State Board of Health
RADIATION CONTROL - X-RAY IMAGING IN THE HEALING ARTS
6 CCR 1007-1 Part 06
[Editor’s Notes follow the text of the rules at the end of this CCR Document.]
_________________________________________________________________________
Adopted by the Board of Health February 21, 2024, effective date April 14, 2024.
PART 6:
X-RAY IMAGING IN THE HEALING ARTS
6.1
Purpose and Scope.
6.1.1
Authority.
6.1.1.1 Rules and regulations set forth herein are adopted pursuant to the provisions of sections
25-1-108, 25-1.5-101(1)(l), and 25-11-104, CRS.
6.1.2
Basis and Purpose.
6.1.2.1 A statement of basis and purpose accompanies this part and changes to this part. A copy
may be obtained from the Department.
6.1.3
Scope.
6.1.3.1 Part 6 establishes requirements, for which a registrant is responsible, for use of
diagnostic and interventional x-ray equipment and imaging systems in the healing arts.
6.1.4
Applicability
6.1.4.1 The provisions of this part are in addition to, and not in substitution for, other applicable
provisions in Part 1, 2, 4, 7, 10, 24 and other parts of these regulations.
6.1.4.2 Part 24 also applies to certain healing arts x-ray imaging registrants.
6.1.4.3 The requirements and provisions of this part apply to each registrant or applicant for
registration subject to this part unless specifically exempted.
6.1.5
Published Material Incorporated by Reference.
6.1.5.1 Throughout this Part 6, federal regulations, state regulations, and standards or guidelines
of outside organizations have been adopted and incorporated by reference. Unless a
prior version of the incorporated material is otherwise specifically indicated, the materials
incorporated by reference cited herein include only those versions that were in effect as
of the most recent effective date of this Part 6 (April 2024), and not later amendments or
editions of the incorporated material.
6.1.5.2 Materials incorporated by reference are available for public inspection, and copies
(including certified copies) can be obtained at reasonable cost, during normal business
hours from the Colorado Department of Public Health and Environment, Hazardous
Materials and Waste Management Division, 4300 Cherry Creek Drive South, Denver,
Colorado 80246. Additionally, https://cdphe.colorado.gov/hm/radregs identifies where the
incorporated federal and state regulations are available to the public on the internet at no
cost. A copy of the materials incorporated in this Part is available for public inspection at
the state publications depository and distribution center.
6.1.5.3 Availability from Source Agencies or Organizations.
(1)
All federal agency regulations incorporated by reference herein are available at
no cost in the online edition of the Code of Federal Regulations (CFR) hosted by
the U.S. Government Printing Office, online at
https://www.govinfo.gov/app/collection/cfr/.
(2)
All state regulations incorporated by reference herein are available at no cost in
the online edition of the Code of Colorado Regulations (CCR) hosted by the
Colorado Secretary of State’s Office, online at
https://www.sos.state.co.us/CCR/NumericalDeptList.do#1000.
(3)
Copies of the standards or guidelines of outside organizations are available
either at no cost or for purchase from the source organizations listed below.
a.
American Association of Physicists in Medicine (AAPM)
1631 Prince Street
Alexandria, VA 22314
Phone 571-298-1300
aapm.org
b.
National Council on Radiation Protection and Measurements (NCRP)
7910 Woodmont Avenue, Suite 400
Bethesda, MD 20814-3095
Phone: 301-657-2652
ncrponline.org
6.2
Definitions.
As used in Part 6, these terms have the definitions set forth as follows:
“AAPM Online Report 03” means “Assessment of Display Performance for Medical Imaging Systems”,
AAPM Online Report No. 03 by Task Group 18 of the American Association of Physicists in Medicine
(April 2005).
“AAPM Report 96” means “The Measurement, Reporting, and Management of Radiation Dose in CT”,
AAPM Report No. 96 by Task Group 23 (CT Dosimetry) of the Radiography and Fluoroscopy
Subcommittee of the Diagnostic Imaging Council CT Committee of the American Association of Physicists
in Medicine (January 2008).
“Alert value” means a dose index that is set by the registrant to trigger an alert to the CT operator prior to
scanning within an ongoing examination. The alert value represents a universal dose index value well
above the registrant’s established range for the examination that warrants more stringent review and
consideration before proceeding.
“Aluminum equivalent” means the thickness of aluminum (type 1100 alloy with a nominal chemical
composition of aluminum 99.00 percent minimum and copper 0.12 percent maximum) affording the same
attenuation, under specified conditions, as the material in question.
“Articulated joint” means a joint between two separate sections of a tabletop which joint provides the
capacity of one of the sections to pivot on the line segment along which the sections join.
“Attenuation block” means a block or stack of type 1100 aluminum alloy, or aluminum having equivalent
attenuation, with dimensions 20 centimeters (cm) or larger by 20 cm or larger by 3.8 cm, that is large
enough to intercept the entire x-ray beam.
“Automatic exposure control” (AEC) means a device which automatically controls one or more technique
factors in order to obtain at the pre-selected location(s) a required quantity of radiation.
“Automatic exposure rate control” (AERC) means a device that automatically controls one or more
technique factors in order to obtain at the pre-selected location(s) a required quantity of radiation per unit
time.
“Automatic film processor” means a device that produces an image from a film-screen system in
mechanical steps with limited human intervention.
“Barrier”. See “protective barrier”.
“Beam axis” means, for purposes of Part 6, a line from the source through the center of the x-ray field.
“Beam-limiting device” means a device that provides a means to restrict the dimensions of the x-ray field.
“Bone densitometry” means a noninvasive measurement of certain physical characteristics of bone that
reflect bone strength. Test results are typically reported as bone mineral content or density and are used
for diagnosing osteoporosis, estimating fracture risk, and monitoring changes in bone mineral content.
“Bone densitometer” means a device intended for medical purposes to measure bone density and mineral
content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This
generic type of device may include signal analysis and display equipment, patient and equipment
supports, component parts, and accessories. A bone densitometer is synonymous with dual-energy x-ray
absorptiometry (DXA) systems.
“C-arm fluoroscope” means a fluoroscopic x-ray system in which the image receptor and the x-ray tube
housing assembly are connected or coordinated to maintain a spatial relationship. Such a system allows
a change in the direction of the beam axis with respect to the patient without moving the patient.
“Cantilevered tabletop” means a tabletop designed such that the unsupported portion can be extended at
least 100 cm beyond the support.
"Cassette holder" means a device, other than a spot-film device, that supports and/or fixes the position of
the image receptor during a radiographic exposure.
“Cephalometric” means imaging equipment or methods that are used for the radiographic visualization
and measurement of the dimensions of the human head.
“Coefficient of variation” (C) means the ratio of the standard deviation to the mean value of a population
of observations. It is estimated using the following equation:
where
s = Estimated standard deviation of the population
= Mean value of observations in sample
= ith observation in sample
n = Number of observations sampled
“Computed radiography” (CR). See “photostimulable storage phosphor system.”
“Computed tomography” (CT) means the production of a tomogram by the acquisition and computer
processing of x-ray transmission data.
"Cone Beam Computed Tomography (CBCT)" means a volumetric imaging modality that uses a two-
dimensional digital flat-panel detector to yield a three dimensional volumetric image in one rotation.
Reconstruction algorithms can be used to generate images of any desired plane.
“Control panel” means that part of the x-ray control upon which are mounted the switches, knobs,
pushbuttons, keypads, touchscreens, and other hardware or software necessary for the operator to
manually select exposure settings.
"Cradle" means:
(1)
A removable device which supports and may restrain a patient above an x-ray table; or
(2)
A device:
(i)
Whose patient support structure is interposed between the patient and the image
receptor during normal use;
(ii)
Which is equipped with means for patient restraint; and
(iii)
Which is capable of rotation about its long (longitudinal) axis.
“CT” (see “computed tomography”).
“CT conditions of operation” means all selectable parameters governing the operation of a CT x-ray
system including, but not limited to, nominal tomographic section thickness, filtration, and the technique
factors as defined in 6.2.
“CT gantry” means the tube housing assemblies, beam-limiting devices, detectors, and the supporting
structures, frames, and covers which hold and/or enclose these components within a computed
tomography system.
"CT number" means the number used to represent the x-ray attenuation associated with each elemental
area of the CT image

where:
k
=
A constant, a normal value of 1,000 when the Houndsfield scale of CT number is
used;
x

=
Linear attenuation coefficient of the material of interest;
w


=
Linear attenuation coefficient of water.
“Dead-man switch” means a switch so constructed that a circuit-closing contact can be maintained only
by continuous pressure on the switch by the operator.
“Detector” (See “Radiation detector”)
“Diagnostic source assembly” means the tube housing assembly with a beam-limiting device attached.
“Diagnostic x-ray system” means an x-ray system designed for irradiation of any part of the human or
animal body for the purpose of diagnosis or visualization.
“Digital radiography” (DR) means an x-ray imaging method (or radiography) which produces a digital
rather than analog image. DR includes both computed radiography and direct digital radiography.
"Direct digital radiography” (DDR; also see CR and DR) means an x-ray imaging method in which a digital
sensor is used to capture an x-ray image.
"Dose area product (DAP) (aka kerma-area product (KAP))" means the product of the air kerma and the
area of the irradiated field and is typically expressed in Gy-cm2, so it does not change with distance from
the x-ray tube.
“Dose profile” means the dose as a function of position along a line.
“Equipment”. See “x-ray equipment”.
“Examination” means performing a procedure, including selection of exposure settings, positioning the x-
ray system and the patient, and initiating and terminating the exposure.
“Facility”, for mammography (to supplement the Part 1 meaning of “facility”), means a hospital, outpatient
Department, clinic, radiology practice, mobile unit, office of a physician, or other facility that conducts
mammography activities, including the following: operation of equipment to produce a mammogram, initial
interpretation of the mammogram, and maintaining viewing conditions for that interpretation.
“Field emission equipment” means equipment that uses an x-ray tube in which electron emission from the
cathode is due solely to the action of an electric field.
“Filter” means material placed in the useful beam to preferentially absorb selected radiations.
“Floor plan” means, for purposes of Part 6, a plan view of the overall layout to scale of a room or group of
rooms, including the location and configuration of any radiation producing machines in each room.
“Fluoroscopic imaging assembly” means a subsystem in which x-ray photons produce a set of
fluoroscopic images or radiographic images recorded from the fluoroscopic image receptor. It includes the
image receptor(s), electrical interlocks, if any, and structural material providing linkage between the image
receptor and diagnostic source assembly.
“Fluoroscopic irradiation time” means the cumulative duration during an examination or procedure of
operator-applied continuous pressure to the device, enabling x-ray tube activation in any fluoroscopic
mode of operation.
"Fluoroscopically-Guided Interventional (FGI) Procedures" means an interventional diagnostic or
therapeutic procedure performed via percutaneous or other access routes, usually with local anesthesia
or intravenous sedation, which uses external ionizing radiation in the form of fluoroscopy to localize or
characterize a lesion, diagnostic site, or treatment site, to monitor the procedure, and to control and
document therapy.
“FGI Procedures Committee” means the representative group of individuals in a FGI facility responsible
for the ongoing review and management of FGI procedures to ensure that exams being performed
achieve the desired diagnostic image quality at the lowest radiation dose possible while properly
exploiting the capabilities of the equipment being used.
“Fluoroscopy” means a technique for generating x-ray images and presenting them simultaneously and
continuously as visible images.
“Focal spot (actual)” means the area projected on the anode of the x-ray tube bombarded by the
electrons accelerated from the cathode and from which the useful beam originates.
“General purpose radiographic x-ray system” means any radiographic x-ray system which, by design, is
not limited to radiographic examination of specific anatomical regions.
“Half-value layer” (HVL) means the thickness of specified material which attenuates the beam of radiation
to an extent such that the air kerma rate (AKR) is reduced by one-half of its original value. In this
definition, the contribution of all scattered radiation, other than any which might be present initially in the
beam concerned, is deemed to be excluded.
“Hand-held x-ray equipment” means a type of portable x-ray equipment that is designed to be held in the
operators hand during operation. See “x-ray equipment”.
“Healing arts screening” means, for purposes of these regulations, the testing or evaluation resulting in
the exposure of any human being to an x-ray machine for the detection or evaluation of health indications
when such a test is not specifically and individually ordered by a licensed physician, chiropractor, dentist,
podiatrist or other person legally authorized to prescribe such a test for the purpose of diagnosis or
treatment.
"Heat unit" means a unit of energy equal to the product of the peak kilovoltage, milliamperes, and
seconds (kVp - mA - second).
“HVL”. See “half-value layer”.
“Image intensifier” means a device, installed in its housing, that instantaneously converts an x-ray pattern
into a corresponding visible light image and electronically amplifies the brightness of that visible image.
“Image receptor” means any device, such as a fluorescent screen, radiographic film, x-ray image
intensifier tube, photostimulable phosphor, or solid-state or gaseous detector that transforms incident x-
ray photons either into a visible image or into another form that can be made into a visible image by
further transformations. In those cases where means are provided to preselect a portion of the image
receptor, the term “image receptor” shall mean the preselected portion of the device.
“Isocenter” means the center of the smallest sphere through which the beam axis passes when the
equipment moves through a full range of rotations about its common center.
“Kerma-area product (KAP)”. See “Dose area product”
“Kilovolts peak”. See “Peak tube potential”.
“kV” means kilovolt.
“kVp”. See “Peak tube potential”.
“kWs” means kilowatt-second.
"Last image hold radiograph" (LIH) means an image obtained either by retaining one or more fluoroscopic
images, which may be temporarily integrated, at the end of a fluoroscopic exposure.
“Lead equivalent” means the thickness of lead affording the same attenuation, under specified conditions,
as the material in question.
“Leakage radiation” means radiation emanating from the diagnostic source assembly except for:
(1)
The useful beam; and
(2)
Radiation produced when the exposure switch or timer is not activated.
"Leakage technique factors" means the technique factors associated with the diagnostic source assembly
which are used in measuring leakage radiation. They are defined as follows:
(1)
For diagnostic source assemblies intended for capacitor energy storage equipment, the
maximum-rated peak tube potential and the maximum-rated number of exposures in an
hour for operation at the maximum-rated peak tube potential with the quantity of charge
per exposure being 10 millicoulombs (or 10 milliampere-seconds) or the minimum
obtainable from the unit, whichever is larger;
(2)
For diagnostic source assemblies intended for field emission equipment rated for pulsed
operation, the maximum-rated peak tube potential and the maximum-rated number of x-
ray pulses in an hour for operation at the maximum-rated peak tube potential; and
(3)
For all other diagnostic source assemblies, the maximum-rated peak tube potential and
the maximum-rated continuous tube current for the maximum-rated peak tube potential.
“Light field” means that area of the intersection of the light beam from the beam-limiting device, and one
of the set of planes parallel to, and including, the plane of the image receptor, whose perimeter is the
locus of points, at which the illumination is one-fourth of the maximum in the intersection.
“Mammogram” means a radiographic image produced through mammography.
“Mammography” means radiography of the breast, but for purposes of this part, does not include:
(1)
Radiography of the breast performed during invasive interventions for localization or
biopsy procedures; or
(2)
Radiography of the breast performed with an investigational mammography device as a
scientific study conducted in accordance with FDA regulations.
“Mammography phantom” means a test object used to simulate radiographic characteristics of
compressed breast tissue and containing components that radiographically model aspects of breast
disease and cancer.
“Manual film developing” means a way to produce an image that requires human intervention to move the
film from developer to fixer to wash.
“mAs” means milliampere-seconds (mAs), a measure of electrical current produced over a set amount of
time via an x-ray tube.
“Mobile x-ray equipment”. See “x-ray equipment”.
“Mode of operation” means, for fluoroscopic systems, a distinct method of fluoroscopy, or radiography
provided by the manufacturer and selected with a set of several technique factors or other control settings
uniquely associated with the mode.
(1)
The set of distinct technique factors and control settings for the mode may be selected by
the operation of a single control.
(2)
Examples of distinct modes of operation include normal fluoroscopy (analog or digital),
high-level control fluoroscopy, cineradiography (analog or digital), digital subtraction
angiography, electronic radiography using the fluoroscopic image receptor, and
photospot recording.
(3)
In a specific mode of operation, certain system variables affecting air kerma, AKR, or
image quality, such as image magnification, x-ray field size, pulse rate, pulse duration,
number of pulses, source-image receptor distance (SID), or optical aperture, may be
adjustable or may vary; their variation per se does not comprise a mode of operation
different from the one that has been selected.
"Multiple tomogram system" means a computed tomography x-ray system which obtains x-ray
transmission data simultaneously during a single scan to produce more than one tomogram.
“NCRP Report 147” means National Council on Radiation Protection and Measurements Report No. 147,
“Structural Shielding Design For Medical Imaging Facilities” (November 2004).
“Noise” in CT means the standard deviation of the fluctuations in CT number expressed as a percentage
of the attenuation coefficient of water. Its estimate (Sn) is calculated using the following expression:

where:
CS = Contrast scale (the change in linear attenuation coefficient per CT number relative to
water).
w

= Linear attenuation coefficient of water.
s = Estimated [S]standard deviation of the CT numbers of picture elements in a specified area of
the CT image.
"Nominal tomographic section thickness" means the measured full width at half-maximum of the
sensitivity profile taken at the center of the cross-sectional volume over which x-ray transmission data are
collected.
"Notification value" means a protocol-specific dose index that is set by the registrant to trigger a
notification to the CT operator prior to scanning when the dose index exceeds the established range for
the examination.
“Optical Density” (OD) equals log (1/transmittance), where the transmittance of the film is the fraction of
incident light transmitted by the film.
“Patient” means a human being or an animal to whom machine-produced radiation is delivered for healing
arts examination, screening, diagnosis, or treatment.
“PBL”. See “positive beam limitation”.
“Peak tube potential” means the maximum value of the potential difference across the x-ray tube during
an exposure.
“Photostimulable storage phosphor” (PSP) means a material used to capture and store radiographic
images in computed radiography systems.
“Picture element” (pixel) means an elemental area of a digitally acquired image.
“PID”. See “position indicating device”.
"Pitch" means the table incrementation, in CT, per x-ray tube rotation, divided by the nominal x-ray beam
width at isocenter.
“Portable x-ray equipment”. See “x-ray equipment”.
“Position indicating device” (PID) means a device on dental x-ray equipment used to indicate the beam
position and to establish a definite source-surface (skin) distance, without regard to whether the device
incorporates or serves as a beam-limiting device.
“Positive beam limitation” (PBL) means the automatic or semi-automatic adjustment of an x-ray beam to
the size of the selected image receptor, whereby exposures cannot be made without such adjustment.
“Primary protective barrier” means the material, excluding filters, placed to attenuate the useful beam for
radiation protection purposes.
“Protective apparel” means a garment made of radiation-absorbing materials used to reduce radiation
exposure to the wearer.
“Protective barrier” means a barrier of radiation absorbing material(s) used to reduce radiation exposure.
"Protocol" means a collection of settings and parameters that fully describe an examination.
"Pulsed mode" means operation of a fluoroscopic x-ray system such that the x-ray tube current is pulsed
by the x-ray control to produce one or more exposure intervals of duration less than one-half second.
“Qualified inspector (QI)” is as defined in Section 2.2 of Part 2 of these regulations.
“Qualified trainer” is as defined in Section 2.2 of Part 2 of these regulations.
“Quality assurance (QA)” means a written monitoring and verification program which uses testing,
auditing and inspection to ensure that deficiencies, deviations, defective equipment, or unsafe practices,
or a combination thereof, relating to the use, disposal, management, or manufacture of radiation devices
are identified, promptly corrected, and reported to the department where required.
“Radiation detector” means a device which in the presence of radiation provides a signal or other
indication suitable for use in measuring one or more quantities of incident radiation.
"Radiation Protocol Committee (RPC)" means the representative group of individuals in a CT facility
responsible for the ongoing review and management of CT protocols to ensure that exams being
performed achieve the desired diagnostic image quality at the lowest radiation dose possible while
properly exploiting the capabilities of the equipment being used.
“Radiation therapy simulation system” means a radiographic or fluoroscopic x-ray system or a computed
tomography system intended for localizing the volume to be exposed during radiation therapy and
confirming the position and size of the therapeutic irradiation field.
“Radiograph” means an image created directly or indirectly by x-rays resulting in a permanent visible
image on film or digital record.
“Radiography” means a technique for generating and recording an x-ray pattern for the purpose of
providing the user with the image(s) after termination of the exposure.
“Reference plane” means a plane which is parallel to and which can be offset (as specified in
manufacturer information provided to users) from the location of the tomographic plane(s).
“Registered medical physicist (RMP)” is as defined in Section 2.2 of Part 2 of these regulations.
“Scan” means the complete process of collecting x-ray transmission data for the production of a
tomogram. Data can be collected simultaneously during a single scan for the production of one or more
tomograms.
"Scan increment" means the amount of relative displacement of the patient with respect to the CT x-ray
system between successive scans measured along the direction of such displacement.
"Scan sequence" means a pre-selected set of two or more scans performed consecutively under pre-
selected CT conditions of operation.
“Scan time” means the time elapsed during the accumulation of x-ray transmission data for a single scan.
“Scatter radiation” means ionizing radiation that, during passage through matter, has been deviated in
direction.
"Sensitivity profile" means the relative response of the CT x-ray system as a function of position along a
line perpendicular to the tomographic plane.
“Shutter” means a device attached to the tube housing assembly that can intercept the entire cross
sectional area of the useful beam and that has a lead equivalency not less than that of the tube housing
assembly.
“SID”. See “source-image receptor distance”.
"Single tomogram system" means a CT x-ray system which obtains x-ray transmission data during a scan
to produce a single tomogram.
"Size-specific dose estimate (SSDE)" means a patient dose estimate which takes into consideration
corrections based on the size of the patient, using linear dimensions measured on the patient or patient
images.
“Source” means the focal spot of the x-ray tube.
“Source-image receptor distance” (SID) means the distance from the source to the center of the input
surface of the image receptor.
“Source-skin distance” (SSD) means the distance from the source to the center of the entrant x-ray field in
the plane tangent to the patient skin surface.
“Spot image” means a radiograph that is made during a fluoroscopic examination to permanently record
conditions that exist during that fluoroscopic procedure.
“Spot-image device” means a device intended to transport and/or position a radiographic image receptor
between the x-ray source and fluoroscopic image receptor. It includes a device intended to hold a
cassette over the input end of the fluoroscopic image receptor for the purpose of producing a radiograph.
A spot-film device is an older type of spot-image device.
“SSD”. See “source-skin distance”.
“Stationary x-ray equipment”. See “x-ray equipment”.
“Stray radiation” means the sum of leakage and scattered radiation.
"Substantial radiation dose level" (SRDL) means an appropriately-selected reference value used to trigger
additional dose-management actions during a procedure and medical follow-up for a radiation level that
might produce a clinically-relevant injury in an average patient. There is no implication that radiation
levels above an SRDL will always cause an injury or that radiation levels below an SRDL will never cause
an injury.
“Technique factor” means an exposure control setting that specifies the peak tube potential in kV and
(1)
For capacitor energy storage equipment, quantity of charge in mAs; or
(2)
For field emission equipment rated for pulsed operation, number of x-ray pulses; or
(3)
For CT systems designed for pulsed operation, scan time in seconds and either:
(a)
Tube current in mA, x-ray pulse width in seconds and the number of x-ray pulses
per scan; or
(b)
The product of tube current, x-ray pulse width, and the number of x-ray pulses in
mAs;
(4)
For CT systems not designed for pulsed operation, either:
(a)
Tube current in mA and scan time in seconds; or
(b)
The product of tube current and exposure time in mAs and the scan time when
the scan time and exposure time are equivalent; and
(5)
For all other equipment, either:
(a)
Tube current in mA and exposure time in seconds; or
(b)
The product of tube current and exposure time in mAs.
“The Report of AAPM Task Group 270” means the report on Display Quality Assurance issued by the
American Association of Physicists in Medicine (AAPM), January 2019.
“Tomogram” means the depiction of the x-ray attenuation properties of a section through the body.
“Tomographic plane” means that geometric plane that is identified as corresponding to the output
tomogram.
“Tomographic section” means the volume of an object whose x-ray attenuation properties are imaged in a
tomogram.
“Tube” means an x-ray tube, unless otherwise specified.
“Tube housing assembly” means the tube housing with tube installed, including high-voltage and/or
filament transformers and other appropriate elements when such are contained within the tube housing.
“Useful beam” means the radiation which passes through the tube housing port and the aperture of the
beam limiting device when the exposure switch or timer is activated.
“Visible area” means that portion of the input surface of the image receptor over which incident x-ray
photons are producing a visible image.
“Volumetric dental imaging system” means an x-ray machine that produces, for oral and maxillofacial
structures, a three-dimensional tomographic data set or a time sequence of three-dimensional
tomographic data sets. A dental x-ray machine only capable of producing a two-dimensional image is not
considered to be a volumetric dental imaging system.
"X-ray control" means a device which controls input power to the x-ray high-voltage generator and/or the
x-ray tube. It includes equipment such as timers, phototimers, automatic brightness stabilizers, and
similar devices, which control the technique factors of an x-ray exposure.
“X-ray exposure control” means a device, switch, button or other similar means by which an operator
initiates and/or terminates the radiation exposure. The x-ray exposure control may include such
associated equipment as timers and back-up timers.
“X-ray equipment” means an x-ray system, subsystem, or component thereof. Types of x-ray equipment
are as follows:
(1)
“Mobile x-ray equipment” means x-ray equipment mounted on a permanent base with
wheels or casters for moving while completely assembled;
(2)
“Portable x-ray equipment” means x-ray equipment designed to be hand-carried;
(3)
“Stationary x-ray equipment” means x-ray equipment that is installed in a fixed location.
(4)
“Hand-held x-ray equipment” means x-ray equipment that is designed to be hand-held
during operation.
“X-ray field” means that area of the intersection of the useful beam and any one of the set of planes
parallel to and including the plane of the image receptor, whose perimeter is the locus of points at which
the air kerma rate is one-fourth of the maximum in the intersection.
“X-ray high-voltage generator” means a device that transforms electrical energy from the potential
supplied by the x-ray exposure control to the tube operating potential. The device may also include
means for transforming alternating current to direct current, filament transformers for the x-ray tube(s),
high-voltage switches, electrical protective devices, and other elements.
“X-ray imaging system” or “x-ray system” means an assemblage of components for the controlled
production of x-rays.
(1)
At a minimum, an x-ray imaging system includes an x-ray high-voltage generator, an x-ray
exposure control, a tube housing assembly, a beam-limiting device, and necessary supporting structures.
(2)
Additional components such as the image receptor(s) that function with the system are
considered integral parts of the system.
“X-ray table” means a patient support device with its patient support structure (tabletop) interposed
between the patient and the image receptor or x-ray tube during radiography and/or fluoroscopy. This
includes, but is not limited to, any stretcher equipped with a radiolucent panel and any table equipped
with a cassette tray (or Bucky), cassette tunnel, fluoroscopic image receptor, or spot-image device
beneath the tabletop.
“X-ray tube” means any electron tube that is designed to be used primarily for the production of x-rays.
“X-ray system”. See “x-ray imaging system”.
GENERAL REGULATORY PROVISIONS
6.3
General and administrative requirements.
6.3.1
Administrative Controls.
6.3.1.1 Each radiation machine used in the healing arts in the State of Colorado shall be
registered with the Department as required by Part 2, Section 2.4 and inspected as
prescribed in Part 2, Section 2.5.
6.3.1.2 Each radiation machine used on humans shall meet the Federal Performance Standards,
Subchapter J - Radiological Health, 21 CFR 1020.30 through 1020.33.
(1)
Diagnostic X-ray systems and their associated components used on humans and
certified pursuant to the Federal X-Ray Equipment Performance Standard (21
CFR 1020.30 through 1020.33) shall be maintained in compliance with applicable
requirements of that standard.
(2)
Diagnostic x-ray components and systems certified in accordance with 21 CFR
Part 1020 shall not be modified such that the component or system fails to
comply with any applicable requirement of 21 CFR Part 1020 or Part 6.
(3)
The owner of a diagnostic x-ray system who uses the system in a professional or
commercial capacity may have the system modified provided the modification
does not result in the failure of the system or component to comply with the
applicable requirements of Part 6 and any modification is completed by a
registered service company in accordance with 6.3.3.1(5).
(a)
The owner who causes such modification need not submit the reports
required by Part 6, provided the owner records the date and the details
of the modification in the system and maintains this information, and
provided the modification of the x-ray system does not result in a failure
to comply with Part 6.
(b)
Registered service companies shall submit to the Department, records of
modifications of the x-ray system, as required by these regulations.
(4)
Limited exemption from this requirement may be granted by the Department for a
radiation machine manufactured prior to August 4, 1974, provided the registrant
demonstrates that such exemption will not result in undue risk.
6.3.1.3 The registrant or the registrant's agent shall use approved providers of services,
consistent with Part 2, Section 2.6., including but not limited to operation of equipment,
inspection of radiation machines and facilities, and assembly, installation, service and/or
calibration of radiation machines.
6.3.1.4 An x-ray imaging system that is found to be non-compliant with the requirements of these
regulations 30 days beyond initial discovery, may continue to be used for up to 90 days
provided:
(1)
The system has not been determined to be unsafe for routine use in accordance
with Appendix 6D;
(2)
Continued use poses no significant radiation risk to patients, members of the
public or employees;
(3)
Does not significantly result in degraded image quality; and
(4)
The registrant obtains in writing, an authorization for continued use from the
Department.
6.3.1.5 An x-ray imaging system that is determined as provided in Appendix 6D to be unsafe for
human, animal, or other use shall not be operated for diagnostic or therapeutic purposes.
6.3.1.6 A radiation machine in the healing arts shall be operated:
(1)
By a physician, chiropractor, dentist, podiatrist or veterinarian who has a current
active State of Colorado license to practice the healing arts and has met the
applicable requirements of Part 2 of the regulations; or
(2)
By an individual authorized by and licensed in accordance with State of Colorado
statutes to engage in the healing arts and has met the applicable requirements of
Part 2 of the regulations; and
(a)
Whose license, licensing body, or licensing regulations and requirements
authorize such operation; and
(b)
Such operation is within the standard and acceptable scope of practice
for the licensed individual; or
(3)
By an individual who is under the general supervision of a licensed individual
authorized in 6.3.1.6(1) or 6.3.1.6(2), where:
(a)
The individual operator being supervised has met the applicable training
requirements of Part 2; and
(b)
Such supervision by a licensed individual is consistent with the
individual’s license, licensing body, regulations, and the standard and
acceptable scope of practice for the supervising individual.
(4)
By an operator who is under the personal supervision of a licensed individual
authorized in 6.3.1.6(1), and where:
(a)
The operator being supervised has met the applicable training
requirements of Part 2, Appendix 2O; and
(b)
Such operation is within the standard and acceptable scope of practice
of the operator being supervised.
6.3.1.7 Exposure under Part 6 of any living human being to the useful beam of an x-ray system
shall be solely for healing arts purposes, or for the purposeful exposure of a living human
research subject in accordance with Part 2, section 2.4.1.3, and only after such exposure
has been authorized by:
(1)
A physician, chiropractor, dentist, or podiatrist who has a current active State of
Colorado license to practice in the healing arts; or
(2)
An individual authorized by and licensed in accordance with State of Colorado
statutes to engage in the healing arts, and:
(a)
Whose license, licensing body, or licensing regulations and requirements
permit authorizing such exposure; and
(b)
Such exposure is within the standard and acceptable scope of practice
for the licensed individual.
6.3.1.8 The requirements of 6.3.1.7 specifically prohibits deliberate exposure for the following
purposes:
(1)
Exposure of an individual for training, demonstration or other non-healing-arts
purposes; and
(2)
Exposure of an individual for the purpose of healing arts screening except as
authorized by the Department in accordance with Section 6.3.3.4
6.3.1.9 Adequate Radiation Safety Training and Experience for a Radiation Machine Operator.
(1)
Each individual who will be operating an x-ray imaging system shall:
(a)
Be adequately instructed in the safe operating procedures;
(b)
Be competent in the safe use of the equipment; and
(c)
Meet each applicable requirement of Part 2, Section 2.6.1.
6.3.1.10
If radioactive materials are also present at the facility, the facility registrant shall
coordinate, as appropriate, requirements under Part 6 with any related requirement of the
radioactive materials license.
6.3.1.11
The registrant shall maintain for inspection, for each x-ray imaging system, the
model and serial number of each tube housing assembly and control panel:
(1)
One unique identification number that designates the entire radiation machine
shall be permanently assigned by the facility registrant to each radiation machine
and provided in all correspondence with the Department.
(a)
If feasible, the identification number shall be the “control serial number”
in Item 4 on U.S. Food and Drug Administration (FDA) Form 2579, or
equivalent.
(2)
If available, the serial number(s) from the manufacturer shall be clearly visible as
a label or stencil on the control panel and on the tube housing assembly.
(a)
Each serial number shall be the same as the corresponding number
found on FDA Form 2579, unless prior written approval is obtained from
the Department.
(3)
If either the control panel or the tube housing assembly serial number from the
manufacturer is used as the one unique identification number that designates the
entire radiation machine, and then subsequently the designated control panel or
the tube housing assembly is replaced, the registrant shall assign a new unique
identification number for the entire radiation machine and immediately provide
that new number to the Department.
6.3.2
General Specifications for Facility and Equipment Design, Configuration and Preparation.
6.3.2.1 Evaluation of Shielding Design Prior to Commencement of Operation.
(1)
The floor plan and equipment configuration of a radiation machine facility shall be
designed to meet all applicable requirements of these regulations and in
particular to preclude an individual from receiving a dose in excess of the limits in
Part 4, Sections 4.6, 4.12, 4.13, 4.14 and 4.15.
(2)
The floor plan and equipment configuration of each radiation machine facility
shall be submitted to a qualified expert for determination of shielding
requirements in accordance with Appendices 6A, 6B and 6C.
(3)
The qualified expert shielding design required by 6.3.2.1 shall be completed prior
to:
(a)
Construction of a new facility;
(b)
Any renovation or modification of an existing facility that has a potential
to reduce the effectiveness of existing shielding from x-ray radiation; or
(c)
Installation of a new radiation machine in an existing facility.
(4)
A qualified expert who completes the shielding design required by 6.3.2.1(2) shall
provide the shielding design to the facility registrant, including the annotated
dimensional drawing specified by 6.3.2.3.
(a)
The shielding design shall meet the requirements of Appendix 6C.
(5)
The facility registrant shall construct the shielding and configure the equipment in
accordance with the recommendation(s) provided by the qualified expert
pursuant to 6.3.2.1(4).
6.3.2.2 Evaluation of Shielding Design After Commencement of Operations.
(1)
A qualified expert shall review and modify a shielding design, consistent with
6.3.2.1 and Appendices 6A, 6B and 6C, whenever:
(a)
A certification evaluation or a survey during operation shows that a dose
in excess of a limit in Part 4 is possible;
(b)
An existing facility is to be modified such that the existing shielding might
be inadequate;
(c)
The primary beam orientation is changed;
(d)
The primary shielding is altered due to the modification or renovation of a
facility;
(e)
Mobile or non-hand-held portable x-ray equipment is used frequently and
regularly in the same area or room.
(f)
Radiation machine workload (for example, mA-minute-per-week
workload) has increased or is projected to increase above that which
was the basis for the original shielding design; or
(g)
The registrant is unable to produce for inspection a written shielding
design completed in accordance with 6.3.2.1 and/or 6.3.2.2.
(2)
If qualified expert analysis of operating conditions required by 6.3.2.2(1) indicates
that an individual might receive a dose in excess of the limits in Part 4, Sections
4.6, 4.12, 4.13, 4.14 or 4.15, then the facility registrant shall modify the shielding
and/or equipment configuration in accordance with the recommendation(s) of the
qualified expert.
6.3.2.3 Except for facilities exempted in 6.3.2.4, the registrant shall retain a copy of a current
dimensional drawing for each room in which a stationary x-ray imaging system is located.
The dimensional drawing shall include the following information:
(1)
Identification and use of each area adjacent to the x-ray room and an estimation
of the extent of occupancy in each such area; and
(2)
Results of calculations (as provided by a qualified expert) indicating the type and
thickness of material(s) in each protective barrier (for example, lead
equivalency):
(a)
After installation and, if possible, prior to commencement of operation,
consistent with 6.3.2.1; and
(b)
Whenever shielding is modified, consistent with 6.3.2.2.
(c)
Calculations should be performed prior to construction. When pre-
construction calculations are not available, other methods must be used
to verify the presence of any necessary shielding.
(3)
If the registrant is unable to produce for inspection the calculation(s) required by
6.3.2.3(2), survey(s) shall be conducted by a qualified expert to determine
radiation levels present under specified test conditions at the operator's position
and at clearly identifiable points outside the room.
(4)
The registrant shall maintain for inspection, for each x-ray imaging system for
which a shielding design is required:
(a)
The installation as-built drawing(s); and
(b)
The signed statement required by Part 2, Section 2.7.1.1 and retained in
accord with Part 2, Section 2.4.1.1, that all floor plan and equipment
configuration specifications in any applicable written shielding designs
required by 6.3.2 were explicitly followed.
6.3.2.4 A facility, or room within a facility, where x-ray imaging is conducted, is exempt from the
requirements of 6.3.2.1, 6.3.2.2, and 6.3.2.3 under the following conditions:
(1)
Only dental intraoral, hand-held intraoral, dental panoramic, mini-c-arm or bone
densitometry x-ray equipment is used in the area or room; or
(2)
Mobile or portable x-ray equipment is used infrequently in the same area or room
and the facility has established a written procedure or policy prescribing any
limitations necessary to demonstrate that such use will preclude any individual
from receiving a dose in excess of the public or occupational dose limits in Part 4
and that such use is consistent with the As Low As Reasonably Achievable
(ALARA) concept of Part 4, Section 4.5.2; or
(3)
Exemption for a particular area or room has been applied for in writing and
granted by the Department.
6.3.3
General Radiation Safety and Control of Radiation Exposure.
The registrant shall be responsible for directing the operation of the x-ray system(s) under their
administrative control and shall assure that the requirements of Parts 1, 2, 4, 6 and 10 are met in
the operation of the x-ray system(s).
6.3.3.1
Consistent with Part 4, Section 4.5.1 of the regulations, each facility registrant
shall have a radiation protection program. In addition to the provisions necessary for
compliance with Part 4, the radiation protection program shall include requirements that:
(1)
The use of ionizing radiation be within the registrant’s scope of practice for
healing arts purposes and shall be performed in accordance with existing laws
and regulations;
(2)
Portable and mobile x-ray equipment requirements.
(a)
Except for dental and veterinary use, portable or mobile x-ray equipment
be used only:
(i)
For examinations where it is impractical to transfer the patient to
a stationary x-ray installation; or
(ii)
When the medical status of the patient prohibits transfer of the
patient to a stationary x-ray installation.
(b)
Each facility develop a written procedure specific to the use of portable
and mobile x-ray systems that prescribes the requirements necessary to
limit an individual from receiving a dose in excess of the applicable public
or occupational dose limits in Part 4 and that such use is consistent with
the As Low As Reasonably Achievable (ALARA) concept in Part 4,
Section 4.5.2.
(c)
The Radiation Safety Officer shall review the implementation of
procedures for portable or mobile x-ray equipment use annually.
(3)
Except for veterinary use, neither the x-ray tube housing nor the collimating
device be held during an exposure with the exception of Department approved
devices specifically designed to be hand-held during operation and in
accordance with Appendix 6E.
(4)
The useful x-ray beam be limited to the area of clinical interest.
(5)
All x-ray equipment be installed by a registered service company except those
systems that do not require a physical installation to become operational.
(a)
For those x-ray systems that do not require a physical installation to
initially operate the machine, the facility registrant be responsible for
submitting the information required by Part 2, Section 2.7.2.1 through
2.7.2.4 to the department. Such systems may include hand-held x-ray
units and certain mobile or portable systems.
(6)
All x-ray equipment be used in accordance with the equipment manufacturer’s
specifications, unless otherwise directed by the licensed practitioner authorized
in 6.3.1.6(1) or (2).
(7)
The registrant use auxiliary equipment designed to minimize human patient and
personnel exposure commensurate with the needed diagnostic information.
The requirements of 6.3.3.1(8) and 6.3.3.1(9) are not applicable to veterinary facilities.
(8)
Consideration be given to selecting the appropriate technique and employing
available dose reduction methods and technologies across all patient sizes and
clinical indications.
(9)
A documented procedure be in place for verification of patient identity and exam
to be performed, including identification of the appropriate body part.
6.3.3.2 Written safety procedures shall be developed and provided for safe operation of each x-
ray imaging system.
(1)
The written safety procedures shall be readily available to each individual
radiation machine operator prior to operating x-ray imaging equipment.
(2)
The operator shall be able to demonstrate familiarity with the procedures
applicable to safe use of the system being operated.
(3)
The procedures shall include:
(a)
Any restriction on the operating technique particular to the system,
consistent with 6.3.3.3;
(b)
Limitation on beam size, to the smallest area that is clinically necessary,
including appropriate collimation:
(i)
For each tube with variable collimation, the collimation procedure
shall specify whether positive beam limitation (PBL) or manual
collimation shall be used; and
(ii)
For tubes collimated manually, all images shall provide a positive
indication of collimation, except when diagnosis might be
compromised;
(c)
Patient holding instructions consistent with 6.3.3.8.
(d)
Requirements and limitations on the use of portable or mobile x-ray
systems consistent with 6.3.3.1(2).
6.3.3.3 To reduce radiation exposure to the minimum that is necessary for general radiographic
systems not equipped or not used with an anatomic programming option, protocols shall
be documented and readily available to the operator.
(1)
Written exam protocol(s) shall be located near each system’s control panel or
available to the operator in digital form.
(a)
The exam protocols shall state the exposure settings to be used
corresponding to the patient’s (adult and pediatric, if appropriate) body
part and anatomical size, or body part thickness, or age (for pediatrics),
including but not limited to:
(i)
Technique factors (kVp, mAs if manual mode is used);
(ii)
Type of image receptor to be used;
(iii)
Type of grid, if any;
(iv)
Source to image receptor distance to be used, except for
intraoral radiography in accordance with 6.7.2.3;
(v)
Mode of operation; and
(vi)
Type and location of placement of patient shielding if used.
(2)
For computed and digital radiography, the exam protocols required by 6.3.3.3(1)
shall:
(a)
Portray how to determine applicable exposure settings in accord with
documented protocol;
(b)
Specify a control range for the exposure indicator in accordance with the
manufacturer’s or RMP recommendation; and
(c)
Specify pediatric protocol for each unit that images pediatric patients.
(3)
The settings to be used during an exposure shall be indicated before the
exposure begins.
(a)
If automatic exposure controls are used, the exposure settings that are
set prior to the exposure shall be indicated.
(b)
The requirement of 6.3.3.3(3) may be met by permanent markings on
equipment having fixed exposure settings.
(4)
The exam protocol shall be revised as necessary whenever a component is
replaced or added.
6.3.3.4 Healing Arts Screening.
(1)
Any person proposing to conduct a healing arts screening program on living
humans shall not initiate such a program without prior approval of the
Department. Authorization for healing arts screening may be granted by the
Department provided the registrant demonstrates that such healing arts
screening will not result in undue risk.
(a)
Each healing arts screening program shall obtain prior written approval
by the Department.
(b)
Each applicant for Department approval of a healing arts screening
program shall submit to the Department a completed Form R-300,
“Application for Registration – Healing Arts Screening,” including as
provided in Part 2, Section 2.4.1.2 all of the information required by
Appendix 6F and/or by Form R-300 and any accompanying instructions,
together with the required fee(s).
(c)
The registrant shall immediately notify the Department if any information
related to the healing arts screening program previously submitted to the
Department becomes invalid or outdated.
(2)
FDA/MQSA-certified facilities that are registered with the department for the use
of dedicated mammographic equipment for mammography screening are
approved for mammography screening only and are considered to have met the
healing arts screening requirements of 6.3.3.4(1).
6.3.3.5 When imaging human patients, the registrant shall restrict the presence of individuals in
the immediate area of the patient being examined to those required or in training for the
medical procedure, or the parent or guardian of a patient – while the x-ray tube is
energized. The following applies to all individuals, other than the patient being examined:
(1)
All persons shall be positioned such that no part of the body will be struck by the
useful beam unless protected by at least 0.5 millimeter lead equivalent material
except where the radiation safety officer has determined that use of protective
equipment is not in the best interest of radiation safety for the patient or
individuals in the immediate area.
(2)
All persons shall be protected from scatter radiation by protective garments,
safety equipment or whole body protective barriers of at least 0.25 millimeter lead
equivalent material except where the radiation safety officer has determined that
use of protective equipment is not in the best interest of radiation safety for the
patient or individuals in the immediate area.
(3)
Instances may warrant having human patients other than the one being
examined in the room during the exam.
(a)
If the procedure results in scatter radiation in excess of 0.02 mSv (2 mR)
in any one hour at the position of these non-imaged patients, they shall
be protected from the scatter radiation by whole body protective barriers
or apparel of at least 0.25 millimeter lead equivalent material or shall be
positioned so that the 0.02 mSv (2 mR) in any one hour limit is met.
6.3.3.6 Each facility shall have a sufficient number of lead equivalent protective apparel,
equipment and shields available to provide the necessary radiation protection to all
individuals who are involved with x-ray operations and who are otherwise not shielded.
(1)
All protective apparel and auxiliary shields shall be evaluated annually for
integrity.
Registrants shall establish a written procedure and criteria for the integrity
evaluation and shall:
(a)
Visually inspect the protective apparel and shields for breaks, tears or
holes that would significantly compromise the protective capability of the
equipment;
(b)
Perform a tactile test by placing the protective apparel on a smooth
surface and feeling for broken or missing shield material.
(2)
Protective garments and shields shall be:
(a)
Clearly labeled with their lead equivalence;
(b)
Hung and not folded to prevent damage, as applicable.
(3)
If results of the integrity test indicate breaks, tears, holes, missing material or
gaps in that would significantly compromise the protective capability of the
material, the protective apparel shall be:
(a)
Removed from service and marked as such; or
(b)
Repaired or replaced as required.
(4)
Records of the integrity check required by 6.3.3.6 shall be maintained by the
registrant for 3 years after the integrity checks are completed.
6.3.3.7 Beam collimation, positioning, and shielding of radiosensitive organs from the useful
beam that will not interfere with the imaging or medical procedure, or is contraindicated
for radiation safety reasons, shall be used to reduce radiation exposure to the patient
whenever possible.
6.3.3.8 In cases where a patient or image receptor requires additional support, mechanical
support devices shall be used whenever possible. If a patient or image receptor must be
provided with additional support during a radiation exposure:
(1)
Written safety procedures, as required by 6.3.3.2, shall indicate the requirements
for selecting a human holder and the procedure the human holder shall follow.
(2)
The human holder shall be instructed in personal radiation safety and protected
as required by 6.3.3.2;
(3)
No individual shall be used routinely to hold the image receptor or patients during
a radiation exposure.
(4)
In those cases where the patient must hold the image receptor, except during
intraoral examinations, any portion of the body other than the area of clinical
interest struck by the useful beam shall be protected by at least 0.5 millimeter
lead equivalent material except where use of protective equipment would
interfere with the examination or is contraindicated for radiation safety reasons.
6.3.3.9 The speed of film, or film-screen combination, imaging plate or receptor and image
processing, shall be the fastest speed or speed equivalent consistent with the diagnostic
objective of the examinations.
6.3.3.10
If anti-scatter grids are used between the patient and the image receptor to
decrease scattered radiation to the image receptor and improve contrast, the grid shall
be:
(1)
Positioned properly, with the tube side facing the correct direction, and centered
to the central ray; and
(2)
Of the proper focal distance for the SID being used.
6.3.3.11
When individual exposure monitoring is required by Part 4, Section 4.18, each
occupationally exposed individual who is associated with the operation of an x-ray
imaging system shall meet the requirements of Part 4, Sections 4.6, 4.10, 4.12, 4.13,
4.14, and 4.18.
(1)
When personnel dosimetric monitoring devices are required, they shall be worn
in accordance with Part 4, Section 4.6.3.
(2)
Each operator of hand-held x-ray equipment shall follow the requirements of
Appendix 6E regarding personnel monitoring devices
(3)
Deliberate exposure of a personnel dosimetric monitoring device to deceptively
indicate a dose delivered to an individual is prohibited.
6.3.4
Maintenance of Records.
6.3.4.1 The registrant shall maintain the following information on each x-ray system for
inspection by the Department as specified below:
(1)
The records in (a) through (d) are required to be retained for 3 years:
(a)
Records of surveys, calibrations, maintenance, and modifications (e.g.,
major software and hardware upgrades) performed on the x-ray
system(s);
(b)
Records of certification evaluations pursuant to 2.5, Department Forms
59-1 and 59-2, and corrective actions for each x-ray imaging system with
the names of persons who performed such services;
(c)
A copy of all correspondence with the Department regarding the x-ray
system.
(d)
Each facility shall maintain a printed or electronic record containing each
patient’s identifier, the type of examination(s), machine operator
identifier, and the date(s) the examination(s) were performed.
The records in (2) are required to be retained for the life of the system:
(2)
Model and serial numbers of all major components, and user’s manuals for those
components, including software.
The records in (3) and (4) are required to be retained for the life of the facility:
(3)
The most recent dimensional drawing and accompanying calculation(s) and/or
survey(s) as provided in 6.3.2.3 for each room in which a stationary x-ray system
is located, except as exempted under 6.3.2.4.
(4)
Consistent with Part 2, Section 2.4.1, and 6.3.2, the registrant shall retain on file
at the facility the most recent shielding design along with installer as-built
drawings.
6.3.5
Quality Assurance (QA) Program.
6.3.5.1 The registrant shall establish and maintain a quality assurance (QA) program. In addition
to the standards in the modality specific sections of Part 6, the registrant shall:
(1)
Maintain documentation of credentials for practitioners, radiation safety officers,
and x-ray operators, as required by Part 2 of the regulations.
(2)
Designate an appropriately trained individual to manage the QA program.
(3)
Establish and maintain written QA and quality control (QC) procedures, including
evaluation frequencies and tolerances or use standards of an appropriate
nationally recognized organization, for example, the American College of
Radiology or American Association of Physicists in Medicine.
(4)
Evaluate image quality by checking each imaging study for artifacts. If an artifact
impacting image interpretation or indicating an imaging system problem is
present, the source shall be identified and appropriate action taken.
(5)
With the exception of Dental facilities performing only intra-oral, panoramic,
cephalometric or volumetric dental imaging, Podiatry facilities, and Veterinary
facilities, perform repeat / reject analysis of radiographic images at least quarterly
following specifications of a nationally recognized organization.
(6)
Perform periodic preventative maintenance on the x-ray systems in accordance
with manufacturer requirements or those of nationally accepted standards.
(7)
Maintain documentation showing the calibration date and serial number for
testing instruments used in determining compliance with the provisions of section
6.3.5. Test instrument calibration frequency shall be consistent with the
regulations or nationally accepted standards.
(8)
Complete and document an annual review of the QA program.
(9)
Retain QA/QC records of evaluations and reviews for no less than three years.
(10)
Follow manufacturer’s recommendations for image processing systems, except
where otherwise specified in the regulations or where it is inconsistent with
nationally accepted standards.
6.3.5.2 Each registrant that uses analog image receptors (e.g., radiographic film) shall have
available suitable equipment for handling and processing radiographic film in accordance
with the following provisions:
Manually developed film:
(1)
Processing tanks shall be constructed of mechanically rigid, corrosion resistant
material; and
(2)
Developing solutions shall be prepared, replenished, and replaced following
manufacturer recommendations.
(3)
The temperature of solutions in the tanks shall be maintained within the range of
60° F to 80° F (16° C to 27° C). Film shall be developed in accordance with the
time-temperature relationships recommended by the film manufacturer, or in the
absence of such recommendations, use the time-temperature chart found in
Appendix 6H;
(4)
Devices shall be utilized which will indicate the actual temperature of the
developer solution and signal the passage of a preset time.
(5)
Measure and log developer temperature each day of use;
(6)
Document in a written log the change of developer chemicals at least every
month.
Automatic processors and other closed processing systems:
(7)
Shall be operated and maintained following manufacturer specifications.
(8)
Films shall be developed in accordance with the time temperature relationships
recommended by the film manufacturer. In the absence of such
recommendations, the film shall be developed using the chart in Appendix 6G.
6.3.5.3 Deviations from the processing requirements of 6.3.5.2 shall be documented by the
registrant in such manner that the requirements are shown to be met or exceeded (e.g.,
extended processing, and special rapid chemistry).
6.3.5.4 Additional requirements for facilities using x-ray film
(1)
All film storage and pass boxes, if provided, shall be so constructed as to exclude
light from the darkroom when cassettes are placed in or removed from the boxes,
and shall incorporate adequate shielding from stray radiation to prevent exposure
of undeveloped film.
(2)
Darkrooms typically used by more than one individual shall be provided a method
to prevent accidental entry while undeveloped films are being handled or
processed.
(3)
Film shall be stored in a cool, dry place and shall be protected from exposure to
stray radiation. Film in open packages shall be stored in a light tight container.
(4)
Film cassettes and intensifying screens shall be inspected periodically and shall
be cleaned and replaced as necessary.
(5)
Outdated x-ray film shall not be used.
(6)
The film and intensifying screen shall be spectrally compatible.
(7)
Facilities shall maintain a light-tight darkroom or closed processing system, use
proper safelighting and safeguards, and evaluate darkroom integrity and daylight
loading systems for film fog every six months and after a change that may impact
film fog or an event that may impact the integrity of the closed processing
system.
(8)
Facilities other than dental, podiatry, and veterinary shall:
(a)
Have a continuous and documented sensitometric quality control
program, including quality control tests for speed, contrast and fog.
These tests shall be performed according to specifications of the
manufacturer, an RMP, or a nationally recognized organization.
(b)
Maintain a light-tight darkroom or processing system and use proper
safelighting and safeguards such that any film type in use exposed in a
cassette to x-radiation sufficient to produce an optical density from 1 to 2
when processed shall not suffer an increase in optical density greater
than 0.1 when exposed in the darkroom for 2 minutes with all safelights
on. If used, daylight film handling boxes shall preclude fogging of the
film.
(c)
Limit the base plus fog of unexposed film to an optical density less than
0.25 when developed by the routine procedure used by the facility.
6.3.5.5 Facilities Using Computed Radiography (CR), Digital Radiography (DR), or Direct Digital
Radiography (DDR).
(1)
Facilities shall establish and follow an image quality control program in accord
with the recommendations of an RMP, the system manufacturer, or a nationally
recognized organization.
(2)
In addition to 6.3.5.5(1), CR facilities shall perform erasure of all CR cassettes, at
least on a weekly basis.
6.3.5.6 The registrant shall ensure that each monitor under the control of the registered facility
used for primary image interpretation is evaluated according to specifications of the
manufacturer and/or a registered medical physicist and/or a nationally recognized
organization, for example, in The Report of AAPM Task Group 270 (January 2019), or
AAPM Online Report OR-03 (April 2005), including but not limited to:
(1)
Frequent careful cleaning of each primary image interpretation workstation and
data acquisition workstation monitor;
(2)
Periodic visual assessment using nationally accepted test patterns appropriate
for the evaluation;
(3)
Verification that monitor calibration conforms with the DICOM Part 14 Grayscale
Standard Display Function, or equivalent:
(a)
Visualization of low contrast patches;
(b)
Visualization of spatial resolution targets;
(c)
Evaluation of ambient light levels;
(d)
Measurement of the luminance from a sufficient number of driving levels;
(e)
Measurements to assure that the luminance for multiple monitors are
within 10% of each other when more than one monitor is being utilized at
a primary image interpretation workstation.
(4)
The requirements of 6.3.5.6(1) through (3) must be completed initially, annually,
and when a monitor is replaced or undergoes a significant repair.
(5)
For monitors used in mammography image interpretation, the applicable monitor
QA requirements of MQSA shall be followed.
6.4
Requirements for use of all diagnostic and interventional x-ray imaging systems.
6.4.1
Administrative Controls.
6.4.1.1 In addition to the general requirements of 6.3, the requirements of 6.4 apply to all
diagnostic and interventional x-ray imaging systems and associated facilities, except as
provided by 6.7.5.1 for dental uses and 6.8.5.1 for veterinary uses.
Additional requirements specific to dental intra-oral, panoramic, cephalometric, and
volumetric dental imaging equipment are included in Section 6.7.
6.4.1.2 Each individual who operates an x-ray imaging system used on living humans shall meet
the applicable radiation safety training and experience requirements of Part 2, Section
2.6.1.
6.4.2
Each diagnostic x-ray imaging system shall meet the following equipment design and
configuration requirements.
6.4.2.1 Warning Label.
(1)
On systems manufactured on or before June 10, 2006, the control panel
containing the main power switch shall bear this or an equivalent warning
statement, or the warning statement in 6.4.2.1(2), legible and accessible to view:
“WARNING: This x-ray unit may be dangerous to patient and operator unless safe
exposure factors and operating instructions are observed.”
(2)
On systems manufactured after June 10, 2006, the control panel containing the
main power switch shall bear the warning statement, legible and accessible to
view:
“WARNING: This x-ray unit may be dangerous to patient and operator unless safe
exposure factors, operating instructions and maintenance schedules are observed.”
6.4.2.2 Battery Charge Indicator.
(1)
On battery-powered x-ray generators, visual means shall be provided on the
control panel to indicate whether the battery is in a state of charge adequate for
proper operation.
6.4.2.3 Leakage Radiation from the Diagnostic Source Assembly.
(1)
The leakage radiation from the diagnostic source assembly measured at a
distance of 1 meter in any direction from the source shall not exceed 0.88
milligray (mGy) air kerma (100 milliroentgen (mR) exposure) in any 1 hour when
the x-ray tube is operated at its leakage technique factors. If the maximum rated
peak tube potential of the tube housing assembly is greater than the maximum
rated peak tube potential for the diagnostic source assembly, positive means
shall be provided to limit the maximum x-ray tube potential to that of the
diagnostic source assembly.
(2)
Compliance shall be determined by measurements averaged over an area of 100
square cm with no linear dimension greater than 20 cm.
6.4.2.4 Radiation from Components Other Than the Diagnostic Source Assembly.
(1)
The radiation emitted by a component other than the diagnostic source assembly
shall not exceed an air kerma of 18 microgray (µGy) (2 milliroentgens (mR)
exposure) in any one hour at 5 cm from any accessible surface of the component
when it is operated in an assembled x-ray system under any conditions for which
it was designed.
(2)
Compliance shall be determined by measurements averaged over an area of 100
square centimeters (cm) with no linear dimension greater than 20 cm.
6.4.2.5 Beam Quality: Half-value Layer
(1)
The half-value layer of the useful beam for a given x-ray tube potential shall not
be less than the values shown in Appendix 6I.
(2)
If it is necessary to determine such half-value layer at an x-ray tube potential that
is not listed in Appendix 6I, linear interpolation or extrapolation is acceptable.
Positive means shall be provided to ensure that at least the minimum filtration
needed to achieve beam quality requirements is in the useful beam during each
exposure. In the case of a system, which is to be operated with more than one
thickness of filtration, this requirement can be met by a filter interlocked with the
kilovoltage selector which will prevent x-ray emissions if the minimum required
filtration is not in place.
(3)
Optional filtration on fluoroscopic systems. Fluoroscopic systems manufactured
on or after June 10, 2006, incorporating an x-ray tube(s) with a continuous output
of 1 kilowatt or more and an anode heat storage capacity of 1 million heat units
or more shall provide the option of adding x-ray filtration to the diagnostic source
assembly in addition to the amount needed to meet the half-value layer
provisions of 6.4.2.5. The selection of this additional x-ray filtration shall be either
at the option of the user or automatic as part of the selected mode of operation. A
means of indicating which combination of additional filtration is in the x-ray beam
shall be provided.
(4)
For capacitor energy storage x-ray equipment still in use, compliance with the
applicable requirements of 6.4.2.5 shall be determined with the system fully
charged and for the highest clinically used mAs.
(a)
Due to reduced image quality and potential for higher patient exposures,
capacitor energy storage x-ray equipment shall no longer be used for
human patient imaging beyond January 1, 2022.
6.4.2.6 Tube Heads.
(1)
The tube housing assembly supports shall be adjusted such that the tube
housing assembly will remain stable during an exposure unless tube housing
movement is a designed function of the x-ray system.
(2)
Where two or more radiographic tubes are controlled by one exposure switch,
the tube or tubes that have been selected shall be clearly indicated prior to
initiation of the exposure. Only the selected tube(s) can be energized.
(a)
This indication shall be both on the x-ray control panel and at or near the
tube housing assembly that has been selected.
(3)
Information displayed at the tube housing assembly meet manufacturer’s
specifications.
6.4.2.7 Locks.
(1)
All position locking, holding, and centering devices on the x-ray system and/or
components shall function as intended.
6.4.2.8 The x-ray control shall provide:
(1)
Visual indication observable at or from the operator's protected position
whenever x-rays are produced; and
(2)
A signal audible to the operator to indicate that the exposure has terminated.
6.5
Requirements for use of a fluoroscopy system.
6.5.1
In addition to the provisions of 6.3 and 6.4, the requirements of 6.5 apply to all fluoroscopic
facilities and equipment used for fluoroscopic imaging or for recording images from the
fluoroscopic image receptor.
6.5.1.1 Only image-intensified or direct-digital receptor fluoroscopic equipment shall be used for
fluoroscopy.
6.5.2
Primary Protective Barrier.
6.5.2.1 Limitation of useful beam.
(1)
The fluoroscopic imaging assembly shall be provided with a primary protective
barrier which intercepts the entire cross section of the useful beam at any SID.
(2)
The x-ray tube used for fluoroscopy shall not produce x-rays unless the barrier is
in position to intercept the entire useful beam.
(3)
The AKR due to transmission through the barrier with the attenuation block in the
useful beam combined with radiation from the fluoroscopic imaging receptor shall
not exceed 3.34x10-3 percent of the entrance AKR, at a distance of 10 cm from
any accessible surface of the fluoroscopic imaging assembly beyond the plane of
the image receptor.
(4)
Radiation therapy simulation systems shall be exempt from 6.5.2.1 provided the
systems are intended only for remote control operation.
6.5.2.2 Measuring compliance.
(1)
The AKR shall be measured in accordance with 6.5.5.
(2)
The AKR due to transmission through the primary barrier combined with radiation
from the fluoroscopic image receptor shall be determined by measurements
averaged over an area of 100 square cm with no linear dimension greater than
20 cm.
(3)
If the source is below the tabletop, the AKR measurement shall be made with the
input surface of the fluoroscopic imaging assembly positioned 30 cm above the
tabletop.
(4)
If the source is above the tabletop and the SID is variable, the AKR
measurement shall be made with the end of the beam-limiting device or spacer
as close to the tabletop as it can be placed, provided that it shall not be closer
than 30 cm.
(5)
Movable grids and compression devices shall be removed from the useful beam
during the measurement.
(6)
For all AKR measurements, the attenuation block shall be positioned in the
useful beam 10 cm from the point of measurement of entrance AKR and between
this point and the input surface of the fluoroscopic imaging assembly.
6.5.3
Field Limitation.
6.5.3.1 Angulation.
(1)
For fluoroscopic equipment manufactured after February 25, 1978, when the
angle between the image receptor and the beam axis of the x-ray beam is
variable, means shall be provided to indicate when the axis of the x-ray beam is
perpendicular to the plane of the image receptor.
(2)
Compliance with 6.5.3.5 and 6.5.3.6 shall be determined with the beam axis
indicated to be perpendicular to the plane of the image receptor.
6.5.3.2 Further means of limitation.
(1)
Means shall be provided to permit further limitation of the x-ray field to sizes
smaller than the limits of 6.5.3.5 and 6.5.3.6.
(2)
Beam-limiting devices manufactured after May 22, 1979, and incorporated in
equipment with a variable SID and/or capability of a visible area of greater than
300 cm2, shall be provided with means for stepless adjustment of the x-ray field.
(3)
Equipment with a fixed SID and the capability of a visible area of no greater than
300 cm2 shall be provided with either:
(a)
Stepless adjustment of the x-ray field; or
(b)
A means to further limit the x-ray field size at the plane of the image
receptor to 125 cm2 or less.
(4)
Stepless adjustment shall, at the greatest SID, provide continuous field sizes
from the maximum obtainable to a field size containable in a square of 5 cm by 5
cm.
(5)
Compliance with 6.5.3.2 shall be determined with the beam axis indicated to be
perpendicular to the plane of the image receptor.
(a)
Measurement shall be made in perpendicular directions corresponding to
the vertical and horizontal directions on the video monitor image.
(b)
For collimating systems that are not circular, measurement shall be
made along the directions closest to the vertical and horizontal direction
on the video monitor image yielding the smallest dimension in each
direction.
6.5.3.3 Spot-image devices.
The following requirements shall apply to spot-image devices, except when the spot-image
device is provided for use with a radiation therapy simulation system:
(1)
Means shall be provided between the source and the patient for adjustment of
the x-ray field size in the plane of the image receptor to the size of that portion of
the image receptor which has been selected on the spot-image selector.
(a)
Such adjustment shall be accomplished automatically when the x-ray
field size in the plane of the image receptor is greater than the selected
portion of the image receptor.
(b)
If the x-ray field size is less than the size of the selected portion of the
image receptor, the field size shall not open automatically to the size of
the selected portion of the image receptor unless the operator has
selected that mode of operation.
(2)
Neither the length nor width of the x-ray field in the plane of the image receptor
shall differ from the corresponding dimensions of the selected portion of the
image receptor by more than 3 percent of the SID when adjusted for full
coverage of the selected portion of the image receptor.
(a)
The sum, without regard to sign, of the length and width differences shall
not exceed 4 percent of the SID.
(b)
On spot-image devices manufactured after February 25, 1978, if the
angle between the plane of the image receptor and beam axis is
variable, means shall be provided to indicate when the axis of the x-ray
beam is perpendicular to the plane of the image receptor, and
compliance shall be determined with the beam axis indicated to be
perpendicular to the plane of the image receptor.
(3)
The center of the x-ray field in the plane of the image receptor shall be aligned
with the center of the selected portion of the image receptor to within 2 percent of
the SID.
(4)
Means shall be provided to reduce the x-ray field size in the plane of the image
receptor to a size smaller than the selected portion of the image receptor such
that:
(a)
For spot-image devices used on fixed-SID fluoroscopic systems which
are not required to, and do not provide stepless adjustment of the x-ray
field, the minimum field size, at the greatest SID, does not exceed 125
square cm; or
(b)
For spot-image devices used on fluoroscopic systems that have a
variable SID and/or stepless adjustment of the field size, the minimum
field size, at the greatest SID, shall be containable in a square of 5 cm by
5 cm.
6.5.3.4 A capability may be provided for overriding the automatic x-ray field size adjustment in
case of system failure.
If it is so provided, a signal visible at the fluoroscopist’s position shall indicate whenever the
automatic x-ray field size adjustment override is engaged. Each such system failure override
switch shall be clearly labeled as follows:
“FOR X-RAY FIELD LIMITATION SYSTEM FAILURE”
6.5.3.5 Fluoroscopy and radiography using the fluoroscopic imaging assembly with inherently
circular image receptors.
(1)
For fluoroscopic equipment manufactured before June 10, 2006, other than
radiation therapy simulation systems, the following applies:
(a)
Neither the length nor width of the x-ray field in the plane of the image
receptor shall exceed that of the visible area of the image receptor by
more than 3 percent of the SID. The sum of the excess length and the
excess width shall be no greater than 4 percent of the SID.
(b)
For rectangular x-ray fields used with circular image receptors, the error
in alignment shall be determined along the length and width dimensions
of the x-ray field which pass through the center of the visible area of the
image receptor.
(2)
For fluoroscopic equipment manufactured on or after June 10, 2006, other than
radiation simulation systems, the maximum area of the x-ray field in the plane of
the image receptor shall conform with one of the following requirements:
(a)
When any linear dimension of the visible area of the image receptor
measured through the center of the visible area is less than or equal to
34 cm in any direction, at least 80 percent of the area of the x-ray field
overlaps the visible area of the image receptor, or
(b)
When any linear dimension of the visible area of the image receptor
measured through the center of the visible area is greater than 34 cm in
any direction, the x-ray field measured along the direction of greatest
misalignment with the visible area of the image receptor does not extend
beyond the edge of the visible area of the image receptor by more than 2
cm.
6.5.3.6 Fluoroscopy and radiography using fluoroscopic imaging assembly with inherently
rectangular image receptors.
For x-ray systems manufactured on or after June 10, 2006, the following applies:
(1)
Neither the length nor width of the x-ray field in the plane of the image receptor
shall exceed that of the visible area of the image receptor by more than 3 percent
of the SID. The sum of the excess length and the excess width shall be no
greater than 4 percent of the SID.
(2)
The error in alignment shall be determined along the length and width
dimensions of the x-ray field which pass through the center of the visible area of
the image receptor.
6.5.3.7 Override capability.
If the fluoroscopic x-ray field size is adjusted automatically as the SID or image receptor size is
changed, a capability may be provided for overriding the automatic adjustment in case of system
failure. If it is so provided, a signal visible at the fluoroscopist’s position shall indicate whenever
the automatic field adjustment is overridden. Each such system failure override switch shall be
clearly labeled as follows:
FOR X-RAY FIELD LIMITATION
SYSTEM FAILURE
6.5.4
Activation of Tube.
6.5.4.1 X-ray production in the fluoroscopic mode shall be controlled by a device which requires
continuous pressure by the operator for the entire time of any exposure.
6.5.4.2 When recording serial radiographic images from the fluoroscopic image receptor, the
operator shall be able to terminate the x-ray exposure(s) at any time, but means may be
provided to permit completion of any single exposure of the series in process.
6.5.5
Air Kerma Rates.
6.5.5.1 Except for fluoroscopic equipment used for radiation therapy simulation purposes, the
following requirements apply to fluoroscopic equipment manufactured before May 19,
1995:
(1)
Equipment provided with automatic exposure rate control (AERC) shall not be
operable at any combination of tube potential and current that will result in an
AKR in excess of 88 mGy per minute (10 R/min exposure rate) at the
measurement point specified in 6.5.5.4, except as specified in 6.5.5.1(5).
(2)
Equipment provided without AERC shall not be operable at any combination of
tube potential and current that will result in an AKR in excess of 44 mGy per
minute (5 R/min exposure rate) at the measurement point specified in 6.5.5.4,
except as specified in 6.5.5.1(5).
(3)
Equipment provided with both an AERC mode and a manual mode shall not be
operable at any combination of tube potential and current that will result in an
AKR in excess of 88 mGy per minute (10 R/min exposure rate) in either mode at
the measurement point specified in 6.5.5.4, except as specified in 6.5.5.1(5).
(4)
Equipment may be modified in accordance with this Part to comply with 6.5.5.2.
When the equipment is modified, it shall bear a label indicating the date of the
modification and the statement:
MODIFIED TO COMPLY WITH 21 CFR 1020.32(H)(2)
(5)
The AKR requirements of 6.5.5.1(1) through (3) are not applicable during:
(a)
Recording of (spot) fluoroscopic images; or
(b)
Operation in high-level control mode(s) as equipped.
6.5.5.2 Except for fluoroscopic equipment used for radiation therapy simulation purposes, the
following requirements apply to fluoroscopic equipment manufactured on or after May 19,
1995:
(1)
Shall be equipped with AERC if operable at any combination of tube potential
and current that results in an AKR greater than 44 mGy per minute (5 R/min
exposure rate) at the measurement point specified in 6.5.5.4. Provision for
manual selection of technique factors may be provided.
(2)
Shall not be operable at any combination of tube potential and current that will
result in an AKR in excess of 88 mGy per minute (10 R/min exposure rate) at the
measurement point specified in 6.5.5.4, except as specified in 6.5.5.2.(3).
(3)
The AKR limits of 6.5.5.2(1) and (2) are not applicable to equipment
manufactured prior to June 10, 2006, during the recording of images from a
fluoroscopic image receptor using photographic film or a video camera when the
x-ray source is operated in a pulsed mode.
(4)
The AKR limits of 6.5.5.2(1) and (2) are not applicable to: equipment
manufactured on or after June 10, 2006:
(a)
During recording of spot images from the fluoroscopic image receptor;
(b)
To images resulting from a last-image-hold feature that are not recorded;
(c)
During operation in high-level control mode(s) as equipped.
6.5.5.3 Except for fluoroscopic equipment used for radiation therapy simulation purposes, the
following requirements apply to fluoroscopy equipment with optional high-level control
(1)
When high-level control is selected and the control is activated, in which case the
equipment shall not be operable at any combination of tube potential and current
that will result in an AKR in excess of 176 mGy per minute (20 R/min exposure
rate) at the measurement point specified in 6.5.5.4.
(2)
Special means of activation of high-level controls shall be required.
(a)
The high-level control shall be operable only when continuous manual
activation is provided by the operator.
(b)
A continuous signal audible to the fluoroscopist shall indicate that the
high-level control is employed.
6.5.5.4 Measuring compliance.
Except for fluoroscopic equipment used for radiation therapy simulation purposes, the
following requirements apply to compliance with 6.5.5.1 through 6.5.5.3 and shall be
determined as follows:
(1)
If the source is below the x-ray table, the AKR shall be measured at 1 cm above
the tabletop or cradle.
(2)
If the source is above the x-ray table, the AKR shall be measured at 30 cm above
the tabletop with the end of the beam-limiting device or spacer positioned as
closely as possible to the point of measurement.
(3)
For a C-arm type of fluoroscope, the AKR shall be measured at 30 cm from the
input surface of the fluoroscopic imaging assembly, with the source positioned at
any available SID, provided that the end of the beam-limiting device or spacer is
no closer than 30 cm from the input surface of the fluoroscopic imaging
assembly.
(4)
For a C-arm type of fluoroscope having an SID less than 45 cm, the AKR shall be
measured at the minimum SSD.
(5)
For a fixed lateral type of fluoroscope, the air kerma rate shall be measured at a
point 15 cm from the centerline of the x-ray table and in the direction of the x-ray
source with the end of the beam-limiting device or spacer positioned as closely
as possible to the point of measurement.
(a)
If the tabletop is movable, it shall be positioned as closely as possible to
the lateral x-ray source, with the end of the beam-limiting device or
spacer no closer than 15 cm to the centerline of the x-ray table.
(6)
For fluoroscopic systems not specifically addressed in 6.5.5.4(1) through
(5) above, the RMP shall determine the measurement point(s)
representing the highest expected dose rate and which is based on
nationally accepted standards and practices.
6.5.6 Indication of potential and current.
6.5.6.1 During fluoroscopy and cinefluorography, x-ray tube potential and current shall be
continuously indicated.
6.5.6.2 Deviation of x-ray tube potential and current from the indicated value shall not exceed the
maximum deviation as stated by the manufacturer.
6.5.7
Source-skin distance.
6.5.7.1 Means shall be provided:
(1)
To limit the source-skin distance to not less than 38 cm on stationary
fluoroscopes and to not less than 30 cm on mobile and portable fluoroscopes.
(2)
In addition, for fluoroscopes intended for specific surgical or interventional
applications that would be prohibited at the source-skin distances specified in
6.5.7.1(1), provisions may be made for operating at shorter source-skin distances
but in no case less than 20 cm, provided a process for such use is justified and
documented in the facility procedures and is periodically reviewed by the FGI
procedures committee or RMP.
6.5.7.2. For stationary, mobile, or portable C-arm fluoroscopic systems manufactured on or after
June 10, 2006, having a maximum source-image receptor distance of less than 45 cm,
means shall be provided to limit the source-skin distance to not less than 19 cm.
(1)
Such systems shall be labeled for extremity use only; and
(2)
For those systems intended for specific surgical or interventional applications that
would be prohibited at the source-skin distance specified in 6.5.7.2, provisions
may be made for operation at shorter source-skin distances but in no case less
than 10 cm, provided a process for such use is justified and documented in the
facility procedures and is periodically reviewed by the FGI procedures committee
or RMP.
6.5.8
Fluoroscopic irradiation time, display, and signal.
6.5.8.1 Fluoroscopic equipment manufactured before June 10, 2006:
(1)
Shall be provided with means to preset the cumulative irradiation time of the
fluoroscopic tube.
(a)
The maximum cumulative time of the timing device shall not exceed 5
minutes without resetting.
(b)
A signal audible to the fluoroscopist shall indicate the completion of any
preset cumulative irradiation time.
(c)
Such signal shall continue to sound while x-rays are produced until the
timing device is reset.
(d)
Fluoroscopic equipment may be modified in accordance with 21 CFR
1020.30(q) to comply with the requirements of 6.5.8.1.
(e)
When the equipment is modified, it shall bear a label indicating the
statement:
MODIFIED TO COMPLY WITH 21 CFR 1020.32(H)(2)
(2)
As an alternative to the requirements of 6.5.8.1, radiation therapy simulation
systems may be provided with a means to indicate the total cumulative exposure
time during which x-rays were produced, and which is capable of being reset
between x-ray examinations.
6.5.8.2. For x-ray controls manufactured on or after June 10, 2006, there shall be provided for
each fluoroscopic tube:
(1)
A display of the fluoroscopic irradiation time at the fluoroscopist’s working
position. This display shall function independently of the audible signal described
in 6.5.8.2(2). The following requirements apply:
(a)
When the x-ray tube is activated, the fluoroscopic irradiation time in
minutes and tenths of minutes shall be continuously displayed and
updated at least once every 6 seconds.
(b)
The fluoroscopic irradiation time shall also be displayed within 6 seconds
of termination of an exposure and remain displayed until reset.
(c)
Means shall be provided to reset the display to zero prior to the
beginning of a new examination or procedure.
(2)
A signal audible to the fluoroscopist shall sound for each passage of 5 minutes of
fluoroscopic irradiation time during an examination or procedure.
(a)
The signal shall sound until manually reset or, if automatically reset, for
at least 2 seconds.
6.5.9
Display of last-image-hold (LIH).
Fluoroscopic equipment manufactured on or after June 10, 2006, shall be equipped with means
to display LIH image following termination of the fluoroscopic exposure.
6.5.9.1 For an LIH image obtained by retaining pretermination fluoroscopic images, if the number
of images and method of combining a predetermined number of images are selectable by
the user, the selection shall be indicated prior to initiation of the fluoroscopic exposure.
6.5.9.2 For an LIH image obtained by initiating a separate radiographic-like exposure at the
termination of fluoroscopic imaging, the technique factors for the LIH image shall be
selectable prior to the fluoroscopic exposure, and the combination selected shall be
indicated prior to initiation of the fluoroscopic exposure.
6.5.9.3 Means shall be provided to clearly indicate to the user whether a displayed image is the
LIH radiograph or fluoroscopy. Display of the LIH radiograph shall be replaced by the
fluoroscopic image concurrently with re-initiation of fluoroscopic exposure, unless
separate displays are provided for the LIH radiograph and fluoroscopic images.
6.5.10 Displays of values of AKR and cumulative air kerma.
Fluoroscopic equipment manufactured on or after June 10, 2006, shall display at the fluoroscopist’s
working position the AKR and cumulative air kerma. The following requirements apply for each x-ray tube
used during an examination or procedure:
6.5.10.1
When the x-ray tube is activated and the number of images produced per unit
time is greater than six images per second, the AKR in mGy/min shall be continuously
displayed and updated at least once every second.
6.5.10.2
The cumulative air kerma in units of mGy shall be displayed either within 5
seconds of termination of an exposure or displayed continuously and updated at least
once every 5 seconds.
6.5.10.3
The display of the AKR shall be clearly distinguishable from the display of the
cumulative air kerma.
6.5.10.4
The AKR and cumulative air kerma shall represent the value for conditions of
free-in- air irradiation at one of the following reference locations specified according to the
type of fluoroscope.
(1)
For fluoroscopes with x-ray source below the x-ray table, x-ray source above the
table, or of fixed lateral type, the reference location shall be the respective
locations specified in 6.5.5.4(1). 6.5.5.4(2), or 6.5.5.4(5).
(2)
For C-arm fluoroscopes, the reference location shall be 15 cm from the isocenter
toward the x-ray source along the beam axis. Alternatively, the reference location
shall be at a point specified by the manufacturer to represent the location of the
intersection of the x-ray beam with the patient’s skin.
6.5.10.5
Means shall be provided to reset to zero the display of cumulative air kerma prior
to the commencement of a new examination or procedure.
6.5.10.6
The displayed AKR and cumulative air kerma shall not deviate from the actual
values by more than ±35 percent over the range of 6 mGy/min and 100 mGy to the
maximum indication of AKR and cumulative air kerma, respectively.
(1)
Compliance shall be determined with an irradiation time greater than 3 seconds.
6.5.11 Protection from scatter radiation.
6.5.11.1
For stationary fluoroscopic systems, ancillary shielding, such as drapes, self-
supporting curtains, or viewing shields, shall be available and used as supplemental
protection for all individuals other than the patient in the room during a fluoroscopy
procedure.
6.5.11.2
Where sterile fields or special procedures prohibit the use of normal protective
barriers or drapes, all of the following conditions shall be met:
(1)
Shielding required under 6.5.11.1 shall be maintained to the degree possible
under clinical conditions;
(2)
All persons, except the patient, in the room where fluoroscopy is performed shall
wear protective apparel (aprons) or shall be positioned behind a stationary or
portable shield that provides a lead equivalent shielding of at least 0.25mm;
(3)
The fluoroscopic field size shall be reduced to the minimum required for the
procedure being performed (area of clinical interest); and
(4)
Operating and safety procedures shall reflect the above conditions, and
fluoroscopy personnel shall exhibit awareness of situations requiring the use
and/or non-use of the protective drapes.
6.5.12 Fluoroscopy specific operator qualifications
6.5.12.1
Operation of a fluoroscopic x-ray system shall be performed under direct
supervision, except where otherwise specified in these regulations.
6.5.12.2
In addition to the applicable sections of these regulations, all persons operating
or supervising the operation of a fluoroscopic x-ray system (including for FGI procedures)
for clinical purposes on living humans shall be limited to persons meeting the applicable
requirements of 6.3.1.6, 6.3.1.9, and Part 2, Section 2.4.5.5, and 2.6.1.5.
6.5.13 Equipment operation
6.5.13.1
Overhead fluoroscopy shall not be used as a positioning tool for general purpose
radiographic examinations.
6.5.13.2
Operators shall be instructed in accordance with Part 2 requirements.
6.5.13.3
Procedure planning for fluoroscopic procedures on pregnant patients shall
include feasible modifications to minimize the dose to the conceptus.
6.5.13.4
Procedure planning for fluoroscopic procedures on pediatric patients shall
include feasible modifications to minimize dose.
6.5.13.5
The facility shall establish a written policy regarding patient dose management in
fluoroscopically guided procedures in conformance with the ACR-AAPM Technical
Standard for Management of the Use of Radiation in Fluoroscopic Procedures (ACR
Resolution 44-2013), NCRP Report 168, or equivalent.
(1)
Consistent with facility policy and procedures, the operator shall use methods
available on the fluoroscopy system to monitor dose during a fluoroscopic
procedure.
(2)
The written policy shall include a requirement to designate a person in the room
to notify the operator that a SRDL or other dose metric value specified in the
facility policy is approaching or has been exceeded.
6.5.14 Registered Medical Physicist evaluations of fluoroscopic equipment.
6.5.14.1
Fluoroscopic x-ray systems shall have a certification evaluation performed by a
RMP under the frequency and conditions specified in Part 2, Section 2.5.
At a minimum these evaluations shall include:
(1)
A measurement of entrance exposure rates that covers a representative sample
of patient thicknesses, including those that are expected to drive the system to
maximum output in all modes clinically used, including fluoroscopy, high-level
control, and acquisition, when available. These measurements shall:
(a)
For systems without automatic exposure control, be made utilizing a
milliamperage and kVp typical of the clinical use of the fluoroscopic
system;
(b)
For systems with automatic exposure control, be made utilizing sufficient
attenuating material in the useful beam to produce a milliamperage and
kVp typical of the clinical use of the fluoroscopic system;
(2)
A measurement and verification of compliance of maximum AKR for fluoroscopy
and high-level control, if available. Measurements shall be made in accordance
with Section 6.5.5.4.
(3)
An evaluation of image quality in the modes necessary to achieve the clinical
imaging task(s).
(4)
An evaluation of the operation of the 5-minute timer, warning lights, interlocks,
and collision sensors.
(5)
An evaluation of the beam quality and collimation in the fluoroscopy mode.
Additional evaluation may be needed where magnification impacts collimation.
(6)
An evaluation of the availability and accuracy of technique indicators and
integrated radiation dose displays.
(7)
An evaluation of changes to the fluoroscopy system impacting radiation safety.
(8)
When operating in the spot image mode, an evaluation of the coefficient of
variation of air kerma for both manual and automatic exposure control systems to
ensure the value does not exceed 0.05.
6.5.14.2
Measurements required in 6.5.14.1 shall be:
(1)
Performed in accordance with manufacturer recommendations or nationally
accepted standards using a calibrated dosimetry system;
(2)
Dosimetry systems used for measurements shall be calibrated in accordance
with manufacturer recommendations or nationally accepted standards not to
exceed 2 years.
(3)
Records indicating the model, serial number and calibration date of equipment
used for dosimetry calibrations on FGI systems shall be maintained for 3 years
for inspection by the Department.
6.5.15 Additional requirements for facilities performing fluoroscopically-guided interventional (FGI)
procedures.
The requirements of 6.5.15 and other requirements associated with an FGI Procedure Committee
in 6.5.15 shall become effective on or after January 1, 2022.
6.5.15.1
A registrant performing FGI procedures shall establish a FGI Procedure
Committee in accordance with the following:
(1)
The registrant may establish a system-wide committee if the registrant has more
than one site;
(2)
Two or more registrants may form a cooperative FGI Procedure Committee as
long as each facility has a representative on the committee; and
(3)
If the registrant has already established a radiation safety committee, the
requirements of 6.5.15 may be delegated to that committee if the members meet
the requirements of 6.5.15.5.
6.5.15.2
At a minimum the FGI Procedure Committee members in 6.5.15.5(1) through (3)
shall meet as often as necessary to conduct business, but at intervals not to exceed 12
months.
6.5.15.3
A record of each FGI Procedure Committee meeting shall include the date,
names of individuals in attendance, minutes of the meeting, and any actions taken. The
registrant shall maintain the record for 3 years for inspection by the Department.
6.5.15.4
Provide an annual report summarizing the details and activities of the FGI
Procedure Committee to the radiation safety committee or radiation safety officer, in the
absence of a radiation safety committee.
6.5.15.5
FGI Procedure Committee members shall include but not be limited to the
following individuals involved in FGI procedures:
(1)
A supervising physician of the healing arts who meet the requirements in
6.3.1.6(1);
(2)
A Registered Medical Physicist;
(3)
A technologist, where applicable
(4)
A licensed individual who meets the requirements of 6.3.1.6(2), where applicable;
and
(5)
Other individuals as deemed necessary by the registrant.
6.5.15.6
Establish and implement FGI procedures
(1)
The registrant shall establish and implement written procedures, or procedures
documented in an electronic recordkeeping system, that include but are not
limited to the following:
(a)
Identification of individuals who are authorized to use fluoroscopic
systems for interventional purposes.
(b)
Methods for patient radiation dose management during FGI procedures.
(c)
Establishing dose metric notification levels for fluoroscopy procedures at
which point the physician, or other authorized operator is notified.
(d)
SRDL values following nationally recognized standards
(e)
Actions to be taken for cases when a SRDL is exceeded which may
include patient follow-up.
(f)
A review of the established processes and procedures at an interval not
to exceed 12 months.
(2)
A record of each procedure developed by the registrant shall be maintained for
inspection by the Department. If the registrant revises a procedure,
documentation shall be maintained that includes the justification for the revision
and the previous procedure for inspection by the Department.
(3)
The FGI Procedure Committee shall review and approve the procedures
developed or modified under 6.5.15.6.
6.5.16 Procedures for maintaining records for fluoroscopic systems
6.5.16.1
A record of radiation output information shall be maintained in the event a dose
reconstruction calculation or estimate is necessary in accordance with established
procedures. The record shall include the following:
(1)
Operator identification;
(2)
Patient identification;
(3)
Type and date of examination;
(4)
Identification of the fluoroscopic system used; and
(5)
Peak skin dose, cumulative air kerma or dose area product used, beam entry
angle(s), and patient position if the information is available on the fluoroscopic
system.
(6)
If the peak skin dose, cumulative air kerma or dose area product are not
displayed on the fluoroscopic system, records shall include other available
information in the event a dose reconstruction calculation or estimate is
necessary in accordance with established procedure or the following as
necessary:
(a)
Fluoroscopic mode, such as, high-level or pulsed mode of operation;
(b)
Cumulative fluoroscopic exposure time; and
(c)
Number of films or recorded exposures.
(7)
The registrant shall maintain records required by 6.5.16.1 for inspection by the
Department for 3 years.
6.6
Requirements for use of general purpose x-ray imaging systems
6.6.1
Administrative controls.
6.6.1.1 The requirements of Section 6.6 apply to all registrants using general diagnostic imaging
systems, excluding the following:
(1)
Fluoroscopy use which is described in 6.5;
(2)
Dental use which is described in 6.7;
(3)
Veterinary use which is described in 6.8;
(4)
Computed tomography use which is described in 6.9;
(5)
Mammography use which is described in 6.10.
6.6.1.2 Certification evaluation (inspection) requirements.
(1)
Within 90 days of initial installation:
(a)
Digital radiographic systems shall have an initial certification evaluation
performed by a RMP;
(b)
Non-digital radiographic systems shall have an initial certification
evaluation performed by a Qualified Inspector authorized for the specific
machine type.
(2)
Periodic certification evaluations shall be performed at the frequency specified in
Part 2, Section 2.5 by Qualified Inspectors authorized for the specific machine
type.
(3)
Testing of display monitors which are under the control of the registrant shall be
performed by or under the supervision of an RMP in accordance with 6.3.5.6.
(4)
Certification evaluations and testing shall follow nationally accepted standards or
those recognized by the Department.
6.6.2
Field limitation and alignment for mobile, portable, and stationary general purpose x-ray systems.
Mobile, portable, and stationary general purpose radiographic x-ray systems shall meet the
following requirements:
6.6.2.1 Variable x-ray field limitation. A means for stepless adjustment of the size of the x-ray
field shall be provided.
(1)
Each dimension of the minimum field size at an SID of 100 cm shall be equal to
or less than 5 cm.
6.6.2.2 Visual definition. Means for visually defining the perimeter of the x-ray field shall be
provided.
(1)
The total misalignment of the edges of the visually defined field with the
respective edges of the x-ray field along either the length or width of the visually
defined field shall not exceed two (2) percent of the distance from the source to
the center of the visually defined field when the surface upon which it appears is
perpendicular to the axis of the x-ray beam.
(2)
A light localizer used to define the x-ray field of a certified system shall provide
illumination sufficient to permit visual determination of the x-ray field under
ambient light conditions of up to 500 lux (46 foot candles).
6.6.2.3 Field indication and alignment on stationary general purpose x-ray equipment. Except
when spot-image devices are in use, stationary general purpose x-ray systems shall
meet the following requirements in addition to those prescribed in 6.6.2.
(1)
Means shall be provided to:
(a)
Indicate when the axis of the x-ray beam is perpendicular to the plane of
the image receptor;
(b)
Align the center of the x-ray field with respect to the center of the image
receptor to within two (2) percent of the SID; and
(c)
Indicate the SID to within two (2) percent.
(2)
The beam-limiting device shall indicate numerically the field size in the plane of
the image receptor to which it is adjusted.
(3)
Indication of field size dimensions and SID's shall be specified in inches and/or
cm, and shall be such that aperture adjustments result in x-ray field dimensions
in the plane of the image receptor that correspond to those indicated by the
beam-limiting device to within two (2) percent of the SID when the beam axis is
indicated to be perpendicular to the plane of the image receptor.
(4)
Compliance measurements will be made at discrete SIDs and image receptor
dimensions in common clinical use or at any other specific dimensions at which
the beam-limiting device or its associated diagnostic x-ray system is uniquely
designed to operate.
6.6.2.4 Field limitation on x-ray equipment other than general purpose radiographic systems.
(1)
X-Ray Systems Designed for One Image Receptor Size.
(a)
Radiographic equipment designed for only one image receptor size at a
fixed SID shall be provided with means to limit the field at the plane of
the image receptor to dimensions no greater than those of the image
receptor, and to align the center of the x-ray field with the center of the
image receptor to within two (2) percent of the SID; or
(b)
Radiographic equipment designed for only one image receptor size at a
fixed SID shall be provided with means to both size and align the x-ray
field such that the x-ray field at the plane of the image receptor does not
extend beyond any edge of the image receptor.
6.6.2.5 Beam Limitation Requirements for Each X-Ray System Not Governed by 6.6.2.1 through
6.6.2.4:
(1)
Which are also designed for use with extraoral image receptors and when used
with an extraoral image receptor shall:
(a)
Be provided with means to limit the x-ray field in the plane of the image
receptor so that such field does not exceed each dimension of the image
receptor by more than two (2) percent of the SID when the axis of the x-
ray beam is perpendicular to the plane of the image receptor; and
(b)
Be provided with means to align the center of the x-ray field with the
center of the image receptor to within two (2) percent of the SID, or
means shall be provided to both size and align the x-ray field such that
the x-ray field at the plane of the image receptor does not extend beyond
any edge of the image receptor.
(2)
The requirements of 6.6.2.5(1) may be met with a system that meets the
requirements for a general purpose x-ray system as specified in 6.6.2 and
6.6.2.3, or, when alignment means are also provided, may be met with either:
(a)
An assortment of removable, fixed-aperture, beam-limiting devices
sufficient to meet the requirement for each combination of image
receptor size and SID for which the unit is designed with each such
device having clear and permanent markings to indicate the image
receptor size and SID for which it is designed; or
(b)
A beam-limiting device having multiple fixed apertures sufficient to meet
the requirement for each combination of image receptor size and SID for
which the unit is designed. Permanent, clearly legible markings shall
indicate the image receptor size and SID for which each aperture is
designed and shall indicate which aperture is in position for use.
6.6.2.6 Positive Beam Limitation (PBL). The requirements of 6.6.2.6 shall apply to radiographic
systems which contain PBL.
(1)
Field size. When a PBL system is provided, it shall prevent x-ray production
when:
(a)
Either the length or width of the x-ray field in the plane of the image
receptor differs from the corresponding image receptor dimension by
more than three (3) percent of the SID; or
(b)
The sum of the length and width differences as stated in 6.6.2.6(1)(a)
without regard to sign exceeds four (4) percent of the SID.
(c)
The beam-limiting device is at a SID for which PBL is not designed for
sizing.
(2)
Conditions for PBL. When provided, the PBL system shall function as described
in 6.6.2.6(1) whenever all the following conditions are met:
(a)
The image receptor is inserted into a permanently mounted cassette
holder;
(b)
The image receptor length and width are less than 50 cm;
(c)
The x-ray beam axis is within ± three (3) degrees of vertical and the SID
is 90 cm to 130 cm inclusive; or the x-ray beam axis is within ± three (3)
degrees of horizontal and the SID is 90 cm to 205 cm inclusive;
(d)
The x-ray beam axis is perpendicular to the plane of the image receptor
to within ± three (3) degrees;
(e)
Neither tomographic nor stereoscopic radiography is being performed;
(3)
Measuring compliance.
(a)
Compliance with the requirements of 6.6.2.6(1) shall be determined
when the equipment indicates that the beam axis is perpendicular to the
plane of the image receptor and the provisions of 6.6.2.6(2) are met; and
(b)
Compliance shall be determined no sooner than 5 seconds after
insertion of the image receptor.
(4)
Override of PBL.
(a)
A capability may be provided for overriding PBL in case of system failure
and for servicing the system.
(b)
This override may be for all SIDs and image receptor sizes.
(c)
A key shall be required for any override capability that is accessible to
the operator.
(i)
It shall not be possible to remove the key while PBL is
overridden.
(ii)
Each such key switch or key shall be clearly and durably labeled
as follows:
“FOR X-RAY FIELD LIMITATION SYSTEM FAILURE”
(d)
The override capability is considered accessible to the operator:
(i)
If it is referenced in the operator's manual, or in other materials
intended for the operator; or
(ii)
If its location is such that the operator would consider it part of
the operational controls.
(e)
Not be used as a substitute for prompt repair.
(5)
Operator initiated undersizing. The PBL system shall be capable of operating
such that, at the discretion of the operator, the size of the field may be made
smaller than the size of the image receptor through stepless adjustment of the
field size.
(a)
Each dimension of the minimum field size at an SID of 100 cm shall be
equal to or less than 5 cm.
(b)
Return to PBL function as described in 6.6.2.6(1) shall occur
automatically upon any change of image receptor size or SID.
(6)
Disabling of PBL. A facility has the option to permanently functionally disable a
PBL system. When this option is chosen, the standards for manual collimation
apply.
6.6.3
Radiation Exposure Control.
6.6.3.1 Exposure initiation
(1)
Means shall be provided to initiate the radiation exposure by a deliberate action
on the part of the operator, such as the depression of a switch. Radiation
exposure shall not be initiated without such an action.
(2)
In addition, it shall not be possible to initiate an exposure when the timer is set to
a "zero" or "off" position if either position is provided.
6.6.3.2 Exposure Indication
(1)
Means shall be provided for visual indication observable at or from the operator’s
protected position whenever x-rays are produced.
(2)
In addition, a signal audible to the operator shall indicate that the exposure has
terminated.
6.6.3.3 Timers.
(1)
Means shall be provided to terminate the exposure at a preset time interval,
preset product of current and time, a preset number of pulses, or a preset
radiation exposure to the image receptor.
(a)
Except during serial radiography, the operator shall be able to terminate
the exposure at any time during an exposure of greater than one-half
second.
(b)
Except during panoramic dental radiography, termination of exposure
shall cause automatic resetting of the timer to its initial setting or to
“zero”.
(c)
During serial radiography, the operator shall be able to terminate the x-
ray exposure(s) at any time, but means may be provided to permit
completion of any single exposure of the series in process.
6.6.3.4 X-ray Control.
(1)
Except for only bone densitometry systems, hand-held intraoral systems, each x-
ray control shall be located in such a way as to meet the following requirements:
(a)
Stationary radiographic systems.
Stationary radiographic systems shall be required to have the x-ray
control, including the exposure switch, permanently mounted in a
protected (shielded) area so that the operator is required to remain in
that protected area during the entire exposure. Design of the operator
protected area shall be consistent with Appendix 6B.
(b)
Mobile and portable systems.
When any one or combination of mobile or portable x-ray systems are:
(i)
Used daily for seven (7) or more consecutive working days in the
same location (a room or area), the system(s) shall meet the
requirements of a stationary system in 6.6.3.4(1)(a), or the
facility shall employ the use of at least one of the items in
6.6.3.4(b)(ii) and establish a written procedure or policy
prescribing any limitations necessary to demonstrate that such
use will preclude any individual from receiving a dose in excess
of the public or occupational dose limits in Part 4 and that such
use is consistent with the As Low As Reasonably Achievable
(ALARA) concept of Part 4, Section 4.5.2;
(ii)
Used daily for less than seven (7) consecutive working days in
the same location (a room or area), shall be provided with at
least one of the following:
1.
A lead-equivalent protective barrier at least 2 meters
(more than 6 feet) high for operator protection during
exposures; or
2.
Means to allow the operator to be at least 2 meters
(more than 6 feet) from the patient, x-ray tube and the
useful beam during the exposure; or
3.
A lead-equivalent protective garment with thyroid
shielding.
(c)
Podiatry facilities shall meet the protection requirements in
6.6.3.4(1)(b)(ii).
(2)
For x-ray equipment capable of displaying technique factors, the technique
factors to be used during an exposure shall be indicated before the exposure
begins.
(a)
When automatic exposure controls are used, the exposure settings that
are set prior to the exposure shall be indicated.
(b)
On equipment having fixed technique factors, the requirement of
6.6.3.4(2)(a) may be met by having permanent markings on the
equipment. Technique factors shall be visible from the operator’s position
except when performing spot imaging during fluoroscopy.
(c)
The accuracy of the indicated kilovoltage peak (kVp) shall meet
manufacturer specifications. In the absence of a manufacturer
specification, kVp accuracy shall be within ±10 percent.
6.6.3.5 Automatic Exposure Controls. When an automatic exposure control is provided:
(1)
Indication shall be made on the control panel when this mode of operation is
selected;
(2)
When the x-ray tube potential is equal to or greater than 51 kilovolts peak (kVp),
the minimum exposure time for field emission equipment rated for pulsed
operation

[Text truncated at 120,000 characters. The full text is on the page linked above.]

## Nearby sections

- [6 CCR 1007-1 RADIATION CONTROL – (Rule Versions Prior to 04/01/2007)](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1.md)
- [6 CCR 1007-1 Part 01 RADIATION CONTROL - GENERAL PROVISIONS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_01.md)
- [6 CCR 1007-1 Part 02 RADIATION CONTROL - REGISTRATION OF RADIATION MACHINES, FACILITIES AND SERVICES](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_02.md)
- [6 CCR 1007-1 Part 03 RADIATION CONTROL - LICENSING OF RADIOACTIVE MATERIAL](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_03.md)
- [6 CCR 1007-1 Part 04 RADIATION CONTROL - STANDARDS FOR PROTECTION AGAINST RADIATION](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_04.md)
- [6 CCR 1007-1 Part 05 RADIATION CONTROL - RADIATION SAFETY REQUIREMENTS FOR INDUSTRIAL RADIOGRAPHIC OPERATIONS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_05.md)
- [6 CCR 1007-1 Part 06 RADIATION CONTROL - X-RAY IMAGING IN THE HEALING ARTS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_06.md)
- [6 CCR 1007-1 Part 07 RADIATION CONTROL - USE OF RADIONUCLIDES IN THE HEALING ARTS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_07.md)
- [6 CCR 1007-1 Part 08 RADIATION CONTROL - RADIATION SAFETY REQUIREMENTS FOR RADIATION GENERATING DEVICES NOT USED IN THE HEALING ARTS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_08.md)
- [6 CCR 1007-1 Part 09 RADIATION CONTROL - RADIATION SAFETY REQUIREMENTS FOR PARTICLE ACCELERATORS NOT USED IN THE HEALING ARTS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_09.md)
- [6 CCR 1007-1 Part 10 RADIATION CONTROL - NOTICES, INSTRUCTIONS, AND REPORTS TO WORKERS: INSPECTIONS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_10.md)
- [6 CCR 1007-1 Part 11 RADIATION CONTROL - SPECIAL LAND OWNERSHIP REQUIREMENTS GENERAL PROVISIONS](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_11.md)
- [6 CCR 1007-1 Part 12 RADIATION CONTROL - FEES FOR RADIATION CONTROL SERVICES](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_12.md)
- [6 CCR 1007-1 Part 13 RADIATION CONTROL - COMPLIANCE ENFORCEMENT](https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_13.md)

---

Source: Frix Law Library, https://www.frixlaw.com/law-library/statutes/STATE_CO_CCR_6_CCR_1007_1_Part_06. Check the current official text before relying on it. Not legal advice.