# EPA Method 23-Determination of Polychlorinated Dibenzo-p-Dioxins and Polychlorinated Dibenzofurans From Stationary Sources

> Briefs, arguments, decisions, and more.

URL: https://www.frixlaw.com/law-library/documents/fr%3A2023-04958

## Record

- **Collection:** Federal Register
- **Document type:** Rule
- **Published:** March 20, 2023
- **Citation:** 88 FR 16732

## Text

ENVIRONMENTAL PROTECTION AGENCY
40 CFR Parts 60, 63, and 266
[EPA-HQ-OAR-2016-0677; FRL-5937-02-OAR]
RIN 2060-AT09
EPA Method 23—Determination of Polychlorinated Dibenzo-p-Dioxins and Polychlorinated Dibenzofurans From Stationary Sources

AGENCY:

Environmental Protection Agency (EPA).

ACTION:

Final rule.

SUMMARY:

This action finalizes editorial and technical revisions to the Environmental Protection Agency's (EPA's) Method 23 (Determination of Polychlorinated Dibenzo-
p
-Dioxins, Polychlorinated Dibenzofurans, and Polycyclic Aromatic Hydrocarbons from Stationary Sources). Final revisions include incorporating true, comprehensive, and stable isotope dilution for quantifying target compounds using corresponding carbon-13 labeled compounds for each target compound including most of the polycyclic aromatic hydrocarbons (PAH) and changing the method quality control from the current prescriptive format to a more flexible performance-based approach with specified performance criteria. We are also finalizing revisions that expand the list of target compounds of Method 23 to include PAH and polychlorinated biphenyls (PCB). The final revisions allow facilities and their test teams flexibility when sampling and measuring polychlorinated dibenzo-
p
-dioxins and polychlorinated dibenzofurans (PCDD/PCDF), PAH, and PCB from stationary sources with a comprehensive isotope dilution method while ensuring that the stack testing community can consistently implement the method across emissions sources and facilities.

DATES:

This final rule is effective on March 20, 2023. The incorporation by reference (IBR) of certain publications listed in the rule is approved by the Director of the Federal Register as of March 20, 2023.

ADDRESSES:

The U.S. Environmental Protection Agency (EPA) has established a docket for this action under Docket ID No. EPA-HQ-OAR-2016-0677. All documents in the docket are listed on the
https://www.regulations.gov
website. Although listed, some information is not publicly available,
e.g.,
Confidential Business Information or other information whose disclosure is restricted by statute. Certain other material, such as copyrighted material, is not placed on the internet and will be publicly available only in hard copy form. Publicly available docket materials are available either electronically through
https://www.regulations.gov
or in hard copy at the EPA Docket Center, WJC West Building, Room 3334, 1301 Constitution Avenue NW, Washington, DC 20004. Out of an abundance of caution for members of the public and our staff, the EPA Docket Center and Reading Room are closed to the public, with limited exceptions, to reduce the risk of transmitting Coronavirus 2019 (COVID-19). Our Docket Center staff will continue to provide remote customer service via email, phone, and webform.

FOR FURTHER INFORMATION CONTACT:

For further questions about this final action, contact Dr. Raymond Merrill, Office of Air Quality Planning and Standards (OAQPS), Air Quality Assessment Division (AQAD), Environmental Protection Agency, Research Triangle Park, NC 27711; mail drop E143-02; telephone number: (919) 541-5225; fax number: (919) 541-0516; email address:
merrill.raymond@epa.gov.

SUPPLEMENTARY INFORMATION:

Preamble acronyms and abbreviations.
We use multiple acronyms in this preamble. While this list may not be exhaustive, to ease the reading of this preamble and for reference purposes, the EPA defines the following terms and acronyms here:

AQAD Air Quality Assessment Division

ASTM American Society for Testing and Materials International

CAA Clean Air Act

CARB California Environmental Protection Agency Air Resources Board

CCV continuing calibration verification

CFR Code of Federal Regulations

EDL estimated detection limit

EPA U.S. Environmental Protection Agency

FR Federal Register

GC gas chromatograph

HRGC high-resolution gas chromatography

HRMS high-resolution mass spectrometry

IBR incorporation by reference

IDC initial demonstration of capability

MDL method detection limit

MS mass spectrometer

NTTAA National Technology Transfer and Advancement Act

OAQPS Office of Air Quality Planning and Standards

OLEM Office of Land and Emergency Management

OMB Office of Management and Budget

OW Office of Water

PAH polycyclic aromatic hydrocarbons

PCB polychlorinated biphenyls

PCDD polychlorinated dibenzo-
p
-dioxins

PCDPE polychlorinated diphenyl ethers

PCDPF polychlorinated dibenzofurans

PRA Paperwork Reduction Act

QCS Quality Control Sample

RFA Regulatory Flexibility Act

RRF relative response factor

SVOC semivolatile organic compounds

SW solid waste

TTN Technology Transfer Network

UMRA Unfunded Mandates Reform Act

Organization of this document.
The information in this preamble is organized as follows:

I. General Information

A. Does this final action apply to me?

B. Where can I get a copy of this document and other related information?

C. Judicial Review

II. Background

III. Incorporation by Reference

IV. Summary of Revisions to Method 23

A. Section 1.0 Scope and Application

B. Section 2.0 Summary of Method

C. Section 3.0 Definitions

D. Section 4.0 Interferences

E. Section 5.0 Safety

F. Section 6.0 Equipment and Supplies

G. Section 7.0 Reagents, Media, and Standards

H. Section 8.0 Sample Collection, Preservation, and Storage

I. Section 9.0 Quality Control

J. Section 10.0 Calibration and Standardization

K. Section 11.0 Analysis Procedure

L. Section 12.0 Data Analysis and Calculations

M. Section 13.0 Method Performance

N. Section 14.0 Pollution Prevention

O. Section 15.0 Waste Management

P. Section 16.0 Bibliography

Q. Section 17.0 Tables, Diagrams, Flow Charts, and Validation Data

V. Summary of Final Revisions Related to 40 CFR Parts 60, 63, and 266

A. 40 CFR Part 60—Standards of Performance for New Stationary Sources

B. 40 CFR Part 63—National Emission Standards for Hazardous Air Pollutants for Source Categories

C. 40 CFR Part 266—Standards for the Management of Specific Hazardous Wastes and Specific Types of Hazardous Waste Management Facilities

VI. Statutory and Executive Order Reviews

A. Executive Order 12866: Regulatory Planning and Review and Executive Order 13563: Improving Regulation and Regulatory Review

B. Paperwork Reduction Act (PRA)

C. Regulatory Flexibility Act (RFA)

D. Unfunded Mandates Reform Act (UMRA)

E. Executive Order 13132: Federalism

F. Executive Order 13175: Consultation and Coordination With Indian Tribal Governments

G. Executive Order 13045: Protection of Children From Environmental Health Risks and Safety Risks

H. Executive Order 13211: Actions That Significantly Affect Energy Supply, Distribution, or Use

I. National Technology Transfer and Advancement Act (NTTAA)

J. Executive Order 12898: Federal Actions To Address Environmental Justice in Minority Populations and Low-Income Populations

K. Congressional Review Act (CRA)

L. Determination Under Clean Air Act Section 307(d)

I. General Information

A. Does this final action apply to me?

The final amendments to Method 23 apply to stationary sources that are subject to certain provisions of 40 CFR parts 60, 62, 63, 79, and 266. The source categories and entities potentially affected are listed in Table 1 of this preamble. This table is not intended to be exhaustive, but rather provides a guide for readers regarding entities likely to be affected by this action. This table lists the types of entities that EPA is now aware could potentially be affected by this action. Other types of entities not listed in the table could also be affected.

Table 1—Potentially Affected Source Categories

Category

NAICS
a

Examples of regulated entities

Industry
332410
Fossil fuel steam generators.

332410
Industrial, commercial, institutional steam generating units.

562213
Municipal Waste Combustors.

322110
Hazardous Waste Combustors.

325211
Polyvinyl Chloride Resins Manufacturing.

327310
Portland cement plants.

324122
Asphalt Shingle and Coating Materials Manufacturing.

331314
Secondary aluminum plants.

327120
Clay Building Material and Refractories Manufacturing.

331410
Nonferrous Metal (except Aluminum) Smelting and Refining.

a
North American Industry Classification System.

If you have any questions regarding the applicability of the final changes to Method 23, contact the person listed in the preceding
FOR FURTHER INFORMATION CONTACT
section.

B. Where can I get a copy of this document and other related information?

The docket number for this action is Docket ID No. EPA-HQ-OAR-2016-0677. In addition to being available in the docket, an electronic copy of the final method revisions is available on the Technology Transfer Network (TTN) website at
https://www.epa.gov/ttn/emc/methods/.
The TTN provides information and technology exchange in various areas of air pollution control.

C. Judicial Review

Under Clean Air Act (CAA) section 307(b)(1), judicial review of this final rule is available only by filing a petition for review in the U.S. Court of Appeals for the District of Columbia Circuit by May 19, 2023. Moreover, under section 307(b)(2) of the CAA, the requirements established by this final rule may not be challenged separately in any civil or criminal proceedings brought by the EPA to enforce these requirements. Section 307(d)(7)(B) of the CAA further provides that “[o]nly an objection to a rule or procedure which was raised with reasonable specificity during the period for public comment (including any public hearing) may be raised during judicial review.” This section also provides a mechanism for the EPA to convene a proceeding for reconsideration, “[i]f the person raising an objection can demonstrate to the EPA that it was impracticable to raise such objection within [the period for public comment] or if the grounds for such objection arose after the period for public comment, (but within the time specified for judicial review) and if such objection is of central relevance to the outcome of the rule.” Any person seeking to make such a demonstration should submit a Petition for Reconsideration to the Office of the Administrator, U.S. EPA, Room 3000, WJC South Building, 1200 Pennsylvania Ave. NW, Washington, DC 20460, with a copy to both the person listed in the preceding
FOR FURTHER INFORMATION CONTACT
section, and the Associate General Counsel for the Air and Radiation Law Office, Office of General Counsel (Mail Code 2344A), U.S. EPA, 1200 Pennsylvania Ave. NW, Washington, DC 20460.

II. Background

The EPA's Method 23 (Determination of Polychlorinated Dibenzo-
p
-Dioxins and Polychlorinated Dibenzofurans from Stationary Sources) is EPA's current reference test method used to determine the amount of polychlorinated dibenzo-
p
-dioxins (PCDD) and polychlorinated dibenzofurans (PCDF) emitted from stationary sources.

The EPA promulgated Method 23 (Appendix A of 40 Code of Federal Regulations (CFR) Part 60, Test Methods) on February 13, 1991 (56 FR 5758). Since promulgation, the ability to measure PCDD and PCDF has evolved as analytical laboratories, EPA, and state entities have developed new standard operating procedures and methods to reflect improvements in sampling and analytical techniques. Examples of newer PCDD/PCDF methods include:

• Office of Land and Emergency Management (OLEM) Solid Waste (SW) SW-846 EPA Method 8290A, Polychlorinated Dibenzo-
p
-Dioxins and Polychlorinated Dibenzofurans (PCDF) by High-Resolution Gas Chromatography/High-Resolution Mass Spectrometry (HRGC/HRMS).

• Office of Water (OW) EPA Method 1613, Tetra- through Octa-Chlorinated Dioxins and Furans by Isotope Dilution HRGC/HRMS.

• California Environmental Protection Agency Air Resources Board (CARB) Method 428, Determination of Polychlorinated Dibenzo-
p
-Dioxin (PCDD), Polychlorinated Dibenzofuran (PCDF), and Polychlorinated Biphenyls Emissions from Stationary Sources.

Beginning in 2016, the EPA held a series of informal discussions with stakeholders to identify technical issues related to the sampling and analysis of PCDD and PCDF and potential revisions to Method 23. The stakeholders consisted of a cross section of interested parties including representatives from state regulatory entities, various EPA offices, analytical laboratories, regulated sources, emission testing firms, analytical standards vendors, instrument vendors, and others with experience in sampling and analysis of PCDD and PCDF and with the equipment, materials, and performance of Method 23 and other PCDD/PCDF methods. In the discussions, EPA also sought stakeholder input regarding their experience combining procedures for sampling and analysis of PCDD and PCDF with procedures for sampling and analysis of PAH and PCB emitted from

stationary sources. The docket contains summaries of the stakeholder discussions. EPA proposed editorial and technical revisions to Method 23 on January 14, 2020 (85 FR 2234). EPA received comments on the proposed revisions to the method and has addressed these in a separate response to comments document, the
Summary of Public Comments and Responses for the Proposed EPA Method 23—Determination of Polychlorinated Dibenzo-p-Dioxins and Polychlorinated Dibenzofurans from Stationary Sources.
This final action summarizes the changes made in response to comments.

III. Incorporation by Reference

The EPA is incorporating by reference American Society for Testing and Materials (ASTM) D6911-15 and ASTM D4840-99(2018)e1 in Method 23. ASTM D6911-15 includes a guide for packaging and shipping environmental samples for laboratory analysis and ASTM D4840-99(2018)e1 includes a standard guide for sample chain-of-custody procedures. These standards were developed and adopted by ASTM International and may be obtained from
https://www.astm.org
or from the American Society for Testing and Materials 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959.

IV. Summary of Revisions to Method 23

In this action, we are finalizing technical revisions and editorial changes to clarify and update the requirements and procedures specified in Method 23 and reformatting the method to conform with the current EPA method format (see
https://www.epa.gov/measurements-modeling/method-development#format
). We are also expanding the applicability of Method 23 to include procedures for sampling and analyzing PAH and PCB. In addition, we are finalizing revisions to various sections of the CFR that either require Method 23 or require the analysis of PCDD/PCDF, PAH, or PCB.

Our intent for the final revisions is to ensure that Method 23 is implemented consistently. EPA has updated the method procedures to include many current best practices. We have added flexibility to the method based on meeting quality control requirements.

The primary focus of the final revisions to Method 23 is to change the method from a prescriptive method to a method which allows users to have flexibility in implementing the method (
e.g.,
choice of gas chromatograph (GC) column, the procedures used for sample cleanup) while still meeting performance criteria that the EPA believes are necessary for demonstrating and documenting the quality of the measurements for the target compounds. The final revisions also address concerns over recovery of target compounds from particulate matter by requiring a pre-extraction filter recovery standard procedure and acceptance criteria for the pre-extraction filter recovery standard recovery as a tool to evaluate filter extraction. These new requirements resolve the concerns that led to the criteria in 40 CFR 63.1208 that required Administrator approval prior to use of Method 23 for measurement of PCDD/PCDF.

The EPA's second focus for the final revisions is to modify the method to allow isotope dilution with isotopically labeled compounds for each target compound. Quantitation is based on isotope dilution, moving from nine to 17 labeled compounds for 17 target toxic 2,3,7,8-substituted PCDD/PCDF. These revisions to the method are possible because additional isotopically labeled standards for the target compounds have become available from vendors since the original promulgation of Method 23. The final revisions eliminate biases with recovery correction based on individual corresponding labeled compounds.

The third major focus for the EPA's final revisions to Method 23 is to include options for combining sampling and analysis of PCDD/PCDF with sampling and analysis of PAH and PCB to allow the measurement of these toxic semivolatile organic compounds (SVOC). Therefore, PCB and PAH were added to the list of target compounds measured by Method 23.

The EPA's final amendments to Method 23 in response to public comments are presented below for each section of Method 23. The proposed revisions to sections of Method 23 that EPA is not changing based on public comments are finalized as proposed. A summary of public comments and our responses are provided in a separate response to comments document in the docket for this action.

A. Section 1.0 Scope and Application

In this action, EPA is renaming Section 1.0 from “Applicability and Principle” to “Scope and Application,” and revising the text to expand the target compounds for Method 23 to include PCB and PAH. We are also adding statements that emphasize the need for working knowledge of the EPA Methods 1 through 5 of Appendices A-1, A-2, and A-3 to 40 CFR part 60, isotope dilution, and the use of high-resolution gas chromatography/high resolution mass spectrometry (HRGC/HRMS) when applying Method 23. We are also adding language to specify that Method 23 is performance-based and allows users to modify parts of the method to overcome interferences or to substitute alternative materials and equipment provided that all performance criteria in the method are met.

B. Section 2.0 Summary of Method

The EPA is renaming Section 2.0 from “Apparatus” to “Summary of Method,” and revising Section 2.0 to provide an overview of the method's sampling and analytical procedures. We are also moving the current language in Section 2.0, which describes the materials needed to conduct Method 23, to a new Section 6.0 (Equipment and Supplies).

C. Section 3.0 Definitions

The current version of Method 23 does not include definitions of key terms and variables used in Method 23. In this action, we are adding a new Section 3.0 titled “Definitions.” We are defining acronyms and technical terms to improve the clarity of the method principles and procedures. We are also moving language from the current Section 3.0 to a new Section 7.0 (Reagents, Media, and Standards).

D. Section 4.0 Interferences

The current version of Method 23 does not discuss the conditions that can potentially interfere with measurements obtained using the method. In this action, we are adding a new Section 4.0 titled “Interferences,” that presents the potential causes and recommendations for avoiding or mitigating interferences or sample contamination. We are stating that enhanced selectivity, or confidence in the data, is based on the fractionation, GC separation, HRMS sensitivity, and monitoring for polychlorinated diphenyl ether (PCDPE) interferences. We are also moving language from the current Section 4.0 to a new Section 8.0 (Sample Collection, Preservation, and Storage).

E. Section 5 Safety

Currently, Method 23 does not provide procedures for safety. In this action, we are adding a new Section 5.0 titled “Safety,” that presents the health hazards and procedures for minimizing risks to field and laboratory personnel when conducting Method 23. We are also moving language from the current Section 5.0 to a new Section 11.0 (Analysis Procedure).

F. Section 6.0 Equipment and Supplies

In this action, we are renumbering and moving the current language in Section 2.0 (Apparatus) to a new

Section 6.0 titled “Equipment and Supplies,” and making clarifying edits and technical revisions to the specifications in Section 6.0. Table 2 of this preamble identifies the new numbering for the subsections currently in Section 2.0 and Table 3 of this preamble identifies new specifications (and the associated subsection) we are including in Section 6.0.

Table 2—Crosswalk for Revisions to Current Method Sections

Description
Current section
Revised section

Filter holder
2.1.1
6.1.3

Condenser
2.1.2
6.1.7

Water circulating bath
2.1.3
6.1.8

Adsorbent module
2.1.4
6.1.9

Fitting caps
2.2.1
6.2.1

Wash bottles
2.2.2
6.2.2

Filter storage container
2.2.4
6.2.4

Field balance
2.2.5
6.2.5

Aluminum foil
2.2.6
6.2.6

Glass sample storage container
2.2.9
6.2.8

Extraction thimble
2.3.4
6.3.3.3

Pasteur pipettes
2.3.5
6.4.1

GC oven
2.3.10.1
6.5.1.1

GC Temperature monitor
2.3.10.2
6.5.1.2

GC Flow system
2.3.10.3
6.5.1.3

Capillary GC column
2.3.10.4
6.5.2

Mass spectrometer (MS)
2.3.11
6.5.3

MS data system
2.3.12
6.5.4

Table 3—Additional Specifications for Section 6.0

Description
Revised section

Probe liner
6.1.2

Filter heating system
6.1.4

Filter temperature sensor
6.1.5

Sample transfer line
6.1.6

Impingers
6.1.10

Soxhlet extraction apparatus
6.3.3.1

Moisture trap of extraction apparatus
6.3.3.2

Heating mantle
6.3.3.4

Kuderna-Danish concentrator
6.3.4

Liquid chromatography columns
6.4.2

GC Injection port
6.5.1.4

PCDD/PCDF GC column
6.5.2.1

PAH GC column
6.5.2.2

PCB GC column
6.5.2.3

In Section 6, we are also finalizing changes to:

• Prohibit the use of brominated flame-retardant coated tape in assembling the sampling train and use of silicon tubing in direct contact with flue gases to avoid sample contamination.

• Revise the specification for a rotary evaporator with a note to use a Kuderna-Danish concentrator for PCB and PAH to avoid the loss of higher vapor pressure target compounds.

• Remove specifications for the graduated cylinder to improve the accuracy of moisture measurements and make Method 23 more consistent with other isokinetic sampling methods.

• Remove the volume requirement for wash bottles to allow greater flexibility in field sample recovery.

We are also moving language from Method 23's current Section 6.0 to new Section 10.0 (Calibration and Standardization).

G. Section 7.0 Reagents, Media, and Standards

In this action, the EPA is renumbering and moving the current language in Section 3.0 (Reagents) to a new Section 7.0 titled “Reagents, Media, and Standards,” and making clarifying edits and technical revisions to the specifications. Table 4 of this preamble identifies the new numbering for the subsections currently in Section 3.0 and Table 5 of this preamble identifies new specifications (and the associated subsection) we are including in Section 7.0.

Table 4—Crosswalk for Revisions to Current Method Sections

Description
Current section
Revised section

Filter
3.1.1
7.1

Adsorbent resin
3.1.2
7.2

Glass wool
3.1.3
7.3

Water
3.1.4
7.4

Silica gel
3.1.5
7.5

Methylene chloride
3.2.2
7.6

Sodium sulfate
3.3.2
7.8.2

Basic alumina
3.3.13
7.8.9.1.2

Silica gel
3.3.14
7.8.9.3

Carbon/Celite®
3.3.17
7.8.9.4

Nitrogen
3.3.18
7.8.10

Table 5—Additional Specifications for Section 7.0

Description
Revised section

High-boiling alkanes used as keeper solvents
7.8.8

Liquid column packing materials
7.8.9

Acidic alumina
7.8.9.1.1

Florisil®
7.8.9.2

Helium
7.9.1

Spiking standards
7.9.2

Pre-sampling adsorbent standard
7.9.3

Pre-extraction filter recovery standard
7.9.4

Pre-extraction standard
7.9.5

Pre-analysis standard
7.9.6

We are replacing the filter precleaning procedures of the current method with specifications for conducting a filter quality control check. We are also deleting unnecessary specifications (presented in Table 6 of this preamble) to reflect modern methods. We are renaming the isotopic spiking standard mixtures to better relate the standards to their use in the final method. We are ensuring that the isotopically labeled spiking standards are named consistently throughout the final method.

Table 6—Deletions of Material Specifications in the Current Method 23

Material
Current section

Chromic acid cleaning solution
3.1.6

Benzene
3.3.7

Ethyl acetate
3.3.8

Cyclohexane
3.3.12

Hydrogen
3.3.19

Internal standard solution
3.3.20

Surrogate standard solution
3.3.21

Recovery standard solution
3.3.22

We are also moving the current Section 7.0 to a new Section 9.0 (Quality Control).

H. Section 8.0 Sample Collection, Preservation, and Storage

In this action, the EPA is renumbering and moving the current language in Section 4.0 (Procedure) to a new Section 8.0 titled “Sample Collection, Preservation, and Storage,” and making clarifying edits and technical revisions to the current procedures for sampling and field sample recovery. The new Section 8.0 also includes added requirements for sample storage conditions and holding times.

Under the sampling procedures of Method 23, we are finalizing revisions to the current requirements in Section 4.1.1 for pretest preparations. Table 7 of this preamble identifies the new numbering to revise and replace the requirements in Section 4.1.

Table 7—Crosswalk for Revisions to Current Method Sections

Description
Current section
Revised section

Glassware cleaning
4.1.1.1
8.1.1.1

Assembling the adsorbent module
4.1.1.2
8.1.1.2

Maintaining the sampling train components
4.1.1.3
8.1.1.3

Silica Gel
4.1.1.4
8.1.1.4

Checking and packing filters
4.1.1.5
8.1.1.5

Field preparation of the sampling train
4.1.3.1
8.1.3.1

Impinger assembly
4.1.3.2
8.1.3.2

Sampling probe and nozzle preparation
4.1.3.4
8.1.3.4

Table 8 of this preamble shows the specifications we are adding to the new Section 8.0. This action finalizes a minimum sample volume and sampling time requirements at each traverse point for continuous industrial processes that align Method 23 with other isokinetic stationary source methods, such as Method 5. The sampling time at each traverse point for batch industrial processes ensure measurements are made for the entire process cycle. The final filter check requirements add details that were absent from the original Method 23 and align the method with the requirements of other isokinetic stationary source methods, such as Methods 5, 26A, and 29, also in Appendix A of this Part. The final adsorbent module orientation requirements clarify the configuration of the adsorbent module to ensure that condensed moisture flows through the module into the water collection impinger. We are adding sampling filter temperature monitoring requirements to align Method 23 with other isokinetic stationary source methods. Also, we are adding adsorbent module temperature monitoring to confirm that the sorbent material was not exposed to elevated temperatures that could bias sample collection and results.

Table 8—Additional Specifications for Section 8.1

Description
Revised section

Minimum sample volume
8.1.2.1

Sampling time for continuous processes
8.1.2.2

Sampling time for batch processes
8.1.2.3

Filter assembly
8.1.3.3

Orientation of the condenser and adsorbent module
8.1.3.4

Monitoring the filter temperature
8.1.5.1

Monitoring the adsorbent module temperature
8.1.5.2

Under sample recovery procedures, we are finalizing technical revisions as shown in Table 9 of this preamble. In this action, we are also adding specifications as shown in Table 10 of this preamble.

Table 9—Crosswalk for Revisions to Current Method Sections

Description
Current section
Revised section

Adsorbent module sample preparation
4.2.2
8.2.5

Preparation of Container No. 2
4.2.3
8.2.6

Rinsing of the filter holder and condenser
4.2.3
8.2.7

Weighing impinger water
4.2.5
8.2.8

Preparation of Container No. 3
4.2.4
8.2.9

Silica gel
4. 2.7
8.2.10

Table 10—Additional Specifications for Section 8.2

Description
Revised section

Conducting a post-test leak check
8.2.1

Storage conditions for Container No. 1
8.2.4

Field sample handling, storage, and transport
8.2.11

Sample chain of custody
8.2.12

In the new Section 8.2.6, acetone and toluene rinses are collected in one bottle rather than separately. New Section 8.2.8 measures moisture by weight rather than by volume.

I. Section 9.0 Quality Control

In this action, the EPA is moving and renumbering the current Section 7.0 (Quality Control) to a new Section 9.0 titled “Quality Control,” and making clarifying and technical revisions to the new Section 9.0. We are adding an introductory note that addresses maintaining, and documenting quality control compliance required in Method 23. We are adding a new subsection that clarifies the recordkeeping and reporting necessary to demonstrate compliance with quality control requirements of this method. We are also adding specifications for conducting pre-sampling, pre-extraction, and pre-analysis standard recoveries of isotopically-labeled standards and adding specifications for:

• Initial demonstration of capability (IDC).

• Quality Control Sample (QCS).

• Method detection limits (MDL).

• Laboratory method blank (LMB).

• Estimated detection limits (EDL).

• Field train proof blank.

It should be noted that the EDLs as proposed remain in the method and are sample specific. It should also be noted that the second source QCS also serves as an initial calibration verification. We are also moving language from the current Section 9.0 to new Section 12.0 (Data Analysis and Calculations).

J. Section 10.0 Calibration and Standardization

In this action, the EPA is renumbering and moving the text in Section 6.0 (Calibration) of the current method to a new Section 10.0 titled “Calibration and Standardization,” and making clarifying and technical revisions to the specifications for calibrating the sampling and the HRGC/HRMS systems. We are adding specifications for tuning the HRMS system, moving the specification for HRMS resolution (currently in Section 5) to this new section, and adding text on the procedures for assessing the relative standard deviation for the mean instrument response factors to bring Method 23 up to date with current laboratory practice. We are also

updating the requirements for ion abundance ratio limits, and resolution checks under the continuing calibration verification to serve as performance indicators for analysis quality. We are adding a specification to prepare the continuing calibration verification (CCV) standard at the same time as the batch of field samples using the same labeled standards. We are also moving language in the current Section 10.0 to a new Section 16.0 (Bibliography).

K. Section 11.0 Analysis Procedure

In this action, the EPA is renumbering and moving the text in Section 5.0 (Analysis) of the current method to a new Section 11.0 titled “Analysis Procedure,” and making clarifying and technical revisions to the current specifications for sample extraction and sample cleanup and fractionation. We are also adding a new subsection describing how sample extract aliquots are prepared for cleanup and analysis.

We are also adding the specifications and recommendations for analysis procedures shown in Table 11 of this preamble.

Table 11—Additional Specifications for Section 11.0

Description
Revised section

Preparing and operating the extraction apparatus
11.1.7 through 11.1.9.

Allow the extraction apparatus to cool
11.2.1.

Initial extract concentration
11.2.2.

Allow the sample extract to cool
11.2.3.

Recommended minimum volume for PCDD/PCDF analysis
11.2.3.

Further concentration of sample (if needed) for cleanup and analysis
11.2.4.

Sample cleanup and fractionation for PAH and PCDPE
11.3.1.

Sample cleanup and fractionation for PCDD/PCDF and PCB
11.3.2.

Addressing unresolved compounds
11.4.1.2.1.

Relative retention time for PCB
11.4.3.4.5.

Chlorodiphenyl ether interference
11.4.3.4.8.

MS lock-mass ions
11.4.3.4.9.

Identification criteria for PAH
11.4.3.4.10.

Calculations of target mass and mass per dry standard cubic meter
11.4.3.5.1 and 11.4.3.5.2.

Quantifying native PCDD/PCDF
11.4.3.5.3.

Reporting options
11.4.3.5.4 through 11.4.3.5.6.

L. Section 12.0 Data Analysis and Calculations

In this action, the EPA is renumbering and moving the current language in Section 9.0 (Calculations) to a new Section 12.0 titled “Data Analysis and Calculations,” and revising the equation variable list. We are revising the equations shown in Table 12 of this preamble.

Table 12—Equation Revisions for Section 12.0

Current equation
Description
Revised section

23-1
Individual relative response factor (RRF) for each compound
12.2

23-2
Amount of individual target compound i in the extract using the RRF of the CCV
12.7

23-4
Recovery of Labeled Compound Standards
12.9

23-7
Estimated detection limit
12.10

23-8
Total concentration
12.11

This section specifies that the CCV RRFs are used to quantify the target compounds rather than the initial calibration RRFs. We are also removing and replacing the current equations in Method 23 with the equations shown in Table 13 of this preamble to accommodate the final changes to the method procedures.

Table 13—Additional Equations for Section 12.0

New equation
Description
Revised section

23-1
Individual compound RRF for each calibration level
12.2

23-2
Individual compound RRF for pre-extraction standard
12.2

23-4
Percent relative standard deviation of the RRFs for a compound over the calibration levels
12.4

23-5
Standard deviation of the RRFs for a compound over the calibration levels
12.5

23-6
Percent difference of the RRF of the continuing calibration verification compared to the average RRF from the initial calibration for each target compound
12.6

23-9
Concentration of the Individual Target Compound or Group i in the Emission Gas
12.8

23-13
Half range for the prediction interval of results
12.12

23-14
Upper limit for the prediction interval of results
12.12

23-15
Lower limit for the prediction interval of results
12.12

M. Section 13.0 Method Performance

In this action, the EPA is adding a new Section 13.0 titled “Method Performance,” that includes the specifications shown in Table 14 of this preamble. The new Section 13 provides the basis for assessing accuracy with LMBs, increases labeled standards, and establishes performance criteria to monitor method performance.

Table 14—Method Performance Specifications for Section 13.0

Description
Revised section

Background assessment of field train proof blank, LMB, and Materials (filters, adsorbent resin, glass wool, etc.)
13.1.

GC column systems used to measure PCDD/PCDF, PAH, and PCB target compounds
13.2 through 13.5.

Detection limits (Method detection limits and Estimated detection limits)
13.6.

Tuning HRGC/HRMS system
13.7.

MS lock-mass ions
13.8.

Initial calibration and continuing calibration verification
13.9 and 13.10.

QCS analysis
13.11.

Identification of target compounds
13.12 and 13.13.

Pre-sampling and pre-extraction standard recovery requirements
13.14 and 13.15.

Pre-analysis standard sensitivity requirements
13.16.

IDC-Lowest calibration concentration, Demonstration of precision, Demonstration of accuracy
13.17.

Modifications of the method
13.18 and 13.19.

N. Section 14.0 Pollution Prevention

In this action, the EPA is adding a new Section 14.0 titled “Pollution Prevention,” that specifies the procedures for minimizing or preventing pollution associated with preparing and using Method 23 standards.

O. Section 15.0 Waste Management

In this action, the EPA is adding a new Section 15.0 titled “Waste Management,” that specifies the laboratory responsibilities for managing the waste streams associated with collecting and analyzing Method 23 samples.

P. Section 16.0 Bibliography

In this action, the EPA is renumbering and moving the current language in Section 10.0 (Bibliography) to a new Section 16.0 titled “Bibliography.” We are deleting previous reference number 3 which is no longer relevant and adding new citations for the following references:

• Fishman, V.N., Martin, G.D. and Lamparski, L.L. Comparison of a variety of gas chromatographic columns with different polarities for the separation of chlorinated dibenzo-p-dioxins and dibenzofurans by high-resolution mass spectrometry. Journal of Chromatography A 1139 (2007) 285-300.

• International Agency for Research on Cancer. Environmental Carcinogens Methods of Analysis and Exposure Measurement, Volume 11—Polychlorinated Dioxins and Dibenzofurans. IARC Scientific Publications No. 108, 1991.

• Stieglitz, L., Zwick, G., Roth, W. Investigation of different treatment techniques for PCDD/PCDF in fly ash. Chemosphere 15: 1135-1140; 1986.

• U.S. Environmental Protection Agency. Method 8290A—Polychlorinated Dibenzo-p-dioxin (PCDDs) and Polychlorinated Dibenzofurans (PCDFs) by High-Resolution Gas Chromatography/High-Resolution Mass Spectrometry (HRGC/HRMS), Revision 1. February 2007. In: Test Methods for Evaluating Solid Waste. Washington, DC. SW-846.

• U.S. Environmental Protection Agency. Office of Air Programs Publication No. APTD-0576: Maintenance, Calibration, and Operation of Isokinetic Source Sampling Equipment. Research Triangle Park, NC. March 1972.

• U.S. Environmental Protection Agency. Method 1625C—Semivolatile Organic Compounds by Isotope Dilution GCMS.

• U.S Environmental Protection Agency. Method 1613B—Tetra- through Octa-Chlorinated Dioxins and Furans by Isotope Dilution HRGC/HRMS.

• U.S. Environmental Protection Agency. Method 1668C—Chlorinated Biphenyl Congeners in Water, Soil, Sediment, Biosolids, and Tissue by HRGC/HRMS.

• Tondeur, Y., Nestrick, T., Silva, Héctor A., Vining, B., Hart, J. Analytical procedures for the determination of polychlorinated-p-dioxins, polychlorinated dibenzofurans, and hexachlorobenzene in pentachlorophenol. Chemosphere Volume 80, Issue 2, June 2010, pages 157-164.

• U.S. Environmental Protection Agency. Definition and Procedure for the Determination of the Method Detection Limit, Revision 2. EPA 821-R-16-006. December 2016.

• Tondeur Y, Niederhut WJ, Missler SR. A hybrid HRGC/MS/MS Method for the Characterization of Tetrachlorodibenzo-p-Dioxins in Environmental Samples; Bio. Med. and Environ. Mass Spectr. 14, pages 449-456, 1987.

• Gianluca R., Mosca S., Guerriero E., Rotatori M. Development of a new automated clean-up system for the simultaneous analysis of polychlorinated dibenzo-p-dioxins (PCDDs), dibenzofurans (PCDFs) and `dioxin-like' polychlorinated biphenyls (dl-PCB) in flue gas emissions by GPC-SPE. J. Environ. Monit. 14, pages 1082-1090, 2012.

• U.S. Environmental Protection Agency. The National Dioxin Air Monitoring Network (NDAMN) Report of the Results of Atmospheric Measurements of Polychlorinated Dibenzo-p-Dioxins (PCDDs), Polychlorinated Dibenzofurans (PCDFs), and Dioxin-like Polychlorinated Biphenyl (PCBs) in Rural and Remote Areas of the United States from June 1998 through November 2004. EPA/600/R-13/183F. August 2013.

• Guo, Y., Kannan, K. Analytical Methods for the Measurement of Legacy and Emerging Persistent Organic Pollutants in Complex Sample Matrices. Comprehensive Analytical Chemistry. Vol. 67. January 2015.

• U.S. Environmental Protection Agency. USEPA Contract Laboratory Program (CLP) National Functional Guidelines for Chlorinated Dibenzo-p-Dioxins (CDDs) and Chlorinated Dibenzofurans (CDFs) Data Review. EPA-540-R-11-016. September 2011.

Q. Section 17.0 Tables, Diagrams, Flow Charts, and Validation Data

In this action, the EPA is adding a new Section 17 titled “Tables, Diagrams, Flow Charts, and Validation Data,” that contains all tables, diagrams, flow charts, and validation data referenced in Method 23. We are revising Figures 23-1 and 23-2 and renaming and/or renumbering the current Method 23 tables as shown in Table 15 of this preamble.

Table 15—Revisions to Method 23 Tables

Current method
Final method

Table 1—Composition of the Sample Fortification and Recovery Standards Solutions
Table 23-7. Concentration of the Sample Fortification for PCDD and PCDF.

Table 2—Composition of the Initial Calibration Solutions
Table 23-11. Concentration of the Initial Calibration Standard Solutions for PCDD and PCDF.

Table 3—Elemental Compositions and Exact Masses of the Ions Monitored by High Resolution Mass Spectrometry for PCDD's and PCDF's
Table 23-4. Elemental Compositions and Exact Masses of the Ions Monitored by High-Resolution Mass Spectrometry for PCDD and PCDF.

Table 4—Acceptable Ranges for Ion-Abundance Ratios of PCDD's and PCDF's
Table 23-15. Recommended Ion Type and Acceptable Ion Abundance Ratios.

Table 5—Minimum Requirements for Initial and Daily Calibration Response Factors
Table 23-14. Minimum Requirements for Initial and Continuing Calibration Response Factors for Isotopically Labeled and Native Compounds.

We are also adding Figure 23-3 (Soxhlet/Dean-Stark Extractor) and Figure 23-4 (Sample Preparation Flow Chart) and adding the tables listed in Table 16 of this preamble.

Table 16—Additional Tables to Method 23

Revised table
Description

23-1

Polychlorinated Dibenzo-
p
-dioxin and Polychlorinated Dibenzofuran Target Analytes.

23-2
Polycyclic Aromatic Hydrocarbon Target Analytes.

23-3
Polychlorinated Biphenyl Target Analytes.

23-5
Elemental Compositions and Exact Masses of the Ions Monitored by High-Resolution Mass Spectrometry for PAH.

23-6
Elemental Compositions and Exact Masses of the Ions Monitored by High-Resolution Mass Spectrometry for PCB.

23-8
Concentration of the Sample Fortification for PAH.

23-9
Concentration of the Sample Fortification for PCB.

23-10
Sample Storage Conditions and Laboratory Hold Times.

23-12
Concentration of the Initial Calibration Standard Solutions for PAH.

23-13
Concentration of the Initial Calibration Standard Solutions for PCB.

23-16
Typical DB5-MS Column Conditions.

23-17
Assignment of Pre-extraction Standards for Quantitation of Target PCB.

23-18
Initial Demonstration of Capability Quality Control (QC) Requirements.

V. Summary of Final Revisions Related to 40 CFR Parts 60, 63, and 266

A. 40 CFR Part 60—Standards of Performance for New Stationary Sources

In 40 CFR 60.17(h), we are incorporating by reference ASTM D4840-99(2018)e1, Standard Guide for Sample Chain-of-Custody Procedures, and amending the reference to ASTM D6911-15, Guide for Packaging and Shipping Environmental Samples for Laboratory Analysis, to include for use in Method 23.

In 40 CFR part 60, subpart CCCC, we are revising 40 CFR 60.2125(g)(2) and (j)(2) to realign the requirement for quantifying isomers to the reorganized Section 11.4.2.4 in the revisions of Method 23.

In 40 CFR part 60, subpart DDDD, we are revising 40 CFR 60.2690(g)(2) and (j)(2) to realign the requirement for identifying isomers to the reorganized Section 11.4.2.4 in the revisions of Method 23.

B. 40 CFR Part 63—National Emission Standards for Hazardous Air Pollutants for Source Categories

In 40 CFR 63.849(a)(13) and (a)(14), we are replacing CARB Method 428 with EPA Method 23 for the measurement of PCB emissions from roof monitors not employing wet roof scrubbers.

In 40 CFR 63.1208(b)(1), we are removing the requirement for administrator's approval to use Method 23 for measuring PCDD/PCDF emissions from hazardous waste combustors.

In 40 CFR 63.1625(b)(10), we are replacing CARB Method 429 with EPA Method 23 for measuring the emissions of PAH from ferromanganese electric arc furnaces.

In Table 3 to Subpart AAAAAAA, we are replacing the requirement for analysis of PAH by SW-846 Method 8270 with a requirement to use EPA Method 23. Specifically, we are deleting “with analysis by SW-846 Method 8270D” in row 6 of Table 3 to Subpart AAAAAAA. Because revisions to Method 23 eliminate the use of methylene chloride in field sampling activities, we are also removing footnote “b” in Table 3 to Subpart AAAAAAA.

C. 40 CFR Part 266—Standards for the Management of Specific Hazardous Wastes and Specific Types of Hazardous Waste Management Facilities

In 40 CFR 266.104, we are adding EPA Method 23 as an alternative to SW-846 Method 0023A. We proposed to make this change to 40 CFR 266.104. In addition to this specific change, we are making a conforming change in 40 CFR part 266 Appendix IX. EPA considers this conforming change a logical outgrowth of the proposed revisions to Method 23.

VI. Statutory and Executive Order Reviews

Additional information about these statutes and Executive Orders can be found at
https://www.epa.gov/laws-regulations/laws-and-executive-orders
.

A. Executive Order 12866: Regulatory Planning and Review and Executive Order 13563: Improving Regulation and Regulatory Review

This action is not a significant regulatory action and was, therefore, not submitted to the Office of Management and Budget (OMB) for review.

B. Paperwork Reduction Act (PRA)

This action does not impose an information collection burden under the PRA. The revisions being promulgated in this action to Method 23 do not add information collection requirements, but make corrections, clarifications, and updates to existing testing methodology.

C. Regulatory Flexibility Act (RFA)

I certify that this action does not have a significant economic impact on a substantial number of small entities under the RFA. This action does not impose any requirements on small entities. The final revisions to Method 23 do not impose any requirements on regulated entities. Rather, the final changes improve the quality of the results when required by other rules to use Method 23. Revisions to Method 23 allow contemporary advances in analysis techniques to be used. Further, the final changes in Method 23 analysis procedures reduce the impact of this method by bringing it into alignment with other agency methods.

D. Unfunded Mandates Reform Act (UMRA)

This action does not contain any unfunded mandate of $100 million or more as described in UMRA, 2 U.S.C. 1531-1538. The action imposes no enforceable duty on any State, local or tribal governments or the private sector.

E. Executive Order 13132: Federalism

This action does not have federalism implications. It will not have substantial direct effects on the states, on the relationship between the national government and the states, or on the distribution of power and responsibilities among the various levels of government.

F. Executive Order 13175: Consultation and Coordination With Indian Tribal Governments

This action does not have tribal implications, as specified in Executive Order 13175. It will not have substantial direct effects on the Indian Tribal Governments, on the relationship between the national government and the Indian Tribal Governments, or on the distribution of power and responsibilities among Indian Tribal Governments and the various levels of government. Thus, Executive Order 13175 does not apply to this action.

G. Executive Order 13045: Protection of Children From Environmental Health Risks and Safety Risks

The EPA interprets Executive Order 13045 as applying only to those regulatory actions that concern environmental health or safety risks that the EPA has reason to believe may disproportionately affect children, per the definition of “covered regulatory action” in Section 2-202 of the Executive Order. This action is not subject to Executive Order 13045 because it does not establish or revise a standard that provides protection to children against environmental health and safety risks.

H. Executive Order 13211: Actions That Significantly Affect Energy Supply, Distribution or Use

This action is not subject to Executive Order 13211, because it is not a significant regulatory action under Executive Order 12866.

I. National Technology Transfer and Advancement Act (NTTAA)

This action involves technical standards. The EPA will use ASTM D6911-15 (Guide for Packaging and Shipping Environmental Samples for Laboratory Analysis) and ASTM D4840-99(2018)e1 (Standard Guide for Sample Chain-of-Custody Procedures). These ASTM standards cover best practices that guide sample shipping and tracking from collection through analysis.

These standards were developed and adopted by ASTM International. The standard may be obtained from
https://www.astm.org
or from the ASTM at 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959.

J. Executive Order 12898: Federal Actions To Address Environmental Justice in Minority Populations and Low-Income Populations

Executive Order 12898 (59 FR 7629, February 16, 1994) directs federal agencies, to the greatest extent practicable and permitted by law, to make environmental justice part of their mission by identifying and addressing, as appropriate, disproportionately high and adverse human health or environmental effects of their programs, policies, and activities on minority populations (people of color) and low-income populations.

The EPA believes that this type of action does not concern human health or environmental conditions and, therefore, cannot be evaluated with respect to potentially disproportionate and adverse effects on people of color, low-income populations and/or Indigenous peoples. This action updates Method 23, which will improve the quality of the results when required by other rules to use Method 23.

K. Congressional Review Act (CRA)

This action is subject to the CRA and the EPA will submit a rule report to each House of the Congress and to the Comptroller General of the United States. This action is not a “major rule” as defined by 5 U.S.C. 804(2).

L. Determination Under Clean Air Act Section 307(d)

This final rule is not subject to the provisions of CAA section 307(d). This final rule does not promulgate any of the actions listed in CAA section 307(d)(1).

List of Subjects

40 CFR Part 60
Environmental protection, Air pollution control, Hazardous air pollutants, Incorporation by reference, Method 23, Polychlorinated biphenyls, Polychlorinated dibenzofurans, Polychlorinated dibenzo-p-dioxins, Polycyclic aromatic compounds, Test methods.

40 CFR Part 63
Environmental protection, Air pollution control, Method 23, New source performance, Polychlorinated biphenyls, Polychlorinated dibenzofurans, Polychlorinated dibenzo-p-dioxins, Polycyclic aromatic hydrocarbons, Test methods.

40 CFR Part 266
Environmental protection, Air pollution control, Hazardous air pollutants, Hazardous waste, Method 23, Polychlorinated biphenyls, Polychlorinated dibenzofurans, Polychlorinated dibenzo-p-dioxins, Polycyclic aromatic hydrocarbons, Test methods, Waste management.

Michael S. Regan,
Administrator.

For the reasons stated in the preamble, the Environmental Protection Agency amends Title 40, Chapter I of the Code of Federal Regulations as follows:

PART 60—STANDARDS OF PERFORMANCE FOR NEW STATIONARY SOURCES

1. The authority citation for part 60 continues to read as follows:

Authority:

42 U.S.C. 7401
et seq.

Subpart A—General Provisions

2. In § 60.17:
a. Redesignate paragraphs (h)(168) through (h)(213) as (h)(169) through (h)(214);
b. Add new paragraph (h)(168); and
c. Revise newly redesignated paragraph (h)(194).
The addition and revision read as follows:

§ 60.17
Incorporations by reference.

(h) * * *

(168) ASTM D4840-99(2018)e1 Standard Guide for Sample Chain-of-Custody Procedures, approved August 2018; IBR approved for Appendix A-7: Method 23.

(194) ASTM D6911-15 Standard Guide for Packaging and Shipping Environmental Samples for Laboratory Analysis, approved January 15, 2015; IBR approved for Appendix A-7: Method 23; Appendix A-8: Method 30B.

Subpart CCCC—Standards of Performance for Commercial and Industrial Solid Waste Incineration Units

3. In § 60.2125, revise paragraphs (g)(2) and (j)(2) to read as follows:

§ 60.2125
How do I conduct the initial and annual performance test?

(g) * * *

(2) Quantify isomers meeting identification criteria in Section 11.4.3.4 of Method 23, regardless of whether the isomers meet identification criteria in Section 11.4.3.4.1 of Method 23. You must quantify the isomers per Section 11.4.3.5 of Method 23. (Note: You may reanalyze the sample aliquot or split to reduce the number of isomers to meet the identification criteria in Section 11.4.3.4 of Method 23.)

(j) * * *

(2) Quantify isomers meeting identification criteria in Section 11.4.3.4 of Method 23, regardless of whether the isomers meet identification Section 11.4.3.4.1 of Method 23. You must quantify the isomers per Section 11.4.3.5 of Method 23. (Note: You may reanalyze the sample aliquot or split to reduce the number of isomers to meet the identification criteria in Section 11.4.3.4 of Method 23.)

Subpart DDDD—Emissions Guidelines and Compliance Times for Commercial and Industrial Solid Waste Incineration Units

4. In § 60.2690, revise paragraphs (g)(2) and (j)(2) to read as follows:

§ 60.2690
How do I conduct the initial and annual performance test?

(g) * * *

(2) Quantify isomers meeting identification criteria in Section 11.4.3.4 of Method 23, regardless of whether the isomers meet identification Section 11.4.3.4.1 of Method 23. You must quantify the isomers per Section 11.4.3.5 of Method 23. (Note: You may reanalyze the sample aliquot or split to reduce the number of isomers to meet the identification criteria in Section 11.4.3.4 of Method 23.)

(j) * * *

(2) Quantify isomers meeting identification criteria in Section 11.4.3.4 of Method 23, regardless of whether the isomers meet identification Section 11.4.3.4.1 of Method 23. You must quantify the isomers per Section 11.4.3.5 of Method 23. (Note: You may reanalyze the sample aliquot or split to reduce the number of isomers to meet the identification criteria in Section 11.4.3.4 of Method 23.); and

5. Revise Method 23 of Appendix A-7 to Part 60 to read as follows:
Appendix A-7 to Part 60—Test Methods 19 Through 25E

Method 23—Determination of Polychlorinated Dibenzo-
p
-Dioxins, Polychlorinated Dibenzofurans, Polychlorinated Biphenyls, and Polycyclic Aromatic Hydrocarbons From Stationary Sources

1.0 Scope and Application

1.1 Applicability.
This method applies to the measurement of polychlorinated dibenzo-
p
-dioxins and polychlorinated dibenzofurans (PCDD/PCDF), polychlorinated biphenyls (PCB), and/or polycyclic aromatic hydrocarbons (PAH) in emissions from stationary sources. Using this method, you can measure these analyte groups individually or in any combination using a single sample acquisition unless otherwise specified in a rule, regulation, or permit. Tables 23-1 through 23-3 of this method list the applicable target analytes for Method 23. If all 209 PCB are analyzed, the 17 toxic PCB congeners should be resolved and reported while the other PCB can be reported as totals by homolog, for example, total trichlorobiphenyl (TrCB).

1.2 Scope.
This method describes the sampling and analytical procedures used to measure selected PCDD and PCDF in stationary sources when required in an applicable subpart. This method also describes how the same sampling and analysis technology can be used to measure selected PCB and PAH from stationary source in combination or as each individual compound class when required in an applicable subpart. However, Method 23 incorporates by reference some of the specifications (
e.g.,
equipment and supplies) and procedures (
e.g.,
sampling and analytical) from other methods in this part that are essential to conducting Method 23. To obtain reliable samples, source sampling teams should be trained and experienced with the following additional EPA test methods: Method 1, Method 2, Method 3, Method 4, and Method 5 of Appendices A-1, A-2, and A-3 to 40 CFR part 60. Laboratory analysis teams should be trained and experienced with Method 1668C (found at:
https://www.epa.gov/sites/production/files/2015-09/documents/method_1668c_2010.pdf)
and Method 1613B of 40 CFR part 136 Appendix A and have a working knowledge of isotope dilution and the use of high-resolution gas chromatography/high-resolution mass spectrometry (HRGC/HRMS).

1.3 The HRGC/HRMS
portions of this method are for use by laboratory analysts experienced with HRGC/HRMS analysis of PCDD, PCDF, PCB, and PAH or under the close supervision of such qualified persons. Each source testing team, including the sampling and laboratory organization(s) that use this method, must demonstrate the ability to generate acceptable results that meet the performance criteria in Section 13 of this method.

1.4 This method is “performance-based” and includes acceptability criteria for assessing sampling and analytical procedures.
Users may modify the method to overcome interferences or to substitute superior materials and equipment, provided that they meet all performance criteria in this method. Section 13 of this method presents requirements for method performance.

2.0 Summary of Method

This method identifies and determines the concentration of specific PCDD, PCDF, PCB, and PAH compounds. Gaseous and particulate bound target pollutants are withdrawn from the gas stream isokinetically and collected in the sample probe, on a glass fiber or quartz filter, and on a packed column of adsorbent material. This method is not intended to differentiate between target compounds in particulate or vapor fractions. The target compounds are extracted from the combined sample collection media. Portions of the extract are chromatographically fractionated to remove interferences, separated into individual compounds or simple mixtures by HRGC, and measured with HRMS. This method uses isotopically labeled standards to improve method accuracy and precision through isotope dilution quantitation.

3.0 Definitions

3.1 Alternate Recovery Standards.
A group of isotopically labeled compounds that is not otherwise designated in this method

for quality control (QC) purposes. Alternate recovery standards can be used to assess the recovery of a compound class relative to any step in the sampling and analysis procedure that is not already assessed as a mandatory part of this method, such as the cleanup step.

3.2 Benzo[a]pyrene Toxic Equivalency Quotient (B[a]P-TEQ).
One of several schemes that express the toxicity for PAH compounds in terms of the most toxic form of PAH, benzo[
a
]pyrene, as specified in applicable regulations, permits, or other requirements.

3.3 Continuing Calibration Verification (CCV) Standard.
A standard prepared at the mid-point concentration of the calibration used to verify the initial calibration. Prepare the CCV standard at the same time as the batch of field samples using the same labeled standards.

3.4 Congener.
An individual compound with a common structure (dioxin, furan, or biphenyl), only differing by the number of chlorine or other substituent attached to the structure.

3.5 Estimated Detection Limit (EDL).
The minimum qualitatively recognizable signal above background for a target compound. The EDL is a detection limit specific to each sample analysis based on the noise signal measured near the retention time of a target compound or target isomer group. Being sample specific, the EDL is affected by sample size, dilution, recoveries of pre-extraction standard, chemical noise from sample extract, electronic noise from instrument, extract aliquot, relative response of instrument, etc.

3.6 Estimated Maximum Possible Concentration (EMPC).
An EMPC is a worst-case estimate of the target compound concentration. Report the results as EMPC when the ion abundance ratio for a target analyte is outside the performance criteria. Calculate the EMPC using both quantitation ions.

3.7 Field Train Proof Blank.
A field train proof blank train is a QC sample to evaluate equipment preparation and potential contamination during sample recovery and consists of a fully assembled train at the sampling site, without actual sampling. The field train proof blank train uses glassware from the same preparation batch as the field samples.

3.8 Homolog.
A compound belonging to a series of compounds with the same general molecular formula, differing from each other by the number of repeating units of chlorine.

3.9 Isomer.
An individual compound with a common structure (dioxin, furan, or biphenyl), only differing by the position of chlorine atoms attached to the structure.

3.10 Isotope Dilution.
A means of determining a naturally occurring (native) compound by reference to the same compound in which one or more atoms has been isotopically enriched.

3.11 Laboratory Method Blank (LMB).
A quality control sample to assess background contamination or interference from media, reagents, equipment, etc. An LMB is prepared in the laboratory, composed of clean sampling media (filter and XAD-2), using same labeled standards, media, reagents, and materials (sodium sulfate, glass wool, etc.) and processed (extraction, fractionations, cleanup) and analyzed using the same procedures as a field sample.

3.12 Polychlorinated Biphenyl (PCB) congeners.
Any or all 209 chlorinated biphenyl congeners. Table 23-3 of this method lists the primary target compounds and Appendix A to this method provides the full list of 209 PCB congeners and isomers.

3.12.1 Monochlorobiphenyl (MoCB). Any or all three monochlorinated biphenyl isomers.

3.12.2 Dichlorobiphenyl (DiCB). Any or all 12 dichlorinated biphenyl isomers.

3.12.3 Trichlorobiphenyl (TrCB). Any or all 24 trichlorinated biphenyl isomers.

3.12.4 Tetrachlorobiphenyl (TeCB). Any or all 42 tetrachlorinated biphenyl isomers.

3.12.5 Pentachlorobiphenyl (PeCB). Any or all 46 pentachlorinated biphenyl isomers.

3.12.6 Hexachlorobiphenyl (HxCB). Any or all 42 hexachlorinated biphenyl isomers.

3.12.7 Heptachlorobiphenyl (HpCB). Any or all 24 heptachlorinated biphenyl isomers.

3.12.8 Octachlorobiphenyl (OcCB). Any or all 12 octachlorinated biphenyl isomers.

3.12.9 Nonachlorobiphenyl (NoCB). Any or all three nonachlorinated biphenyl isomers.

3.12.10 Decachlorobiphenyl (DeCB). Biphenyl fully chlorinated with 10 chlorine atom substituents replacing hydrogen in the parent compound.

3.13 Polychlorinated dibenzo-p-dioxin (PCDD) congeners.
Any or all 75 chlorinated dibenzo-
p
-dioxin congeners. There are seven 2,3,7,8 substituted PCDD congeners and four PCDD homolog groups listed in Table 23-1 of this method. This method does not measure mono- through tri-PCDD and includes non-2,3,7,8 substituted congeners in the total homolog categories.

3.13.1 Tetrachlorodibenzo-
p
-dioxin (TeCDD
).
Any or all 22 tetrachlorinated dibenzo-
p
-dioxin isomers.

3.13.2 Pentachlorodibenzo-
p
-dioxin (PeCDD). Any or all 14 pentachlorinated dibenzo-
p
-dioxin isomers.

3.13.3 Hexachlorodibenzo-
p
-dioxin (HxCDD). Any or all 10 hexachlorinated dibenzo-
p
-dioxin isomers.

3.13.4 Heptachlorodibenzo-
p
-dioxin (HpCDD). Any or all two heptachlorinated dibenzo-
p
-dioxin isomers.

3.13.5 Octachlorodibenzo-
p
-dioxin (OCDD). Dibenzodioxin fully chlorinated with eight chlorine atom substituents replacing hydrogen in the parent compound.

3.14 Polychlorinated dibenzofuran (PCDF) congeners.
Any or all chlorinated dibenzofuran congeners. There are ten 2,3,7,8 substituted PCDF congeners and four PCDF homolog groups listed in Table 23-1 of this method. This method does not measure mono- through tri-PCDF and includes non-2,3,7,8 substituted congeners in the total homolog categories.

3.14.1 Tetrachlorodibenzofuran (TeCDF). Any or all 38 tetrachlorinated dibenzofuran isomers.

3.14.2 Pentachlorodibenzofuran (PeCDF). Any or all 28 pentachlorinated dibenzofuran isomers.

3.14.3 Hexachlorodibenzofuran (HxCDF). Any or all 16 hexachlorinated dibenzofuran isomers.

3.14.4 Heptachlordibenzofuran (HpCDF). Any or all four heptachlorinated dibenzofuran isomers.

3.14.5 Octachlorodibenzofuran (OCDF). Dibenzofuran fully chlorinated with eight chlorine atom substituents replacing hydrogen in the parent compound.

3.15 Polychlorinated diphenyl ethers (PCDPE).
Any or all chlorinated substituted diphenyl ethers.

3.15.1 Hexachlorodiphenyl ether (HxCDPE). Any or all 42 hexachlorinated diphenyl ether isomers.

3.15.2 Heptachlorodiphenyl ether (HpCDPE). Any or all 24 heptachlorinated diphenyl ether isomers.

3.15.3 Octachlorodiphenyl ether (OCDPE). Any or all 12 octachlorinated diphenyl ether isomers.

3.15.4 Nonachlorodiphenyl ether (NCDPE). Any or all three nonachlorinated diphenyl ether isomers.

3.15.5 Decachlorodiphenyl ether (DCDPE).

3.16 Polycyclic Aromatic Hydrocarbons (PAH).
Any or all aromatic compounds with two or more fused six-member rings. Table 23-2 of this method lists the target PAH compounds for this method. You may add and analyze additional PAH compounds by adding the appropriate
13
C isotopically labeled compound to the pre-extraction standard mixture and by following the other requirements for target PAH compounds in this method.

3.17 Pre-analysis Standard.
A group of isotopically labeled compounds added at a known amount immediately prior to analysis and used to monitor instrument response, injection errors, instrument drift and to determine the recovery of the pre-extraction standard compounds. Add pre-analysis standard to every sample (including blank, QC samples, and calibration solutions) at a known amount.

3.18 Pre-extraction Filter Recovery Standard.
A group of isotopically labeled compounds added at a known amount to the filter used to indicate the extraction efficiency of the filter media. Add pre-extraction filter recovery standard to the filter samples just prior extraction. The pre-extraction filter recovery standard is not used for quantitating or recovery correction.

3.19 Pre-extraction Standard.
A group of isotopically labeled compounds added in a known amount to the XAD-2 adsorbent resin of each sample immediately before extraction and used for quantitation of target and other labeled compounds to correct for extraction, cleanup, and concentration recovery. These isotopically labeled compounds constitute a matrix spike of the resin. Add pre-extraction standard to every sample at the same level (including blank, QC samples, and calibration solutions).

3.20 Pre-sampling Adsorbent Standard.
A group of isotopically labeled compounds added in a known amount to the XAD-2 adsorbent prior to sampling used to monitor sampling aspects of the method.

3.21 Pre-transport Standard.
Spiking compound from the list of alternative recovery standards that can be added by the laboratory to the sample shipping containers used to transport field equipment rinse and

recovery samples prior to sampling. The measured concentration of the pre-transport recovery standard provides a quality check on potential probe rinse sample spillage or mishandling after sample collection and during shipping.

3.22 Quality Control Sample (QCS).
A mid-level standard prepared from a second source standard or prepared from a source of standards different from the source of calibration standards. The purpose of the QCS is to verify the integrity of the primary calibration standards. A QCS is analyzed during the initial demonstration of capability (IDC) and following each initial calibration (at a minimum quarterly) thereafter.

3.23 Relative Response Factor (RRF).
The response of the mass spectrometer (MS) to a known amount of an analyte relative to a known amount of an isotopically labeled standard.

3.24 2,3,7,8-Tetrachlorodibenzo-p-dioxin Toxic Equivalency Quotient (2,3,7,8-TeCDD TEQ).
A procedure that expresses the toxicity of PCDD, PCDF, and PCB in terms of the most toxic dioxin, as specified in applicable regulations, permits, or other requirements.

4.0 Interferences

Despite interferences, confidence of the data is based on the enhanced selectivity of fractionation, gas chromatograph (GC) separation and detector resolving power, the QC check ions, and monitoring PCDPE.

4.1 PCB and PCDPE have similar molecular weight and chromatographic properties to PCDD and PCDF.
PCB may produce fragment ions at interfering mass-to-charge ratios (m/z) when losing chlorine (Cl
2
) or 2 Cl
2
during ionization processes. With HRMS, GC separation, and fractionation, PCB should not pose a problem for PCDD/PCDF identification and quantitation. PCDPE, when losing Cl
2,
also produce interfering m/z values in the PCDF homolog group with two fewer chlorine atoms (
i.e.,
an octachlorinated PCDPE can interfere with a hexachlorinated PCDF). The latter interferences are potentially detected by monitoring an m/z corresponding to the potentially interfering PCDPE; however, the fragmentation patterns of all PCDPE may not be known, complicating any attempt to quantify the extent of ether interference.

Note:
Consider monitoring 328 m/z if high levels of PCB are expected.

4.2 Very high amounts of other organic compounds in the matrix may interfere with the analysis.
This method provides examples of column-chromatographic cleanup as procedures to reduce, but not necessarily eliminate, matrix effects due to high concentrations of organic compounds (International Agency for Research on Cancer 1991).

4.3 Target compound contaminants
or related organics in solvents, reagents, glassware, isotopically labeled spiking standards, and other sample processing hardware are potential method interferences. Routinely evaluate all these materials to demonstrate that they are either free from interferences under the conditions of the analysis, or that the interference does not compromise the quality of the analysis results. Evaluate chemical interference through the preparation and analysis of an LMB. Use high purity reagents, solvents, and standards to minimize interferences in sample analysis.

4.4 PAH
are subject to degradation when exposed to ultraviolet light. Take precautions to shield samples from sunlight or fluorescent light sources during sample collection, recovery, extraction, cleanup, and concentration.

5.0 Safety

Note:
Develop a strict laboratory safety program for the handling of PCDD, PCDF, PCB, and/or PAH.

5.1 Compounds in the PCDD and PCDF
classes such as 2,3,7,8-TeCDD are aneugenic, carcinogenic, and teratogenic in laboratory animal studies. Other PCDD and PCDF containing chlorine atoms in positions 2,3,7,8 have toxicities comparable to that of 2,3,7,8-TeCDD.

5.2 PCB and benzo[a]pyrene
are classified as known or suspected human or mammalian carcinogens. Be aware of the potential for inhalation and ingestion exposure to laboratory analysts.

5.3
This method recommends that the laboratory purchase dilute standard solutions of the analytes required for this method. However, if preparing primary solutions, use a hood or glove box. Laboratory personnel handling primary solutions should wear personal protective equipment including a toxic gas respirator mask fitted with charcoal filters approved by the National Institute for Occupational Safety and Health (NIOSH)/Mine Safety Health Administration (MSHA) to prevent the inhalation of airborne particulates if not working in an approved hood or glove box.

5.4
The toxicity or carcinogenicity of other reagents or chemicals used in this method is not precisely defined. However, treat each chemical as a potential health hazard and minimize exposure to these chemicals. The laboratory is responsible for maintaining a current awareness file of Occupational Safety and Health Administration (OSHA) regulations regarding the safe handling of the chemicals specified in this method. Ensure that a reference file or list of internet sites that contain safety data sheets (SDS) is available to all personnel involved in the sampling and chemical analysis of samples known or suspected to contain PCDD, PCDF, PCB, and PAH.

6.0 Equipment and Supplies

Note:
Brand names, suppliers, and part numbers are for illustration purposes only and no endorsement is implied. Apparatus and materials other than those specified in this method may achieve equivalent performance. Meeting the performance requirements of this method is the responsibility of the source testing team and laboratory team.

6.1 Sampling Apparatus.
Figure 23-1 of this method shows a schematic of the Method 23 sampling train. Do not use sealing greases or brominated flame retardant-coated tape in assembling the train. Do not use silicon tubing in direct contact with flue gases. The train is identical to that described in Section 6.1.1 of Method 5 of Appendix A-3 to 40 CFR part 60 with the following additions:

6.1.1 Nozzle. The nozzle must be made of quartz, borosilicate glass, or titanium. Stainless steel nozzles should not be used.

6.1.2 Probe Liner. Use either polytetrafluoroethylene (PTFE), borosilicate, or quartz glass probe liners with a heating system capable of maintaining a probe gas temperature of 120 ± 14 °C (248 ± 25 °F) during sampling, or such other temperature as specified by an applicable subpart of the standards or as approved by the Administrator. Use a PTFE ferrule or single-use PTFE coated O-ring to achieve the seal at the nozzle end of the probe for stack temperatures up to about 300 °C (572 °F). Use a quartz glass liner and integrated quartz nozzle for stack temperatures between 300 and 1,200 °C (572 and 2,192 °F).

6.1.3 Filter Holder. Use a filter holder of borosilicate glass with a PTFE frit or PTFE-coated wire filter support. The holder design should provide a positive seal against leakage from the outside or around the filter. The holder should be durable, easy to load, leak-free in normal applications, and positioned immediately following the probe and cyclone bypass (or cyclone, if used) with the active side of the filter perpendicular to the source of the flow.

6.1.4 Filter Heating System. Use any heating system capable of monitoring and maintaining the temperature around the filter to ensure that the sample gas temperature exiting the filter is 120 ± 14 °C (248 ± 25 °F) during sampling or such other temperature as specified by an applicable subpart of the standards or approved by the Administrator for a particular application.

6.1.5 Filter Temperature Sensor. Install a temperature sensor capable of measuring temperature to within ±3 °C (5.4 °F) so that the sensing tip protrudes at least 1.3 centimeters (cm) (1-2 in.) into the sample gas exiting the filter. Encase the sensing tip of the sensor in glass or PTFE, if needed.

6.1.6 Sample Transfer Line. The sample transfer line transports gaseous emissions from the heated filter holder to the condenser and must be heat traced and constructed of glass or PTFE with connecting fittings that form leak-free, vacuum-tight connections without using sealing greases or tapes. Keep the sample transfer lines as short as possible and maintain the lines at a temperature of 120 °C ± 14 °C (248 °F ± 25 °F) using active heating when necessary. Orient the sample transfer lines with the downstream end lower than the upstream end so that any condensate will flow away from the filter and into the condenser.

6.1.7 Condenser. Glass, water-jacketed, coil-type with compatible fittings. Orient the condenser to cause moisture to flow down to the adsorbent module to facilitate condensate drainage. Figure 23-2 of this method shows a schematic diagram of the condenser.

6.1.8 Water Circulating Bath. Use a bath pump circulating system capable of providing chilled water flow to the condenser and adsorbent module water jackets. Typically, a submersible pump is placed in the impinger ice water bath to circulate the ice water contained in the bath. Verify the function of this system by

measuring the gas temperature at the entrance to the adsorbent module. Maintain this temperature at <20 °C (68 °F).

6.1.9 Adsorbent Module. Use a water-jacketed glass container to hold up to 40 grams (g) of the solid adsorbent. Figure 23-2 of this method shows a schematic diagram of the adsorbent module. Other physical configurations of the adsorbent resin module/condenser assembly are acceptable if the configuration contains the requisite amount of solid adsorbent and maintains the minimum length-to-width adsorbent bed ratio of two-to-one. Orient the adsorbent module vertically to facilitate condensate drainage. The connecting fittings must form leak-free, vacuum-tight seals. Include a coarse glass frit in the adsorbent module to retain the adsorbent.

6.1.10 Impingers. Use five impingers connected in series with leak-free ground glass fittings or any similar leak-free noncontaminating fittings. The first impinger must be a short-stem (water-dropout) design or equivalent. The second, fourth, and fifth impingers must be of the Greenburg-Smith design, modified by replacing the tip with a 1.3 cm (
1/2
in.) inside diameter (ID) glass tube extending to approximately 1.3 cm (
1/2
in.) from the bottom of the flask. The third impinger must be of the Greenburg-Smith design with the standard tip. The second and third impingers must contain known quantities of water, and the fifth impinger must contain a known weight of silica gel or equivalent desiccant. Alternatively, you may omit the first impinger if you do not expect excess moisture in the sample gas.

6.2 Sample Recovery Equipment.

6.2.1 Fitting Caps. Use leak-free ground glass fittings or any similar leak-free non-contaminating fitting to cap the sections of the sampling train exposed to the sample gas. Alternatively, use PTFE tape or contaminant-free aluminum foil for this purpose (see Section 6.2.6 of this method).

6.2.2 Wash Bottles. Use PTFE bottles.

6.2.3 Probe-Liner, Probe-Nozzle, and Filter-Holder Brushes. Use inert bristle brushes with precleaned stainless steel or PTFE handles. Extensions of the probe brush must be made of stainless steel or PTFE and be at least as long as the probe. Use brushes that are properly sized and shaped to remove accumulated material from the nozzle and probe liner if used.

6.2.4 Filter Storage Container. Use a sealed filter holder, wide-mouth amber glass jar with PTFE-lined cap, or glass petri dish sealed with PTFE tape. Purchase precleaned amber glass jars and petri dishes, or clean according to the glassware cleaning procedures listed in Section 8.1.1.1 of this method.

6.2.5 Field Balance. Use a weighing device capable of measurements to an accuracy of 0.5 g.

6.2.6 Aluminum Foil. Use heavy duty aluminum foil cleaned by rinsing three times with hexane or toluene and stored in a pre-cleaned glass petri dish or glass jar. Do not use aluminum foil to wrap or contact filter samples due to the possibility of reaction between the sample and the aluminum.

6.2.7 Silica Adsorbent Storage Container. Use an air-tight container to store silica gel.

6.2.8 Glass Sample Storage Container. Recover samples in amber glass bottles, 500- or 1000-milliliters (mL) with leak-free PTFE-lined caps. Either purchase precleaned bottles or clean containers according to glassware cleaning procedures listed in Section 8.1.1.1 of this method.

6.3 Sample Extraction Equipment.

6.3.1 Sample Container. Use 125- and 250-mL amber glass bottles with PTFE-lined caps.

6.3.2 Test Tubes. Use glass test tubes or small (
e.g.,
5 to 10 mL) amber vials.

6.3.3 Soxhlet/Dean-Stark Extraction Apparatus.

6.3.3.1 Soxhlet Apparatus. Use 200-mL capacity thimble holder capable of holding 43 × 123-millimeter (mm) extraction thimbles, with receiving flask (typically round-bottom).

6.3.3.2 Moisture Trap. Use Dean-Stark or Barret with fluoropolymer stopcock trap to fit between the Soxhlet extractor body and the condenser as shown in Figure 23-3 of this method.

Note:
Dean-Stark or Barret traps are used to remove water with extraction solvents that are less dense and insoluble in water.

6.3.3.3 Extraction Thimble. Use quartz, glass, or glass fiber thimble, typically 43 × 123 mm to fit Soxhlet apparatus. The use of cellulose thimbles for sample extraction in this method is prohibited.

6.3.3.4 Heating Mantle. Use a hemispherical shaped heating mantle to fit round-bottom flask.

6.3.4 Kuderna-Danish (KD) Concentrator. Use an apparatus consisting of a three-ball Snyder column, a flask with leak-free joint to accept the three-ball Snyder column at the top, a leak-free joint to receive a graduated concentration tube at the bottom and a heating mantle.

Note:
Rotary evaporation has only been demonstrated when analyzing PCDD/PCDF. The KD with Snyder column is recommended when analyzing for PAH and/or PCB to avoid evaporation loss resulting in failed performance criteria for pre-extraction spike recovery.

6.3.5 Nitrogen Evaporative Concentrator. Use a nitrogen evaporative concentrator equipped with a water bath with the temperature controlled in the range of 30 to 60 °C (86 to 140 °F) (N-Evap Organomation Associates, Inc., South Berlin, MA, or equivalent).

6.3.6 Separatory Funnels. Use glass or PTFE 2-liter separatory funnels.

6.4 Glass Liquid Chromatography Columns.

6.4.1 Pasteur Pipettes. Use disposable pipettes, or glass serological pipettes typically 150 mm long × 6 mm ID.

6.4.2 Liquid Chromatography Columns. 200 to 300 mm long × 20 mm ID with 250-mL reservoir.

6.5 Analytical Equipment.

6.5.1 Gas Chromatograph. Use a gas chromatograph consisting of the following components:

6.5.1.1 GC Oven. Use an oven capable of maintaining the separation column at the proper operating temperature ± 1.0 °C (1.8 °F) and performing programmed increases in temperature at rates of at least 40 °C/min with isothermal hold.

6.5.1.2 GC Temperature Monitor. Use a temperature monitor to measure column oven temperature to ± 1.0 °C (1.8 °F).

6.5.1.3 GC Flow System. Use an electronic pressure control or equivalent gas metering system to control carrier gas flow or pressure.

6.5.1.4 GC Injection Port. Use a split/splitless injection port in the splitless mode or on-column injection port for the capillary column.

6.5.2 Capillary GC Column. Use different columns for the analysis of the different target compound classes in this method, if needed. Perform the resolution checks in Sections 10.2.3.5 and 10.2.3.6 of this method to document the required resolution. Compound separation must meet the resolution specifications in Section 10.2.3.5 of this method and the identification specifications found in Section 11.4.3.4 of this method.

6.5.2.1 PCDD/PCDF Column. Gas chromatographic columns used to measure PCDD/PCDF should be capable of achieving separation of the 17 PCDD/PCDF target compounds from the nearest eluting target compound(s). The valley height resolution between 2,3,7,8-substituted TeCDD and TeCDF and the nearest eluting isomers must not exceed 25% of the taller of the two peaks. The valley height resolution between all other target PCDD/PCDF compounds and the nearest eluting targets (or interference) must not exceed 40% of the taller of the two peaks.

Note:
Fishman, et al. (see Section 16.3 of this method) demonstrated that all TEF isomers can be fully differentiated from closely eluting isomers using either of two sets of non-polar and polar stationary phase combinations. One set consisted of 5% phenyl methylpolysiloxane (DB-5, HP-5MS, Rtx-5MS, Equity-5) and 50% cyanopropylmethyl, 50% phenylmethylsiloxane (DB-225, SP 2331) GC columns and the other set consisted of 5% phenyl, 94% methyl, 1% vinyl silicone bonded-phase (DB-5MS, ZB-5MS, VF-5MS, CP-Sil 8 CB LowBleed/MS) with 50% cyanopropylmethyl, 50% phenylmethylsiloxane (SP-2331).

6.5.2.2 PAH Column. Use column systems for measuring PAH that can achieve separation of anthracene and phenanthrene at m/z 178 such that the valley between the peaks does not exceed 50% of the taller of the two peaks, and benzo[
b
]fluoranthene and benzo[
k
]fluoranthene such that the valley between the peaks is less than 60% of the height of the taller peak. These requirements are achievable using a 30-m narrow bore (0.25 mm ID) 5% phenyl polysilphenylene-siloxane (BPX5 or equivalent) bonded-phase, fused-silica capillary column.

6.5.2.3 PCB Column. Use column systems for measuring PCB that can achieve unique resolution and identification of the toxics for determination of a TEQ
PCB
using toxic equivalency factors (TEF). Resolution is shown by a valley between the peaks not exceeding 40% of the taller of the two peaks. Isomers may be unresolved if they have the same TEF and RRF and if these unresolved isomers are uniquely resolved from all other congeners. These requirements are achievable using several 30-meter (m) narrow

bore (0.25 mm ID) columns including 8% phenyl polycarborane-siloxane (HT8), DB-XLB, and poly (50% n-octyl/50% methyl siloxane) (SPB-Octyl). Quantification of unresolved isomers should use the nearest eluting target PCB pre-extraction standard in Appendix A of this method, unless otherwise specified in applicable rule, regulation, or permit.

Note:
If all 209 PCB are analyzed the 17 toxic PCB congeners should be resolved and reported while the other PCB can be reported as totals by homolog, for example, total TrCB.

6.5.3 Mass Spectrometer. Instrument employing 28 to 70 electron volt ionization. The instrument and data system must be capable of repetitive monitoring of at least 12 exact m/z values with a mass resolution defined in Section 10.2.1 within the measurement mass range. The recommended lock-mass ions to be used for mass drift correction are presented in Tables 23-4, 23-5, and 23-6 of this method for PCDD/PCDF, PAH, and PCB, respectively, as applicable to target analytes. Mass drifts of 5 parts per million (ppm) or more can have serious effects on instrument performance.

6.5.4 Mass Spectrometer Data System. Use a data system compatible with the mass spectrometer and capable of sequencing and monitoring multiple groups of selected ions.

6.5.5 Analytical Balance. Use an analytical balance to measure within 0.1 milligram (mg).

7.0 Reagents, Media, and Standards

7.1 Filter. Glass fiber filters, without organic binder, exhibiting at least 99.95% efficiency (<0.05% penetration) on 0.3-micron dioctyl phthalate smoke particles.

7.1.1 Conduct a QC check on the filter lot prior to the field test to demonstrate that filters are free from contamination or interference by extracting and analyzing a minimum of three filters from each lot as follows. Spike with pre-extraction and pre-extraction filter recovery standards for target compounds to be measured and extract each filter separately with toluene as described in Section 11 of this method. After extraction, remove the filters and the solvent from the filters under clean conditions (
e.g.,
a clean nitrogen stream). Analyze the extracts according to the procedures in Section 11 of this method, including adding pre-analysis standard. This filter check analysis must meet the performance requirements in Section 13.1 of this method. Ongoing analysis of LMB can be used to fulfill this check. If criteria are not met for target compounds, repeat with additional filters from the lot or evaluate another lot.

7.2 Adsorbent Resin. Amberlite® XAD-2 resin.
All adsorbent resin must meet the cleanliness criteria described for LMB in Section 13.1 of this method following the same extraction, concentration, cleanup, and analysis steps as field samples. This method recommends using the procedures provided in Appendix B to this method to clean the resin before use, if needed. However, this method allows alternative cleanup procedures that use automated extraction equipment if the adsorbent meets the required performance criteria described for LMB in Section 13.1 of this method.

7.2.1 Conduct a QC check on the cleaned adsorbent lot or batch following the extraction and analyses procedures in Section 11 of this method, including adding applicable labeled standards. The cleaned adsorbent must meet the criteria described for LMB in Section 13.1 of this method. An LMB conducted with an adsorbent lot or batch can serve this purpose.

7.2.2 Storage. Store adsorbent in a solvent-rinsed nonporous clean container and secure lid.

7.3 Glass Wool.
Clean the glass wool to meet the specifications in Section 13.1 of this method. Glass wool is dried of the solvent and stored in a clean glass container with a PTFE-lined screw cap.

7.4 Water.
Use deionized or distilled water meeting requirements in Section 13.1 of this method and store in its original container or in a clean glass container with a PTFE-lined screw cap.

7.5 Silica Gel.
Indicating type for sampling, 6-16 mesh. If previously used, dry at 175 °C (347 °F) for two hours. Use new silica gel as received. As an alternative, use other types of desiccants (equivalent or better), subject to the approval of the Administrator.

7.6 Methylene Chloride.
Pesticide grade or better.

7.7 Sample Recovery Reagents.

7.7.1 Acetone. Pesticide grade or better.

7.7.2 Toluene. Pesticide grade or better.

7.8 Sample Extraction and Cleanup.

7.8.1 Potassium Hydroxide. American Chemical Society (ACS) grade, 2% (weight/volume) in water.

7.8.2 Sodium Sulfate. Granulated or powdered, reagent grade. Evaluate for cleanliness prior to use with an LMB. The LMB must meet the requirements in Section 13.1 of this method for target compounds. Store in a clean glass container with a PTFE-lined screw cap.

7.8.3 Sulfuric Acid. Reagent grade.

7.8.4 Sodium Hydroxide. 1.0 N. Weigh 40 g of sodium hydroxide into a 1-liter volumetric flask. Dilute to 1 liter with water.

7.8.5 Hexane. Pesticide grade or better.

7.8.6 Methanol. Pesticide grade or better.

7.8.7 Toluene. Pesticide grade or better.

7.8.8 High-Boiling Alkanes Used as Keeper Solvents (
e.g.,
tetradecane, nonane, decane). Pesticide grade.
Note:
Lower homologous series alkanes (nonane or decane) are necessary for higher volatility targets such as MoCB and naphthalene to maintain retention during concentration procedures. However, do not take samples to dryness when using these lower alkane homologs.

7.8.9 Liquid Column Chromatography Packing Materials. Use the following column chromatography packing materials, as needed, to prepare sample extracts by fractionation and removal of interferences. Commercially prepacked cleaning columns may be available for this purpose. The liquid column chromatography packing materials must be adequate to clean the samples to be fit for purpose and meet the performance criteria of this method. All procedures for preparing column chromatography packing materials are recommendations shown to meet the performance specifications required for the recovery of labeled compounds described in Section 13 of this method.

7.8.9.1 Alumina. Use either acidic or basic alumina in the cleanup of sample extracts. Use the same type of alumina for all samples in an analytical sequence, including those used to demonstrate LMB performance.

7.8.9.1.1 Acidic Alumina (Sigma-Aldrich® 199966 or equivalent). Brockmann activity grade 1, 100-200 mesh. Prior to use, activate the alumina by heating for 12 hours at 130 °C (266 °F). Store in a desiccator. You may use pre-activated alumina purchased from a supplier as received.

7.8.9.1.2 Basic Alumina (Sigma-Aldrich® 19943 or equivalent). Brockmann activity grade 1. Activate by heating to 600 °C (1,112 °F) for a minimum of 24 hours. Do not heat to over 700 °C (1,292 °F) because this can lead to reduced capacity for retaining the target compounds. Store at 130 °C (266 °F) in a covered flask. Recommended storage time for acidic alumina is up to five days from baking. Use prepacked alumina columns immediately after opening the vacuum-sealed pouch or container.

7.8.9.2 Florisil®. Activated, 60-100 mesh recommended. Heat previously activated Florisil® in a glass container loosely covered with aluminum foil in an oven at 130 to 150 °C (266 to 302 °F) for a minimum of 24 hours. Allow to cool and store activated Florisil® silica in a desiccator.

7.8.9.3 Silica Gel. Use either activated, acid- or base-coated silica gel in the cleanup of sample extracts. Use the same type of silica gel for all samples in an analytical sequence, including those used to demonstrate LMB performance.

7.8.9.3.1 Activated Silica Gel. Supelco® 1-3651, Bio-Sil® A, 100-200 mesh (or equivalent). Prior to use, silica gel should be activated by solvent rinsing and heat activation. It is recommended to rinse with methylene chloride and activate the silica gel by heating for at least 1 hour at 180 °C (356 °F). After allowing to cool, rinse the silica gel sequentially with methanol and toluene. Heat the rinsed silica gel at 50 °C (122 °F) for 10 minutes, then increase the temperature gradually to 180 °C (356 °F) over 25 minutes and maintain the gel at this temperature for 90 minutes. Allow to cool in a desiccator to room temperature and store in a glass container with a PTFE-lined screw cap. Alternative conditioning procedure may be used if the performance criteria in Section 13.1 are met for target compounds.

7.8.9.3.2 Acidic Silica Gel (30% weight/weight). Combine 100 g of activated silica gel with 44 g of concentrated sulfuric acid in a clean screw-capped glass container and agitate thoroughly. Disperse the solids with a stirring rod until obtaining a uniform mixture of acid-coated silica gel. Store the mixture in a glass container with a PTFE-lined screw cap.

7.8.9.3.3 Basic Silica Gel. Combine 30 g of 1 N sodium hydroxide with 100 g of activated silica gel in a clean screw-capped glass container and agitate thoroughly. Disperse solids with a stirring rod until obtaining a uniform mixture of base-coated silica gel. Store the mixture in glass container with a PTFE-lined screw cap.

7.8.9.4 Carbon/Celite® 545 (or equivalent solid support). Use of a carbon-based column

cleanup material (
e.g.,
one of the many including for example Carbopack® B or C) to further remove non-planar impurities from the samples prior to analysis may be necessary. You must evaluate alternative carbon-based sorbents for this purpose prior to their use. An 18% weight/weight mixture of Carbopack® C and Celite® 545 has been used for this purpose and should be activated at 130 °C (266 °F) for a minimum of 6 hours. Allow to cool and store this mixture in a desiccator.

7.8.10 Nitrogen. 99.999% (ultra-high) purity.

7.9 Sample Analysis.

7.9.1 Helium. 99.999% (ultra-high) purity.

7.9.2 Spiking Standards. Prepare spiking standards quantitatively at a convenient concentration (
e.g.,
10 nanograms (ng)/mL) or use commercial standards if available, to enable accurate spiking of a labeled standard at various stages of the sample and extract preparation. You may adjust the sample fortification concentrations from those recommended in Tables 23-7, 23-8, and 23-9 of this method to accommodate the concentration of target compounds anticipated in samples if the performance criteria in Section 13 of this method are met.

Note:
When adjusting the fortification concentrations in the final sample extract, consider variables such as the aliquot of extract used and injection volume of samples and calibration.

7.9.3 Pre-Sampling Adsorbent Standard. Prepare stock standard solutions in nonane to enable spiking so that the isotopically labeled compounds in the final sample extract are at the concentration shown under the heading “Pre-sampling Adsorbent Standard” in Tables 23-7, 23-8, and 23-9 of this method, for applicable target compound classes.

7.9.4 Pre-extraction Filter Recovery Standard. Prepare stock standard solutions in nonane to enable spiking so that the isotopically labeled compounds in the final sample extract are at the concentration shown under the heading “Pre-extraction Filter Recovery Standard” in Tables 23-7, 23-8, and 23-9 of this method, for applicable target compound classes.

7.9.5 Pre-extraction Standard. Prepare stock standard solutions in nonane to enable spiking so that the isotopically labeled compounds in the final sample extract are at the concentration shown under the heading “Pre-extraction Standard” in Tables 23-7, 23-8, and 23-9 of this method, for applicable target compound classes.

7.9.6 Pre-analysis Standard. Prepare stock standard solutions in nonane to enable spiking so that the isotopically labeled compounds in the final sample extract are at the concentration shown under the heading “Pre-analysis Standard” in Tables 23-7, 23-8, and 23-9 of this method, for applicable target compound classes.

8.0 Sample Collection, Preservation, and Storage

8.1 Sampling.
This method involves collection and recovery of trace concentrations of target semivolatile organic compounds. Therefore, field sampling and recovery staff should be trained and experienced in the best practices for handling and using organic solvents in field environments to recover and protect samples from contamination.

8.1.1 Pretest Preparation.

8.1.1.1 Cleaning Glassware. Clean glassware thoroughly before using. This section provides a recommended procedure, but any protocol that consistently results in contamination-free glassware meeting the LMB criteria in Section 13.1 of this method is acceptable.

8.1.1.1.1 Soak all glassware in hot soapy water (Alconox® or equivalent).

8.1.1.1.2 Rinse with hot tap water.

8.1.1.1.3 Rinse with deionized/distilled water.

8.1.1.1.4 Rinse with methanol.

8.1.1.1.5 Rinse with toluene.

8.1.1.1.6 Baking glassware up to 400 °C (752 °F) for a minimum of 2 hours may be necessary to remove contaminants or interferents from particularly dirty samples. Allow glassware to cool after baking.

Note:
Repeated baking of glassware may cause active sites on the glass surface that may irreversibly adsorb target compounds.

8.1.1.1.7 Cover glassware openings with clean glass fitting caps or cleaned aluminum foil (see Section 6.2.6 of this method).

8.1.1.1.8 Rinse glassware immediately before use with acetone and toluene.

Note:
To prepare heavily soiled glassware, remove surface residuals from the glassware by soaking in hot soapy water, rinsing with hot water, then soaking with a non-chromic acid oxidizing cleaning reagent in a strong acid (
e.g.,
NOCHROMIX® prepared according to manufacturer's directions). After the acid soak, rinse with hot water and repeat the cleaning procedures in Section 8.1.1.1 of this method.

8.1.1.2 Adsorbent Module. Load the modules in a clean area to avoid contamination. Fill a module with 20 to 40 g of XAD-2. Spike modules before the sampling event, but do not spike the modules in the field. Add the pre-sampling adsorbent standard to the top quarter of the adsorbent bed rather than onto the top or bottom of the adsorbent bed. Add sufficient spike (picograms (pg)/module) to result in the final sample theoretical concentrations specified in Tables 23-7, 23-8, and 23-9 of this method for PCDD/PCDF, PAH, and PCB, respectively, and to be above the lowest calibration concentration to ensure the standard recovery is quantitative. For samples with known or anticipated target compound concentration significantly higher or lower than the specified concentration in these tables, adjust the pre-sampling adsorbent standard concentration to the expected native compound concentration, but no less than 10 times the method detection limit (MDL). Follow the XAD-2 with cleaned glass wool and tightly cap both ends of the module. For analysis that includes PAH, use spiked modules within 14 days of preparation. See Table 23-10 of this method for storage conditions.

8.1.1.3 Sampling Train. Figure 23-1 of this method shows the complete sampling train. Follow the best practices by maintaining all sampling train components according to the procedure described in APTD-0576 Maintenance, Calibration, and Operation of Isokinetic Source-sampling Equipment (U.S. EPA 1972).

8.1.1.4 Silica Gel. Weigh several 200 to 300 g portions of silica gel in an air-tight container to the nearest 0.5 g. Record the total weight of the silica gel plus container, on the outside of each container. As an alternative, directly weigh the silica gel in its impinger or sampling holder just prior to sampling.

8.1.1.5 Filter. Check each filter against light for irregularities and flaws or pinhole leaks. Pack the filters flat in a clean glass container. Do not mark filters with ink or any other contaminating substance.

8.1.2 Preliminary Determinations. Use the procedures specified in Section 8.2 of Method 5 of Appendix A-3 to 40 CFR part 60.

8.1.2.1 Sample Volume. Unless otherwise specified in an applicable rule, regulation, or permit, sample for a minimum of 2 minutes at each traverse point. This method recommends sampling a minimum of 2.5 dry standard cubic meters (dscm).

8.1.2.2 For continuously operating processes, use the same sampling time at each traverse point. To avoid timekeeping errors, use an integer, or an integer plus one-half minute, for each traverse point.

8.1.2.3 For batch processes, determine the minimum operating cycle duration, dividing the sampling time evenly between the required numbers of traverse points. After sampling all traverse points once, sample each point again for the same duration of time per sampling point in reverse order until the operating cycle is completed. Sample all traverse points at least once during each test run.

8.1.3 Preparation of Sampling Train.

8.1.3.1 During field preparation and assembly of the sampling train, keep all train openings where contamination can enter sealed until just prior to assembly or until sampling is about to begin. To protect the adsorbent module from radiant heat and sunlight, you must wrap the module with aluminum foil or other suitable material capable of shielding the module from light. The XAD-2 adsorbent resin temperature must never exceed 50 °C (122 °F) because thermal decomposition will occur. Clean and prepare a complete set of sampling train components that will contact the sample for each sampling run, including one complete set to be used as a field train proof blank as a tool to evaluate equipment preparation and potential contamination during sample recovery as described in Section 9.6 of this method.

8.1.3.2 Place approximately 100 mL of water in the second and third impingers but leave the first and fourth impingers empty. Transfer approximately 200 g or more of silica gel from its container to the fifth impinger. Weigh each impinger and the adsorbent module, including the fitting caps, to the nearest 0.5 g using the field balance and record the weight for moisture determination. Remove the aluminum foil from the adsorbent module before weighing. Keep the module out of direct sunlight and rewrap the module with foil immediately after recording the module weight.

8.1.3.3 Using tweezers or clean disposable surgical gloves, place a filter in the filter holder. Be sure that the filter is properly centered, and the gasket properly placed, to prevent the sample gas stream from circumventing the filter. Check the filter for tears after completing the assembly.

8.1.3.4 Prepare the inside of the sampling probe and nozzle by brushing each component while rinsing three times each with acetone and toluene. Install the selected nozzle, using the connecting systems described in Section 6.1.2 of this method. Mark the probe with heat resistant tape or by some other method to denote the proper distance into the stack or duct for each sampling point. Assemble the train as shown in Figure 23-1 of this method. Orient the adsorbent module vertically so condensed moisture drains into the first impinger. See APTD-0576 Maintenance, Calibration, and Operation of Isokinetic Source-sampling Equipment (U.S. EPA 1972) for details.

8.1.3.5 Turn on the recirculation pump to the adsorbent module and condenser coil and begin monitoring the temperature of the gas entering the adsorbent module. Ensure proper temperature of the gas entering the adsorbent module before proceeding.

8.1.4 Leak-Check Procedure. Same as Section 8.4 of Method 5 of Appendix A-3 to 40 CFR part 60.

8.1.5 Sampling Train Operation. Same as Sections 8.5.1 through 8.5.9 of Method 5 of Appendix A-3 to 40 CFR part 60.

8.1.5.1 Monitor the filter temperature with a sensor and record the filter temperature during sampling to ensure a sample gas temperature exiting the filter of 120 °C ± 14 °C (248 °F ± 25 °F), or such other temperature as specified by an applicable subpart of the standards or approved by the Administrator for an application of this method.

8.1.5.2 During testing, you must record the temperature of the gas entering the XAD-2 adsorbent module. The temperature of the gas must not exceed 20 °C (68 °F) for efficient capture of the target compounds.

8.2 Sample Recovery.
Begin the cleanup procedure as soon as the probe is removed from the stack at the end of the sampling period. Seal the nozzle end of the sampling probe with PTFE tape or clean (
e.g.,
toluene rinsed) aluminum foil.

8.2.1 When the probe can be safely handled, wipe off all external particulate matter near the tip of the probe. Conduct a post-test leak check. Remove the probe from the train and close off both ends with PTFE tape or clean aluminum foil. Seal off the inlet to the train with PTFE tape, a ground glass cap, or clean aluminum foil.

8.2.2 Transfer the probe and impinger assembly to the cleanup area. This method recommends cleaning and enclosing this area to minimize the chances of losing or contaminating the sample. To avoid sample contamination and unnecessary exposure to toxic chemicals, smoking or eating in the sample recovery area shall not be allowed.

8.2.3 Inspect the train prior to and during disassembly. Note and record any abnormal conditions (
e.g.,
broken filters, colored impinger liquid). Recover and prepare samples for shipping as follows in Sections 8.2.4 through 8.2.12 of this method.

8.2.4 Container No. 1. Either seal the filter holder or carefully remove the filter from the filter holder and place it in its identified container. If it is necessary to remove the filter, use a pair of cleaned tweezers to handle the filter. If necessary, fold the filter such that the particulate cake is inside the fold. Carefully transfer to the container any particulate matter and filter fibers that adhere to the filter holder gasket by using a dry inert bristle brush and a sharp-edged blade. Seal the container and store cool (≤20 °C, 68 °F) for transport to the laboratory.

8.2.5 Adsorbent Module Sample. Remove the module from the train, tightly cover both ends with fitting caps and PTFE tape, remove the foil, drain the recirculating water from the module, weigh and record the module weight, and label the adsorbent module. Moisture measurement in the field using the Method 23 train requires weighing the adsorbent module before sampling and after sampling as part of the sample recovery.

8.2.6 Container No. 2. Quantitatively recover material deposited in the nozzle, the front half of the filter holder, and the cyclone, if used, by brushing while rinsing three times with acetone followed by three rinses with toluene. Collect all the rinses in Container No. 2.

8.2.7 Rinse the back half of the filter holder three times with acetone followed by three rinses with toluene. Rinse the sample transfer line between the filter and the condenser three times with acetone followed by three rinses with toluene. If using a separate condenser and adsorbent module, rinse the condenser three times with acetone followed by three rinses with toluene. Collect all the rinses in Container No. 2 and mark the level of the liquid on the container.

8.2.8 Moisture Weight. Weigh the adsorbent module, impingers, and silica gel impinger to within ±0.5 g using the field balance and record the weights. This information is required to calculate the moisture content of the effluent gas. For PCDD/PCDF-only measurements, discard the liquid after measuring and recording the weight.

8.2.9 Container No. 3. You must save and analyze impinger water samples if PAH and/or PCB are the target compounds. Quantitatively recover impinger water samples for analysis if PAH and/or PCB are the target compounds by rinsing three times with acetone followed by three rinses with toluene. Collect impinger water and rinses in Container No. 3 and mark the level of the liquid on the container.

8.2.10 Silica Gel. Note the color of the indicating silica gel to determine if it has been completely spent and report its condition on the field data sheet.

8.2.11 Field Sample Handling, Preservation, Storage, and Transport. Store all field samples temporarily in cool (≤20 °C, 68 °F) and dark conditions prior to transport to the laboratory. Ship samples cool (≤20 °C, 68 °F), shielded from ultraviolet light. In addition, follow the procedures in American Society for Testing and Materials (ASTM) D6911-15 (Guide for Packaging and Shipping Environmental Samples for Laboratory Analysis) for all samples, where appropriate. To avoid contamination of the samples, pay special attention to cleanliness during transport, field handling, sampling, recovery, and laboratory analysis, as well as during preparation of the adsorbent cartridges.

8.2.12 Sample Custody. Proper procedures and documentation for sample chain of custody are critical to ensuring data integrity. Follow the chain of custody procedures in ASTM D4840-99(2018)e1 (Standard Guide for Sample Chain-of-Custody Procedures) for all samples (including field samples and blanks).

8.3 Sample Storage Conditions and Laboratory Hold Times.

8.3.1 Table 23-10 of this method summarizes the sample storage conditions and laboratory hold times.

8.3.2 Store sampling train rinses and filter samples in the dark at the storage conditions in Table 23-10 from the time the laboratory receives the samples until analysis.

8.3.3 You may store adsorbent samples for PCDD/PCDF or PCB analysis prior to extraction in the dark at 6 °C (43 °F) or less for up to one year from the time the laboratory receives the samples.

Note:
The hold times listed in this method for adsorbent samples for PCDD/PCDF and PCB are recommendations as these compounds are very stable under the conditions listed in this section.

8.3.4 Protect adsorbent samples destined for PAH analysis from ultraviolet light. You may store adsorbent samples for PAH analysis in the dark at 6 °C (43 °F) or less for up to 30 days from the time the laboratory receives the samples.

8.3.5 Analyze PAH extracts within 40 days of extraction.

8.3.6 You may store sample aliquots including archived extracts of PCDD/PCDF, PAH and/or PCB samples in the dark at −10 °C (14 °F) or less for up to one year. Sample extracts must not be stored with pierced septa.

Note:
The hold times listed in this method for sample aliquots for PCDD/PCDF and PCB are recommendations as these compounds are very stable under the conditions listed in this section.

9.0 Quality Control

Note:
In recognition of advances that are occurring in sampling and analytical technology, and to allow the test team to overcome analyte sensitivity and matrix interferences, this method allows certain options to increase sample collection volume and to improve separations and the quality of the analysis results for target analytes. It is the laboratory's responsibility to establish the conditions for optimum sample extraction, cleanup, and concentration to meet the performance criteria in this method. However, you may not change the fundamental sampling and analysis techniques, isokinetic sampling with an adsorbent collection media followed by sample extraction, and HRMS detection and isotopic dilution quantification procedures. Section 13 of this method specifies the performance criteria to ensure that options employed for a sample set and analytes of interest are equal to or better than the

specificity of the techniques in this method. The minimum requirements of this method consist of the initial demonstration of capability (IDC) and ongoing QC requirements. The analysis team shall perform an IDC to demonstrate acceptable accuracy and precision with this method as described in Section 9.3. The ongoing QC includes performing CCVs and LMBs to evaluate an individual laboratory's performance against the criteria in this method. The method includes analysis of samples spiked with labeled compounds to evaluate and document data quality. Laboratory performance is compared to established performance criteria to determine if the results of analyses meet the performance characteristics and requirements of the method.

9.1 Record and report data and information
that will allow an independent reviewer to validate the determination of each target compound concentration. Record and report the data as described in Sections 9.1.1 through 9.1.7 of this method and performance criteria results required in Section 13 of this method.

9.1.1 Sample numbers and other sample identifiers. Each sample must have a unique identifier.

9.1.2 Field sample volume.

9.1.3 Field sampling date.

9.1.4 Extraction dates.

9.1.5 Analysis dates and times.

9.1.6 Analysis sequence/run chronology.

9.1.7 Quantitation Reports.

9.1.7.1 This method does not consider EMPC-flagged data to be zero concentrations. Calculate and report the EMPC concentrations.

9.1.7.2 In determining compliance with any PCDD and PCDF standard developed using zero for values that are below the EDL of the method, including federal emission standards using Method 23 promulgated under 40 CFR parts 60 and 63 prior to March 20, 2023, use zero for the determination of total and weighted concentrations when the target compound is not detected. For all other circumstances, unless otherwise specified in applicable regulations, permits, or other requirements, when a target compound is measured at or below EDL, use EDL as the concentration for calculating compliance.

9.1.7.3 For each sample you must report EDLs, MDLs, LMBs and Field Train Proof Blank results and target compound analysis results.

9.2 Isotopically Labeled Standard Recovery.

9.2.1 Pre-sampling Adsorbent Standard and Pre-extraction Filter Recovery Standard Recoveries. Pre-sampling adsorbent standard and pre-extraction filter recovery standard recoveries must demonstrate on a per sample basis that recovery of the labeled standard achieved the requirements in Section 13 of this method. Recoveries below the acceptable range for the pre-sampling adsorbent standard may be an indication of breakthrough in the sampling train.

9.2.1.1 If the pre-sampling adsorbent standard average percent recovery is below 70%, the sampling run is not valid, and the stack test must be repeated. As an alternative, you do not have to repeat the stack test for invalid analyses if the pre-sampling adsorbent standard average percent recovery is 25% or more and you divide the final results by the fraction of the pre-sampling adsorbent standard average percent recovery.

9.2.1.2 If the percent recovery of all the pre-extraction filter recovery standard compounds is below 70%, you may reanalyze the sample. If the recovery is still below the limit, the filter sampling extraction is not valid, and you must repeat the stack or vent sampling and subsequent analysis.

9.2.2 Pre-extraction Standard Recoveries. Pre-extraction standard recoveries must demonstrate on a per sample basis that recovery of the labeled standard achieved the requirements in Section 13.15 of this method. If the recovery criteria are not met, you may reanalyze the sample. If the recovery criteria are still not met, the sampling run is not valid, and the stack test must be repeated. Recoveries outside the acceptable range for pre-extraction standard are an indication that sample preparation procedures did not adequately address sample and or sample matrix processing to recover native target compounds.

9.2.3 Pre-analysis Standard Response. Pre-analysis standard recoveries must demonstrate on a per sample basis that adequate labeled standard signal meets the requirements in Section 13.16 of this method. Add pre-analysis standard to every sample (including blanks, QC samples, and calibration solutions) in a known concentration. If the prepared samples do not meet the pre-analysis standard response criteria, you may reanalyze and/or prepare and analyze archive samples to attempt meeting requirements for the compounds that do not meet the pre-analysis standard response criteria. Poor sensitivity compared to initial calibration response may indicate injection errors or instrument drift.

9.3 Initial Demonstration of Capability (IDC).
The IDC must be successfully performed prior to analyzing field samples by meeting the QC requirements in Table 23-18. The IDC must be repeated if changes are made to analytical parameters not previously validated during the IDC. This may include, for example, changing the sample volume, selecting alternate quantitation ions, extending the calibration range, adding additional pre-analysis standard, or adding additional pre-extraction standard. The same calibration range used during the IDC must be used for the analysis of field samples.

9.3.1 Perform initial calibration following the procedures in Section 10. The lowest calibration standard used to establish the initial calibration must not be less than three times the MDL. The initial calibration must meet performance criteria in Section 13.9.

9.3.2 Lowest Calibration Concentration Confirmation. Establish a target concentration for the lowest calibration standard based on the intended use of the method. The lowest calibration concentration may be established by a laboratory or programmatic lowest quantitative reporting requirement. The laboratory calibration curve must be set at or below this level. Perform seven replicate analyses of a calibration sample prepared at proposed lowest calibration concentration. The replicate analyses of the lowest calibration concentrations standards must meet the criteria in Sections 13.9 and 13.17.1.

Note:
Consider that establishing the lowest calibration concentration too low may cause repeated failure of ongoing QC requirements.

9.3.3 Calculate Lowest Calibration Statistics. Calculate the mean and standard deviation for each analyte in these replicates (those used in Section 9.3.2). Determine the Half Range for the Prediction Interval of Results (HRPIR) using Equation 23-13. Calculate the Upper and Lower Limits for the Prediction Interval of Results (PIR) with Equations 23-14 and 23-15.

9.3.4 Lowest Calibration Point Acceptance Criteria. The laboratory's ability to measure analyte concentrations down to the lowest calibration point is confirmed if the criteria presented in Section 13.17.1 are met. If these criteria are not met, the lowest calibration point as been set too low and must be confirmed at a higher concentration.

9.3.5 Demonstration of Low System Background. Analyze an LMB after the highest standard in the calibration range. If an automated extraction system is used, an LMB must be extracted on each port. Performance criteria are presented in Section 13.1. Note: When using automated systems, the same systems must be used for samples and QC samples, such as blanks and resin checks.

9.3.6 Initial Calibration Verification. A QCS must be analyzed during the IDC, and then following each initial calibration thereafter (at a minimum quarterly). A QCS is a mid-level standard prepared from a second source standard or prepared from a source of standards different from the source of calibration standards. The purpose of the QCS is to verify the integrity of the primary calibration standards. The acceptance criterion is presented in Section 13.11.

9.3.7 MDL. Perform an MDL determination using a minimum of seven spiked combined filter/sorbent media samples, spiked within 2 to 10 times of the expected MDL, and seven LMBs (combined filter/sorbent media) through all the steps of the method following the requirements in 40 CFR part 136 Appendix B. Confirm target compounds meet the qualitative identification criteria in Sections 13.12 and 13.13. The criteria for the MDL determination are presented in Section 13.6.1 of this method.

9.3.8 MDL Confirmation. Confirm newly determined MDLs by preparing a low-level spiked combined filter/sorbent media sample by spiking the sorbent with native target compounds at 1 to 5 times the MDL and pre-extraction standard at the concentration used to analyze field samples and analyze. The criterion for the MDL confirmation is presente

[Text truncated at 120,000 characters. The full text is on the page linked above.]

---