance Interlocks.
All entrances into the shielded room shall be provided with interlocks. After an interlock has been
interrupted, broken, or tripped, it shall be possible to cause x-rays to be produced again only from
the control panel. Interlocks shall not be used to shut off the x-ray equipment except in an
emergency or during testing.
8.9.3
Entrance Warning Devices.
All entrances into the shielded room shall be provided with a conspicuously visible warning
device, which need not be flashing or rotating but which operates only when radiation is being
produced. The warning device shall be labeled in accordance with H.6a.
8.9.4
Room Warning Lights.
The interior of the shielded room shall be provided with flashing or rotating warning lights that
operate when, and only when, radiation is being produced. These lights shall be positioned so
that they can be observed from any position or orientation within the room. The lights shall be
posted indicating the meaning of the warning signal and instructions on what to do; the posting
shall be legible, conspicuous, and accessible to view.
8.9.5
Audible Room Warning Device.
An audible warning signal within the room shall be actuated for at least ten (10) seconds
immediately prior to the first initiation of radiation after the closing of any opening that can admit
personnel. The registrant shall post the meaning of the warning signal and instructions on what to
do; the posting shall be legible, conspicuous, and accessible to view.
8.9.6
Emergency Shut-off.
If dose rates exceed the High Radiation Area limits as defined in Part 1 of these regulations,
emergency shut-off switches shall be located within the high radiation areas so as to be
accessible to individuals therein within 10 seconds. These switches and their mode of operation
shall be identified by a conspicuously posted sign adjacent to the switch. The emergency shut-off
switches shall include a manual reset that must be reset at the switch before x-rays can again be
produced from the control panel. After an emergency shut-off switch has been activated, it shall
be possible to produce x-rays again only from the control panel.
8.9.7
Separate Electrical Systems.
The interlock system and the emergency shut-off system shall be separate electrical and/or
mechanical systems.
8.9.8
Exit from Shielded Room.
A person within the room housing a RGD shall be able to exit at all times.
8.9.9
Entry into the Shielded Room.
A.
After each exposure and before entry of any personnel, a survey shall be performed upon
entry to the shielded room to determine that the RGD is no longer producing radiation.
B.
Personnel devices providing an audible signal when activated by radiation will be
acceptable for the survey requirement of 8.9.9.A.
1.
Proper operation of the audible detection device shall be checked daily and a
record maintained of this check.
2.
The audible device shall be designed so as to clearly indicate entry into a 2
mrem (0.02 mSv) per hour or greater radiation field.
3.
All personnel working with the RGD shall be provided with such a device.
C.
Stationary area monitors providing an audible signal when activated by radiation will be
acceptable for the survey requirement of 8.9.9.A.
1.
Proper operation of the stationary detection device shall be checked daily and a
record maintained of this check.
2.
The stationary device shall be designed so as to clearly indicate entry into a 2
mrem (0.02 mSv) per hour or greater radiation field.
3.
Stationary area monitors shall be calibrated annually to determine that the
audible signal operates at a 2 mrem (0.02 mSv) per hour radiation field.
8.9.10 Personnel Monitoring.
All personnel associated with the x-ray equipment shall be provided with personnel monitoring
devices that shall be calibrated for the x-ray energies being utilized. Records of personnel
exposure shall be maintained.
8.9.11 Training.
No registrant shall permit any individual to operate a RGD in a shielded room until such individual
has received a copy of, instruction in, and demonstrated an understanding of, operating and
emergency procedures for the unit and competence in its use. Records shall be maintained of all
operator training.
8.9.12 Control Panel Security.
The equipment control panel shall be provided with a locking device to prevent unauthorized use.
Such locking device shall, when locked, prevent the production of radiation by the equipment.
8.9.13 Malfunctions.
If a safety or warning device malfunctions, the control panel shall be locked in the “off” position.
The control panel shall not be used, except as may be necessary for repair or replacement of the
malfunctioning safety or warning device, until the safety or warning device is functioning properly.
8.10
Reserved.
8.11
RGDs Used in Human Body Security Screening or Vehicle Screening for Public Protection.
The requirements in this section are in addition to the General Requirements in 8.4, 8.5, and 8.6.
A person must request Department approval for a RGD to be used for Human Body Security
Screening or Vehicle Screening involving intended exposure of human occupants to the primary
beam for public protection purposes. Such persons shall submit the appropriate form or submit in
writing the following information to the Department for evaluation and approval, and demonstrate
how the dose limits in 8.11 will be met:
8.11.1 Efficacy Evaluation.
An evaluation of all known alternate methods that could achieve the goals of the security
screening program, and why these methods will not be used in preference to the proposed
approach utilizing ionizing radiation.
8.11.2 Equipment Evaluation.
RGDs used for human body security screening shall be evaluated every 12 months by a Qualified
Inspector for radiation dose and optimization of image quality. Evaluation of optimization of image
quality shall be done in accord with the system manufacturer’s specifications, recommendations,
and quality assurance procedures or, in the absence of manufacturer provided information on
image quality, the recommendations of a nationally recognized organization or Qualified
Inspector.
8.11.3 Dose Limits for General-Use Systems.
For general-use screening systems, where the system is used without regard to the number of
individuals scanned or number of scans per individual in a year, an effective dose for a single
complete screening shall be limited to 25 µrem (0.25 µSv).
8.11.4 Dose Limits for Limited-Use Systems.
For limited-use screening systems, where the system is capable of operation greater than 25
µrem (0.25 µSv) per screening, and is used with discretion, the effective dose per screening shall
be less than or equal to 1 mrem (0.01 mSv).
8.11.5 Dose Limits for Repeat Security Screenings.
Individuals subject to repeat security screening at a single venue shall not receive an effective
dose greater than 25 mrem (0.25 mSv) in any one year at the registrant’s facility.
8.11.6 Requirements for vehicle or container screening systems.
A.
When the procedures for operation of a RGD used for security screening of vehicles
includes knowingly exposing human occupants to the primary beam when screening
vehicles, structures or containers, the system shall be subject to the same requirements
as general-use or limited-use systems as provided in 8.11.1 through 8.11.5.
B.
If the requirements in 8.11.3 through 8.11.5. cannot be met if vehicle occupants are
knowingly exposed to the primary beam of a security screening system, then there shall
be means to assure the occupied portion of the vehicle is outside of the scan area while
the primary beam is emitted or procedures shall be established and implemented to
assure that no occupants are present in the vehicle during screening.
C.
The effective dose to an individual for a single inadvertent exposure to the primary beam
shall not exceed 500 mrem (5 mSv) and should not exceed 100 mrem (1 mSv). The
reliability of the procedure used to assure that there are no occupants of a vehicle to be
scanned shall be commensurate with the potential severity of an inadvertent exposure. If
the 500 mrem (5 mSv) limit cannot be assured, a pre-screening with a mode or system
which can meet the limits in 8.11.3 through 8.11.6 shall be used to verify there are no
occupants in the vehicle being examined.

## Nearby sections

- [6 CCR 1007-1 RADIATION CONTROL – (Rule Versions Prior to 04/01/2007)](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1.md)
- [6 CCR 1007-1 Part 01 RADIATION CONTROL - GENERAL PROVISIONS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_01.md)
- [6 CCR 1007-1 Part 02 RADIATION CONTROL - REGISTRATION OF RADIATION MACHINES, FACILITIES AND SERVICES](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_02.md)
- [6 CCR 1007-1 Part 03 RADIATION CONTROL - LICENSING OF RADIOACTIVE MATERIAL](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_03.md)
- [6 CCR 1007-1 Part 04 RADIATION CONTROL - STANDARDS FOR PROTECTION AGAINST RADIATION](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_04.md)
- [6 CCR 1007-1 Part 05 RADIATION CONTROL - RADIATION SAFETY REQUIREMENTS FOR INDUSTRIAL RADIOGRAPHIC OPERATIONS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_05.md)
- [6 CCR 1007-1 Part 06 RADIATION CONTROL - X-RAY IMAGING IN THE HEALING ARTS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_06.md)
- [6 CCR 1007-1 Part 07 RADIATION CONTROL - USE OF RADIONUCLIDES IN THE HEALING ARTS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_07.md)
- [6 CCR 1007-1 Part 08 RADIATION CONTROL - RADIATION SAFETY REQUIREMENTS FOR RADIATION GENERATING DEVICES NOT USED IN THE HEALING ARTS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_08.md)
- [6 CCR 1007-1 Part 09 RADIATION CONTROL - RADIATION SAFETY REQUIREMENTS FOR PARTICLE ACCELERATORS NOT USED IN THE HEALING ARTS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_09.md)
- [6 CCR 1007-1 Part 10 RADIATION CONTROL - NOTICES, INSTRUCTIONS, AND REPORTS TO WORKERS: INSPECTIONS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_10.md)
- [6 CCR 1007-1 Part 11 RADIATION CONTROL - SPECIAL LAND OWNERSHIP REQUIREMENTS GENERAL PROVISIONS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_11.md)
- [6 CCR 1007-1 Part 12 RADIATION CONTROL - FEES FOR RADIATION CONTROL SERVICES](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_12.md)
- [6 CCR 1007-1 Part 13 RADIATION CONTROL - COMPLIANCE ENFORCEMENT](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_13.md)

---

Source: Frix Law Library, https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_08. Check the current official text before relying on it. Not legal advice.